manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Nikon
  6. •
  7. Measuring Instrument
  8. •
  9. Nikon NR-5500 Operating instructions

Nikon NR-5500 Operating instructions

Nikon
Repair
Manual
NIKON
CORPORATION
,
JAPAN
TOKYO
ptical
System
n-n
C
3
]
NR-5
50
0
Repair
tool
list
J
-
3
5
0
5
6
Centering
too
J
-
3
8
0
7
5
Centering
tool
attachment
J-38054A
Optical
adjustment
TV
tool
J-38054B
TV
camera
(NTSC
type
&
C
mount)
Sketch
Remarks
Use
with
J-38075
Centering
tool
attachment
Use
with
J-35056
Centering
tool
Useful
for
NR-5100
&
NRK-8000
For
C
mount
TV
camera
Use
with
J-38054A
Optical
adjustment
TV
tool
With
DC12V
power
supply
cord
k
BNC
cable
J
-
3
8
0
5
7
LED
iIluminater
Use
with
J-38075
Centering
tool
attachment
A
pair
of
Infrared
ray
LED
for
TV
tool
Visibler
ray
LED
for
eyepiece
(3
-
l]
Repair
ManuaI
Sketch
Remarks
J
-
3
8
0
5
5
Supplement
lens
set
For
Slit
diaphragm
k
Chopper
slit
Use
with
J-38054A
Optical
adjustment
TV
tool
or
J-35050
Eyepiece
with
reticle
J
-
3
5
0
5
8
Model
eye
attachint
tool
J-35058A
Model
eye
adapter
J-
3
5
0
5
0
Eyepiece
with
reticle
(15x)
J
-
3
5
0
5
7
Eyepiece
retaining
ring
Use
with
J-35056
Centering
tool
Use
with
J-35058
Model
eye
attaching
tool
Use
with
J-35057
Eyepiece
retaining
ring
and
J-35056
Centering
tool
Use
with
J-35050
Eyepiece
with
reticle
and
J-35056
Centering
tool
C3
-
21
Repair
ManuaI
Tool
number
Name
J
-
3
8
0
5
3
Centering
telescope
Sketch
J
-
3
9
0
1
1
Inspecting
telescope
holder
J
-
3
9
0
1
4
Inspecting
telescope
J
-
3
9
0
1
5
Inspecting
lens
(—
10Dpt.)
J
-
3
9
0
1
6
Inspecting
lens
(-MODpt.)
Remarks
Use
with
J-35057
Eyepiece
retaining
ring
and
J-35056
Centering
tool
Use
with
J-35056
Centering
tool
A:for
J-39014
Inspecting
tele.
B:for
J-35058A
Model
eye
ada.
C:For
J-39011A
Holder
Use
with
J-39011A
Inspecting
telescope
holder
A
and
J-35056
Centering
tool
Use
with
J-3
9
11A
Inspecting
telescope
holder
A
Use
with
J-39011A
Inspecting
telescope
holder
A
[3-31
Repair
Manua
I
Tool
number
Name
J
-
3
5
0
6
6
Blackout
curtain
Sketch
Remarks
J
-
3
9
0
2
9
Model
eye
for
Main
LED
power
check
J
-
3
9
0
3
3
Model
eye
set
J-38052A
High
power
model
eye
±22.
OODpt.
J
-
3
8
0
5
6
Cly.
model
eye
Use
with
J-35058A
Model
eye
adapter
and
J-35058
Model
eye
attaching
tool
Also
use
with
J-35058A
Model
eye
adapter
and
J-35058
Model
eye
attaching
tool
Use
with
J-35058A
Model
eye
adapter
and
J-35058
Model
eye
attaching
tool
[3-43
Repa
i
r
ManuaI
TEST
PROGRAM
[6]
TEST
PROGRAM
1
.
Entering
the
Test
Program
Mode
(T)
After
turning
on
the
Main
switch,
press
the
[PRINT]
key
while
the
sound
“do,
mi,
so,
do”
is
being
heard.
(2)
After
you
hear
the
sound
“do,
mi,
so,
do,
do,
so,
mi,
do”
followed
by
three
beeps,
press
the
measurement
button
three
times
in
succession.
(3)
The
NR
-
5500
will
enter
the
test
program
mode
and
display
the
test
program
select
menu
screen.
(The
NR
-
5500
will
display
the
ordinary
measurement
mode
screen
if
you
do
not
press
the
key
at
the
proper
time.)
RTST
0
1
0
5
0
A
0
F
EXIT
0
2
0
6
0
B
1
0
0
3
0
7
0
C
1
1
VER
1.0
0
0
4
0
8
0D
1
2
0
5
0
9
0
E
1
3
ROM
version
number
2
.
Exiting
the
test
mode
Press
the
[EXIT]
key
in
the
refractometer
test
mode
select
screen,
and
the
NR
-
5500
will
