KPS DCM3000 User manual

DCM3000
SKU: KPSDCM3000CBINT
TRMS clamp meter 1000A AC
Instructions manual EN

- 13 -
DCM3000 EN
TABLE OF CONTENTS
1. Safety information.......................................................................................14
2. List of contents............................................................................................15
3. Description...................................................................................................16
4. Operating Guidance ....................................................................................17
5. Maintenance.................................................................................................21
6. Specications ..............................................................................................22
We MGL INTERNATIONAL wants to thank
you for choosing our product. Please
thoroughly go through this quick guide
before operation and keep it well for
future reference.
The user manual can be found:
• On our website:
www.kps-intl.com
• Through the QR code below

- 14 -
DCM3000 EN
1. Safety information
1.1. Preliminary
WARNING
Be extremely careful when using this meter. Improper use of this device can result in electric
shock or destruction of the meter. Take all normal safety precautions and follow the safeguards
suggested in this manual. To exploit full functionality of the meter and ensure safe operation,
please read carefully and follow the indications in this manual. If the equipment is used in a
manner not specied by the manufacturer, the protection provided by the equipment may be
impaired.
This meter is designed and manufactured according to safety requirements of EN 61010-1, EN
61010-2-032, EN 61010-2-033 concerning electronic measuring instruments with a measurement
CAT III 1000V, CAT IV 600V and pollution degree 2 and safety requirements for hand-held clamps
for electrical measurement and test.
•When using the meter, the user should comply with standards safety rules:
-General shock protection
-Prevent misusing the meter
•Please check for damage during the transportation after receiving the meter.
•If the meter is stored and shipped under hard conditions, please conrm if the meter is damaged.
•Probe shuld be in good conditions. Before use, please check whether the probe insulation is
damaged and if the metal wire is bare.
•Use the probe table provided with the meter to ensure safety, if necessary, replace the probe with
another identical probe or one with the same level of performance.
1.2. Usage
•When using, select the right function and measuring range.
•Don´t measure by exceeding indication value stated in each measuring range.
•When measuring a circuit with the meter connected, do not contact with probe tip (metal part).
•When measuring, if the voltage to be measured is more than 60VDC or 30VAC (TRMS), always
keep your ngers behind nger protection device
•Do not measure voltage greater than AC 750.
•In the manual measuring range mode, when measuring an unknown value, select the highest
measuring range rst.
•Before rotating conversion switch to change measuring function, remove probe from the circuit to
be measured.
•Don´t mesure resistor, diode and circuit connected to power.
•During the test of currents, resistors, diodes and circuit continuity, be careful to avoid connecting
the meter to a voltage source.
•Do not use the meter in explosive gas, vapor or dusty enviroments.
•If you nd any abnormal phenomena or failure on the meter, stop using the meter.
•Unless the meter bottom case and the battery cover are completely fastened, do not use the meter.
•Don’t store or use the meter in the conditions of direct sunlight, high temperature and high humidity.

- 15 -
DCM3000 EN
2. List of components
Digital clamp meter PouchSilicone test leads 1,5m
3xAAA 1.5V batteries Instructions manual Calibration guarantee

- 16 -
15
DCM3000 EN
3. Description
3.1. Part name
(1) Current clamp head: used for current measurement.
(2) Clamp head light
(3) Panel
(4) Trigger
(5) Function choice button (FUNC)
(6) Relative measurement button
(7) Range selection button
(8) LCD display
(9) Common end jack
(10) Resistance, voltage, diode and continuity input
jack
(11) Reading hold/Back light button (B.L/ HOLD)
(12) Maximum/minimum choice button (MAX/MIN)
(13) Transfer switch
(14) NCV indicator
(15) Protective Barrier (to warn the operator of the limit
of safe access)
1
2
3
4
5
6
8
910
11
12
13
14
7

- 17 -
DCM3000 EN
3.2 Switch, buttons and input Jack description
Used for Reading hold or back light control
Used for switching manual measuring range state.
Used for entering relative Measurement state
Used for maximun/mínimum Measurement function switch
Used for measuring function switch
Transfer switch: used for selecting function and measuring range
OFF position used for shutting off the power.
NCV position used for Non-Contact Voltage detection
COM Jack: V, R, diode,circuit connection common wire connecting terminal
INPUT Jack: V, R, diode,circuit connection input wire connecting terminal
4. OPERATING GUIDANCE
4.1 Reading Hold
In the process of measurement, if reading hold is required, press “HOLD/B.L” key, the value on the
display will be locked. Press “HOLD/B.L” key again to cancel reading hold state.
4.2 Manual Measuring Range
RAN key is automatic/manual measuring range key to trigger mode. The preset one is automatic
measuring range. Press to switch to manual measuring range. In the manual measuring range
mode, click once to change to upper range. Continue to the top range, then continue to press this
key to change to the bottom range. If this key is pressed more than 2 sec, it will switch back to
automatic measuring range state.
4.3 Max/Min Measurement Choice
1) Press “MAX/MIN” key to enter MAX mode, and always keep measurement maximum value;
press “MAX/MIN” key again, the meter will enter minimum value measurement state; press
“MAX/MIN” key for the third time, the meter will display the difference of máximum and minimum
value; press “MAX/MIN” key to repeat the above operations by recycling.
2) After entering MAX or MIN mode, it will automatically save the measured maximum or minimum
value.
3) If the user presses “MAX/MIN” key more than 2 sec, the meter will restore normal measuring
range.
Note:

