iM3 GS Deluxe START User manual

GS Deluxe START
Dental System
Manual
iM3 - The Veterinary Dental Company
iM3 Inc iM3 Pty Ltd
12414 95th Street 21 Chaplin Drive
Vancouver WA 98682 USA Lane Cove NSW 2066 Australia
Tel (360) 254 2981 Fax (360) 254 2940 Tel (02) 9420 5766 Fax: (02) 9420 5677
iM3 Dental Limited
Unit 9, Block 4, City North Business Park
Stamullen, Co Meath
K32 V634
Ireland
Tel Main office: +353 16911277
Tel UK Direct : 01423 224297
www.im3vet.com 2015

2
TABLE OF CONTENTS
Table of contents 2
iM3 Unique Design Features......................................................................................................... 3
iM3 GS DeluxeControl Panel Guide............................................................................................ 4
Rear of iM3 GS Deluxe & Ultrasonic Scaler coolant outlet…………………….……………. 5
Getting Started …………………………………………………………………………………..6-7
iM3 Distilled Water and CLS Enviro Solution Supply.............................................................. 8
iM3 Filtration System.................................................................................................................... 9
3 Way Air Water Syringe.............................................................................................................. 10
iM3 Advantage Low Speed Handpiece and attachments …….................................................. 11-13
High Speed Handpiece ............................................................................. 14
iM3 Dental System Bur Kits and Rodent Kits………................................................................ 15
Handpiece Pressure Adjustment.................................................................................................. 16
Handpiece Lubrication ................................................................................................................. 17
Handpiece Sterilization ………………………………………………........................................ 18
Handpiece Maintenance................................................................................................................ 19
Silent Hurricane Oil Free Compressor ……………...................................................................
Operating Instructions..............................................................................................................
Identifying Air Leaks.................................................................................................................
Compressor Component Details……………………………………………………………...
Adjusting Regulator Pressure...................................................................................................
20
20
21
22
Silent Hurricane Compressor Maintenance................................................................................ 23
Trouble Shooting;
Cartridge Toggle switches ……………………………………………………………………
High Speed Handpiece ……….................................................................................................
Low Speed Handpiece ..............................................................................................................
24
24
25
iM3 Operator Safety
Im3 Dental equipment should be operated by a qualified person—for animal use only
For operator safety, it is recommended that safety goggles
or visor, face mask and surgical gloves be worn
during all procedures.
Visit Our Web Site: www.im3vet.com

3
Unique Design Features
iM3 Dental Systems are designed with unique and practical features to make your work
even more satisfying and productive. Please note; some features listed below, may be
optional depending on the system you purchase.
Control Top
The Corian® top is a non-porous work surface which is easy to sanitize. It incorporates a groove
around the edge to prevent burs falling on the floor during procedures. The control box is made from
powder coated aluminum that won’t warp or deteriorate with age.
Color Coded System
All supply lines on iM3 dental systems are color coded to simplify service and
maintenance requirements.
Maintenance Instructions
All maintenance instructions are conveniently printed on the card hanging from the unit, or printed
on the back of the dental machine ensuring the system is kept in perfect working order.
Power Board
A four socket electric power board is conveniently located on the side of the control panel, featuring
an IEC female socket. (Optional on the GS, standard on the Deluxe)
Adjustable Stainless Steel Stand
The iM3 stand is made from high quality stainless steel and is height adjustable for work and ease
of storage. (Optional on the GS, standard on the Deluxe)
Wall Arm
An air operated approx 54”/1.50m long arm is available
Ultrasonic Scaler Coolant Outlet
The ultrasonic scaler coolant outlet is mounted conveniently on the rear of the control box, to allow
the use of the water or system to supply the coolant to your ultrasonic dental scaler.
CLS Enviro System
CLS Enviro System is an ingenious occupational health and patient safety system. CLS is designed
to help reduce the risk of atmospheric pollution and the transfer of viruses, such as FIV from patient
to patient and reduce unpleasant mouth odors.
Oil Free Silent Hurricane Compressor
Mounts underneath the unit on the stainless steel stand—supplying air wherever the unit is to be
used, and is considered more convenient than a noisy compressor which is usually situated in an-
other room and requires a length of air line. iM3’s has produced an Oil free compressor - low
maintenance, oil free air.

