1
4000 COUNTS AUTO RANGE
INTELLIGENT DIGITAL AC/DC CLAMP MULTIMETER
OPERATION MANUAL
This LCD Auto Range Intelligent Digital AC/DC clamp multimeter is a portable,
4000 counts multimeter. It is ideally suited for field, laboratory, shop and home
applications. The meter can be connected with mobile phone by Bluetooth,
and display on phone by APP, you can remote monitoring the measurement
condition, the distance control is 10~15m.
1. SAFETY INFORMATION
The following safety information must be observed to insure maximum
personal safety during the operation at this meter.
1) When measuring voltage ensure that instrument is not switched to the
current range, resistance range, diode and continuity range, capacitance range
or temperature range.
2) Use extreme care when measuring voltage above 50V. especially from
sources where high energy is existed.
3) Avoid making connections to "live" circuits whenever possible.
4) Before making resistance measurements, diode or continuity test,
capacitance test or temperature test, ensure that the circuit under test is
de-energized.
5) Always ensure that the correct function and range is selected.
6) Extreme care should be taken when using the instrument to conjunction with
a current transformer connected to the terminals if an open circuit occurs.
7) Ensure that the test leads and probes are in good condition with no damage
to the insulation.
8) Take care not to exceed the over-load limits as given in the specifications.
9) Before opening the cover of the battery cabinet to replace batteries.
disconnect the test leads from any external circuit, set the selector switch to
"OFF" position.
10) Keep the fingers after the protection ring when measuring through the
instrument lead.
11) Change the battery when the symbol “ ” appears to avoid incorrect data.
2. Panel Layout
1) Rotary Switch: Use this switch to select functions and ranges.
2) D.HOLD/ Back Light key: In any range, push the key, the present display
value will be locked and the " H " symbol will appear, push it again to exit.
Press “D.HOLD” button more than 2 seconds, the back light will light, press it
more than 2 seconds again, the back light will light off.
3) SELECT key: This key work on the " " range. Push the key to
choose resistance, capacitance, diode or continuity test. Press “SELECT” in
“V”range, choose AC voltage or DC voltage test.
4) RANGE Key: Press the “RANGE” key, the meter enters manual range mode,
press it more than 2 seconds again, return to auto mode.
5) Hz/Duty Key: On “ACV/ACA” or “Hz” range, press the “Hz/Duty” key, you
can choose the Frequency or Duty Cycle measurement.
6) REL Key: Press the “REL” key, the meter enters relative measuring mode,
“REL” is displayed on the LCD and the present reading becomes the reference
value and displayed on the display. Relative measurement
REL△=measurement value-Reference value. Press it again to exit.
7) LCD display: 4000 counts digit, full function symbol display.
8) T+VΩ Input Jack、COM Input Jack、T- Input Jack
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: 4000 counts digit LCD with a max. reading of 4000.
Range control: Auto range control.
Polarity: Automatic negative polarity indication.
Zero adjustment: Automatic.
Over range indication: Only the "OL" display.
Low Battery Indication: “ ” displayed
Safety Standards: The meter is up to the standards of IEC1010 Double
Insulation, Pollution Degree 2, Overvoltage Category Ⅱ.
Clamp opening size: 45mm.
Operating Environment: Temperature 0~40℃, humidity<80%RH.
Storage Environment: Temperature -20~60℃, humidity<90%RH.
Power supply: 9V Zinc-carbon battery.
Dimension: 225(H)×77(W)×45(D)mm.
Weight: Approx. 330g (including batteries).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ±(% of reading +number in last digit) at 23±5℃,≤70%RH.
2.2.1 DC Voltage
Range
Accuracy
Resolution
400mV
±(0.5% of rdg + 2 digits)
0.1mV
4V
1mV
40V
10mV
400V
100mV
1000V
±(0.8% of rdg + 2 digits)
1V
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ, More than 100MΩ on 400mV scale
2.2.2 AC Voltage
Range
Accuracy
Resolution
Frequency
4V
±(0.8% of rdg + 3 digits)
1mV
40~400Hz
40V
10mV
400V
100mV
750V
±(1.2% of rdg + 3 digits)
1V
Average sensing, calibrated to rms of sine wave
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ.
2.2.3 DC Current
Range
Accuracy
Resolution
200A
±(3.0% of rdg + 10 digits)
0.1A
1000A
0~800A
±(3.5% of rdg + 10 digits)
1A
800~1000A
±(6.5% of rdg + 10 digits)
Overload protection: 1000Arms within 60 seconds
2.2.4 AC Current
Range
Accuracy
Resolution
Frequency
400A
±(3.0% of rdg + 10 digits)
0.1A
50~60Hz
1000A
0~800A
±(3.5% of rdg + 10 digits)
