HBM AED9101D User manual

English
Digital Transducer Electronics
AED9101D
Basic device for AD103C
with RS‐232, RS‐422, RS‐485
Operating Manual
I3563-1.0 en

AED9101D
2I3563-1.0 en
1 Safety instructions 3........................................
2 Markings used 7............................................
2.1 Symbols on the device 7......................................
2.2 The marking used in this document 7...........................
3 Introduction 9..............................................
3.1 Scope of supply 9............................................
3.2 Application 9................................................
4 Design and function 11......................................
5 Mechanical installation 13....................................
6 Electrical connection 14......................................
6.1 Transducer connection 15......................................
6.1.1 Six-wire circuit (standard mode of operation) 15...................
6.1.2 Four-wire circuit 16............................................
6.2 Connecting the supply voltage 18...............................
6.3 Connection to a PC 19........................................
6.3.1 Connection via RS‐232 21.....................................
6.3.2 Connection via RS‐422 22.....................................
6.3.3 Connection via RS‐485 22.....................................
6.4 Using several AEDs (bus mode with RS‐485) 23..................
6.5 Connecting the diagnostic bus 25...............................
6.6 Trigger input 27...............................................
6.7 Cable connection via the PG gland 28...........................
6.8 Ground (GND) and shield wiring 29.............................
7 Selling on, waste disposal and environmental protection 31....
8 Index 33....................................................

AED9101D
I3563-1.0 en 3
1 Safety instructions
Appropriate use
The device is to be used exclusively for measurement
tasks and directly related control tasks within the applica-
tion limits detailed in the specifications. Use for any pur-
pose other than the above is deemed to be non-desig-
nated use.
Any person instructed to carry out installation, commis-
sioning or operation of the device must have read and
understood the Operating Manual and in particular the
technical safety instructions.
In the interests of safety, the device should only be oper-
ated by qualified personnel and as described in the Oper-
ating Manual. It is also essential to comply with the legal
and safety requirements for the application concerned
during use. The same applies to the use of accessories.
The device is not intended for use as a safety compon-
ent. Please also follow the instructions in the "Additional
safety precautions" section. Proper and safe operation
requires proper transportation, correct storage, siting and
mounting, and careful operation.
Operating conditions
SProtect the device from direct contact with water.
SProtect the device from moisture and weather such as
rain or snow. The protection class of the device is
IP65 (DIN EN 60529).
SDo not expose the device to direct sunlight.
SComply with the maximum permissible ambient tem-
peratures and the data on maximum humidity as
stated in the specifications.

AED9101D
4I3563-1.0 en
SThe device must not be modified from the design or
safety engineering point of view except with our
express agreement. In particular, any repair or solder-
ing work on motherboards (replacement of compon-
ents) is prohibited. When exchanging complete mod-
ules, use only original parts from HBM.
SThe device is delivered from the factory with a fixed
hardware and software configuration. Changes can
only be made within the possibilities documented in
the manuals.
SThe device is maintenance free.
SPlease note the following points when cleaning the
housing:
- Disconnect the device from all current and voltage
supplies before cleaning it.
- Clean the housing with a soft, slightly damp (not
wet!) cloth. Never use solvent as this could damage
the labeling or the housing.
- When cleaning, ensure that no liquid gets into the
device or connections.
SIn accordance with national and local environmental
protection and material recovery and recycling regula-
tions, old equipment that can no longer be used must
be disposed of separately and not with normal house-
hold garbage, see Chapter 7, page 31.
Qualified personnel
Qualified persons means persons entrusted with the
installation, fitting, commissioning and operation of the
product who possess the appropriate qualifications for
their function.