exit
the
test
mode
and
enter
the
measurement
mode
with
the
ordinary
measurement
mode
screen.
C
6
-
13
Repair
Manual
3
.
Refractometer
Test
Items
TEST
PROGRAM
Check
chopper
frequency
and
R/L.
FOG
and
AX
sensors.
Turn
on
M.
LED.
Rotate
IRM.
Not
used.
Adjust
alignment
mark.
Turn
on
C.
LED.
Turn
on
M.
LED.
Not
used.
Measure
count
offset
value.
Measure
CHA-FIL.
CHB-FIL
peak
values.
Not
used.
Check
monitor
background
color.
Test
printer
operation.
Adjust
automatic
fogging
target
position.
Adjust
key
display
position.
Making
of
model
eye
count
table.
Adjust
alignment
offset.
Check
a
measurement
system.
Initialize
memory.
Check
touch
panel
switches.
Adjust
cylindrical
axis
offset.
Check
1C
card
reader/writer.
Selection
model
eye
mode
and
human
eye
mode.
CHOP=524
±1.0
Light
occlusion
=
LO
MLED
ON:
M.
LED
lights
up
continuously.
IRM
ROT:
Rotates
continuously.
TOT=80—85:
Dust-proof
glass
is
instal
led.
TOT=90—95:
Dust-proof
glass
is
not
installed.
CLED
ON:
C.
LED
lights
up
continuously.
MLED
ON:
M.
LED
lights
up
continuously.
PHASE
COUNT
OFFSET=-200~+200
CHA=CHB=150
—200
0
V=000
Measurement
switch
[PRINT]
key
+:
PRINT
switch
-:
Joy
stickCmeasurement
button)
Use
the
model
eye
set
(-18D-
+22D).
Set
to
T
=
R±10.
Set
to
X
=
R±
10.
Set
to
Y
=
R±10.
R=1
2
0-1
3
5
A
=
B
=
1
5
0
-
2
0
0
0
=
000
[X]
[Y]
=
60
when
checking
[X]
[Y]
=
30
when
adjusting
When
EEPROM
is
replaced.
Verify
that
the
switches
can
be
turned
on
and
off
normally.
Use
J38056
cylindrical
model
eye.
Valid
only
when
an
1C
card
reader/writer
is
built-in.
HUMN
EYE
=
Human
eye
MODL
EYE
=
Model
eye
+
1.00-1.00
Dpt.
VR3
GAIN
VR6
TOT
VR5
XAL
VR4
YAL
VR7
CAL
C6-
2]
Repair
Manual
6
.
Test
Item
Description
TEST
PROGRAM
®
TEST
-
00
:
Displaying
the
chopper
frequency
and
its
variation,
as
well
as
R/L,
FOG
and
AX
sensor
status
information
TEST-0
0
CH0P=5
2
4.
1
000.
2
R/L
=
L
O
FOG
=
L
O
AX
=LO
Press
this
key
to
quit
TEST-00.
Chopper
frequency:
CH0P=524,1
000.2
-Chopper
frequency
variation
Chopper
frequency:
OK
if
524.0
±
1.0
is
obtained.
•
R/L,
FOG
and
AX
sensors
L
O
(Low)
:
The
sensor
light
is
cut
by
the
sensor
plate.
H
I
(High)
:
The
sensor
light
is
not
cut
by
the
sensor
plate.
(2)
TEST
-
01
:
Turning
on
the
main
LED
and
rotating
the
IRM
(image
rotator
motor)
continuously
END
TEST-01
MLED
1
RM
OFF
STOP
Main
LED
-
-
IR
motor
Press
this
key
to
quit
TEST-01.
•
Turning
on
the
main
LED
(M.
LED)
Press
the
[MLED
OFF]
key
to
change
the
display
into
[MLED
ON],
and
the
M.
LED
will
turn
on
continuously.
Press
the
[MLED
ON]
key
again
to
change
the
display
into
[MLED
OFF],
and
the
M.
LED
will
go
off.
Keep
the
LED
on
for
as
short
a
time
as
possible
because
extended
LED
lighting
will
shorten
the
LED
life.
[6-31
Repair
Manual
TEST
PROGRAM
♦
Continuous
IRM
rotation
Press
the
[IRM
STOP]
key
to
change
the
display
into
[IRM
ROT],
and
the
IR
motor
will
rotate
continuously.
Press
the
[IRM
ROT]
key
again
to
change
the
display
into
[IRM
STOP],
and
the
IR
motor
will
stop.
*
*
*
When
a
motor
have
been
stopped
and
noised
because
of
the
depth
of
gearing,
set
the
power
to
OFF
once
and
start
to
rotate
again.
*
*
*
After
turning
on
the
M.
LED
and
rotating
the
IR
motor,
make
optical
adjustments
on
the
IR
(image
rotator),
