2014.11.20 10 S
Caution
• Never clean the handpiece (EX-5B, EX-6B, EX-6D) in boiling water.
• Do not wipe with, or clean or immerse in, high acid water or
sterilizing solutions.
Caution
• Be sure to hold the handpiece firmly to prevent the handpiece from
slipping out by the spray pressure when lubricating.
• Supply lubricant until it comes out of the handpiece nose (for
approx. 2 seconds).
• Keep the PANA SPRAY Plus / PANA SPRAY can upright.
CAUTION
• Make sure to align the ●mark with " " mark. If those marks are
not completely aligned, it may cause handpiece overheating or
patient burn injuries.
• Always insert the bur fully into the chuck. Insufficient insertion of the
bur could cause premature failure of the bearing.
• Remove the bur only after the handpiece has completely stopped
rotating.
• Do not rotate the handpiece when the Bur Lock Ring is turned to
" " (OPEN) position or a bur is not inserted.
• Always keep the bur shank clean.
• Entry of debris into the chuck, via the bur shank, could cause bur
rotation slip and also prevent the bur from being securely located in
the chuck.
• Do not exceed the bur speed recommended by the bur
manufacturer.
• Do not exceed maximum bur length recommended by the
handpiece manufacturer.
• Do not apply excess pressure to the bur as it may break or bend or
become difficult to remove.
• DO NOT use burs with problems listed below as the bur may break
or disengage from the chuck.
- Bent, deformed, anisomerous (worn), rusted, broken, deficient bur.
- Bur which is cracked on the edge or axis.
- Non-ISO standard, or tampered bur.
(1) Cleaning
Wipe clean the handpiece sheath with alcohol-dipped soft tissue or
cotton. Never clean the handpiece in chemicals, with wire brushes, or
in ultrasonic cleaners.
(2) Sterilization
Steam autoclave is recommended.
Sterilization is required first time you use and after each patient.
Autoclaving procedure
1) Clean the handpiece sheath as described above.
2) Lubricate the handpiece with PANA SPRAY Plus / PANA SPRAY. (Ref.
4. Lubrication)
3) Place handpiece into autoclaving pouch and seal it in accordance with
instructions on the pouch.
4) Autoclavable up to max. 135°C (275°F).
ex.) Autoclave for 20 min. at 121°C, or 15 min. at 132°C.
5) Keep the handpiece in the autoclave pouch to keep it clean until you
use it.
* Sterilization at 121°C for more than 15 minutes is recommended by
EN13060 or EN ISO17665-1.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system,
refer to the system instructions.
5. Cleaning and Sterilization
Apply PANA SPRAY Plus / PANA SPRAY or NSK automatic handpiece
cleaning and lubrication system after each use and/or before autoclaving.
PANA SPRAY Plus / PANA SPRAY
1) Mount the E-type spray nozzle into the spray port.
2) Shake the can 3-4 times to well mix lubricant and propellant.
3) Insert the E-type spray nozzle in the rear of the handpiece and spray for
approximately 2-3 seconds until the oil comes out of the handpiece
head or nose.
4. Lubrication
Fig.2
Fig.3
E-type Spray Nozzle Handpiece
PANA SPRAY Plus / PANA SPRAYPANA SPRAY Plus / PANA SPRAY
The products listed below are washable with Thermo Disinfector.
EX-5BM, EX-6BM, EX-6DM
Open
Close
Chuck Open
Chuck Close