Mounting Instructions
Montageanleitung
Notice de montage

English Seite 3 - 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Deutsch Page 19 - 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Français Page 35 - 52. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3
C9B
A2088-3.0 en/de/fr HBM
Contents Page
English
Safety information 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Scope of supply 8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 General application instructions 8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Construction and mode of operation 9. . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Transducer 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 SG cover 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Conditions on site 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 Ambient temperature 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Moisture and corrosion protection 10. . . . . . . . . . . . . . . . . . . . . . . . . .
4.3 Deposits 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4 External pressure 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Mechanical installation 12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Important precautions during installation 12. . . . . . . . . . . . . . . . . . . .
5.2 General installation guidelines 12. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Installation 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6 Electrical connection 14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1 Connection with four‐wire configuration 14. . . . . . . . . . . . . . . . . . . . .
6.2 Shortening the cable 14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3 Cable extension 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 EMC protection 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Specifications (VDI/VDE 2638) 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Dimensions 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

C9B
4
A2088-3.0 en/de/frHBM
Safety information
Designated use
The force transducers in the type series C9B are solely designed for measur
ing static and dynamic compressive forces within the load limits specified by
the technical data for the respective maximum capacities. Any other use is not
the designated use.
To ensure safe operation, the regulations in the assembly and operating in
structions, together with the following safety rules and regulations, and the
data specified in the technical data sheets, must be complied with. It is also
essential to observe the applicable legal and safety regulations for the ap
plication concerned.
The force transducers are not intended for use as safety components. Please
also refer to the section “Additional safety precautions”. Proper and safe op
eration of the force transducer requires proper transportation, correct storage,
siting and mounting, and careful operation.
Loading capacity limits
The data in the technical data sheet must be complied with when using the
force transducer. In particular, the respective maximum loads specified must
never be exceeded. The following limits set out in the technical data sheets
must not be exceeded
•Limit forces
•Lateral limit forces
•Breaking forces
•Permissible dynamic loads
•Temperature limits
•Limits of electrical loading capacity
Please note that when several force transducers are interconnected, the load/
force distribution is not always uniform.
Use as a machine element
The force transducers can be used as machine elements. When used in this
manner, it must be noted that, to favor greater sensitivity, force transducers
cannot be designed with the safety factors usual in mechanical engineering.
Please refer here to the section “Loading capacity limits” and the technical
data.

5
C9B
A2088-3.0 en/de/fr HBM
Accident prevention
The prevailing accident prevention regulations must be taken into account,
even though the nominal (rated) force values in the destructive range are well
in excess of the full scale value.
Additional safety precautions
The force transducers cannot (as passive transducers) implement any (safety‐
relevant) cutoffs. This requires additional components and constructive mea
sures for which the installer and operator of the plant is responsible.
In cases where a breakage or malfunction of the force transducer would
cause injury to persons or damage to equipment, the user must take appropri
ate additional safety measures that meet at least the requirements of applica
ble safety and accident prevention regulations (e.g. automatic emergency
shutdown, overload protection, catch straps or chains, or other fall protection).
The layout of the electronics conditioning the measurement signal should be
such that measurement signal failure does not cause damage.
General dangers of failing to follow the safety instructions
The force transducers are state‐of‐the‐art and reliable. Transducers can give
rise to remaining dangers if they are incorrectly operated or inappropriately
mounted, installed and operated by untrained personnel. Every person in
volved with siting, starting‐up, operating or repairing a force transducer must
have read and understood the mounting instructions and in particular the
technical safety instructions. The force transducers can be damaged or de
stroyed by non‐designated use of the force transducer or by non‐compliance
with the mounting and operating instructions, these safety instructions or any
other applicable safety regulations (BG safety and accident prevention regula
tions) when using the force transducers. Force transducers can break, partic
ularly in the case of overloading. The breakage of a force transducer can also
cause damage to property or injury to persons in the vicinity of the force trans
ducer.
If force transducers are not used according to their designated use, or if the
safety instructions or specifications in the mounting‐ and operating instruc
tions are ignored, it is also possible that the force transducer may fail or mal
function, with the result that persons or property may be affected (due to the
loads acting on or being monitored by the force transducer).
The scope of supply and performance of the transducer covers only a small
area of force measurement technology, as measurements with (resistive)
strain gage sensors presuppose the use of electronic signal processing. In
addition, equipment planners, installers and operators should plan, implement
and respond to the safety engineering considerations of force measurement

