6. Mechanical Connection
Before installation
Please ensure that the actual flow-volume is in agreement with the measuring
range of the unit. The measuring range can be read from the model label.
Attention! On over-stepping of measuring range (more than 20 %), may
cause damage to bearings and significant measurement errors may result.
Make sure that the permitted max. operating pressure and temperature of the
device are not violated.
Remove all the transportation-safety locks etc. and ensure, that there exist no
packing-material parts inside the unit.
These units can be mounted without regard to position. Thereby, the flow must
always be in the direction of arrow, and the front-side of the unit remains in the
vertical plain.
In particular, attention must be paid to flow from top to bottom, such that the
flow chamber is completely filled with the liquid. Additional inlet and outlet
lengths are not required.
The sealing of connection screw joints is succeeded through PTFE tape or
similar material.
Sealing of units with flange connections succeeds by means of a suitable flat
seal, provided by the customer.
During installation of the unit, attention must be paid to see that no high
pressure or tensile loads exert on the connection screw joints.We recommend
fastening an inlet and outlet pipe mechanically at a distance of approx. 50 mm
from the connection screw joint.
If possible, after mechanical installation, check whether the sealing of
connection-joints/piping is adequate
Note! The union fittings (connection screw joints) of the device must be
absolutely countered when screwing in by means of a suitable fork wrench.
Otherwise voltages will be transferred to the transducer housing, which can
lead to destruction of the device.