4
iM3 GS Deluxe START Control Panel
(A) Corian ® work surface with recessed magnetized bur holder, to prevent burs falling on the floor.
(B) Handpiece Selector switch for the selection of low speed or High Speed Handpiece.
(C) Low Speed Handpiece – used for polishing, and low speed cutting – 20,000 RPM.
(D) 3 way syringe – air, water, or mist when both buttons are pressed together.
(E) Coolant bottle pressure switch.
(F) Pressure gauge for setting handpiece pressure.
(G) Coolant Flow Control – regulates the amount of liquid to the high speed bur.
(H) High Speed Handpiece – for high speed cutting
(I) The amber bottle allows the use of CLS, which is light sensitive (contains chlorhexidine)
(J) iM3 R&R oscillating prophy angle L7465
Maintenance Instructions are printed on the back of the machine.
Choice of Stand – Wall mounted air activated arm, Stainless Steel Height adjustable stand and Stainless
Steel Height adjustable Stand with iM3 silent hurricane oil free compressor.
iM3 is a licensed trademark of iM3 Inc. Corian ® is a registered trademark of the DUPONT Company
iM3 Dental Systems
BE
C
D
F
G
H
A
I
J

5
REAR OF iM3 GS Deluxe START
With the coolant bottle pressure switch (F - Page 4) in the “ON” position, the bottle will be
pressurized to 35PSI (approx. 2.5 bar) to supply water/CLS to the 3 way syringe and the High
Speed Handpiece. It must be switched to the off position to release the pressure before re filling
the bottle.
(1) Ultra Sonic Coolant outlet – This will supply pressurized water/CLS to your ultrasonic dental
scaler – A male fitting can be supplied by iM3 for your ultrasonic scaler water line. DO NOT use
any coolant other than distilled water or CLS – any other solution may cause a blockage and will
void your warranty.
(2) Amber coolant bottle for the use of CLS or distilled water in the GS Deluxe.
(1) Ultra Sonic
Coolant Outlet
(2) PET Pressure Coolant
Bottle amber - for CLS
solution
iM3 Dental Systems
Machine maintenance
instructions printed on back
of machine
Main power supply.
IEC Male power cord is
required.
Power board overload
protector located inside
unit.

6
Getting Started
1. Attach all four castor wheels to the stainless steel base.
2. Loosen the stand height adjustment knob and lift the dental control box to working height.
Tighten the stand height adjustment knob to secure.
4. Fill the amber plastic bottle with the CLS Concentrate Solution (8.5oz or 250 ml) and
dilute with distilled water (25.4oz or 750ml) to make a 33.82 oz or 1 litre solution. Screw the
bottle into the holder with the green tubing. Do not over tighten.
5. Place the 3 way air/water syringe in its handpiece holder—center position.
6. Connect the handpieces by screwing them to the grey handpiece line (see picture below) and
place in correct handpiece holder labeled on the dental system. Page 4 (D) and (I)
7. Check the main dental machine air line
(grey in colour) is connected to the quick disconnect fitting on the compressor outlet tap (on/off)
(See page 23)
8. Plug the unit into an electrical outlet with power cord supplied and turn power board ON.
(see page 5)
iM3 Dental Systems