1A
800~1000A
±(6.5% of rdg + 10 digits)
Average sensing, calibrated to rms of sine wave
Overload protection: 1000Arms within 60 seconds
2.2.5 Resistance
Range
Accuracy
Resolution
400Ω
±(1.0% of rdg + 3 digits)
0.1Ω
4kΩ
±(1.0% of rdg + 2 digits)
1Ω
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
±(1.5% of rdg + 3 digits)
10kΩ
Overload protection: 250V DC/250Vrms AC
2.2.6 Capacitance
Range
Accuracy
Resolution
51.2nF
±(3.0% of rdg + 10 digits)
10PF
512nF
±(2.5% of rdg + 5 digits)
100PF
5.12μF
1nF
51.2μF
10nF
100μF
±(5.0% of rdg + 10 digits)
100nF
Overload protection: 250V DC/250Vrms AC
2.2.7 Frequency
Range
Accuracy
Resolution
Sensitivity
5.12Hz
± (0.1% of rdg + 5 digits)
0.001Hz
Sine wave 0.6V~
10V rms
(5.12MHz: 1.5V~
10V rms)
If input voltage
over range, need
adjust
51.2Hz
0.01Hz
512Hz
0.1Hz
5.12kHz
1Hz
5.12kHz
10Hz
512kHz
100Hz
5.12MHz
1kHz
Overload protection: 250V DC/250Vrms AC
Duty cycle: 0.1%~99.9%
2.2.8 Temperature (NiCr-NiSi sensor)
Range
Accuracy
Resolution
℃
-20~150℃
± ( 3℃+ 1digit )
1℃
150~1000℃
± ( 3% of rdg + 2digits )
Overload protection: 250V DC/250Vrms AC
2
2.2.9 Diode and Audible continuity test
Range
Description
Test Condition
Display read approximately
forward voltage of diode
Forward DC current
approx. 0.4mA
Reversed DC voltage
approx. 1.5V
Built-in buzzer sounds if
resistance is less than 100Ω
Open circuit voltage
approx. 0.5V
Overload protection: 250V DC or AC rms
3. MEASURING INSTRUCTION
3.1 DC &AC Voltage measurement
1) Connect the black test lead to "COM" socket and red test lead to the "VΩHz"
socket.
2) Set the selector switch to desired "V" position, it shows symbol for testing
DC voltage, if you want to test AC voltage, push “SELECT” button switch.
3) Connect test leads across the source or load under measurement.
4) You can get reading from LCD. The polarity of the red lead connection will
be indicated along with the DC voltage value.
3.2 DC Current Measurement
1) Set the selector switch to desired "200A “or "1000A " position.
2) Press the "REL" the display show "0".
Note: As the jaw core may remain some magnetic force after using for a while.
If the display can not reach "0" When press the "REL”, please take following
process to correct it:
A. To change the direction of the measured DC current.
B. Open the JAWS several times.
3) Open the clamp by pressing the jaw-opening handle and insert the cable
(one cable only) to be measured into the jaw.
4) Close the clamp and get the reading from the LCD panel.
Note: Before this measurement, disconnect the test lead with the meter for
safety. Non unit sign display of DCA.
3.3 AC Current measurement
1) Set selector switch to desired "400/1000A~" position.
2) Open the clamp by pressing the jaw-opening handle and insert the cable
(one cable only) to be measured into the jaw.
3) Close the clamp and get the reading from the LCD panel.
Note:
a) Before this measurement, disconnect the test lead with the meter for safety.
b) In same occasion that the reading is hard to read, push the D.HOLD button
and read the result later.
3.4 Resistance measurement
1) Connect the black test lead to "COM" socket and red test lead to the "VΩHz"
socket.
2) Set the selector switch to desired "Ω" position, the present function is
resistance measurement, if it is other function, push the SELECT to select
resistance measurement.
3) Connect the probes across circuit to be tested.
4) Read the result from the LCD panel.
Caution: Ensure that the circuit to be tested is "dead".
Max. input over-load: 250V rms<10sec
3.5 Capacitance measurement
1) Connect the black test lead to "COM" socket and red test lead to the "VΩHz"
socket.
2) Set the selector switch to desired " " position, push “SELECT” to
choose Capacitance measurement.
3) Connect the probes to the capacitance to be tested.
4) Read the result from the LCD panel.
Caution:
a) Capacitors should be discharged before being tested.
b) This device adopts charging mode to measure capacitance, so when testing
large capacitance, it will take longer time before the final indication, and the
larger capacitor, the longer the time (For 100uF range, it will take about 15
seconds).
c) When testing small capacitance, to assure the measurement accuracy, first
press "REL", then go on measuring.
Max. input over-load: 250V rms<10sec
3.6 Frequency measurement
1) Connect the black test lead to "COM" socket and red test lead to the "VΩHz"
socket.
2) Set the selector switch to desired "Hz" position.
3) Connect the probes to the point of measurement and read the frequency
from the display.
Max. input over-load: 250V rms<10sec
3.7 Temperature measurement