C9B
6
A2088-3.0 en/de/frHBM
technology in such a way as to minimize residual dangers. Prevailing regula
tions must be complied with at all times.
Markings used in this document
The marking below warns of a potentially dangerous situation in which failure
to comply with safety requirements can result in death or serious physical in
jury.
WARNING
Description of a potentially dangerous situation
Measures to avoid/prevent the danger
The marking below warns of a potentially dangerous situation in which failure
to comply with safety requirements can result in slight or moderate physical
injury.
CAUTION
Description of a potentially dangerous situation
Measures to avoid/prevent the danger
The marking below warns of a situation in which failure to comply with safety
requirements could lead to damage to property.
NOTE
Description of a situation that could lead to damage to property
The marking below draws your attention to important information about the
product or about handling the product.
Important
Important information
The marking below contains application tips or other information that is useful
to you.

7
C9B
A2088-3.0 en/de/fr HBM
Tip
Information/Application instructions
Conversions and modifications
The transducer must not be modified from the design or safety engineering
point of view except with our express agreement. Any modification shall ex
clude all liability on our part for any damage resulting therefrom.
Maintenance
The C9B force transducer is maintenance‐free.
Disposal
In accordance with national and local environmental protection and material
recovery and recycling regulations, old transducers that can no longer be
used must be disposed of separately and not with normal household garbage.
If you need more information about waste disposal, please contact your local
authorities or the dealer from whom you purchased the product.
Qualified personnel
Qualified personnel means persons entrusted with siting, mounting, starting
up and operating the product, who possess the appropriate qualifications for
their function.
This includes people who meet at least one of the three following require
ments:
- Knowledge of the safety concepts of automation technology is a require
ment and as project personnel, you must be familiar with these concepts.
- As automation plant operating personnel, you have been instructed how to
handle the machinery. You are familiar with the operation of the equipment
and technologies described in this documentation.
- As commissioning engineers or service engineers, you have successfully
completed the training to qualify you to repair the automation systems. You
are also authorized to activate, ground and label circuits and equipment in
accordance with safety engineering standards.
It is also essential to comply with the legal and safety requirements for the ap
plication concerned during use. The same applies to the use of accessories.
The force transducer must only be installed by qualified personnel, strictly in
accordance with the specifications and with the safety requirements and regu
lations listed below.

C9B
8
A2088-3.0 en/de/frHBM
1 Scope of supply
•1 Force transducer C9B
•1 Mounting instructions
•1 Manufacturing certificate
2 General application instructions
Force transducers in the C9B type series are suitable for measuring compres
sive forces. Because they provide highly accurate static and dynamic force
measurements, they must be handled very carefully. Particular care must be
taken when transporting and installing the devices. Dropping or knocking the
transducers may cause permanent damage.
The relatively thin housing top and base must not be damaged, otherwise the
transducer will become useless, see section 3.2 on Page 9.
The specifications list the permissible limits for mechanical, thermal and elec
trical stress. It is essential that these are taken into account when planning the
measuring set‐up, during installation and ultimately, during operation.

9
C9B
A2088-3.0 en/de/fr HBM
3 Construction and mode of operation
3.1 Transducer
The measuring body is a membrane that is mounted on four strain gages
(SG) so that two are stretched and two compressed. Force is applied to the
measuring body of the C9B via a ball‐shaped load button (see also Page 13).
The membrane with the SG is elastically deformed by the force acting in the
measuring direction. The strain gauge changes its ohmic resistance in propor
tion to its change in length and so unbalances the Wheatstone bridge. If there
is a bridge excitation voltage, the circuit produces an output signal propor
tional to the change in resistance and thus also proportional to the applied
force.
3.2 SG cover
To protect the SG, the C9B force transducers are welded on the base, and on
the top with some types, with a thin metal plate cover (Fig 3.1). This proce
dure offers the SG a high protection against environmental influences.
Fig 3.1: SG protection
NOTE
The relatively thin housing cover (specially within the measuring range up to
200 N) and base (all measuring ranges) of the transducer must not be
stressed. The specifications list the permissible limits for mechanical, thermal
and electrical stress. They must be complied with, otherwise the transducer
can be destroyed.

