
10
01. Temperature Display (°C): This is the display where the sterilization chamber
temperature and the temperature set values are shown on. It displays the set
value during programming and the sterilization chamber temperature any time
during the run.
02. Time Display : This is the display where the sterilization time is counted down
after the set temperature has been reached. It also displays the set value during
programming.
03. Temperature Set Key : This key is used to set the sterilization temperature. It is
also used to check the set value of temperature during the operation.
04. Time Set Key : The key is used to set the sterilization time. It is also used to
check the set value of time during the operation.
05. Value Decrease Key : This key is pushed to decrease the programming values.
06. Value Increase Key : This key is pushed to increase the programming values.
07. Water Filling Led : This led is on while the chamber is being filled with water
autmatically.
08. Preparation Led : This led is on until the set temperature is reached. The
sterilization does not start while this led is on.
09. Sterilization Led : This led is on while sterilization phase is being carried out.
The led turns off automatically as the sterilization phase ends according to
programmed sterilization time.
10. Steam Discharge Led : This led turns on when the steam in the sterilization
chamber starts to be discharged. The drying led accompanies this led for the last
three minutes of the discharge cycle.
11. Drying Led : This led turns on while the drying cycle is being carried out.
12. Cycle Ended Led : This led turns on to show that the cycle is completed.
13. Start Key and Led : This button is pushed to start the program.
14. Stop Key and Led : This button is pushed to stop the program manually.
15. Audible Alarm Mute Button : This button is pushed,
•to mute the audible alarms and to cancel the alarm codes,
•to turn off “Cycle Ended” led
•to turn back the sterilizer to the stand-by position from OFF position which occurs
after the power cut.
16. Lid Open Led : This led is on when the sterilization chamber lid is open or if it is
not closed properly. The program doesn’t start until the lid is properly closed.