CAUTION
• When operating the handpiece always consider the safety of the patient.
• Check for vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop using the handpiece
immediately and contact dealer.
• Should the handpiece function abnormally, cease operation immediately and return the handpiece to an authorized dealer for repair.
• Depressing the push-button while the handpiece is rotating will result in OVERHEATING of the handpiece head. Special caution must be exercised
during use to keep cheek tissue AWAY from the push-button of the handpiece. Contact with cheek tissue may cause the push-button to depress and
burn injury to the patient may occur.
• Do not applying excessive power in the direction which bur removes. When the bur drops into patient’s oral cavity, they may drink by mistake and get
injured.
• This handpiece have built-in dynamo. Consumption Air generates electricity and LED become luminous. Therefore, do not perform autoclave
sterilization which it heat and cools rapidly.
• Do not allow any impact on to the Handpiece. Do not drop the Handpiece.
• Do not look into LED light directly.
CAUTION
• Ensure connect “Coupling and Hose” or “Handpiece and Coupling” firmly.
• Do not pull the Connector Ring while the handpiece is under air pressure. It may jump the handpiece out from the connector because of the
air pressure.
• Do not exceed the recommended air pressure. If it exceeds the pressure, it may bur jump out from the handpiece.
• Ensure clean, dry air supply. Moisture and debris in air supply may cause the motor to malfunction.
CAUTION
• Do not connect or disconnect the handpiece or bur until the motor has completely stopped.
• When used for crown removal and other procedures that would involve heavy vibration, the bur may become difficult to remove. Push the
button firmly, and forcibly pull out the bur with a pair of pliers. Apply small amount of spray lubricant into the chuck. This can be prevented if
the bur is repositioned a few times during the procedure.
• When abnormalities are in LED (dark, do not light on or flashing or so), contact dealer to repair.
• Do not attempt to disassemble the handpiece or tamper with the mechanism except NSK recommend on this Operation Manual.
• Always keep bur shanks clean. Dirt or debris in a chuck could cause poor bur concentricity or low chuck retention force.
• Do not exceed the drive motor speed recommended the bur manufacturers. If the rotary speed which the manufacturer of bur nominates is lower than
spec of handpiece, do not use it.
• Do not connect or disconnect the handpiece or bur until the motor has completely stopped.
• NSK has not recommended the repair and disassembly of the cartridge without exception. Handpiece may perform abnormally (abnormal noise or
abnormal vibration). Damage, failure or accidents are outside of our guarantee.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
• Use eye protectors, mask for safety when operating this handpiece.
• User is responsible for operation, maintenance and safety.
• If the handpiece has not been used for a long period, check for noise, vibration, and overheating before use.
• Treatment under the sufficient irrigation water(Spraying water).
1. Specification
Intended to use: This handpiece is designed only for clinical dental use by qualified personnel.
3. Check before treatment
Follow the check below, before use. If any abnormalities are found, stop using the handpiece immediately and contact the authorized NSK dealer.
1) Insert the Bur. (Do not exceed maximum rotation speed for any item.)
2) Check that there is no abnormal rotation, vibration, noise and overheating.
3) If there are no immediate apparent abnormalities, continue to rotate the handpiece for about 1 minute. Then check that there is no abnormal
heating on the outer surface.
4) Operate it if there are not abnormalities above.
CAUTION
• Do not use non-standard burs. The ISO standard shank diameter is φ1.59 - φ1.60 mm.
• Do not exceed the bur speed recommended by the bur manufacturer. Moreover, do not use the bur which it has lower rotation speed.
• Do not use long surgical burs. Do not use burs longer than 26mm.
• Do not use short shank burs. (M600LG M4, M600LG B2, M600LG QD)
• Always keep the bur shank clean. Entry of hard debris into the chuck via the bur shank could cause rotation slip and also prevent the bur from
being securely located in the chuck.
• Fully depress the Push Button and insert the bur into the chuck until it is secure.
• Test that the bur is secure by gently pulling and pushing the bur without depressing the Push Button.
4. Insertion /Removal of the Bur
2. Connecting/Disconnecting Handpiece
M500LG M4/B2, M600LG M4/B2
2-1 Connecting
1) Insert the handpiece into the hose connector.
2) Tighten the Hose Nut. (Fig.1)
2-2 Disconnect
Loosen the Hose Nut and pull out the Hose.
M500LG QD, M600LG QD
2-3 Connecting
1) Insert the Coupling into the Hose Connector.
