manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. MadgeTech
  6. •
  7. Data Logger
  8. •
  9. MadgeTech HITEMP140-1 Specification sheet

MadgeTech HITEMP140-1 Specification sheet

HiTemp140 Series
High Temperature Data Loggers
HiTemp140-1 HiTemp140-2 HiTemp140-5.25 HiTemp140-7
IFC406 (Sold Separately)
IFC400 (Sold Separately)
PRODUCT
USER GUIDE
To view the full MadgeTech product line,
visit our website at madgetech.com.
Product User Guide | 2
HiTemp140 Series
PRODUCT USER GUIDE
Product Overview
The HiTemp140 is a rugged, high precision, temperature
data logger that is built for use with autoclaves and harsh
environments. This stainless steel device is submersible,
can withstand temperatures up to +140 °C (+284 °F) and
has an accuracy of ±0.1 °C (±0.18 °F).
The HiTemp140 can store up to 32,256 readings, and features
a rigid external probe capable of measuring extended
temperatures, up to +260 °C (+500 °F). Custom probe lengths
up to 7 inches are available. The device records date and time
stamped readings, and has non-volatile solid state memory
that will retain data even if the battery becomes discharged.
The HiTemp140 can be used in both wet and dry applications
up to +140 °C (+284 °F) indefinitely.
Water Resistance
The HiTemp140 is fully submersible and is rated IP68. It
can be placed in environments with up to 230 feet (70 m)
of water.
Installation Guide
Installing the Software
The Software can be downloaded from the MadgeTech
website at madgetech.com. Follow the instructions
provided in the Installation Wizard.
Installing the Docking Station
IFC400 or IFC406 (sold separately) — Follow the
instructions provided in the Installation Wizard to install
the USB Interface Drivers. Drivers can also be downloaded
from the MadgeTech website at madgetech.com.
Device Operation
Connecting and Starting the Data Logger
1. Once the software is installed and running, plug the
interface cable into the docking station.
2. Connect the USB end of the interface cable into an
open USB port on the computer.
3. Place the data logger into the docking station.
4. The data logger will automatically appear under
Connected Devices within the software.
5. For most applications, select Custom Start from
the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the
data logging application and click Start. (Quick Start
applies the most recent custom start options, Batch
Start is used for managing multiple loggers at once,
Real Time Start stores the dataset as it records while
connected to the logger.)
6. The status of the device will change to Running or
Waiting to Start, depending upon your start method.
7. Disconnect the data logger from the interface cable
and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is
reached or the device is stopped, unless user selectable memory wrap
is enabled. At this point the device cannot be restarted until it has been
re-armed by the computer.
Downloading Data from a Data Logger
1. Place the logger into the docking station..
2. Highlight the data logger in the Connected Devices
list. Click Stop on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click Download.
4. Downloading will offload and save all the recorded
data to the PC.
Product User Guide | 3
HiTemp140 Series
PRODUCT USER GUIDE
Device Operation (cont’d)
Set Password
To password protect the device so that others cannot
start, stop or reset the device:
1. In the Connected Devices panel, click the device desired.
2. On the Device Tab, in the Information Group, click
Properties. Or, right-click the device and select
Properties in the context menu.
3. On the General Tab, click Set Password.
4. Enter and confirm the password in the box that
appears, then select OK.
Device Maintenance
Battery Replacement
Materials: ER14250-SM
1. Unscrew the bottom of the logger
and remove the battery.
2. Place the new battery into the logger.
Note the polarity of the battery. It is
important to insert the battery with
positive polarity pointing upward
towards the probe. Failure to do so
could result in product inoperability
or potential explosion if exposed to high temperatures.
3. Screw the cover back onto the logger.
Recalibration
MadgeTech recommends annual recalibration. To send
devices back for calibration, visit madgetech.com.
Note: This product is rated for use up to 140 °C
(284 °F). Please heed the battery warning. The
product will explode if exposed to temperatures
above 140 °C (284 °F).
HiTemp140 Series
NEED HELP?
DOC-1211036-00 | REV 4 2022.07.05
6 Warner Road, Warner, NH 03278
(603) 456-2011
info@madgetech.com
madgetech.com
Product Support & Troubleshooting
• Visit our Resource Library online at madgetech.com/resources.
• Contact our friendly Customer Support Team at (603) 456-2011 or support@madgetech.com.
MadgeTech 4 Software Support
• Refer to the built-in help section of the MadgeTech 4 Software.
• Download the MadgeTech 4 Software Manual at madgetech.com.
• Contact our friendly Customer Support Team at (603) 456-2011 or support@madgetech.com.
Ordering Information
HITEMP140-1 PN 900140-00 High Temperature Data Logger with a 1 inch probe
HITEMP140-2 PN 900153-00 High Temperature Data Logger with a 2 inch probe
HITEMP140-2-TD PN 900165-00 High Temperature Data Logger with a 2 inch transitional diameter probe
HITEMP140-5.25 PN 900173-00 High Temperature Data Logger with a 5.25 inch probe
HITEMP140-5.25-TD PN 900180-00 High Temperature Data Logger with a 5.25 inch transitional diameter probe
HITEMP140-7 PN 900198-00 High Temperature Data Logger with a 7 inch probe
IFC400 PN 900319-00 Docking station with USB cable
IFC406 PN 900325-00 6 Port, Multiplexer docking station with USB cable
ER14250-SM
Formerly ER14250MR-145
PN 900097-00 Replacement battery for the HiTemp140
For Quantity Discounts call (603) 456-2011 or email [email protected]om