AED9101D
I3563-1.0 en 5
This includes people who meet at least one of the three
following requirements:
SKnowledge of the safety concepts of measurement
and automation technology is a requirement and as
project personnel, they must be familiar with these
concepts.
SAs measurement or automation plant operating per-
sonnel, they have been instructed how to handle the
machinery. They are familiar with the operation of the
equipment and technologies described in this docu-
mentation.
SAs commissioning engineers or service engineers,
they have successfully completed the training to qual-
ify them to repair the automation systems. They are
also authorized to activate, ground and label circuits
and equipment in accordance with safety engineering
standards.
Working safely
SThe device must not be directly connected to the
power supply system. The supply voltage must be
between 10 and 30 VDC.
SError messages should only be acknowledged once
the cause of the error is removed and no further
danger exists.
SAutomation equipment and devices must be designed
in such a way that adequate protection or locking
against unintentional actuation is provided (e.g.
access checks, password protection, etc.).
SFor those devices operating in networks, safety pre-
cautions must be taken both in terms of hardware and
software, so that a line break or other interruptions to

AED9101D
6I3563-1.0 en
signal transmission do not cause undefined states or
loss of data in the automation device.
SAfter making settings and carrying out activities that
are password-protected, ensure that any controls that
may be connected remain in a safe condition until the
switching performance of the device has been tested.
Additional safety precautions
Additional safety precautions to meet the requirements of
the relevant national and local accident prevention regu-
lations must be taken in plants where malfunctions could
cause major damage, loss of data or even personal
injury.
The scope of supply and performance of the device cov-
ers only a small area of measurement technology. Before
starting up the device in a system, a project planning and
risk analysis must first be implemented, taking into
account all the safety aspects of measurement and auto-
mation technology so that residual dangers are minim-
ized. This particularly concerns personal and machine
protection. In the event of a fault, the relevant precau-
tions must establish safe operating conditions.
General dangers of failing to follow the safety
instructions
The device is state of the art and reliable. The device
may give rise to residual dangers if it is inappropriately
installed and operated.

AED9101D
I3563-1.0 en 7
2 Markings used
2.1 Symbols on the device
CE mark
The CE mark enables the manufacturer to guarantee that
the product complies with the requirements of the relev-
ant EC directives (the Declaration of Conformity can be
found on the HBM website (www.hbm.com) under HBM-
doc).
Statutory waste disposal mark
In accordance with national and local environmental pro-
tection and material recovery and recycling regulations,
old devices that can no longer be used must be disposed
of separately and not with normal household garbage.
Also see Section7 on page31.
2.2 The marking used in this document
Important instructions for your safety are specifically
identified. It is essential to follow these instructions in
order to prevent accidents and damage to property.
Symbol Meaning
Note This marking draws your attention to a situation in
which failure to comply with safety requirements
could lead to damage to property.
Important
This marking draws your attention to important
information about the product or about handling the
product.

AED9101D
8I3563-1.0 en
Symbol Meaning
Tip
This marking indicates application tips or other
information that is useful to you.
Emphasis
See …
Italics are used to emphasize and highlight text and
identify references to sections, diagrams, or external
documents and files.
STP; Bold italics indicate inputs or commands sent to the
device.

AED9101D
I3563-1.0 en 9
3 Introduction
3.1 Scope of supply
SQuick start guide
SAED9101D (basic device)
A PDF of this operating manual is available on the HBM
website (www.hbm.com) under HBMdoc at Digital weigh
ing electronics.
3.2 Application
AED9101D digital transducer electronics are part of the
AED component family that digitally conditions signals
from mechanical transducers and networks them with
bus capability. They include digital amplifier boards, basic
devices with serial interfaces, and intelligent sensors with
integrated signal conditioning.
Using AED9101D digital transducer electronics, you can
connect SG1 transducers in a full bridge circuit directly to
a computer or a PLC. This enables you to configure com-
plete measuring chains quickly and with little extra work.
The AED9101D basic device takes AD101B or AD103C
amplifier boards. It provides mechanical protection,
shields the amplifier board (EMC protection) and allows
you to select RS-422 (factory setting), RS-485 or RS-232
serial interfaces. An amplifier board is not included in the
scope of supply of the basic device, and must be ordered
separately.
1) Strain gauges

AED9101D
10 I3563-1.0 en
AED9101D transducer electronics is shortened in the
following text to AED (digital transducer electronics).
The AED_Panel32 PC software is available to facilitate
parameter settings, to display dynamic measurement
signals and for comprehensive analysis of the system.
Download your (free) software from the HBM website, at
Digital weighing electronics: www.hbm.com/HBM Soft-
ware. The AED commands are described in the online
Help of this program.
The DWS2103 digital indicator can be connected to all
AED basic devices and supports all the implemented
functions.
Tip
All the factory settings are stored in the amplifier, where
they are power failsafe and cannot be changed. The fact
ory settings can be restored at any time, using the TDD0;
command. Further information can be found in the online
Help of the AED_Panel32 program.

