  9. Hanna Instruments HI 991002 User manual

Hanna Instruments HI 991002 User manual

Instruction Manual
HI 991002
ExtendedRangeWaterproof
pH/ORP/TemperatureMeter
Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
Please read this instruction manual care-
fully before using the instrument. If you need
additional technical information, do not hesi-
tate to e-mail us at [email protected]
This instrument is in compliance with the
directives.
All Hanna Instruments meters are warranted
for two years against defects in workmanship
and materials when used for their intended pur-
pose and maintained according to instructions.
The probes are warranted for a period of
six months.
This warranty is limited to repair or replacement
free of charge. Damage due to accidents, mis-
use, tampering or lack of prescribed maintenance
are not covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of pur-
chase, serial number and the nature of the prob-
lem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instru-
ments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service depart-
ment and then send it with shipment costs pre-
paid. When shipping any instrument, make sure it
is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent of
the copyright owner.
WARRANTY
PRELIMINARY EXAMINATION
Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your dealer.
Note: Save all packing material until you
are sure that the instrument func-
tions correctly. All defective items
must be returned in the original
packing together with the supplied
accessories.
HI991002 is a portable, microprocessor-
based pH/ORP/temperature meter.
Main features include: extended pH, mV
and temperature ranges; waterproof and
compact casing; large dual-line display; low
battery detection; automatic pH calibration
at one or two points with two memorized
buffer sets (standard or NIST); selectable
temperature unit (°C or °F).
The pH electrode has a built-in tempera-
ture sensor for simultaneous pH (or pH-
mV) and temperature readings, and also
contains a mini amplifier to render mea-
surements impervious to noise and elec-
trical interferences.
Each meter is supplied complete with:
•HI1296D pH/temperature probe with DIN
connector and 1 m (3.3’) cable
•calibration & cleaning solution sachets
•3x1.5V AA alkaline batteries
• instruction manual
• rugged carrying case.
GENERAL DESCRIPTION
SPECIFICATIONS
Range (*) -2.00 to 16.00 pH
±1999 mV
-5.0 to 105.0°C / 23.0 to 221.0°F
Resolution 0.01 pH
1 mV
0.1°C / 0.1°F
Accuracy ±0.02 pH
(@20°C/68°F) ±2 mV
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140°F; ±2°F outside
TypicalEMC ±0.02 pH
Deviation ±2 mV
±0.2°C or ±0.4°F
Temperature Automatic for pH readings
Compensation
pHCalibration Automatic, 1 or 2 point
with 2 sets of memorized buffers
(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)
ORPCalibration Factory calibrated
Probe (included) HI 1296D
amplified pH/temperature
probewith DINconnector and1 m(3.3’) cable
Battery Type 3x 1.5VAA/IEC LR6
Battery Life Approx. 1500 hours
Auto-off After 8 minutes of non-use
Environment 0 to 50°C (32 to 122°F);
RH max 100%
Dimensions
150 x 80 x 36 mm (5.9 x 3.1 x 1.4”)
Weight 210 g (7.4 oz.)
(*) the temperature range is limited to 80°C (176°F) if
using the HI1296D or HI1297D probes.
To clean the meter, use water only.
HI 1296D Combination amplified pH/
temperature probe with DIN
connector and 1 m (3.3’) cable
HI 1297D Combination amplified pH/
ORP/temperature probe with
DIN and 1 m (3.3’) cable
HI 7004L pH 4.01 buffer solution, 500 mL
HI 7006L pH 6.86 buffer solution, 500 mL
HI 7007L pH 7.01 buffer solution, 500 mL
HI 7009L pH 9.18 buffer solution, 500 mL
HI 7010L pH 10.01 solution, 500 mL
HI 70300L Storage solution, 500 mL
HI 7061L Cleaning solution, 500 mL
HI 710007 Blue shockproof rubber boot
HI 710008 Orangeshockproofrubber boot
HI 721312 Rugged carrying case
HI 76405 Electrode holder
ACCESSORIES
LCD DESCRIPTION
IST991002R2 07/05
1. Automatic Temperature Compensation
indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display
www.hannainst.com
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
To connect the probe
With the meter turned off, connect the HI1296D
probe to the DIN socket on the top of the meter by
aligning the pins and pushing in the plug. Tighten
the nut to ensure a good connection. Remove the
protective cap from the probe before taking any
measurements.
To turn meter ON and check battery status
Press the ON/OFF/MODE button until the display
lights up. At start-up, all the LCD segments are
displayed for 1 second, then the percent indica-
tion of the remaining battery life is shown for an-
other second (E.g. % 100 BATT). The meter then
enters the normal measuring mode.
Note:If thedisplay needs tobe checked, keep the
ON button pressed while turning the meter on.
The meter will display all segments as long as
the button is pressed.
To select the measurement range
While in normal measurement mode, press
the SET/HOLD button to select pH or mV (pH-
mV or ORP, depending on probe) measure-
ment range.
To freeze the display
While in measurement mode, press the SET/
HOLD button until HOLD appears on the second-
ary display; the reading will be frozen on the LCD
(E.g.pH5.73 HOLD). Pressanybutton to returnto
normal mode.
To turn the meter OFF
While in normal measurement mode, press the
ON/OFF/MODE button. OFF will appear on the
secondary display. Release the button.
Note: The meter is provided with an acoustic sig-
