
3
accessories. In the US, for a description of entrapment hazards, vulnerable patient profile and guidance to further
reduce entrapment risks, refer to FDA’s Hospital Bed System Dimensional and Assessment Guidance To Reduce
Entrapment. Outside the US, consult the local Competent Authority or Government Agency for Medical Device Safety
for specific local guidance. Consult a caregiver and carefully consider the use of bolsters, positioning aids or floor
pads, especially with confused, restless or agitated patients. It is recommended that side rails (if used) be locked in the
full upright position when the patient is unattended. Make sure a capable patient knows how to get out of bed safely
(and, if necessary, how to release the side rails) in case of fire or other emergency. Monitor patients frequently to
guard against patient entrapment.
It is recommended that electrically operated beds conform to IEC 60601-2-38. Medical Electrical Equipment Part 2:
particular requirements for the safety of electrically operated hospital beds.
NOTE: Side rails must be in the full upright and locked position before activating turn assist.
Exit Alarm ‑ Activation of the bed exit alarm is recommended whenever a patient is unattended. Be sure to
reactivate the bed exit alarm each time the patient returns to bed.
Ambulatory Patient Entrance / Exit ‑ Caregiver should always aid patient in exiting the bed. Lower the patient
surface completely during assisted patient entrance and exit. With caregiver assistance, ambulatory patients should
always use the patient exit button to properly position the bed and then exit at the foot of the bed.
Full Trendelenburg ‑ Use full Trendelenburg to shift patient weight back and ensure against sliding when the
footboard is released or when the head is elevated. The bed will automatically move into Trendelenburg as the head
is raised.
Adjust Side Rails to Narrow Door Position ‑ After narrow door passage, reposition the side rails to full up position
unless decision has been made to not use side rails.
Extension Pack Clip Engagement ‑ When extension packs are used, ensure that the clip at the bottom of each
extension pack fully engages side rail shaft.
I.V. and Drainage Tubes ‑ Tubes should always have slack for Turn Assist and other patient movements.
Blank Display ‑ If the main control panel remains blank, call for service immediately. In the interim, use auxiliary
bed controls (including CPR) in absence of facility protocol. Auxiliary controls are located in the head end of the base
cover, on the patient’s right side.
Fluids ‑ Avoid spilling fluids on bed controls. If spills do occur, unplug the unit and clean fluid from the bed wearing
rubber gloves to avoid any possibility of shock. Once fluid is removed, plug the power cord into a wall outlet and
check operation of components in area of spill.
NOTE: Fluids remaining on controls can cause corrosion, which may cause components to fail or operate erratically,
possibly producing potential hazards for patient and caregiver.
Lock Outs ‑ Lock out of therapy functions and air functions from the main control panel and lock out of bed
functions by the service switches should be used at the staff’s discretion to ensure against unauthorized tampering
with system settings.
Moving Parts ‑ Powered bed mechanisms can cause serious injury. Keep all equipment, tubes and lines, loose
clothing, hair and parts of the body away from moving parts and pinch points.