and
check
the
accuracy
of
the
rotation
and
the
position
of
the
photo
detector
unit.
Connect
the
oscilloscpe
to
TP18
(XAL)
and
TP17
(YAL).
(3)
TEST
-03
:
Model
eye
alignment
and
adjusting
the
alignment
mark
position
I-Alignment
I
position
display
T
E
ST-0
3
XAL
:+
0
0
0
YAL
:+0
0
0
TOT
:
0
0
0
END
-Alignment
mark
fyH
SET
Press
this
key
to
quit
TEST-03.
•
Adjusting
the
model
eye
position
Adjust
the
position
of
the
model
eye
tool
installed
on
the
centering
tool
so
that
the
following
readings
are
obtained:
XAL
=
+000
YAL
=
+000
TOT
=
093
approx.
(OK
if
read-out
is
between
60
and
105)
Adjust
the
center
position
roughly
while
referring
to
the
position
of
the
asterisk
“
*
”
mark
in
the
alignment
position
display
located
at
the
upper
left
corner.
(Set
the
focus
point
to
the
fifth
graduation
of
the
scale
in
the
model
eye
attaching
tool.)
Finally,
install
the
dust
-
proof
glass
and
adjust
VR3
GAIN
until
TOT=080
~085
reading
is
obtained.
•
Adjusting
the
alignment
mark
position
After
adjusting
the
model
eye
position,
do
the
following
if
the
blinking
dot
of
the
M.LED
is
misaligned
slightly
from
the
alignment
mark:
Move
the
alignment
mark
vertically
with
the
[U]
and
[D]
keys
and
horizontally
with
the
[L]
and
[R]
keys
until
the
blinking
dot
of
the
M.
LED
falls
at
the
center
of
the
alignment
mark.
After
the
adjustment
is
complete,
turn
DIP
switch
No.
5
located
on
the
right
side
of
the
Main
body
to
the
ON
position
and
press
the
[SET]
key.
The
alignment
mark
position
is
then
stored
in
the
EEPROM
and
the
message
“SET
DIP
SW5
TO
OFF”
is
displayed.
Return
DIP
switch
No.
5
to
the
OFF
position.
[6-4]
Repair
Manual
_
TEST
PROGRAM
@
TEST
-
04
:
Turning
on
the
C.
LED
(Calibration
LED)
and
M.
LED
(Main
LED)
Press
this
key
to
quit
TEST-04.
•
Turning
on
the
C.
LED:
Press
the
[CLED
OFF]
key
to
change
the
display
into
[CLED
ON],
and
the
C.
LED
will
light
up.
This
mode
is
used
to
check
the
CHA
and
CHB
signals.
•
Turning
on
the
M.
LED:
Press
the
[MLED
OFF]
key
to
change
the
display
into
[MLED
ON],
and
the
M.
LED
will
light
up
cotinuously.
This
mode
is
used
to
adjust
the
position
of
the
M.
LED.
(5)
TEST
-
06
:
Displaying
the
CHA
and
CHB
phase
count
offsets
and
the
CHA
-
FIL
and
CHB
-
FIL
signal
peak
values
Press
this
key
to
quit
TEST-06.
•
Displaying
the
CHA
and
CHB
phase
count
offsets
This
test
displays
the
digital
read
-
out
of
the
phase
difference
offset
after
A/D
conversion
is
made
when
the
C.
LED
is
on.
Make
sure
that
the
following
readings
are
observed:
Offset:
-
00200
to
+00200
;
Variation:
0010
or
less
•
Displaying
the
CHA
-
FIL
and
CHB
-
FIL
signal
peak
values
This
test
displays
the
digital
read
-
out
of
the
peak
values
of
the
CHA
-
FIL
and
CHB
-
FIL
signals
after
A/D
conversion
is
made
when
the
C.
LED
is
on.
CHA:
150
to
200
;
CHB:
150
to
200
Adjust
VR7
CAL
until
the
reading
of
the
CHA
-
FIL
and
CHB
-
FIL
peak
values
falls
between
150
to
200.
(it
is
advisable
that
this
adjustment
be
carried
out
during
TEST
-
0D)
•
0V
display:
Make
sure
that
the
level
measured
is
000.
If
read
-
out
is
not
000,
the
A/D
converter
is
damaged.
C6
-
51
Repair
Manual
©
TEST
-
08
:
Checking
the
monitor
background
color
TEST
PROGRAM
END
TEST-0
8
PUSH
START
KEY
Press
this
key