C9B
10
A2088-3.0 en/de/frHBM
4 Conditions on site
Protect the transducer from weather conditions such as rain, snow, ice and
salt water.
4.1 Ambient temperature
The temperature effects on the zero signal and on the sensitivity are compen
sated.
To obtain optimum measurement results, the nominal (rated) temperature
range must be complied with. Constant, or very slowly changing, tempera
tures are best. Temperature‐related measurement errors are caused by heat
ing on one side (e.g. radiant heat) or by cooling. A radiation shield and all‐
round thermal insulation produces noticeable improvements, but must not be
allowed to set up a force shunt.
Tip
Hand warmth can cause a one‐sided warming of the transducer. Such temper
ature gradients can change, e.g. the zero point of the transducer. We there
fore recommend, after such warming, to wait approx. 15 minutes before start
ing measurements so that the thermal equilibrium can be restored.
4.2 Moisture and corrosion protection
Series C9B force transducers are hermetically encapsulated and are therefore
very insensitive to moisture. The transducers reach the protection class IP67
per DIN EN 60259 (test conditions: 0.5 hours under 1m water column). Nev
ertheless, the force transducers must be protected against permanent mois
ture influence.
The housing of the transducer is completely made of stainless steel. The
transducer must be protected against chemicals that could attack the trans
ducer body steel or the cable (polyurethane, PUR).
With stainless steel force transducers, please note that acids and all materials
which release ions will also attack stainless steels and their welded seams.
Should there be any corrosion, this could cause the force transducer to fail. In
this case, appropriate protective measures must be implemented.

11
C9B
A2088-3.0 en/de/fr HBM
NOTE
Moisture must not be allowed to penetrate the free end of the connection
cable. Otherwise the characteristic values of the transducer could change and
therefore cause measurement errors.
4.3 Deposits
Dust, dirt and other foreign matter must not be allowed to accumulate suffi
ciently to divert some of the measuring force onto the housing, thus invalidat
ing the measured value (force shunt).
4.4 External pressure
The external air pressure for the transducer with measuring ranges w500N
must lie between 0 and 5bar. Please note that pressure fluctuations can shift
the zero point.
Nominal (rated) force N kN
500200*) 0.5 1 2 5 10 20 50
Zero point variation with
an ambient pressure vari
ation of 10mbar (relative
to nominal (rated) force)
% 0.01 0.01 0.006 0.01 0.004 0.002 0.002 0.001
*) when used in pressure range 800 - 1200mbar

C9B
12
A2088-3.0 en/de/frHBM
5 Mechanical installation
5.1 Important precautions during installation
•Handle the transducer carefully.
•Do not permit welding currents to flow over the transducer. If this danger
exists, you must shunt the transducer electrically with a suitable
low‐resistance connection. HBM offers e.g. the highly flexible EEK ground
cable for this purpose, which can be screwed on, both above and below the
transducer.
•Ensure that the transducer cannot be overloaded.
WARNING
There is a danger of the transducer breaking if it is overloaded. This can
cause danger for the operating personnel of the system in which the
transducer is installed.
Implement appropriate safety measures to avoid overloads or to protect
against resulting dangers.
5.2 General installation guidelines
The forces to be measured must act on the transducer as accurately as pos
sible in the direction of measurement. Torsional and bending moments, ec
centric loading and lateral forces may produce measurement errors and de
stroy the transducer, if limit values are exceeded. Interference effects must be
absorbed by suitable construction elements, whereby these elements must
not absorb any loads in the measuring direction of the transducer.
Important
The cable fastening side of the transducer should always be connected dir
ectly with the rigid customer‐side force transfer areas. Ensure that the cable is
laid so that, where possible, no force shunt is caused by the cable (e.g.
through the weight or stiffness of the cable).

C9B
14
A2088-3.0 en/de/frHBM
6 Electrical connection
The following can be connected for measurement signal conditioning:
•Carrier‐frequency amplifier
•DC amplifier
designed for strain gage measurement systems.
The C9B force transducer is delivered with a four‐wire configuration.
6.1 Connection with four‐wire configuration
The transducer is delivered with a 1.5 m cable with free ends as standard.
1: white Measurement signal (+)
2: black Excitation voltage (-)
4: red Measurement signal (-)
3: blue Excitation voltage (+)
Cable shield, connected
to the housing
Fig 6.1: Transducer with a four‐core connection cable
With this cable assignment, the output voltage at the measuring amplifier is
positive when the transducer is loaded with compressive forces. If a negative
output voltage is required at the amplifier when the transducer is under com
pressive loading, the two measurement signal leads must be transposed.
The connection cable shielding is connected to the transducer housing. Trans
ducers with free cable ends must be fitted with connectors complying with the
EMC guidelines. The shielding must be connected extensively. With other
connection techniques, an EMC‐proof screen should be applied in the wire
area and this screen should also be connected extensively (see also HBM
Greenline Information, brochure i1577).
6.2 Shortening the cable
The cable can be shortened, influence on the temperature coefficients of the
characteristic values or the sensitivity itself is low.

