Mars1717XF User Manual Safety Information
8
Copyright © iRay Technology Co., Ltd.
1.2.5 Battery Pack
If the enclosure is broken or overheating, or emits unusual odors, smoke, or leaks any-
thing. Avoid contact with any material leaking from the battery pack.
If any liquid comes into contact with your skin or eyes, wash the affected area with clean
running water and seek immediate medical attention.
If this device is not in use for a long period of time (more than 5 days), store it with the
battery pack moved. Make sure to recharge it to 30%~50% every 3 months, and 50%~
70% every 6 months to maintain its best performance.
Do not hit or squeeze the battery. Do not insert any object into the battery or use any
device to pry open the battery pack case.
Attempting to doing so will damage the battery case, which may cause the battery to release
toxic and hazardous substances, resulting in injuries such as electric shock, burns, or fire, and
render the battery unusable.
Keep the battery away from fire, do not use it at high temperature, and do not charge the
battery near flammable materials.
Do not invert positive and negative poles.
Do not touch the battery with metal to prevent short circuit.
Do not use non-standard batteries.
Do not change the internal structure of the battery.
Do not immerse the battery in water or other liquids, keep it dry when using the battery.
Do not use batteries not provided by iRay.
Do not charge damaged batteries or charge batteries with damaged chargers.
Do not use the battery charger around the patient.
Please remove the battery when the detector is not used for a long time.
1.2.6 Maintenance and Inspection
Be sure to turn off the power of detector when the inspections indicated in this manual
are going to be performed. If the detector is powered by AC power supply, turn off the
power switch and/or unplug the AC power cord. If the detector is powered by a battery,
remove the battery.
Ignoring this warning may result in an explosion, fire, or electric shock, which may result in
personal injury, death, or substantial product damage.
NEVER use alcohol, ether and other flammable cleaning agent to clean the equipment
for the sake of safety. NEVER use methanol, benzene, acid, alkali or other corrosive liq-
uids to clean the equipment.
The X-ray detector must be repaired by X-ray detector manufacturer authorized person-
nel only.
Ignoring this warning may result in explosion, fire, electric shock, or unknown hazards, which
may result in severe personal injury, death, or substantial product damage.
PROHIBITED
!