2) Tighten the Hose Nut.
3) Pull the Connector Ring.
4) Push into the handpiece into Coupling.
2-4 Disconnect
Slide back the Connector Ring as shown in Fig.2.
SUPPLY AIR PRESSURE Setting
Set the Air pressure at the “Air Check point” on Fig. 1 and Fig. 2 The setting values are on the specification
table. Do not exceed the recommended air pressure by using Multi Gauge.
Multi Gauge: Midwest 4 Hole Order code Z109400 / Borden2/3 Hole Order code Z109200
4-1 To inserts the Bur
1) Insert the bur until it stops.( )
2) Depress the Push Button.( )
3) Insert the bur into the chuck until it is stops.( )
4) Release the button.
4-2 To remove the Bur
Depress the Push Button firmly and remove the bur.( )
CAUTION
• Be sure to hold the handpiece firmly to prevent the handpiece from slipping out of hand by the spray pressure when lubricating.
• Supply lubricant until it comes out of the handpiece head.(at least 2 seconds)
• Hold the spray can upright.
9. Non-Retraction Valve (QD Coupling)
To prevent oral fluids drawn into the water line, QD coupling has Non-Retraction valve. If water is
beginning to leaking from handpiece, replace non-retraction valve.
1) Remove the back-end gasket.
2) Replace Non-Retraction Valve.
10. Warranty
NSK warrants our product against poor manufacturing, poor workmanship and defects in materials. NSK reserves the right to analyze and determine
the cause of any problem. Warranty is voided should the handpiece be not used correctly. Replacement parts are available for seven years beyond
discontinuation of the model.
5-4 Lubrication
Lubricate by PANA SPRAY Plus
1) Mount the Arrow Head Spray Nozzle into the Spray Port.(Fig.7)
2) Remove the Bur from the handpiece.
3) Insert the Arrow Head Spray Nozzle in the drive Air Tube of the
handpiece and spray for approximately 2-3 seconds.
1) Mount the Spray Nozzle into the Spray Port.(Fig.8)
2) Remove the bur from the handpiece.
3) Insert the Spray Nozzle in rear of the handpiece and spray for approximately 2-3 seconds.
5. Maintenance
After the treatment of each patient, perform maintenance below.
5-1 Cleaning (Clean-Head System)
1) Scrub dirt and debris from the handpiece.(Fig.4)
2) Half fill a cup with clean water.
3) Rotate the handpiece and immerse half of the handpice head. (Fig. 5)
4) Rotate and stop intermittently the handpiece about 2 to 3 seconds. (About 3 times)
5) By the function of the clean head system, the clean head hole can be washed.
6) Stop the handpiece and wipe it dry. If the dirt could not remove from the hole, remove the cartridge and clean by brush.
7) Perform lubrication (4-3 Lubrication) and sterilization. (5-4 Lubrication, 5-5 Sterilization)
5-5 Sterilization
Autoclave sterilization is recommended. Sterilization is required first time you use and after each patient as noted below.
This handpiece and motor can be autoclavable up to Max.135°C.
Lubricate by Care3 plus
NSK Care3 Plus automatic handpiece cleaning and lubrication system.
After connecting the handpiece to the correct adaptor, activate the Care3 System per the Care3 System Operation Manual.
Autoclave Procedure:
1) Clean and lubricate the handpiece.
2) Insert into an autoclave pouch. Seal the pouch.
3) Autoclavable up to max. 135˚C.
ex.) Autoclave for 3 min. at 134°C.
4) Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C.
NSK recommends Class B sterilization.
