5
• Protect the device and accessories from bumping and im-
pacting�
• Check the device and accessories for damage before each
use� Do not use damaged components�
• Safe operation is only guaranteed with the accessories de-
scribed in the ”Accessories” section�
• The socket for the mains cord must be easily accessible�
• Socket must be earthed (protective ground contact)�
• The voltage stated on the type plate must correspond to
the mains voltage�
• The device can only be disconnected from the mains sup-
ply by pulling out the mains plug or the connector plug�
• Disconnect the mains plug before attaching or changing
any accessories�
• Disconnect the mains plug before cleaning, maintenance
and transportation of the unit�
•
The device must only be opened by trained specialists,
even during repair� The device must be unplugged from
the power supply before opening� Live parts inside the
device may still be live for some time after unplugging
from the power supply�
Coverings or parts that can be
removed from the device with-
out the aid of any tools must be
put back on the device again to ensure safe operation,
for example to keep foreign objects and liquids, etc� from
getting into the device�
• The device may only be used as prescribed and as de-
scribed in the operating instructions� This includes opera-
tion by instructed specialist personnel�
• When using critical or hazardous materials in your pro-
cesses, IKA recommends to use additional appropriate
measures to ensure safety in the experiment� For exam-
ple, users can implement measures that inhibit fire or ex-
plosions or comprehensive monitoring equipment�
• Process pathogenic material only in closed vessels under a
suitable fume hood� Please contact IKA application sup-
port if you have any question�
If the OFF switch is not within
reach when device is operating, an
EMERGENCY STOP switch that
can be easily accessed must be installed in the work area�
•
A laboratory circulator heats and circulates fluid according
to specified parameters� This involves hazards due to high
temperatures, fire and general hazards due to the device
of electrical energy� The user is largely protected by the
device of relevant standards� Further hazard sources may
arise due to the type of tempering fluid, e�g� by exceeding
or undercutting certain temperature thresholds or by the
breakage of the container and reaction with the heat car-
rier fluid� It is not possible to consider all eventualities� They
remain largely subject to the judgment and responsibility
of the operator� For this reason, it may become necessary
for user to take other precautionary safety measures�
NOTICE
WARNING
WARNING
DANGER
• Insucient ventilation may result in the formation of explo-
sive mixtures� Only use the device in well ventilated areas�
The safety circuit (safe tempera-
ture) must be adjusted so that
the maximum permissible tem-
perature cannot be exceeded even in the event of a fault�
Check the safe temperature circuit on a regular basis (see
section “Setting the safety temperature“) �
• Securely fix the ICC immersion circulator for use at the
bath, so that it cannot tip over�
• When device is used for external circulation, extra precau-
tion must be taken for hot fluid leakage due to damaged
hose�
• Use suitable hoses for connection�
• Secure hoses and tubes against slippage and avoid kinks�
• Check hoses, tubes and bath at regular intervals for pos-
sible material fatigue (cracks/leaks)�
• Mains cable should not get in contact with hot parts and
fluids�
• If you are using plastic bath, observe the permitted work-
ing temperature range and fluids�
Do not start up the device if:
- It is damaged or leaking
-
Cable (not only supply cable) is damaged�
• Be careful when filling a hot bath�
At high operating temperatures,
the temperature of housing
parts, surfaces and tubes can ex-
ceed 70 ° C�
It is dangerous to touch the
heater� The temperature of the
heater can be very high�
• After a power failure during operation, the device may
start automatically (depending on operating mode)�
• Transport the device with care�
• Do not transport or empty the bath while it is still hot�
This may result in accidents, especially scalding injuries�
In order to prevent the power
cable from falling into the fluid,
the mains cable must be secured
with the cable clamp (13, see Fig. 2)�
WARNING
WARNING
NOTICE