to
quit
TEST-08.
•
The
background
color
of
the
monitor
should
change
on
each
press
of
the
measurement
button.
|—*
Transparent
(video)
-
►
Black
-
►
White
-
Transparent:
Raw
signal
from
the
camera.
Black
:
The
screen
background
is
set
to
black
by
the
black
level
signal.
White
:
The
screen
background
is
set
to
white
by
the
white
level
signal
(characters
vanished
in
the
background).
©TEST-09
:
Testing
the
printer
Press
this
key
to
quit
TEST-09
•
Press
the
[PRINT]
key,
and
the
NR
-
5500
will
print
8
lines
of
data.
This
test
is
used
to
check
the
printer
(e.g.,
for
missing
characters)
and
to
check
the
print
darkness.
To
adjust
the
print
darkness,
use
the
VR1
on
the
CPU
board,
and
then
turn
the
Main
switch
off
and
on
to
make
the
new
setting
valid.
(6-61
Repair
Manual
TEST
PROGRAM
(8)
TEST
-
OA
:
Adjusting
the
automatic
fogging
mechanism
Press
this
key
to
quit
TEST-OA.
♦
Positioning
the
automatic
fogging
target
Mount
J39014
subjective
inspection
telescope
and
adjust
the
amount
of
offset
to
the
O.OODpt.
position
of
the
fogging
target.
While
observing
the
fogging
target
with
the
J39014,
move
it
with
the
[PRINT]
key
and
measurement
button
(Joy
stick)
until
the
target
position
becomes
clearly
visible.
Select
the
0.12
Dpt.
step
with
the
[STEP]
key
and
adjust
the
offset.
[PRINT]
key
:
Moves
the
fogging
target
in
the
plus
diopter
direction.
Measurement
button
:
Moves
the
fogging
target
in
the
minus
diopter
direction.
(Joy
stick)
[STEP]
key
:
Adjust
the
increment
by
which
the
fogging
target
moves.
The
distance
of
the
fogging
target
travel
is
the
offset
amount
(number
of
pulses)
and
is
displayed
in
the
format
“OFFSET
-***.”
After
establishing
the
offset,
turn
DIP
switch
No.
5
located
on
the
right
side
of
the
Main
body
to
the
ON
position
and
press
the
[SET]
key.
The
offset
amount
is
then
stored
in
the
EEPROM
and
the
message
“SET
DIP
SW5
TO
OFF”
is
displayed.
Return
the
DIP
switch
No.
5
to
the
OFF
position.
•
Checking
the
fogging
target
movement
(The
[PRINT]
key
and
measurement
button
can
not
be
used)
[
-
8D]
key:
Pressed
to
move
the
fogging
target
to
the
-
8.00Dpt.
position.
[+14D]
key:
Pressed
to
move
the
fogging
target
to
the
+14.00Dpt.
position.
C6
—
73
Repair
Manual
TEST
PROGRAM
(9)
TEST
-
OB
:
Adjusting
the
key
display
positions
Press
this
key
to
quit
TEST-OB.
•
Adjust
the
discrepancy
between
the
working
area
of
the
keys
on
the
monitor
touch
panel
and
the
key
position
displayed
on
the
screen.
[U]
and
[D]
keys:
Move
the
key
display
vertically.
[L]
and
[R]
keys:
Move
the
key
display
horizontally.
After
adjusting
the
key
display
position,
turn
DIP
switch
No.
5
located
on
the
right
side
of
the
Main
body
to
the
ON
position
and
press
the
[SET]
key.
The
offset
amount
is
then
stored
in
the
EEPROM
and
the
message
“SET
DIP
SW5
TO
OFF”
is
displayed.
Return
DIP
switch
No.
5
to
the
OFF
position.
©
TEST
-
PC
:
Making
of
the
model
eye
tool
count
table
Alignment
position
display
T
E
ST-0
C
X
A
L
:
+
0
0
0
YAL
:
+
0
0
0
TOT:
000
-Alignment
mark
H-H
SET
-
2
6
D
0
0
0
0
DATA
DATA
+
1
-1
-Press
this
key
to
quit
TEST-OC.
When
it
is
necessary
to
remeasure
the
count
table
data
for
the
model
eye
after
making
optical
adjustments
on
the
auto
refractometer
unit,