7
iM3 Dental Systems
Getting Started
9. Turn on the compressor and allow to pressurize, (see page 21.)
10. Make sure your compressor is set to deliver the correct working pressure. The air pressure
regulator gauge should read 80 psi (5.5Bar). If adjustment is required, please see page 27
for compressor pressure adjustment instructions. The pressure is set at the iM3 factory.
11. Place the foot control pedal on the floor in front of the unit. The foot pedal operates the High
and Low speed handpieces when depressed.
12. Check handpiece operation:
a) Select HS on the Toggle, Remove High Speed Handpiece from holder and depress
handpiece foot control pedal—handpiece should now operate. Page 4 (C)
b) Select LS and Remove Low Speed Handpiece from holder and depress handpiece foot
control pedal—handpiece should now operate. Page 4 (C)
13. Adjust individual handpiece pressure if required, (see page 16). Each handpiece—Low
speed and high speed has a specific pressure requirement.
14. Pressurize water system by selecting the coolant bottle pressure switch to ON, located on
the front of the control panel. (see page 4 F)
Important. When filling distilled water and/or CLS Enviro bottles de-pressurize the
water system, by switching the coolant bottle pressure switch to off. Check water and air
supply to the 3 way air water syringe. Depress the right hand button for air and left hand
button for water or both together for a mist.
15. Water volume. Remove high speed handpiece, from holder and depress foot control pedal.
Adjust water spray volume to a fine mist by rotating the flow control knob. (see page 4 “H”)
Please observe manufacturers handpiece maintenance instructions for
handpiece longevity.

8
iM3 Distilled Water and CLS Enviro Solution Supply
Each iM3 dental system is supplied with an amber water bottle which should be filled only
with distilled water to assist in preventing mineral or bacteria build up in the handpieces.
This water bottle supplies the high speed handpieces, 3 way air water syringe and ultra
sonic scaler coolant outlet .
The CLS solution contains chlorhexidine which is light sensitive and should be kept in the
amber bottle provided.
The coolant system for your iM3 dental system, has been preset at 25 psi.
Refilling Coolant Bottles
To refill the distilled water or CLS solution, flip the coolant pressure switch (p. 4 F) on your
dental system to OFF - depressurizing the system. After the system has been
depressurized, unscrew the bottle and refill with either distilled water or CLS Solution. After
filling the coolant bottle screw it back into the holder and flip the coolant switch to ON,
repressurizing the system.
DO NOT OVERTIGHTEN THE BOTTLES.
Please Note: When the day’s dental procedures are complete, flip the ON/OFF coolant
switch to the OFF position, depressurizing the coolant bottles for storage.
Here’s what Dr Jeanie Hawkins, DVM, Diplomat AVDC had to say about CLS
Solution and the Auto Flush System…..
“Having CLS Solution at my fingertips to flush periodontal pockets is a real plus. iM3® is the only
Company to address the problem of cleaning the airlines and high speed handpieces.”
iM3 Dental Systems
iM3 CLS solution 250ml - Ordering Code C3130

9
iM3 Dental Systems
iM3 Filtration System CHECK
Water and CLS Enviro Solution Supply
Inline filters in the supply bottles help prevent particles from entering the system and
blocking handpieces. These filters are located in the distilled water and CLS Enviro bottles.
The inline filters will need to be checked and replaced if they appear dirty.
The use of distilled water is mandatory in the water bottles, as local water may clog the
system and can corrode the brass valves in the system, apart from the risk of a bacterial
build up.
Note: Failure to use distilled water will
void the iM3 machine warranty.
Note: The use of other chlorhexidine solutions may cause crystal deposits
in handpieces and valves. Use of such solutions will void the warranty.
Do not dilute CLS beyond the recommended label instructions, as it may
result in crystal deposits forming and block handpieces and valves. Chlorine
in water will cause chlorhexidine to flocculate—use only distilled water.