AED9101D
I3563-1.0 en 11
4 Design and function
The AED basic device extends the functionality of AD
amplifier boards and provides:
Smechanical protection (IP65)
Sthe voltage supply for the amplifier board and trans-
ducer excitation
Sbridge excitation for transducers with a total bridge
resistance of 40 to 4000
Sa choice of RS-422, RS-485 and RS-232 serial inter-
faces
San EMC-tested combination of basic device and
AD101B or AD103C amplifier boards
SDiagnostic bus
The amplifier board is a plug-in board, with two plug con-
nections for inserting into the carrier board of the basic
device. The basic device contains terminals for trans-
ducer, power supply and interface connection, slide
switches for interface selection and the voltage stabilizer.
The connection cables exit the housing via PG glands on
the side.

AED9101D
12 I3563-1.0 en
1
1 Transducer
connection
2 Interface settings
3RS‐485 bus
termination
4 AD10x amplifier
connection
5 Terminals for
interface, supply
voltage, trigger input
and diagnostic bus
2
3
4
5
Fig. 4.1 Mechanical construction

AED9101D
I3563-1.0 en 13
5 Mechanical installation
Conditions on site
SProtect the device from direct contact with water.
SProtect the device from moisture and weather such as
rain or snow. The protection class of the device is
IP65 (DIN EN 60529).
SDo not expose the device to direct sunlight
SComply with the maximum permissible ambient tem-
peratures and the data on maximum humidity as
stated in the specifications.
Mounting position
The device can be mounted in any position.
Installation
Mount the device with two bolts with an outside diameter
less than 4.4mm. Remove the lid of the housing to reach
the mounting holes.

AED9101D
14 I3563-1.0 en
6 Electrical connection
Note
Electronic components are sensitive to electrostatic dis
charge (ESD). So you must discharge your own static
electricity before opening the device. We recommend
wearing an antistatic band (a conductive wristband) and
using a conductive base.
The required power supply is an extra-low voltage (10 ...
30 VDC ) with safe disconnection from the mains.
Use shielded cables for the connecting cables to trans-
ducers and interfaces and connect the shield extensively
to the device housings on both sides, not to the measure-
ment ground or the supply voltage GND (use shielded
terminals if necessary). Also see Section 6.8 on page 29.
The cables for the supply voltage and the digital I/Os only
need to be shielded if the cables are longer than 30 m, or
are routed outside closed buildings (as per EN 61326-1).
Please ensure when making connections that the cable
wires do not project beyond the connection terminals
(risk of short circuit) and do not lie on the amplifier board
(risk of interference coupling). Please also make sure
that the cable shield is properly connected to the PG
gland, see Section 6.7 on page 28.
Tip
A connection diagram is attached inside the lid of the
each AED9101D basic device.

AED9101D
I3563-1.0 en 15
6.1 Transducer connection
You can connect the SG transducers in a full bridge cir-
cuit with a total bridge resistance of RB = 40 ...
4000 . With a transducer resistance of more than
1000 , increased noise (measurement ripple) is to be
expected. Transducer excitation is supplied in the
AED9101D basic device at 5VDC (bridge excitation
voltage).
When connecting several transducers, use the HBM
VKKx junction box.
6.1.1 Sixwire circuit (standard mode of operation)
AED9101D
HBM color code:
BU = blue GN = green
RD = red WH = white
GY = gray BK = black
BU (3)
GN (3')
RD (4)
WH (1)
GY (2')
BK (2)
TE1 TE4
1
7
2
16
1
15
Fig. 6.1 Connecting transducers with a six-wire configuration
in the AED9101D basic device

AED9101D
16 I3563-1.0 en
Bridge excitation voltage (+)
Sense lead (+)
Measurement signal (-)
Measurement signal (+)
Sense lead (-)
Bridge excitation voltage (-)
Cable shield
BU (blue)
GN (green)
RD (red)
WH (white)
GY (gray)
BK (black)
Cable braid
HBM
color code Connector
3
3'
4
1
2'
2
Housing
Fig. 6.2 Transducer connections for a sixwire configuration
(HBM color and connection coding)
The six-wire connection eliminates the effect of cable
resistance. If you are using several transducers, make
the six-wire connection up to the junction box.
6.1.2 Fourwire circuit
You have two connection options for transducers in a
four-wire configuration:
1. Connection without an extension cable; the sense
leads are bridged at the AED (Fig. 6.3).
2. Connection via a 6-wire extension cable; the sense
leads are bridged in the transducer plug (Fig. 6.4).