M500LG M4
Mini Mini
ISO9168 Type 2 4 hole
380,000 - 450,000 min-1
0.20MPa(2.0kgf/cm2) - 0.25MPa(2.5kgf/cm2)
Quattro Spray
Push Button Chuck
ISO 1797-1 Type3
φ1.59-1.60mm Short Shank Bur
9.0mm
21mm
φ2mm
45±5NL/min
M500LG B2
ISO9168 Type 1 2 hole
40±5NL/min
M500LG QD
M600LG M4
Standard Standard Mini Standard
330,000 - 430,000 min-1 380,000 - 450,000 min-1 330,000 - 430,000 min-1 380,000 - 450,000 min-1 330,000 - 430,000 min-1
ISO1797-1 Type3
φ1.59-1.60mm Standard Bur ISO 1797-1 Type3
φ1.59-1.60mm Short Shank Bur ISO1797-1 Type3
φ1.59-1.60mm Standard Bur ISO 1797-1 Type3
φ1.59-1.60mm Short Shank Bur ISO1797-1 Type3
φ1.59-1.60mm Standard Bur
10.5mm 9.0mm 10.5mm 9.0mm 10.5mm
25mm 21mm 25mm 21mm 25mm
M600LG B2 M600LG QD
QD Coupling
Model
Head Type
Connection Type
Rotation Speed
Drive Air Pressure
Water Spray Type
Chuck Type
Bur Type
Min. Chucking Length
Max. Bur Length
Max. Working Part
Diameter
Max. Air Consumption
(0.25MPa)
Head Diameter x Height) φ10.6mm x 12.4mm φ12.1mm x 13.3mm φ10.6mm x 12.4mm φ12.1mm x 13.3mm
45±5NL/min 50±5NL/min
φ10.6mm x 12.4mm φ12.1mm x 13.3mm
Water Pressure
Chip Air Pressure
0.05MPa(0.5kgf/cm2) - 0.20MPa(2.0kgf/cm2)
Optic
0.15MPa(1.5kgf/cm2) - 0.25MPa(2.5kgf/cm2)-0.15MPa(1.5kgf/cm2) - 0.25MPa(2.5kgf/cm2)
Glass Rod
Gasket Non-Retraction Valve
Drive Air Tube
MidWest
4 hole Bordent
2 hole
Handpiece
Arrow Head Spray Nozzle
PANA SPRAY plus/PANA SPRAY
Fig. 7
M500LG QD, M600LG QD
M500LG M4/B2, M600LG M4/B2
Fig. 13
Fig. 3
CAUTION
Use only clean water to maintain the Clean Head System.
Fig. 4
CAUTION
• To prevent infectiveness, Autoclave the handpiece every patient.
• Do not autoclave the handpiece, even when it is bagged, together with other instruments, from which chemicals are not sufficiently cleaned.
This is to avoid discoloration and/or damage to the plating of the handpiece.
• In case the sterilizer chamber temperature could exceed 135°C during the dry cycle, skip the dry cycle.
• Do not heat or cool the handpiece quickly. Rapid change in temperature could break the glass rod or dynamo or LED parts.
6. Changing Turbine Cartridge
1) Mount a dummy bur in the chuck.
2) Turn the Wrench counter clockwise and remove the Head Cap by Head Cap Wrench.
3) Push up the dummy bur to remove Cartridge.
4) Clean the head interior. (Wipe extra oil after lubrication)
5) Align the Alignment Pin and insert Cartridge. (Fig. 8)
6) Turn the Wrench clockwise and mount the Head Cap. (Fig. 9)
* Replacement Cartridge: SX-MU03 Mini Cartridge Order Code P1001250
SX-SU03 Standard Cartridge Order Code P1002250
2) Take out the cleaning wire from the back end of the brush holder. Insert the wire straight into the spray ports with caution. Do not forcibly insert the
wire into the port. Damaged or disoriented port could cause the spray diverge or directed away from the bur, and the cooling efficiency deteriorates.
3) After use, clean the cleaning wire. Push into the brush holder with the pointed end of the wire inward.
7.Spray Port Cleaning
When spray port are clogged, or spray does not exit evenly from three ports, clean the ports as follows:
1) Remove the bur from the chuck. (Refer to “4. Insertion/Removal the Bur” and “6.
Replacing the Cartridge”)
8. Replacing the O-rings (QD Coupling)
Worn or damaged O-ring may cause air or water to leak.
When water is found in the exhaust air or water leak from the coupled point, replace four O-rings as shown in Fig.10.
Replacement O-ring Set (4pcs) : Order Code Y900592
Fig. 12
O-Ring
Clean Head Hole
Clean Head HoleClean Head Hole
Fig. 5
CAUTION
Do not use a pointed tool or sharp edged to clean the Cellular Glass Optic Rod. They may damage the mirror finish and reduce the light
transmission efficiency. If Cellular Glass Optic Rod end has become damaged, return it to dealer for repair.
5-3 Cleaning
Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth. Do not
use a wire brush.
* Refer the manual of Thermo-Disinfector for performing it.
This handpiece can be washed via Thermo Disinfector.
5-2 Cleaning (Glass Rod)
Wipe clean the Cellular Glass Optic Rod entry point and exit point with an alcohol-immersed cotton swab. Remove all debris and oil.(Fig.6)