perform
measurements
on
the
model
eye
tool
concerned
and
restore
the
new
count
table
to
memory
(EEPROM).
Selecting
the
model
eye
[INCR]
key
:
Changes
the
power
of
the
model
eye
tool
displayed
on
the
[SET]
key
by
2.00Dpt.
to
plus
side.
[DCR]
key
:
Changes
the
power
of
the
model
eye
tool
displayed
in
the
[SET]
key
by
2.00Dpt.
to
minus
side.
The
NR
-
5500
can
show
model
eye
tool
power
from
-
26.00Dpt.
to
+26.00Dpt.
in
2.00Dpt.
step.
r->-26.00<—>-24.00-—>-22.00-—>-20.00-—
.
0
.
0
0-^
I—>+26.00
<—>+24.00-—>+22.00-—>+20.00-—
.
+
2
.
00^
C6
-
8]
Repair
Manual
TEST
PROGRAM
•
Model
eye
alignment
Place
the
model
eye
tool
of
the
selected
power
on
the
centering
tool
and
align
it.
Focusing
:
Adjust
the
focus
by
pushing
or
pulling
the
J35058
model
eye
attaching
tool
which
installs
the
model
eye.
Alignment:
While
observing
the
alignment
mark,
make
a
rough
alignment
of
the
model
eye
position
by
turning
the
centering
knob
of
the
centering
tool
until
the
asterisk
*
mark
is
positioned
near
the
center
of
the
display.
Then
make
a
fine
alignment
while
observing
the
“XAL:
”
and
“
YAL:
”
displays
to
obtain
“XAL:
+0.00"
and
“YAL:
+0.00”
respectively.
Make
sure
that
“TOT:
09
3
or
a
similar
value
is
also
displayed.
•
Measuring
the
model
eye
count
When
you
press
the
measurement
button,
the
image
rotator
rotates
and
the
count
is
automatically
measured.
The
measured
count
is
displayed
in
the
[SET]
key.
Use
the
[DATA]
key
to
change
the
displayed
count
or
manually
set
the
count.
[DATA
+1]
:
Increments
the
counter
by
1.
[DATA
-1]
:
Decrements
the
counter
by
1.
•
Saving
the
model
eye
count
After
establishing
the
count,
turn
DIP
switch
No.
5
located
on
the
right
side
of
the
Main
body
to
the
ON
position
and
press
the
[SET]
key.
The
count
is
then
stored
in
EEPROM
and
two
short
beeps
are
heards.
*
Press
the
[INCR]
key
and
select
-18.00Dpt..
(-26.00Dpt.,-24.00Dpt.,
-22.00Dpt.,+24.00Dpt.
and
+26.00Dpt.
are
out
of
measurement
range.
So
they
are
not
necessary
to
calibrate.)
Start
counting
at
-
18.00Dpt.
After
the
storage
of
the
count
for
-
18.00Dpt.
is
complete,
press
the
[INCR]
key
and
select
-
16.00Dpt.
Then
measure
and
store
the
count
for
-
16.00Dpt.
Continue
measurement
sequentially
up
to
the
measurement
of
+22.00Dpt..
After
storing
all
counts
for
measured
model
eyes,
turn
DIP
switch
No.
5
to
the
OFF
position.
Press
the
[PRINT]
key,
and
the
count
values
stored
in
the
EEPROM
will
be
printed.
AX
OFFSET
=
67-
FOG
OFFSET
=
+0074
—
***
DP
DATA
***
DP
COUNT
D1FF.
-26D:
0000
+0050
-24D:
0050
—
+0020
-22D:
0070
+0030
-20D:
0100
TEST
-11
Cylindrical
axis
offset
TEST
-
0A
Fogging
target
offset
Count
for
-24.00Dpt.
Difference
between
counts
for
-22.00Dpt.
and
-20.00Dpt.
C6
-
9)
Repair
Manual
TEST
PROGRAM
©TEST-OD
:
Alignment
offset
adjustment
Press
this
key
to
quit
TEST-OD.
•
Adjusting
the
alignment
signal
offsets
(with
no
room
lamp)
“T”:
Offset
of
the
TOT
signal
caused
by
the
amplifier
system
Adjust
VR6
TOT
to
set
“
R
±
10”.
“X”:
Offset
of
the
XAL
signal
caused
by
the
amplifier
system
Adjust
VR5
XAL
to
set
“
R
±10
”.
“
Y”:
Offset
of
the
YAL
signal
caused
by
the
amplifier
system
Adjust
VR4
YAL
to
set
“
R
±10
”.
“R”:
Reference
voltage
level
for
the
alignment
signal
(2.5V)