10
3 Way Air/Water Syringe
Right-hand button for air (a) and the left-hand button for water (b) and both for a fine mist
(a & b).
The 3 Way Syringe is ideal for flushing away debris during prophys or drying areas during
endodontic procedures.
Tip Removal
The tip of the 3 Way Syringe may be removed for cleaning and autoclaving.
Simply depress the outer ring housing (c) and remove the tip. (See diagram below).
iM3 Dental Systems
Ultrasonic Scaler Coolant Outlet
The Ultra Sonic coolant outlet is located on the back of the control box. Page 5
This outlet provides coolant under pressure for an ultrasonic scaler.
(a)
(b)
(c)
(a) Air
(b) Water
(c) Outer Ring Housing

11
iM3 Dental Systems
Operation Recommendations
Air pressure The dental unit air pressure
should be adjusted to be 40 psi
See page 16
Lubrication To assure optimal
performance and life the
handpiece should be
lubricated daily with MD-30 oil
lubricant. Instructions
for lubrication are found on page 20.
Air supply Moist or dirty air from the
compressor can cause damage.
Check the regulator filter regularly.
If dirt or moisture is present, replace
the regulator filter and bleed the
system, to remove water or oil.
Installation Instructions
1. Preset the air pressure to 40 psi
(see page 16)
1. Connect the low speed handpiece to the
grey handpiece line
2. Fit the 4:1 Advantage straight nose cone.
Attach the Blue R&R oscillating prophy head
3. Depress the foot control fully to operate the
low speed handpiece
Note: Polishing should be carried out at 5,000 rpm or less. This can be
accomplished by running the motor at full revs (20,000 Rpm) and use a 4:1
reduction head, to achieve 5,000 revs, maintaining torque.
iM3 Advantage Low Speed Handpiece L6200
The low speed handpiece features include a light weight balanced profile and
smooth swivel action. Its vane type power system is quiet and vibration free. The
motor accommodates both doriot or ISO heads and its powerful torque is suitable
for all low speed procedures. The Low speed Motor operates at 20,000 RPM with
delivery pressure of 40 psi.

12
iM3 Dental Systems
iM3 Low Speed Handpiece Attachments
(1) 4:1 Advantage Straight Nose Cone Attachment L6250
The reducing head is designed for polishing (prophy). The speed of the polishing
cup should be maintained at less than 5,000 r.p.m. The reducing attachment
therefore allows the motor speed and torque to be maintained while reducing the
polishing speed. iM3 disposable oscillating prophy angles are fitted to the SNC.
They won’t trap muzzle hair, or generate excessive heat.
(iM3 R&R prophy cup Code L7465)
(2) 1:1 Straight Nose Cone Attachment Blue Band
For (HP burs) and the use with the iM3 rodent kit.
This includes a soft tissue protector with diamond
bur and diamond disc.
(3) Diamond Saw Attachment
This saw blade may be single or double sided
and operates in the 1:1 attachment blue band
straight nose cone on the low speed handpiece.
It should always be used with a guard.
(4) Prophy Paste
iM3 recommends the use of iM3 R&R individual tub
prophy paste. The paste contains no fluoride or
flavor oils.
It is a medium grit suitable for use with the R&R prophy cups.

13
iM3 Dental Systems
Advantage Low Speed Motor and 4:1 straight Nose Cone
See operating instructions supplied with iM3 low speed handpiece
Fitting R&R Prophy Cups (L7465)
Place the R&R prophy angle over the
chuck housing on the 4:1 SNC mak-
ing sure the slot in the R&R lines up
with the small screw in the side of the
chuck housing.
See picture to right.

14
High Speed Handpiece
The High Speed Handpiece is designed to operate at a pressure of 40 psi at the handpiece
and at speeds of up to 350,000 rpm. A very light touch is necessary as handpieces
operates with very low torque and high speed. Pressure on the handpiece of 30gms
(1ounce) or more, will cause the bur to stall.
Note: All handpieces should be lubricated prior to use with iM3 MD-30
iM3 Dental Systems