1) Connect the black test lead of the sensor to "T-" socket and the red test
lead to the "T+" socket.
2) Set the selector switch to " " position.
3) Put the sensor probe into the temperature field under measurement.
4) Read the result from the LCD panel.
Max. input over-load: 250V rms<10sec
A. The temperature function shows the random number at ordinary times,
must insert the thermocouple in temperature test hole while examining
temperature.
B. This meter in closure WRNM-010 type contact thermocouple limit
temperature is 250 ℃(300 ℃shortly ).
C. Please don't change the thermocouple at will, otherwise we can't
guarantee to measure accuracy.
D. Please don’t importing the voltage in the temperature function.
3.8 Diode & Audible continuity Testing
1) Connect the black test lead to "COM" socket and red test lead to the "VΩHz"
socket.
2) Set the rotary switch at the “ ” range position, push “SELECT” to
choose Diode or Audible continuity measurement.
3) On diode range, connect the test leads across the diode under
measurement, display shows the approx. forward voltage of this diode.
4) On Audible continuity range, connect the test leads to two point of circuit,
if the resistance is lower than approx. 100Ω, the buzzer sounds.
NOTE: Make sure the power is cut off and all capacitors need to be discharged
under this measurement.
Max. input over-load: 250V rms<10sec
3.9 Connect to mobile phone APP
The meter has serial data output function. It can be connected with mobile
phone by Bluetooth, so the measured data can be recorded, analyzed, and
processed by mobile phone APP. Before use this function, you need install the
mobile phone APP by scan the QR code.
NOTE: The mobile phone APP can be installed in iphone 4S iOS or
android 4.30 system and up.
1) Make sure the “CLAMP METER” mobile phone APP successfully installed
before any measurement.
2) Open the Bluetooth of the mobile phone, Run the “CLAMP METER” mobile
phone APP to enter the main interface.
3) Turn on the meter, the meter will connect to mobile phone automatically,
when the “Start” key change to green, click the “Start” key to measure and view
the synchronic data or graph in the mobile phone APP.
4) The MAX or MIN value and corresponding time will display on the mobile
phone APP, and the average value in a period of time from start measurement
will display too.
5) Press the “Reset” key to reset measurement, the old data will be cleared
and resume data recording. Press the “Stop” key to stop connection.
6) Press the “Data” key to view the data and time, Press the “Graph” key to
view the graph, Press the “Save” key to save the data or the graph. Press the
“Home” key to return the main interface.
7) More information about the mobile phone APP, please refer to the Help topic
including in the APP.
4. CARE AND MAINTENANCE
4.1 CARING FOR YOUR MULTIMETER
Your Digital Multi meter is an example of superior design and craftsmanship.
The following suggestions will help you care for the multi meter so you can
enjoy it for years.
1) Keep the multi meter dry. If it gets wet, wipe it dry immediately. Liquids can
contain minerals that can corrode electronic circuits.
2) Use and store the multi meter only in normal temperature environments.
Temperature extremes can shorten the life of electronic devices, damage batteries
and distort or melt plastic parts.
3) Handle the multi meter gently and carefully. Dropping it can damage the
circuit boards and cause and can accuse the multi meter to work improperly.
4) When take current measurement, keep the cable at the center of the clamp
will get more accurate test result.
5) Keep the multi meter away from dust and dirt, which can cause premature
wear of parts.
6) Wipe the multi meter with a damp cloth occasionally to keep it looking new.
Do not use harsh chemicals, cleaning solvents, or strong detergents to clean
the multi meter.
7) Use only fresh batteries of the required size and type. Always remove old or
weak batteries. They can leak chemicals that destroy electronic circuits.
8) Please take out the battery when not using for a long time.
4.2 9Volt battery replacement
1) Ensure the instrument is not connected to any external circuit. Set the
selector switch to "OFF" position and remove the test leads from the terminals.
2) Open the cover of the battery cabinet by a screwdriver.
3) Replace the old batteries with the same type batteries.
4) Close the battery cabinet cover and fasten the screw.
Above picture and content just for your reference. Please be subject to
the actual products if anything different or updated. Please pardon for
not informing in advance.

