TEE-SH-e-0810 3
1 Features
•Full Tare
•Display segment test function
•Negative value indication
•Stable indicator
•Low battery indicator
•Auto zero
•Auto off function
•Overload protection
•Two types of Digital Auto Calibration
•Solder pads to prevent end-user internal calibration
•Operated by 2 x AA alkaline batteries
2 Calibration Procedure (CAL)
1. Turn scale on and allow the unit to acclimatize and stabilize for 1 minute.
2. Press and hold the [Light/Cal] key for approx. 3 seconds. Display will show [CAL]
and then the required calibration weight.
3. Gently place the correct calibration weight on the center on the weighing pan.
4. Wait until display shows [ F] then turn off.
In case [ E] is displayed instead of [ F], this indicates a calibration
procedure error, unstable or wrong calibration weight applied for calibration.
Turn the scale off and on again and repeat the procedure.
3 Internal Calibration Procedure
1. Remove two screws of the rear housing and open it.
2. Connect J1 on the main board by soldering the pads together.
3. Place scale on a hard level surface. Turn scale on.
Display shall show the internal counts.
4. The internal counts shall fall in the range from 3 000 to 5 000. In case out of this
range, connect J8 left side or right side to increase or reduce the reading.
Connecting or disconnecting pads on J5, J6 or J7 can fine tune the reading.
5. Press [On/Zero/Off] key once. Display will show [CAL] and then the required calibration
weight. Place the corresponding calibration weight on the center of the weighing pan
and wait until display shows [ F] and then turn off.
6. Disconnect J1. Calibration completed.
7. Turn scale on and check the accuracy at different weight.
8. Install rear housing of the scale.
In case [CAL E] is display instead of [CAL F], this indicates a calibration procedure error
or wrong weight applied for calibration.
Turn the scale off and then on and repeat the procedure.