15
iM3 Autoclavable Bur Kits
There are three Bur Kits in the iM3 range, the F.G. Dog (Friction Grip), F.G. Cat and H.P.
(Handpiece Burs). Each kit contains 10 burs of various sizes and applications. Each bur
has been selected by a specialist Veterinary Dentist. iM3 Bur Kits come complete with a
laminated suggested application guide. Each bur holder carries individual re-order codes
and can be ordered in packs of five. The blue holder is autoclavable and prevents the burs
from falling out.
iM3 Dental Systems
Feline FG Bur kit
(Order code) D7530
Canine FG Bur kit
(Order code) D7510
HP Bur kit
(Order code) D7540
44.5mm long
The iM3 Rabbit/Rodent Soft tissue kit comes complete with Long diamond bur and soft
tissue protector and diamond disc with safety shield. Complete Kit (optional)
Code: D2245
The soft tissue kit should be used on an iM3 1:1 straight nose cone - Blue

16
iM3 Handpiece Pressure Adjustment
Handpiece Pressure Adjustment
Both the High Speed and Low speed handpiece should be set to a specific pressure. The Handpiece
pressure adjustment Knobs for the High Speed and Low Speed handpieces are located underneath
the dental machine behind the grey tubing for each handpiece. The adjustment knobs are a small
silver knurled knob (see picture below).
Turn your unit on and select the Handpiece (high or low speed) to be adjusted, remove the handpiece
and depress the foot pedal to operate the handpiece. Turn the adjustment knob until the desired
pressure is displayed on the handpiece pressure gauge picture 2 below.
Low
Speed
Handpiece
40 p.s.i. - Pre set by iM3
High Speed Handpiece
40 p.s.i. Pre set by iM3
iM3 Dental Systems
3 way syringe hold-
er
Low speed hand-
piece holder
Handpiece pressure
adjustment knob
PRESSURE IS READ
ON THE HANDPIECE
PRESSURE GAUGE

17
iM3 Dental Systems
Lubrication of HS & LS Handpieces L5200 & L6200
Refer to separate handpiece manual.
All handpieces should be lubricated regularly. We recommend the use of MD-30
handpiece oil before use and after 30 minutes of running time.
The foot pedal should be depressed for 30 seconds after oiling to ensure
the lubricant will reach the turbines and working portions of the handpiece.
The lubricant is placed in the smaller of the 2 large holes.
“Air entry hole” at the base of the handpiece.
NOTE:
DON’T FORGET TO DEPRESS
THE FOOT CONTROL AFTER
OILING TO ENSURE LUBRICANT
REACHES THE TURBINES.
Maintenance instructions are printed on
the back of the dental machine

18
iM3 Dental Systems
Sterilization
Refer to handpiece operating manual.
HIGH SPEED HANDPIECE
Handpieces should be thoroughly cleaned to remove any foreign particles, eg; blood, saliva
etc. Make sure that the turbine cartridge and head cavity are clean before autoclaving. If
autoclaved with bloody particles in the turbine housing, autoclaving will cause particles to
be scorched and damage to the turbine may result.
The exterior of the handpiece may be cleaned with any good surgical disinfectant. Be sure
to rinse and dry thoroughly. Lubricate handpiece turbine with MD-30 Lubricant.
Refer to handpiece operating manual.
LOW SPEED HANDPIECE
Prepare handpiece and attachments for sterilization by carefully cleaning the exterior
surface. Thoroughly scrub, rinse and clean away any residual solution and particles.
Remove moisture with a towel or use air from the 3 way air water syringe.
1. Prior to sterilization, lubricate the handpiece using MD-30 Lubricant.
HANDPIECES: Apply oil through the handpiece air inlet tube. (see page 17)
ATTACHMENT: Remove attachment from handpiece and oil through both end openings.
Note: After lubricating items, assemble and operate for a few seconds to expel excess
lubricant.
2. Seal items in sterilization tubing bags. Make sure to remove bur from handpiece prior to sealing.
3. Place handpiece or attachments in sterilization trays of dry heat sterilizer or steam autoclave. Do
not exceed temperatures of 275F or 135C. Sterilize according to manufacturers instructions.
4. After sterilization cycle is completed, allow handpieces and attachments to cool down then
lubricate handpiece and attachments thoroughly. Make sure to use a different container after
sterilization cycle, to prevent recontaminating.
Note: Lubricate handpieces using MD-30 Lubricant after cooling down, do not allow hand-
piece to sit in sealing bag overnight or for a long period. Immediately operate handpiece as
described in the following step.
5. After lubricating items, insert bur into handpieces and operate for a few seconds to
remove excess lubricant.
Do not autoclave the high speed handpiece with a bur in the chuck.

