5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5508 Rev.3 P. 1 / 4
INSTRUCTION MANUAL
THERMOCOUPLE TRANSMITTER
(PC programmable) MODEL W2XT
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner (body + base socket + CJC sensor)......(1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
The unit is programmable using the PC Configurator Soft-
ware. For detailed information on the PC configuration,
refer to the W2CFG users manual. The W2CFG PC Con-
figurator Software is downloadable at M-System’s web site:
http://www.m-system.co.jp.
POINTS OF CAUTION
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
100 – 240V AC rating: 85 – 264V, 47 – 66 Hz, approx. 5 – 7VA
24V DC rating: 24V ±10%, approx. 3W
■GENERAL PRECAUTIONS
• To protect very delicate components contained inside
the unit against damage from static electricity, wear a
grounded wrist strap when handling them. If you do not
have one, touch both of your hands to a safely grounded
object or to a metal object.
• Ensure that the power supply and input signal are
switched off before you plug in or remove the unit.
• Before you remove the unit from its base socket or mount
it, turn off the power supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with ca-
bles, etc.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 20 minutes is required for
satisfying complete performance described in the data
sheet.
• With voltage output, do not leave the output terminals
shortcircuited for an extended time period. The unit is
designed to endure it without breakdown, however, it may
shorten appropriate life duration.
COMPONENT IDENTIFICATION
CJC Sensor
Fixing Screw
Connection
Diagram
Specifications
Body Base Socket
INSTALLATION
Loosen the fixing screw at the front of the unit in order to
separate the body from the base socket.
DIN Rail
35mm wide
Spring Loaded
DIN Rail Adaptor
■DIN RAIL MOUNTING
Set the base socket so that
its DIN rail adaptor is at
the bottom. Position the
upper hook at the rear side
of base socket on the DIN
rail and push in the lower.
When removing the socket,
push down the DIN rail
adaptor utilizing a minus
screwdriver and pull.
■WALL MOUNTING
Refer to “EXTERNAL DIMENSIONS.”
W2XT
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5508 Rev.3 P. 2 / 4
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
4
8 7
10
21.5 (.87)
88.5 (3.48)
84 (3.31)10 (.39)
22 (.87)
59 (2.32)
DIN RAIL
35mm wide
[4 (.16)]
11
–
M3
SCREW
72 (2.83)
29.5 (1.16)
53 (2.09)
2
–
4.2x5
(.17x.20)
MTG HOLE
6 (.24) deep
114 (4.49)
11
56
123
9
6 (.23)
• When mounting, no extra space is needed between units.
4
(.16)
■CONNECTION DIAGRAM
+
-
1
4
2
+
−
U(+)
V(–)
POWER
10
11
+
–
OUTPUT 1
7
8
+
–
3
6
OUTPUT 2
The section enclosed by broken line is only with 2nd output option.
CONFIGURATOR JACK
CJC
sensor
T/C
comp. leadwire
■TERMINAL ASSIGNMENTS
1
6
8 7
1011
3
9
4
21
5
CJC SENSOR
(model: CJM)
WIRING INSTRUCTIONS
■SCREW TERMINAL
Torque: 0.8 N·m
■SOLDERLESS TERMINAL
Refer to the drawing below for recommended ring tongue
terminal size. Spade tongue type is also applicable.
Applicable wire size: 0.25 to 1.65 mm2(AWG 22 to 16)
Recommended manufacturer: Japan Solderless Terminal
MFG.Co.Ltd, Nichifu Co.,ltd
6 (.24) max
3.3 (.13) max
mm (inch)
W2XT
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5508 Rev.3 P. 3 / 4
EXTERNAL VIEWS
LED1 (LD1)
LED2 (LD2)
LED3 (LD3)
Configurator Jack
■ FRONT VIEW ■ SIDE VIEW
SW1SW2
Output 2
Configuration DIP SW
Output 1
Configuration DIP SW
OFF
ON
1234
OFF
ON
1234
*Provided only for two-output type.
*
OUTPUT RANGING
The internal DIP switch setting is required to select output
types before setting a precise output range using PC Con-
figurator Software (model: W2CFG).
For detailed information on the PC configuration, refer to
the W2CFG users manual.
Table 1. DIP switch setting: Output 1 type
OUTPUT 1 TYPE
SW1-1 SW1-2 SW1-3 SW1-4
0 – 20 mA DC OFF OFF ON OFF
-10 – +10 V DC OFF ON OFF ON
-5 – +5 V DC ON OFF OFF ON
Table 2. DIP switch setting: Output 2 type
OUTPUT 2 TYPE
SW2-1 SW2-2 SW2-3 SW2-4
0 – 20 mA DC OFF OFF ON OFF
-10 – +10 V DC OFF ON OFF ON
-5 – +5 V DC ON OFF OFF ON
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Check DIP switch setting.
3) Power input voltage: Check voltage across the terminal
10 – 11 with a multimeter.
4) Input: Check that the input signal is within 0 – 100% of
the full-scale.
If the thermocouple or its extension wires are broken, the
output goes over 100% (below 0% with downscale) due to
burnout function. Confirm the status indicator LED pat-
tern and check leadwires in such a case.
5) Output: Check that the load resistance meets the de-
scribed specifications.
6) LED: Confirm that the LEDs indicate the normal operat-
ing mode.
W2XT
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5508 Rev.3 P. 4 / 4
STATUS INDICATOR LED
Combinations of the three front LEDs (LD1, LD2, LD3) indicate the transmitter’s operating status by different blinking pat-
terns.
Examples are shown below.
LD1
LD2
LD3
A
LD1
LD2
LD3
G
LD1
LD2
LD3
R
NORMAL OPERATION (RUN) MODE
ERROR MODE
Parameter Error
Error occurred when reading parameters from the
internal memory.
LD1
LD2
LD3
A
A
GOverrange Input
The input is below -15% or above 115%.
LD1
LD2
LD3
R
G
GG A
Burnout
PC Configuration RUN Mode
The transmitter is configured via PC and
is in normal operating conditions.
AD Conversion Error
Internal AD conversion error.
*1. LD1 may be one of the
LD1
LD2
LD3
A
GOutput Limit
The output is at the upper or the lower limit.
Amber LED
Green LED
Red LED
OFF
ON
Blink
A
G
R
*1
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit using the PC Configurator
Software (model: W2CFG).
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.