Other HoldPeak Multimeter manuals

HoldPeak 770A User manual

HoldPeak

HoldPeak 770A User manual

HoldPeak 770A User manual

HoldPeak

HoldPeak 770A User manual

HoldPeak 770D User manual

HoldPeak

HoldPeak 770D User manual

HoldPeak 870E Installation manual

HoldPeak

HoldPeak 870E Installation manual

HoldPeak 6300 User manual

HoldPeak

HoldPeak 6300 User manual

HoldPeak 870CR User manual

HoldPeak

HoldPeak 870CR User manual

HoldPeak 6208 User manual

HoldPeak

HoldPeak 6208 User manual

HoldPeak 760J Installation manual

HoldPeak

HoldPeak 760J Installation manual

HoldPeak 4201 User manual

HoldPeak

HoldPeak 4201 User manual

HoldPeak HP-90E User manual

HoldPeak

HoldPeak HP-90E User manual

HoldPeak HP-760E User manual

HoldPeak

HoldPeak HP-760E User manual

HoldPeak HP-770H User manual

HoldPeak

HoldPeak HP-770H User manual

HoldPeak HP-37B User manual

HoldPeak

HoldPeak HP-37B User manual

HoldPeak 36 k User manual

HoldPeak

HoldPeak 36 k User manual

HoldPeak 6205 User manual

HoldPeak

HoldPeak 6205 User manual

HoldPeak HP-90F User manual

HoldPeak

HoldPeak HP-90F User manual

HoldPeak 770HD User manual

HoldPeak

HoldPeak 770HD User manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 850E User manual

HoldPeak

HoldPeak 850E User manual

HoldPeak 90B User manual

HoldPeak

HoldPeak 90B User manual

HoldPeak HP-36C User manual

HoldPeak

HoldPeak HP-36C User manual

HoldPeak 90F User manual

HoldPeak

HoldPeak 90F User manual

HoldPeak 860D User manual

HoldPeak

HoldPeak 860D User manual

HoldPeak HP-37A User manual

HoldPeak

HoldPeak HP-37A User manual

Popular Multimeter manuals by other brands

Siemens SENTRON 7KT1681 manual

Siemens

Siemens SENTRON 7KT1681 manual

Elma Instruments Elma 319s user manual

Elma Instruments

Elma Instruments Elma 319s user manual

Quadlogic MiniCloset-5c installation manual

Quadlogic

Quadlogic MiniCloset-5c installation manual

Mastech M266 Series user manual

Mastech

Mastech M266 Series user manual

Fluke 8060A instruction manual

Fluke

Fluke 8060A instruction manual

Omega HHM14 user manual

Omega

Omega HHM14 user manual

Multimetrix DMM 107 user manual

Multimetrix

Multimetrix DMM 107 user manual

All-sun EM3672 user manual

All-sun

All-sun EM3672 user manual

Fieldpiece LT17 Operator's manual

Fieldpiece

Fieldpiece LT17 Operator's manual

E-one e+erne VC60B user manual

E-one

E-one e+erne VC60B user manual

Wavetek XP Series user manual

Wavetek

Wavetek XP Series user manual

HT HT77N user manual

HT

HT HT77N user manual

Velleman DVM1000 user manual

Velleman

Velleman DVM1000 user manual

Clarke CDM80 operating manual

Clarke

Clarke CDM80 operating manual

Fluke 87 V/AN Service manual

Fluke

Fluke 87 V/AN Service manual

Format 55790010 instruction manual

Format

Format 55790010 instruction manual

Velleman DVM2000 user manual

Velleman

Velleman DVM2000 user manual

Projecta DT200 manual

Projecta

Projecta DT200 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.