19
Handpiece Maintenance
All Handpieces have a working life which can be extended by maintaining the
correct operating pressures and speeds as well as adequate lubrication.
The working life of most Handpieces turbines is 12-24 months, however, this will
depend upon the degree of care taken and the amount of use.
Preventative maintenance will ensure a working Handpiece at all times.
iM3 Dental Systems
Handpiece Daily Weekly Monthly Yearly
iM3 Low
Speed
Handpiece
Oil in smaller of
2 big holes.
Depress the
foot pedal for
10 seconds to
ensure the oil
reaches the
Turbine
(page 17)
Oil AM or PM
before or after
use.
2-3 drops
Use iM3 MD-30
Lubricant.
Replace turbine
and ‘o’ rings
and sealing
gasket. If worn
iM3 High
Speed
Handpiece as
per the instruc-
tions for the
Low speed
above.
Page 17.
Oil AM or PM
before or after
use.
2-3 drops
Oil with MD-30
Lubricant.
Replace turbine
and ‘o’ rings
and sealing
gasket. If worn
Prophy Head
R&R Oscillat-
ing
Dispose of prophy
head between pa-
tients to prevent
cross contamina-
tion

20
Silent Hurricane Compressor
Oil Free
The compressor is perhaps the most crucial part of any dental system as it supplies the
valuable air needed to run handpieces and pressurize the water and cooling system.
The regulator/pressure gauge should be set to read 70 p.s.i./5 BAR. The tank pressure
gauge should read 110 p.s.i./7.5 BAR. See page 21
Operating the Silent Hurricane Oil Free Compressor
1. Connect the unit to the appropriate voltage.
2. Check that the air tank on/off valve is closed. See page 23.
3. Switch unit on at the mains (power outlet).
4. Turn unit on at switch situated on top of the pressure switch. See page 21.
6. Motor should now run and will stop running when 110 p.s.i./7.5 BAR. has been reached.
7. Connect the grey air line from dental unit to the compressor and turn air tank valve to
the ON position. See page 23
Identifying Air Leaks
To ensure the Silent Hurricane Compressor Motor is not subjected to higher than normal
operating temperatures, each iM3 Dental unit should be checked for air leaks at regular
intervals, the easiest way to check for air leaks is to pump the compressor tank up to 110
p.s.i./7.5 BAR. and turn the motor off. Make sure the air line to the dental head is
connected and the tank valve left in the open position with the pressure switch for the water
bottles also turned on.
Once the tank has been pumped to 110 p.s.i./7.5 BAR., check the pressure in the tank 10
minutes later. If the pressure has dropped more than 10% during that time then the unit is
leaking air which should be identified and remedied to ensure the compressor motor does
not overheat due to excessive running caused by the air leak.
Model: Silent Hurricane Oil Free compressor
Voltage: 220-240 50Hz
900W
Maximum operating Pressure 110psi (770KPa)
Manufactured By: iM3 Pty Ltd, Sydney
Australia.
iM3 Dental Systems
Table of contents
Other iM3 Dental Equipment manuals
Popular Dental Equipment manuals by other brands

DENTAURUM
DENTAURUM palexx click Instructions for use

Biolase
Biolase Waterlase iPlus user manual

Olsen
Olsen Hemodialysis Chair Quick operation guide

NSK
NSK Ti-max Z45L Operation manual

KMD technology
KMD technology Gema Scaler Sonic operating instructions

DENTAURUM
DENTAURUM palex screws Instructions for use