Instruction Manual 6. Calibration and Maintenance
■Stabilizationtimeof30minfromtheinitialsupplyofoperationpowertothesensorforthestabilizationofthesensor.
Calibration and test must be performed approx. 30 min after when the sensor has been stabilized.
6.1. 4-20mA Output Diagnosis
■Inastatewheregasdetectorandreceiverareconnected,whenthe
voltage is measured by connecting a multimeter to TP3 (+mA) and
TP2 (-mA) of AMP PCB, current of 4~20 mA can be confirmed.
6.2. 4mA Adjustment (ZERO Calibration)
■Checkconnectionbetweenterminalblock'CN1'andthereceiver
then confirm voltage input of 18 ~ 31 V.
■ChangeDMMtocurrentmeasuringmode,thenconnect(+)and(-)terminalsorDMMtoTP3(+mA)andTP2(-mA)
of AMP PCB, respectively.
■Injectcleanairor100%nitrogenataflowrateof500mL/minfor2min.
■Whenzerogasisinjected,4mAcurrentisoutputted.Iftheoutputisnot4mA,adjustthepotentiometer'VR2(SPAN)'
left and right to set 12 mA.
■Whenitisrotatedclockwise,thecurrentincreasesandtheoppositedecreasesthecurrent.
6.3. SPAN Calibration
■Checkconnectionbetweenterminalblock'CN1'andthereceiverthenconfirmvoltageinputof18~31V.
■ChangeDMMtocurrentmeasuringmode,thenconnect(+)and(-)terminalsorDMMtoTP5(+mA)andTP6(-mA)
of AMP PCB, respectively.
■Injectthestandardgas(airbalance)ataflowrateof500mL/minfor2min.
■Currentoutputproportionaltothestandardgasinjectedtothesensorisdisplayed.Iftheoutputisnotcorrectadjust
thepotentiometer'VR1(SPAN'leftandrighttocalibratecurrentoutput.
■Whenitisrotatedclockwise,thecurrentincreasesandtheoppositedecreasesthecurrent.
[Figure 8. 4-20mA Output Test Terminal]
6. Calibration and Maintenance
[Figure 9. ZERO Calibration related Parts]
■Ex.)OutputCalculationMethodfor
Range : 0-150ppm Calibration gas : 100ppm NH3/N2balance
Output signal : 4-20mA Test point signal: TP3, TP2
■Ex.)OutputCalculationMethodfor
Range : 0-150ppm Standard Gas : 100ppm
Output signal : 4-20mA Test point signal TP3, TP2
(200 - 40) × 100ppm NH3(Standard Gas) +40=147=14.7mA
150ppm NH3(Measuring Range)
(200 - 40) × 100ppm CO (Standard Gas) +40=147=14.7mA
150ppm CO (Measuring Range)
[Figure 10. Span Calibration related Parts]