Usage
Thesystemstartsfromapressurereference
valuewhencalculatingtheindividualpressure
setting.Whentheuserislyingonthemattress
thedifferencesbetweenthe“unloaded”andthe
“loaded”pressurearecomparedtoamatrixand
1of5pressurestepsisselected.Whenadjusting
thepressuretheautosettingsareoverruledbythe
manualsettings.
5.3Usingthemattresssystem
EstablishingPressure(supine/faceupposition)
1.Thesystemissetupandinatedinstaticmode.
2.Whenthesystemisready,theautosensingfunction
willindicatewitha“beep”thattheusercanlaydown
onthemattress.
3.After10-15secondsthesystemhasmeasuredthe
impacttheuserhaveonthesystemandcalculateda
pressuresetting.
4.Thesystemisnowreadytouseandwillturnto
alternatingtherapyunlessotherselectionsaremade.
5.Iffurtheradjustmentsareneeded,thepressuresettings
canbealteredwiththepressurearrowbuttons.
6.Assesswhetherthepatientiscomfortableandthe
systemisfunctioningcorrectlybyperforminga
“bottomingout”test.
Thesystemrunsonthechosensettingbutwhenthe
userchangesandstaysinanewpositionthepressure
inthemattresscanadjust(higher/lower).Thesensor
inthecontrolunitwillreactandadjustthepressure
ifrequiredtocompensateforthenewposition.
“BottomingOut”Test
Whenalteringthepressuresetting,ensurethepatientisnot
“bottomingout”(insufcientlysupportedbytheaircellsand
thereforecomingincontactwithbedbase).
1.Ensuresystemisinalternationmodebutisnot
undergoinganalternation.
2.Withthepatientlyinginasupineposition,unziptop
coverjustpastsacral(bottom)region.
3.Slideyourhandalongadeatedcellunderthepatients
sacralarea(bottom).Theinnerstaticcellwillremain
inatedbutyourhandshouldslideeasilybetween
patientandbase.
4.Ifhandcanpassunderpatientthenpatientisadequately
suspendedandpressurecanbelowered.
5.Repeat“bottomingout”testafterpressurehasbeen
lowered.
Intheeventofasystemmalfunction,thealarmwill
activateandpressureLEDswillash.
EstablishingPressure(inclinedposition)
Thesystemrunsonthechosensettingbutwhenthe
userchangesandstaysinanewpositionthepressure
inthemattresscanadjust(higher/lower).Thesensor
inthecontrolunitwillreactandadjustthepressure
ifrequiredtocompensateforthenewposition.
IMPORTANT!
–Returntotheoriginalpressuresettingwhen
thepatientreturnstothesupineposition,and
performa“bottomingout”test.
–Waitaminimumof10-12minutesbetween
pressureadjustmentandpatientassessment,as
itmaytakeafullcycleforthesystemtoadjust
toanynewsetting.
5.4HeelCellZones-SoftCloudPro
ForSoftCloudPro,theveheelcellsincorporateanoption
whichallowsthecarertopermanentlyofoadtheheelof
theuser,idealforthemostvulnerableuserssuchasthose
withlimbischaemiaorexistingwounds.
CAUTION!
Insufcientpressurerelief
–Deatingmorethanoneadjacentaircellcan
affectthesupportoftheuserduringthenormal
alternatingsupportcycle.Thisfunctionshould
onlybeusedfortemporaryprocedures.
AB
1.OpenthepushbuttonapAatthefootendofthe
mattresstoaccessthe5releaseconnectors.
2.DisconnectthechosenconnectorBforpermanent
ofoading.
•Topconnectorisclosesttotheknee.
•Connector4–2isbelowthekneeandgoes
towardstheheel.
•Bottomconnectoristhelastheelcellclosest
totheendofthemattress.
Itisnotrecommendedtodeatemorethanone
aircelldirectlybeneaththeareawhererelief
isneeded.
3.Reconnecttheconnector(s)todisengagepermanent
ofoading.Theaircellwillbeinatedagainandreturn
tonormaloperationmode.
4.ClosethepushbuttonapA.
5.5CPRProcedure
1630732-C11