Make
sure
that
the
reference
voltage
is
between
120
and
135
(not
adjustable).
•
Adjusting
the
peak
value
of
CHA
-
FIL
and
CHB
-
FIL.
“A”:
Peak
value
of
the
CHA
-
FIL
signal
when
the
C
LED
is
on.
Make
sure
that
the
peak
value
is
between
150
and
200.
“B”:
Peak
value
of
the
CHB
-
FIL
signal
when
the
C
LED
is
on.
Make
sure
that
the
peak
value
is
between
150
and
200.
*
To
adjust
“A”
and
“B”
use
VR7
CAL
to
measure
the
peak
values
for
both
channels
so
that
they
fall
within
the
permitted
value
range
(same
as
in
TEST-06).
•
0V
display
“0”:
Make
sure
that
“000”
is
displayed.
[6-10]
Repair
Manual
TEST
PROGRAM
©
TEST-OE
:
Checking
the
measuring
optical
system
adjustment
Press
this
key
to
quit
TEST-OE.
•
TEST
-
OE
can
be
used
to
test
the
accuracy
of
adjustments
made
on
the
measuring
optical
system
(Image
rotator
and
photo
sensor)
•
Model
eye
alignment
Place
the
model
eye
tool
on
the
centering
tool
and
align
it.
Focusing
:
Adjust
the
focus
by
pushing
or
pulling
the
J35058
model
eye
attaching
tool
which
installs
the
model
eye.
Alignment:
While
observing
the
alignment
mark,
make
a
rough
alignment
of
the
model
eye
position
by
turning
the
centering
knob
of
the
cenrering
tool
until
the
asterisk
“
*
”
mark
is
positioned
near
the
center
of
the
display.
Then
make
a
fine
alignment
while
observing
the
“XAL:
”
and
“YAL:”
displays
to
obtain
“XAL:
+0.00”
and
“YAL:
+0.00”
respectively.
Make
sure
that
“TOT:
093”
or
a
similar
value
is
also
displayed.
If
this
value
is
not
displayed,
adjust
the
VR3
GAIN
control.
•
Measurement
When
you
press
the
measurement
button,
the
Image
rotator
rotates
and
measurement
starts.
The
asterisk
mark
in
the
alignment
mark
moves
around
and
the
“XAL:
”
and
“YAL:
”
values
change
as
the
Image
rotator
rotates.
•
Displaying
the
measurement
results
The
NR
-
5500
displays
the
total
of
the
alignment
offsets
measured
at
72
measuring
points
at
the
5
degrees
steps.
[X][Y]:
Totals
of
alignment
offsets
in
X
and
Y
directions
At
checking
:
50
or
less
At
adjustment:
Adjust
so
that
the
total
will
be
30
or
less.
X
:
Total
of
alignment
offsets
in
X
direction
only
Y
:
Total
of
alignment
offsets
in
Y
direction
only
*
Calculation
formula
[X]
[YJ
=
-rr-r
X
(
|
X
L
J
+
|
Y
i
|
)
0
4
i»l
X
=
77
Z
Xi
(Signed)
6
4
i-i
Y
=
-TT-r
X
Y
i
(Signed)
6
4
i-i
(6-1
n
Repair
Manual
TEST
PROGRAM
©
TEST
-
OF:
Initializing
memory
(EEPROM)
Press
this
key
to
quit
TEST-OF.
♦
This
test
initializes
the
EEPROM
that
stores
the
model
eye
count
tables
and
offset
values
by
storing
their
system
standard
values.
This
test
must
always
be
conducted
whenever
the
EEPROM
is
replaced
with
a
new
one.
*
Accidental
execution
of
this
test
will
overwrite
all
existing
data
with
standard
values,
resulting
in
accuracy
deterioration
in
any
subsequent
measurements.
•
Initializing
EEPROM
Press
the
measurement
button,
and
the
confirmation
screen
will
appear.
[YES]
key
:
Press
this
key
after
setting
DIP
switch
No.
5
on
the
right
side
of
the
main
body
to
ON
to
start
initializing
the
EEPROM.
When
initialization
is
finished,
’’SET
DIP
SW5
TO
OFF”
is
displayed.
Return
DIP
SW
No.5
to
the
OFF
position.
[NO]
key
:
Press
this
key
to
cancel
the
initialization
test.
The
NR
-
5500
will
then
quit
TEST
-
OF.
[6-1