AED9101D
I3563-1.0 en 17
AED9101D
HBM color code:
BU = blue GN = green
RD = red WH = white
GY = gray BK = black
GN (3')
RD (4)
WH (1)
GY (2')
BK (2)
BU (3)
TE1 TE4
12
16
1
15
7
Fig. 6.3 Connecting transducers with a fourwire
configuration without an extension cable
Plug connection
Cable to the transducer
electronics
BU
RD
WH
BK
BU
GN
RD
WH
GY
BK
Cable shield
Transducer
Fig. 6.4 Connecting transducers with a fourwire
configuration via a 6wire extension cable
When transducers are in a four-wire configuration, the
connections for the sense leads must be connected to

AED9101D
18 I3563-1.0 en
the relevant connections for the bridge excitation voltage,
i.e. 2' to 2 and 3' to 3 (Fig. 6.4 or Fig. 6.3).
6.2 Connecting the supply voltage
The required voltage supply is an extra-low voltage (10 ...
30 VDC) with safe disconnection from the mains. The
maximum current consumption is:
Imax +100mA )bridgeexcitationvoltageUB+5V
bridgeresistanceRB
The actual current consumption depends on the level of
the voltage supply and falls if the supply voltage is
higher.
Uv +10...30 V
GND
WH
HBM color code:
RD = red
WH = white
RD
AED9101D
TE1 TE4
12
16
1
15
7
Fig. 6.5 Connecting the voltage supply
The voltage supply leads can be routed together with the
interface leads in one cable, or you can use a separate
cable. If several AEDs are being supplied by one cable,
note the voltage drop over the cable, so that the supply
voltage to the AEDs does not get too low. The voltage

AED9101D
I3563-1.0 en 19
drop depends on the supply current and on the line res-
istance.
6.3 Connection to a PC
The basic device has an interface that can be operated in
different configurations. You can select one of them for
communication and data transmission. The choice is
made via several DIP switches (S1 to S6).
AED9101D Bus termination
Equivalent interface cable
designations:
TA = T- TB = T+
RA = R- RB = R+
RS‐232
RXD
TXD
DTR
GND
RS‐422
TB
TA
RB
RA
RS-485
4-wire
TB
TA
RB
RA
RS-485
2-wire
TB/RB
TA/RA
S1
S2
S3
S4
S5
S6
TE1 TE4
12
16
1
15
7
Fig. 6.6 Pin assignment and switch positions for the various
interfaces
Again use shielded cables for the interface cables, with
the cable shield connected to the AED housing via the
PG gland (see Sections 6.7 and 6.8, starting on page
28). The voltage supply can also be routed via this cable,
which means that a 5 or 6-wire, shielded cable is
required, depending on the interface. EMC reasons make
it an advantage to use a double-shielded cable, such as
HBM type 4-3301.0071, at 3 2 0.14 m2.

AED9101D
20 I3563-1.0 en
Interface configuration, DIP
switches S1 to S6 Bus termination
Fig. 6.7 DIP switch positions for the interface configuration
The interface cable shield must not be connected to GND
or to the supply ground, but to the housing of the AED
(see Section 6.8 on page 29).
The sections that follow contain more connection inform-
ation.
Note
With longer cables (over 30m, only possible with RS‐422
and RS‐485), there is a risk that the individual bus nodes
will have different ground potentials. The interface allows
a maximum common-mode range of -7 012 V, related
to GND. If these values could possibly be exceeded, a
separate line must be used to establish potential equaliz
ation between the bus nodes.
A flexible cable with a conductor cross-section of at least
10mm2 is the best choice for potential equalization.
Table of contents
Other HBM Transducer manuals