nal feature, which can be disabled using the
switch located in the battery compartment.
Note: When the meter detects the absence of a
temperature probe at its input, the Automatic
Temperature Compensation is turned off, and
the meter uses a default value of 25°C (77°F)
for the temperature measurement and com-
pensation. In this condition, the secondary LCD
shows 25.0°C (77.0°F) blinking. When a probe
is connected, the meter automatically returns
to theATC mode, theATC tag is turned on, and
the temperature is shown on the secondary
display.
OPERATIONAL GUIDE METER SETUPpH MEASUREMENT & CALIBRATION
• Make sure the probe is connected.
• Set the meter to mV range.
• Place the probe into the sample to be tested.
• Stir briefly and wait until the symbol on the top
left of the LCD is turned off.
• If using supplied HI1296D probe, the meter will
show the mV value equivalent to the pH and the
temperature of the sample; if using optional
HI1297D probe, the meter will show the ORP
(mV) value and the temperature of the sample.
Note: The ORP range is factory calibrated.
Note: ATC tag turns off because the mV readings
are not temperature compensated.
ORP MEASUREMENT
BATTERY REPLACEMENT
Setup mode allows the selection of the tempera-
ture unit and of the pH buffer set.
To enter the Setup mode, press and hold the ON/
OFF/MODEbutton until CAL on the secondary dis-
playis replaced byTEMP and thecurrent tempera-
tureunit (E.g.TEMP °C).Then:
•
for °C/°F selection
, use the SET/HOLD button.
After the temperature unit has been selected,
press ON/OFF/MODE to enter the buffer set se-
lection mode; press ON/OFF/MODE twice to re-
turn to the normal measuring mode.
•
to change the calibration buffer set
, after setting
the temperature unit, the meter will show the
currentbuffer set: “pH 7.01 BUFF” (for 4.01/7.01/
10.01) or “pH 6.86 BUFF” (for 4.01/6.86/9.18).
Change the set with the SET/HOLD button, then
press ON/OFF/MODE to return to normal mode.
• Make sure the meter has been calibrated before
use.
• If the probe is dry, soak it in HI70300 storage
solution for one hour to reactivate it.
• Place the tip of the probe into the sample to be
tested.
• Stir briefly and wait until the symbol on the top
left of the LCD is turned off.
• The LCD shows the pH value (automatically
compensated for temperature) on the primary
LCD, while the secondary LCD shows the tem-
perature of the solution.
• If measurements are taken in different samples
successively, rinse the probe tip thoroughly to
eliminate cross-contamination. After cleaning,
rinse the probe tip with some of the sample to be
measured.
pH calibration
• Fromnormalmode, pressandhold the ON/OFF/
MODEbuttonuntilOFFonthesecondarydisplay
is replaced by CAL. Release the button.
• The LCD enters the calibration mode, display-
ing “pH 7.01 USE” (or “pH 6.86 USE” if the NIST
buffersetwas selected).After1secondthemeter
activates the automatic buffer recognition fea-
ture. If a valid buffer is detected, then its value is
shown on the primary display, and REC appears
on the secondary LCD. If no valid buffer is de-
tected, the meter keeps the USE indication ac-
tive for 12 seconds, and then replaces it with
WRNG, indicating that the sample being mea-
sured is not a valid buffer.
• Fora
single-pointcalibration
withbufferspH4.01,
9.18 or 10.01, the meter automatically accepts
the calibration when the reading is stable; the
meter will show on the primary display the ac-
cepted buffer, with the message “OK 1” on the
secondary display, and an audible signal is pro-
duced. After 1 second the meter automatically
returns to the normal measuring mode.
If a single-point calibration with buffers pH 7.01
or 6.86 is desired, then after the calibration point
has been accepted the ON/OFF/MODE button
must be pressed in order to return to the normal
measuring mode. After the button is pressed,
the meter shows "7.01" (or "6.86") - "OK 1", and
an audible signal is produced.
After 1 second, the meter automatically returns
to the normal measuring mode.
Note: It is always recommended to carry out a two-
point calibration for better accuracy.
• For a
two-point calibration
, place the probe in pH
7.01(or pH 6.86) buffer.Afterthecalibration point
has been accepted, the “pH 4.01 USE” mes-
sage appears. The message is held for 12 sec-
onds, unless a valid buffer is recognized. If no
valid buffer is recognized, then the WRNG mes-
sage is shown. If a valid buffer (pH 4.01, pH
10.01 or pH 9.18) is detected, then the meter
completes the calibration procedure. When the
buffer is accepted, the LCD shows the accepted
value with the “OK 2” message on the second-
ary display. The meter then returns to the normal
measuring mode.
Note:Whenthecalibrationprocedureiscompleted,
the CAL tag is turned on.
To quit calibration and to reset to the default
values
• After entering the calibration mode and before
the first point is accepted, it is possible to quit the
procedure and return to the last calibration data
bypressingtheON/OFF/MODEbutton.Thesec-
ondary LCD displays ESC for 1 second and the
meter returns to normal mode.
• To reset to the default values and clear a previ-
ous calibration, press the SET/HOLD button af-
ter entering the calibration mode and before the
first point is accepted.
The secondary LCD displays CLR for 1 second,
the meter resets to the default calibration and
the “CAL” tag on the LCD disappears.
The meter displays the remaining battery per-
centage when turned on. When the level is be-
low 5%, the