2]
Repair
Manual
TEST
PROGRAM
©
TEST
-10:
Making
functional
c
hecks
on
the
touch
pa
nel
keys
—
Press
this
key
to
quit
TEST-10.
•
This
tests
permits
the
user
to
check
the
touch
panel
keys
for
normal
operation.
♦
Checking
the
key
function
Press
the
key
to
be
checked.
The
key
is
normal
if
the
corresponding
display
switches
from
OFF
to
ON.
The
END
key
is
normal
when
the
NR
-
5500
quits
TEST
-10
normally.
*
TEST
-
0B
may
also
be
executed
in
conjunction
with
this
test
to
check
the
position
of
the
key
display
frame.
©
TEST
-11:
Adjusting
the
cylindrical
axis
offset
•
Model
eye
alignment
adjustment
Place
J38056
astigmatism
model
eye
tool
on
the
chin
rest.
Focusing
:
Focus
the
blinking
dot
of
the
M
LED
by
sliding
the
Main
body
forwards
or
backwards.
Positioning:
Move
the
Main
body
by
sliding
and
elevating
until
the
blinking
dot
of
the
M
LED
is
positioned
at
the
center
of
the
alignment
mark.
*
Perform
focusing
and
position
alignment
simultaneously.
Before
installing
the
astigmatism
model
eye
tool,
clean
it
with
cleaner
liquid.
♦
Measuring
the
cylindrical
axis
Press
the
measurement
button
to
start
measurement.
The
measured
cylindrical
axis
is
displayed
in
the
“
AXIS=”
field.
OFF
OFF
OFF
END
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
[6-133
Repair
Manual
TEST
PROGRAM
•
Adjusting
the
cylindrical
axis
Using
the
[INCR]
and
[DCR]
keys,
adjust
the
cylindrical
axis
displayed
in
the
“
AXIS=”
field
matches
the
cylindrical
axis
value
displayed
on
the
mounting
holder
of
J38056
astigmatism
model
eye
tool.
[INCR]
key:
Increases
the
cylindrical
axis
by
one.
The
value
displayed
in
the
“OFFSET^”
field
is
also
increased.
[DCR]
key
:
Decreases
the
cylindrical
axis
by
one.
The
value
displayed
in
the
“OFFSET=”
field
is
also
decreased.
*
*
*
Set
a
cylindrical
axis
sign
to
+C.
•
Saving
the
offset
value
When
the
offset
value
displayed
in
the
“OFFSET=”
field
simultaneously
during
the
cylindrical
axis
adjustment
is
established,
set
DIP
switch
No.
5
to
ON
and
press
the
[SET]
key.
The
offset
value
will
then
be
stored
in
the
EEPROM
and
two
short
beeps
will
be
generated.
Then
reset
DIP
switch
No.
5
to
OFF.
©
TEST
-13
:
Selection
of
the
model
eye
mode
and
human
eye
mode
and
parallel
shifting
of
the
spherical
power
Press
this
key
to
quit
TEST-13.
•
Selection
of
the
model
eye
mode
and
human
eye
mode
[HUMN
EYE]
:
Must
be
selected
to
measure
the
human
eye
and
model
eye
provided
with
the
instrument
(normal
measurement
mqde).
(MODL
EYE]
:
Must
be
selected
to
measure
the
model
eye
tools
during
measurement
accuracy
test.
Press
the
[End]
key
to
return
to
the
refractometer
test
program
select
screen
and
press
the
[EXIT]
key
to
enter
the
normal
measurement
mode
screen.
The
“R/L”
display
on
the
normal
measurement
mode
screen
will
turn
to
"M".
Now
you
can
measure
the
model
eye
tool
accurately.
*
After
the
model
eye
tool
measurement
is
complete,
turn
the
Main
switch
off
and
on
to
return
to
the
normal
human
eye
measurement
mode.
•
Parallel
shifting
of
the
spherical
power
Use
this
test
to
shift
the
measured
spherical
power
parallel.
[+000]
key:
Used
to
select
the
shifting
amount.
Each
press
of
this
key
changes
the
power
displayed
on
the
key
top.
[6-141
Repair
Manual