This manual suits for next models

9

Other MadgeTech Data Logger manuals

MadgeTech RTDTemp2000 User manual

MadgeTech

MadgeTech RTDTemp2000 User manual

MadgeTech IC-RFCO2RH Specification sheet

MadgeTech

MadgeTech IC-RFCO2RH Specification sheet

MadgeTech PRHTemp2000 User manual

MadgeTech

MadgeTech PRHTemp2000 User manual

MadgeTech QuadState User manual

MadgeTech

MadgeTech QuadState User manual

MadgeTech Cryo-Temp Specification sheet

MadgeTech

MadgeTech Cryo-Temp Specification sheet

MadgeTech Event101A Specification sheet

MadgeTech

MadgeTech Event101A Specification sheet

MadgeTech RFCO2RHTemp2000A Specification sheet

MadgeTech

MadgeTech RFCO2RHTemp2000A Specification sheet

MadgeTech HiTemp140-5.25-TD Specification sheet

MadgeTech

MadgeTech HiTemp140-5.25-TD Specification sheet

MadgeTech CurrentX Series Specification sheet

MadgeTech

MadgeTech CurrentX Series Specification sheet

MadgeTech Bridge101A-30 User manual

MadgeTech

MadgeTech Bridge101A-30 User manual

MadgeTech RHTemp1000IS-KR Specification sheet

MadgeTech

MadgeTech RHTemp1000IS-KR Specification sheet

MadgeTech CryoTemp Specification sheet

MadgeTech

MadgeTech CryoTemp Specification sheet

MadgeTech PR140 Specification sheet

MadgeTech

MadgeTech PR140 Specification sheet

MadgeTech X Series Specification sheet

MadgeTech

MadgeTech X Series Specification sheet

MadgeTech Therm-A-lert Specification sheet

MadgeTech

MadgeTech Therm-A-lert Specification sheet

MadgeTech HiTemp140-M12 Specification sheet

MadgeTech

MadgeTech HiTemp140-M12 Specification sheet

MadgeTech pHTemp2000 Specification sheet

MadgeTech

MadgeTech pHTemp2000 Specification sheet

MadgeTech RTDTemp2000 User manual

MadgeTech

MadgeTech RTDTemp2000 User manual

MadgeTech DHS User manual

MadgeTech

MadgeTech DHS User manual

MadgeTech Volt101A-30V Specification sheet

MadgeTech

MadgeTech Volt101A-30V Specification sheet

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech HiTemp140-PT Specification sheet

MadgeTech

MadgeTech HiTemp140-PT Specification sheet

MadgeTech Pulse101A User manual

MadgeTech

MadgeTech Pulse101A User manual

MadgeTech TEMP1000EX User manual

MadgeTech

MadgeTech TEMP1000EX User manual

Popular Data Logger manuals by other brands

TempSen Tempod 30 user manual

TempSen

TempSen Tempod 30 user manual

Vantage Hearth CL1 quick start

Vantage Hearth

Vantage Hearth CL1 quick start

Hoymiles DTU-Lite-S user manual

Hoymiles

Hoymiles DTU-Lite-S user manual

RS PRO RS-1700 instruction manual

RS PRO

RS PRO RS-1700 instruction manual

T&D RTR500BW Read first

T&D

T&D RTR500BW Read first

Sonifex Net-Log User handbook

Sonifex

Sonifex Net-Log User handbook

Acksen AL-2VA User instructions

Acksen

Acksen AL-2VA User instructions

Data Harvest EASYSENSE V-Hub Reference manual

Data Harvest

Data Harvest EASYSENSE V-Hub Reference manual

ADLINK Technology USBDAQ-9100MS user guide

ADLINK Technology

ADLINK Technology USBDAQ-9100MS user guide

SwitchDin Droplet Quick installation guide

SwitchDin

SwitchDin Droplet Quick installation guide

T&D TR-71W quick start guide

T&D

T&D TR-71W quick start guide

Hobo U12 quick start guide

Hobo

Hobo U12 quick start guide

berlinger Fridge-tag 2 L Setup

berlinger

berlinger Fridge-tag 2 L Setup

National Instruments NI 6110 CALIBRATION PROCEDURE

National Instruments

National Instruments NI 6110 CALIBRATION PROCEDURE

DeltaTRAK FlashLink 40516 quick start guide

DeltaTRAK

DeltaTRAK FlashLink 40516 quick start guide

T&D RTR-500DC Setup guide

T&D

T&D RTR-500DC Setup guide

Leadtek LR9553D instruction manual

Leadtek

Leadtek LR9553D instruction manual

Escort iMiniPlus PDF user guide

Escort

Escort iMiniPlus PDF user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.