15
C9B
A2088-3.0 en/de/fr HBM
Important
Balancing resistors (for sensitivity adjustment) are mounted on a PCB inside a
colored shrinkdown plastic tubing on transducers with nominal (rated) load of
50 N, 100 N and 200 N. If the cable is to be shortened, you must solder the
PCB back into the connection cable according to the labels on the PCB.
6.3 Cable extension
Only use shielded, low‐capacitance measurement cables for extension. En
sure that connection is perfect, with a low contact resistance. We recommend
implementing extensions using six‐wire configuration to exclude changes to
the sensitivity.
(red)
(black)
(blue)
(white)
Excitation voltage (+)
Excitation voltage (-)
Measurement signal (-)
Measurement signal (+)
Sense lead (-)
Sense lead (+)
Cable extension
(Six‐wire configuration)
Force transducer
(Four‐wire configuration)
Shielding
Amplifier
(Six‐wire configuration)
Fig 6.2: Transducer connection to amplifier with six‐wire configuration
6.4 EMC protection
Electrical and magnetic fields often induce interference voltages in the mea
suring circuit. Therefore:
•Use shielded, low‐capacitance measurement cables only (HBM cables
fulfill both conditions).
•Do not route the measurement cables parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with e.g. steel
conduit.
•Avoid stray fields from transformers, motors and contact switches.
•Do not ground the transducer, amplifier and indicator more than once.
•Connect all devices in the measurement chain to the same grounded
conductor.

C9B
16
A2088-3.0 en/de/frHBM
7 Specifications (VDI/VDE 2638 standards)
Type C9B
Nominal (rated) force Fnom N 50 100 200
kN 0.5 1 2 5 10 20 50
Nominal (rated) sensitivity Cnom mV/V v1
Accuracy class 0.5
Relative sensitivity error dC%v1
Relative reproducibility error
with unchanging mounting
position brg %v"0.5
Zero signal error ds, 0mV/V "0.075 "0.2
Relative reversibility error
(at 0.5 Fnom)ν0.5 %v"0.5
Relative linearity error dlin %v"0.5
Relative creep over 30min dcrf+E %v"0.2
Effect of temperature on the
sensitivity per 10 K
in the nominal (rated)
temperature range
in the operating temp. range
TKC%
%
v"0.5
v"0.8
Effect of temperature on the
zero signal per 10 K
in the nominal (rated)
temperature range
in the operating temp. range
TK0%
%
v"0.5
v"0.8
Output resistance RoΩ300 …400 t350
Input resistance RiΩu345 300 …400
Insulation resistance Ris GΩu109
Reference excitation voltage Uref V 5
Operating range of excitation
voltage BU, GV0.5 …12
Reference temperature Tref °C +23
Nominal (rated) temperature
range BT,nom °C-10 …+70
Operating temperature range BT, G°C-30 …+85
Storage temperature range BT, S°C-30 …+85
Max. operating force (FG)% v.
Fnom
300 120
Breaking force (FB)u400
Static lateral force limit 1) (FQ) 100 40
Nominal (rated)
displacement"15% snom mm t0.1 0.04 0.06 0.09 0.11 0.13
Fundamental resonance
frequency "15% fGkHz 7.3 10 15.7 3.5 5 7 13 15.1 20 12
Permissible oscillatory stress
(vibration bandwidth per
DIN50100) Frb % v.
Fnom 70 40
Weight, approx. g 55 65 260
Degree of protection per DIN EN 60529 IP67
Cable length m 1.5
1) referred to a force application point 2mm above membrane

17
C9B
A2088-3.0 en/de/fr HBM
8 Dimensions
R
∅U
H
B
T
J
G
A
Y
X
Cable ∅3; 1.5 m long
Min.
bending
radius R10
Dimensions in mm (1 mm = 0.03937 inches)
Nominal (rated
force) C9B A‐0.1 B G±0.1 H J R T U‐0.1 X Y
50 …200N 26 15 20.5 3xM3 6 20 2.5 5.5 approx.
10.5 approx.
5.5
0.5 …20 kN 26 13 22.75 3xM2 3.5 40 1 8 approx.1
0.5 approx.
5.5
50kN 46 28 40 4xM4 6 80 8 16 approx.1
0.5 approx.
5.5

