is released from the mixing head, operation will stop. When set to the ON position, the
mixer will operate continuously until the switch is moved to either the OFF or TOUCH
position.
The speed control, located on the right side of the panel, is continuously adjustable.
Oscillating speed increases at a non-uniform rate to a maximum of 3,400rpm as the
knob is turned clockwise. Before turning on the mixer, set the speed control to the
minimum.
To begin operation, move the power switch to either the TOUCH or ON position and turn
the speed control to the desired setting. If the power switch is in the ON position, the
shaking head will begin moving immediately. If the power switch is in the TOUCH
position, operation will begin when an object is placed on the shaking head and slight
pressure is applied to depress the head. Shaking will cease when the object is removed
and pressure released.
Safety Precautions
The following precautions should be taken when operating or working near the Vortex
Mixer:
• CAUTION; A PINCH POINT EXISTS BETWEEN CUP HEAD AND ENCLOSURE.
• Never operate the unit without a shakinghead securelyattached.
• Alwayswear shatterproofeye protection.
• Donotuseor mix solventsand flammables on or near the mixer.
• Shake hazardous samples in appropriatecontainmentvessels.
• Repair should be performedonly byqualified individuals
• Protection impairment may resultif used in a manner not specified by the manufacturer.
Optional Accessories
Instruction for using accessories:
1 – Turn power off and rotate speed knob to minimum
2 – Hold the unit down and lift the mixing head
3 – Apply the desired optional accessory and load it in a balanced manner
4 – Turn on the unit
5 – Slowly rotate the speed knob clockwise until the desired mixing level is achieved
Important: If the unit starts to walk, reduce speed accordingly
Optional shaking heads are as follows:
S0200-21 Microtube head, foam, for 80 tubes
S0200-22 Microplate head, foam, for 1 microplate or 64 x 0.2mL tubes
S0200-23 10-13mm and 15mL tube head, for 16 tubes
S0200-24 50mL tube head, for 6 tubes
S0200-25 Head for 12 x 1.5ml tubes, held horizontally
S0200-26 Head for 4 x 15 ml tubes, held horizontally
S0200-27 Head for 2 x 50 ml tubes, held horizontally
Contact your Labnet dealer for information about ordering additional heads