- 18 -
DCM3000 EN
1) For using the maximum/minimum value measurement mode, the manual range mode must be
active.
4.4 Function Switch
1) In the voltage mode, press “FUNC” button for changing to DC or AC current type.
4.5 REL Measurement
1) REL button is relative value Measurement button. The current display value can be stored in the
memory as reference value. When the user measures later, the display value is the difference
for input value minus reference value. ie. RELΔ(current reading)= Input value - Reference value.
4.6 Back Light And Clamp Head Light
1) In the process of measurement, if ambient light is too dark to read, press “B.L/HOLD” key for two
seconds to turn on the display backlight and the clamp head light. Both lights will automatically
turn off after about 30 seconds.
2) During this period, pressing “B.L/ HOLD” key for two seconds will turn off the display backlight
and the clamp head light.
Note:
When battery voltage ≤ 3.6V, the LCD displays “ “ (undervoltage) symbol.When the user uses the
backlight, the battery voltage drops below 3.6V, due to high working current. The “ “
symbol may
appear, and Measurement accuracy is not guaranteed. Continue to use the meter normally without
using backlight. Do not replace the battery until “ “ symbol shows under normal conditions.
4.7 Automatic Power-Off
1) If there is no operation during any 30 minutes after turning the machine on, the meter will enter
suspended stated, automatically powering off to save the battery.
2) After automatic power-off, press any key, the meter will turn on again.
3) If the user holds “FUNC” key when powering on, it will cancel automatic power-off function.
4.8 Measurement preparation
1) Turn the transfer switch to turn on the power. When battery voltage is low (about ≤ 3.6V, LCD
displays “ “ symbol, replace the battery.
2) “” symbol means that input voltage or current should not be more than the specied value,
which is to protect the internal line from damage.
3) Place transfer switch to required measuring function and range.
4) When connecting line, rst connect the common test line, then connect charged test line. When
removing line, remove charged test line rst.

- 19 -
DCM3000 EN
4.9 Current measurement
WARNING
Electric shock Hazard.
Remove the probe from the meter before mesuring with current clamp
1) Measuring switch is placed to position A. At this time, the meter is in AC current measurement
state. Choose appropriate measuring range.
2) Hold the trigger, open clamp head, clip one lead of measurement circuit to be tested in the clamp.
3) Read the current value on the LCD display.
Note:
1) Clamping two or more lines of circuit to be tested simultaneously will not get the correct easuring
results
2) To get accurate Reading, connect the wire to be tested at the center of current clamp
3) “” symbol indicates that máximum input AC current is 1000A
4.10 Voltage measurement
WARNING
Electric shock hazard.
Pay special attention to avoid shock when measuring high voltage.
Don’t input voltage more than 750Vac T-RMS.
1) Insert black probe to COM jack, insert red probe to INPUT jack, choose appropiated measuring
range.
2) Place transfer switch to AC voltage or position. At this time, the meter is in the DC
voltage measurement state. To measure AC voltage, press FUNC button to enter AC voltage
measurement state
3) Connect the probe with voltage source or both ends of load in parallel for measurement.
4) Read the voltage on the LCD.
Note:
1) In the small voltage measuring range, if the probe is not connected with the circuit to be tested, the
meter may have uctuating readings, which is normal and caused by the meter´s high sensitivity.
When the meter is connected with the circuit under test, you will get actual measured value
2) On the relative measurement mode, automatic measuring range is invalid. Manual measuring
range must be selected.
3) “” symbol indicates that máximum input voltage is 1000VDC / 750VAC. Maximum input
voltage at mV is 600mVac/dc
4) If the readings measured by the meter is more than 600VAC(RMS), it will be send out “beep”
alarm.