Other Nikon Measuring Instrument manuals

Nikon 8431 - Monarch - Dot User manual

Nikon

Nikon 8431 - Monarch - Dot User manual

Nikon LASER 50 User manual

Nikon

Nikon LASER 50 User manual

Nikon CoolShot 80 i VR User manual

Nikon

Nikon CoolShot 80 i VR User manual

Nikon MONARCH 3000 STABILIZED User manual

Nikon

Nikon MONARCH 3000 STABILIZED User manual

Nikon Nivo 2.C User manual

Nikon

Nikon Nivo 2.C User manual

Nikon ProStaff LASER 440 User manual

Nikon

Nikon ProStaff LASER 440 User manual

Nikon Aculon User manual

Nikon

Nikon Aculon User manual

Nikon Laser 550 User manual

Nikon

Nikon Laser 550 User manual

Nikon LASER 30 User manual

Nikon

Nikon LASER 30 User manual

Nikon 7487 - Team Realtree Laser 1200 Rangefinder User manual

Nikon

Nikon 7487 - Team Realtree Laser 1200 Rangefinder User manual

Nikon Forestry Pro II User manual

Nikon

Nikon Forestry Pro II User manual

Nikon COOLSHOT PRO II STABILIZED User manual

Nikon

Nikon COOLSHOT PRO II STABILIZED User manual

Nikon Coolshot 40i User manual

Nikon

Nikon Coolshot 40i User manual

Nikon Retinomax User manual

Nikon

Nikon Retinomax User manual

Nikon PROSTAFF 3i User manual

Nikon

Nikon PROSTAFF 3i User manual

Nikon Callaway ID Tech User manual

Nikon

Nikon Callaway ID Tech User manual

Nikon Callaway LR550 User manual

Nikon

Nikon Callaway LR550 User manual

Nikon NPL-302 Series User manual

Nikon

Nikon NPL-302 Series User manual

Nikon XF SERIES User manual

Nikon

Nikon XF SERIES User manual

Nikon Forestry Pro User manual

Nikon

Nikon Forestry Pro User manual

Nikon MONARCH 2000 User manual

Nikon

Nikon MONARCH 2000 User manual

Nikon Prostaff 1000i User manual

Nikon

Nikon Prostaff 1000i User manual

Nikon Forestry Pro User manual

Nikon

Nikon Forestry Pro User manual

Nikon DTM-350 User manual

Nikon

Nikon DTM-350 User manual

Popular Measuring Instrument manuals by other brands

Design Technology Tiny-Tach 2A operating instructions

Design Technology

Design Technology Tiny-Tach 2A operating instructions

mru SYNGAS VARIOluxx user manual

mru

mru SYNGAS VARIOluxx user manual

AFRISO Unitel operating instructions

AFRISO

AFRISO Unitel operating instructions

Flow vision FC50-EX-CA Instructions for installation and adjustment

Flow vision

Flow vision FC50-EX-CA Instructions for installation and adjustment

immerSUN iLink Installation & user guide

immerSUN

immerSUN iLink Installation & user guide

Sylvac EVO Bluetooth quick start guide

Sylvac

Sylvac EVO Bluetooth quick start guide

Unitek Miyachi LM45 Nd manual

Unitek Miyachi

Unitek Miyachi LM45 Nd manual

Measurement Computing DPG Series instruction manual

Measurement Computing

Measurement Computing DPG Series instruction manual

Tyredog TD2700F Adding a Relay / Programming Sensors

Tyredog

Tyredog TD2700F Adding a Relay / Programming Sensors

Moeller ZEV-XSW-820 installation instructions

Moeller

Moeller ZEV-XSW-820 installation instructions

Omron Z300 V3 - SETUP Setup manual

Omron

Omron Z300 V3 - SETUP Setup manual

MaxiTrack Sensors InTransit-2 instruction manual

MaxiTrack Sensors

MaxiTrack Sensors InTransit-2 instruction manual

Precision Digital Corporation PD6900-GP instruction manual

Precision Digital Corporation

Precision Digital Corporation PD6900-GP instruction manual

Siemens Sitrans F Series operating instructions

Siemens

Siemens Sitrans F Series operating instructions

wtw pH 330i operating manual

wtw

wtw pH 330i operating manual

Actel Silicon Explorer quick start

Actel

Actel Silicon Explorer quick start

Desco 19494 Operation and maintenance instructions

Desco

Desco 19494 Operation and maintenance instructions

AR DC7420 Operating and service manual

AR

AR DC7420 Operating and service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.