Other M-system Transmitter manuals

M-system Mini-M M2YV User manual

M-system

M-system Mini-M M2YV User manual

M-system VOS2T-R User manual

M-system

M-system VOS2T-R User manual

M-system W2XM User manual

M-system

M-system W2XM User manual

M-system M2LR User manual

M-system

M-system M2LR User manual

M-system M5PA-CA-R User manual

M-system

M-system M5PA-CA-R User manual

M-system M5XPA User manual

M-system

M-system M5XPA User manual

M-system M5VF User manual

M-system

M-system M5VF User manual

M-system Mini-M M2MS User manual

M-system

M-system Mini-M M2MS User manual

M-system Mini-M M2XRP2 User manual

M-system

M-system Mini-M M2XRP2 User manual

M-system KCNA User manual

M-system

M-system KCNA User manual

M-system Mini-M M2LCS User manual

M-system

M-system Mini-M M2LCS User manual

M-system M6SPA User manual

M-system

M-system M6SPA User manual

M-system LC User manual

M-system

M-system LC User manual

M-system Mini-M M2XM2 User manual

M-system

M-system Mini-M M2XM2 User manual

M-system 10TG User manual

M-system

M-system 10TG User manual

M-system MXV User manual

M-system

M-system MXV User manual

M-system Mini-MW Series User manual

M-system

M-system Mini-MW Series User manual

M-system M3LR User manual

M-system

M-system M3LR User manual

M-system 10TS User manual

M-system

M-system 10TS User manual

M-system CTH User manual

M-system

M-system CTH User manual

M-system 27RS User manual

M-system

M-system 27RS User manual

M-system M80YS User manual

M-system

M-system M80YS User manual

M-system M3LM User manual

M-system

M-system M3LM User manual

M-system 10VK User manual

M-system

M-system 10VK User manual

Popular Transmitter manuals by other brands

Honeywell 5802mn - Ademco Wireless Emergency... Setup guide

Honeywell

Honeywell 5802mn - Ademco Wireless Emergency... Setup guide

Care Call CCF3A-2219-EU quick start guide

Care Call

Care Call CCF3A-2219-EU quick start guide

Evolved Sonic ES2 STREAM 2 System user manual

Evolved Sonic

Evolved Sonic ES2 STREAM 2 System user manual

RKI 65-2335RK Operator's manual

RKI

RKI 65-2335RK Operator's manual

enika P8 T 4a Cobra quick start guide

enika

enika P8 T 4a Cobra quick start guide

Vega VEGABAR 87 Quick setup guide

Vega

Vega VEGABAR 87 Quick setup guide

Kramer TWA user manual

Kramer

Kramer TWA user manual

Zoom 4350 quick start

Zoom

Zoom 4350 quick start

Dwyer Instruments 605 Installation and operating instructions

Dwyer Instruments

Dwyer Instruments 605 Installation and operating instructions

Novus TXMINI-M12 operating manual

Novus

Novus TXMINI-M12 operating manual

Marmitek BOOMBOOM 55 user manual

Marmitek

Marmitek BOOMBOOM 55 user manual

KBC eCopper EERF1-LS1-T-IN-B quick start guide

KBC

KBC eCopper EERF1-LS1-T-IN-B quick start guide

VESALA Sonde PL18-05 user manual

VESALA

VESALA Sonde PL18-05 user manual

GEIGER GFB01 operating instructions

GEIGER

GEIGER GFB01 operating instructions

Neon DS-TT-02 Series product manual

Neon

Neon DS-TT-02 Series product manual

Danfoss MBS 1700 installation guide

Danfoss

Danfoss MBS 1700 installation guide

Lamtec LT3-F quick reference

Lamtec

Lamtec LT3-F quick reference

Swann AMU Powercon Operator's manual

Swann

Swann AMU Powercon Operator's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.