- 20 -
DCM3000 EN
4.11 Resistance Test
WARNING
Electric shock hazard.
When measuring circuit impedance, determine that the power supply is disconnected and the
capacitor in the circuit is completely discharged.
1) Insert black probe to COM jack, insert red probe to INPUT jack.
2) Place measuring range switch to position “Ω”. At this time, the meter is in the measurement state.
3) Connect the probe to the both ends of resistor or circuit to be tested for measurement.
4) LCD will show readings.
Note:
1) When the input end is open, LCD shows “OL” outrange state. .
2) When the resistance to be tested > 1M, the meter reading will stabilize after a few seconds, which
is normal for high resistance readings
4.12 Diode Test
1) Insert black probe to COM Jack, insert red probe to INPUT Jack
2) Measuring switch is placed to position
3) Connect the red probe to diode anode and connect the black probe to diode cathode to make
test.
4) Read on the LCD
Note:
1) What the meter shows is approximation of diode forward voltage drop.
2) If the probe has reverse connection or the probe is open, the LCD will show “OL”.
4.13 Circuit Continuty Test
WARNING
Electric shock hazard.
When measuring circuit continuity, determine that the power supply is disconnected and the
capacitor in the circuit is completely discharges.
1) Insert black probe to COM Jack, insert the red probe to INPUT Jack
2) Measuring switch is placed to position
3) Connect the probe to the both ends of circuit to be tested for the Measurement
4) If the resistance of circuit under test is less than 50Ω, the meter´s built-in buzzer may sound.
5) Read the circuit resistance value on the LCD

- 21 -
DCM3000 EN
4.14 NCV Measurement
1) Turn the meter to NCV mode
2) Place the meter top close to the conductor. When test voltage is greater than 110Vac (T-RMS),
the meter´s LED indicator will blink and the buzzer will give an alternating high-low alarm sound.
Note:
1) Even there is no indication, voltage may exist still. Don’t use non-contact voltage detector to
judge whether there is voltage in the wire. Detection operation could be affected by socket
design, insulation thickness, type and other factors.
2) When inputting voltage on the meter input terminal, due to the existence of the induced voltage,
voltage induction indicator also may light.
3) External sources of interference (such as ashlight, motor, etc.) may incorrectly trigger non-
contact voltage detection.
5. MAINTENANCE
5.1 Replace Battery
WARNING
To avoid electric shock, make sure that the test leads have been clearly move away from the
circuit under measurement before opening the battery cover.
1) If the symbol “ “ appears, it means that the batteriers should be replaced.
2) Loosen the screw of the battery cover and remove it.
3) Replace the used battery with a new one.
4) Return the battery cover and tighten the screw.
Note: The battery polarity can´t be reversed
5.2 Replace probe
Replace test lead if leads become damaged or worn.
WARNING
Use meet EN 61010-031 standard, rated CAT IV600V, or better test leads
WARNING
To avoid electric shock, make sure the probes are disconnected from the measured circuit before
removing the rear cover. Make sure the rear cover is tightly screwed before using the instrument.

- 22 -
DCM3000 EN
6. SPECIFICATIONS
Measuring range Resolution Accuracy
AC CURRENT
60A 0.01A
±(2.0% reading + 8 digits)600A 0.1A
1000A 1A
DC VOLTAGE
60mV 0.01mV
±(0.5% reading + 5 digits)
600mV 0.1mV
6V 0.001V
60V 0.01V
600V 0.1V
1000V 1V ±(0.5% reading + 5 digits)
AC VOLTAGE
60mV 0.01mV
±(0.6% reading + 5 digits)
600mV 0.1mV
6V 0.001V
60V 0.01V
600V 0.1V
750V 1V ±(0.6% reading + 5 digits)
CONTINUITY 0.1Ω
If the resistance of the circuit
under test is less than 50 Ω,
the meter´s built-in buzzer may
sound
RESISTANCE
600Ω 0.1Ω
±(1.0% reading + 5 digits)
6kΩ 0.001kΩ
60kΩ 0.01kΩ
600kΩ 0.1kΩ
6MΩ 0.001MΩ
20MΩ 0.01MΩ ±(1.2% reading + 3 digits)
DIODE 0.001V Display approximate diode
forward voltage value
Note: For more detailed specications, please check the user manual (see the instructions to nd it on page 13)
Other manuals for DCM3000
2
This manual suits for next models
1
Table of contents
Other KPS Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Hanna Instruments
Hanna Instruments HI 93702 instruction manual

NIVELCO
NIVELCO NIPRESS D-400 Series user manual

AmpliVox
AmpliVox Otowave 202-H operating manual

Orange Electronic
Orange Electronic P451 manual

Dwyer Instruments
Dwyer Instruments MW-1 quick start guide

Akron
Akron MERCURYMASTER 3446 Operating & maintenance instructions

PQ Plus
PQ Plus MMU 3 Quick start manual

WILDLIFE
WILDLIFE AquaPulse Impulse Counter quick start guide

Tektronix
Tektronix P6209 instruction manual

Gasmet
Gasmet Alnab CMM Installation instructions and operating manual

Yamato
Yamato TA300 instruction manual

Michell Instruments
Michell Instruments QMA401 user manual