symbol on the bottom left of
the LCD blinks to indicate a low battery condi-
tion. If the battery level is low enough to cause
erroneous readings, the Battery Error Preven-
tion System (BEPS) turns the meter off.
Unscrew the 4 screws located on the back of
the meter and carefully replace the 3AAbatter-
ies located in the battery compartment, while
paying attention to their polarity. Reattach the
back making sure that the gasket is in place
and tighten the 4 screws to ensure a water-
tight seal.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments Alkalinity HI775 User manual

Hanna Instruments

Hanna Instruments Alkalinity HI775 User manual

Hanna Instruments HI 4521 User manual

Hanna Instruments

Hanna Instruments HI 4521 User manual

Hanna Instruments HI 140 Series User manual

Hanna Instruments

Hanna Instruments HI 140 Series User manual

Hanna Instruments HI 96706C User manual

Hanna Instruments

Hanna Instruments HI 96706C User manual

Hanna Instruments HI 991001 User manual

Hanna Instruments

Hanna Instruments HI 991001 User manual

Hanna Instruments HI 96721C User manual

Hanna Instruments

Hanna Instruments HI 96721C User manual

Hanna Instruments HI708 User manual

Hanna Instruments

Hanna Instruments HI708 User manual

Hanna Instruments HI 4105 User manual

Hanna Instruments

Hanna Instruments HI 4105 User manual

Hanna Instruments HI 9813-5 User manual

Hanna Instruments

Hanna Instruments HI 9813-5 User manual

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments Checker HI 770 User manual

Hanna Instruments

Hanna Instruments Checker HI 770 User manual

Hanna Instruments HI 8010 User manual

Hanna Instruments

Hanna Instruments HI 8010 User manual

Hanna Instruments HI 93712 User manual

Hanna Instruments

Hanna Instruments HI 93712 User manual

Hanna Instruments HI 991002 User manual

Hanna Instruments

Hanna Instruments HI 991002 User manual

Hanna Instruments HI 253 User manual

Hanna Instruments

Hanna Instruments HI 253 User manual

Hanna Instruments HI 96770 User manual

Hanna Instruments

Hanna Instruments HI 96770 User manual

Hanna Instruments HI97702 User manual

Hanna Instruments

Hanna Instruments HI97702 User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments pH 209V User manual

Hanna Instruments

Hanna Instruments pH 209V User manual

Hanna Instruments HI 84184 User manual

Hanna Instruments

Hanna Instruments HI 84184 User manual

Hanna Instruments HI 4111 User manual

Hanna Instruments

Hanna Instruments HI 4111 User manual

Hanna Instruments HI 8510 User manual

Hanna Instruments

Hanna Instruments HI 8510 User manual

Hanna Instruments HI 723 User manual

Hanna Instruments

Hanna Instruments HI 723 User manual

Hanna Instruments HI 96734C User manual

Hanna Instruments

Hanna Instruments HI 96734C User manual

Popular Measuring Instrument manuals by other brands

Bosch CU-304-0503 user manual

Bosch

Bosch CU-304-0503 user manual

Sibata Aquab AQ-202 Operation manual

Sibata

Sibata Aquab AQ-202 Operation manual

Lincoln Electric D Series manual

Lincoln Electric

Lincoln Electric D Series manual

YOKOGAWA 2038 user manual

YOKOGAWA

YOKOGAWA 2038 user manual

Brodie BiRotor B064A Installation & operation manual

Brodie

Brodie BiRotor B064A Installation & operation manual

Unico SQ Series Service manual

Unico

Unico SQ Series Service manual

Fieldpiece SMAN SM380VINT Operator's manual

Fieldpiece

Fieldpiece SMAN SM380VINT Operator's manual

493K K-METRON User instructions

493K

493K K-METRON User instructions

Vernier Go Direct GDX-CA quick start guide

Vernier

Vernier Go Direct GDX-CA quick start guide

Heathkit Heathkit HM-102 Operation manual

Heathkit

Heathkit Heathkit HM-102 Operation manual

EINHELL 11037 operating instructions

EINHELL

EINHELL 11037 operating instructions

Tektronix A622 instructions

Tektronix

Tektronix A622 instructions

Anritsu MD1230B quick start guide

Anritsu

Anritsu MD1230B quick start guide

WAGNER MMC 205 instruction manual

WAGNER

WAGNER MMC 205 instruction manual

Polaroid E235GW user guide

Polaroid

Polaroid E235GW user guide

Keithley 1920 instruction manual

Keithley

Keithley 1920 instruction manual

Leica Sprinter 250M quick guide

Leica

Leica Sprinter 250M quick guide

nal von minden Reactif Touch manual

nal von minden

nal von minden Reactif Touch manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.