C9B
18
A2088-3.0 en/de/frHBM

19
C9B
A2088-3.0 en/de/fr HBM
Inhalt Seite
Deutsch
Sicherheitshinweise 20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Lieferumfang 25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Allgemeine Anwendungshinweise 25. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Aufbau und Wirkungsweise 26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Aufnehmer 26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 DMS‐Abdeckung 26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Bedingungen am Einbauort 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 Umgebungstemperatur 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Feuchtigkeits‐ und Korrosionsschutz 27. . . . . . . . . . . . . . . . . . . . . . .
4.3 Ablagerungen 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4 Außendruck 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Mechanischer Einbau 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Wichtige Vorkehrungen beim Einbau 29. . . . . . . . . . . . . . . . . . . . . . .
5.2 Allgemeine Einbaurichtlinien 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Einbau 30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6 Elektrischer Anschluss 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1 Anschluss in Vierleiter‐Technik 31. . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Kabelkürzung 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3 Kabelverlängerung 32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 EMV‐Schutz 32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Technische Daten (VDI/VDE 2638) 33. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Abmessungen 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

C9B
20
A2088-3.0 en/de/frHBM
Sicherheitshinweise
Bestimmungsgemäßer Gebrauch
Die Kraftaufnehmer der Typenreihe C9B sind ausschließlich für die Messung
statischer und dynamischer Druckkräfte im Rahmen der durch die
technischen Daten spezifizierten Belastungsgrenzen konzipiert. Jeder andere
Gebrauch ist nicht bestimmungsgemäß.
Zur Gewährleistung eines sicheren Betriebes sind die Vorschriften der
Montage‐ und Betriebsanleitung sowie die nachfolgenden Sicherheitsbestim
mungen und die in den technischen Datenblättern mitgeteilten Daten unbe
dingt zu beachten. Zusätzlich sind die für den jeweiligen Anwendungsfall zu
beachtenden Rechts‐ und Sicherheitsvorschriften zu beachten.
Die Kraftaufnehmer sind nicht zum Einsatz als Sicherheitsbauteile bestimmt.
Bitte beachten Sie hierzu den Abschnitt „Zusätzliche Sicherheitsvorkeh
rungen“. Der einwandfreie und sichere Betrieb der Kraftaufnehmer setzt sach
gemäßen Transport, fachgerechte Lagerung, Aufstellung und Montage sowie
sorgfältige Bedienung voraus.
Belastbarkeitsgrenzen
Beim Einsatz der Kraftaufnehmer sind die Angaben in den technischen
Datenblättern unbedingt zu beachten. Insbesondere dürfen die jeweils
angegebenen Maximalbelastungen keinesfalls überschritten werden. Nicht
überschritten werden dürfen die in den technischen Datenblättern angege
benen
•Grenzkräfte
•Grenzquerkräfte
•Bruchkräfte
•Zulässigen dynamische Belastungen
•Temperaturgrenzen
•Grenzen der elektrischen Belastbarkeit
Beachten Sie bei der Zusammenschaltung mehrerer Kraftaufnehmer, dass die
Last-/Kraftverteilung nicht immer gleichmäßig ist.
Einsatz als Maschinenelemente
Die Kraftaufnehmer können als Maschinenelemente eingesetzt werden. Bei
dieser Verwendung ist zu beachten, dass die Kraftaufnehmer zu Gunsten
einer hohen Messempfindlichkeit nicht mit den im Maschinenbau üblichen
Sicherheitsfaktoren konstruiert wurden. Beachten Sie hierzu den Abschnitt
„Belastbarkeitsgrenzen“ und die technischen Daten.
Table of contents
Languages:
Other HBM Transducer manuals
Popular Transducer manuals by other brands

WHD
WHD Exciter X.25-4 quick start guide

HPS
HPS mks 999 Quattro Series Operation and maintenance manual

CONTINENTAL CONTROL SYSTEMS
CONTINENTAL CONTROL SYSTEMS WattNode Pulse Quick install guide

Gossen MetraWatt
Gossen MetraWatt SINEAX DME 4 Series operating instructions

Simrad
Simrad 200-7C - REV B datasheet

Daiichi Electronics
Daiichi Electronics VETT2-91A instruction manual