PKP DS02.1 User manual

Instruction Manual
DS02
Miniature Variable Area Flowmeter and Switch
- All Metal for high pressures -
PKP Prozessmesstechnik GmbH
Borsigstraße 24
D-65205 Wiesbaden- ordenstadt
Tel.: ++49-(0)6122-7055-0
Fax: ++49-(0)6122-7055-50
Email: info@pkp.de

Table of ontents
Safety Information..................................................................................................................2
Overview................................................................................................................................4
Installation..............................................................................................................................5
Connecting devices equipped with Reed switches................................................................7
Electrical connection..............................................................................................................9
Contacts for DS02.1 / DS02.2:............................................................................................10
Contacts for DS02.2 / DS02.3 / DS02.4..............................................................................12
Setting the switch point........................................................................................................14
Troubleshooting guide.........................................................................................................15
Maintenance / Maintenance plan.........................................................................................16
Degree of protection (IP-Code)............................................................................................16
Returns.................................................................................................................................16
Safety Information
General Instructions
To ensure safe operation, the device should only be operated according to the specifica -
tions in the instruction manual. The requisite Health & Safety regulations for a given
application must also be observed. This statement also applies to the use of accessories.
Every person who is commissioned with the initiation or operation of this device must have
read and understood the operating instructions and in particular the safety instructions!
The liability of the manufacturer expires in the event of damage due to improper use, non-
observance of this operating manual, use of insufficiently qualified personnel and unau-
thorized modification of the device.
Proper Usage
The variable area flowmeters of the DS02 series are used exclusively for monitoring and
indicating continuous flows of gaseous and liquid media that do not attack the materials
used. All other usage is regarded as being improper and outside the scope of the device.
Max. Pressure:
DS02.1 DS02.2 DS02.3/4
Liquids: 200 bar (brass)
300 bar (VA)
300 bar (brass)
350 bar (VA)
300 bar (brass)
350 bar (VA)
Gases: 300 bar (brass)
350 bar (VA)
300 bar (brass)
350 bar (VA)
250 bar (brass)
300 bar (VA)
Pressure loss: DS02.1: 0,02–0,2 bar
DS02.2: 0,02–0,3 bar
DS02./4: 0,02–0,4 bar
DS02 Instruction manual 03/2020 page 2

Max. Medium-Temperature: 100 °C (optional 160 °C) for liquids
120 °C (optional 160 °C) for gases
Ex-devices: acc. ATEX-declaration
In particular, applications in which shock loads occur (for example, pulsed operation)
should be discussed and checked in advance with our technical staff.
In particular applications in which shock loads occur (e.g. intermittent operation) should be
discussed and checked in advance with our technical staff.
Do not use flow monitors with fast-switching valves.
The series DS02 flow meter devices should not be deployed as the sole agents to prevent
dangerous conditions occurring in plant or machinery. Machinery and plant need to be
designed in such a manner that faulty conditions and malfunctions do not arise that could
pose a safety risk for operators.
Dangerous substances
For dangerous media such as e.g. Oxygen, Acetylene, flammable or toxic substances as
well as refrigeration systems, compressors, etc. must comply with the relevant regulations
beyond the general rules.
Qualified Personnel
The DS02 devices may only be installed by trained, qualified personnel who are able to
mount the devices correctly. Qualified personnel are persons, who are familiar with
assembling, installation, placing in service and operating these devices and who are suit-
ably trained and qualified.
Inward Monitoring
Please check directly after delivery the device for any transport damages and deficiencies.
Additional with reference to the accompanying delivery note the number of parts must be
checked.
Claims for replacement or goods which relate to transport damage can only be considered
valid if the delivery company is notified without delay.
DS02 Instruction manual 03/2020 page 3

Overview
1: Process connection (outlet)
2: Device body
3: Switch contact and female socket or switch contact
with cable
4: Switch point adjustment scale
5: Display housing with display scale and pointer
mechanism
6: Process connection (inlet)
Switch poi t adjustme t scale:
A scale is applied to the device body, to which the
desired switch point can be adjusted.
Display scale:
On the device, a display housing is mounted with integrated
indicator scale and pointer movement. The pointer on the
scale indicates the current flow value.
DS02 Instruction manual 03/2020 page 4

Installation
Process connection:
The following requirements must be observed to avoid damage to the flow monitor or the
system:
a process connection suitable for the device must be provided by the customer
check connection size and screw-in depth
use suitable sealants (liquid sealants damage the flow monitor if they run in)
seal properly
Ambient conditions:
do not install the flow monitor as a supporting part in a pipe system
the medium must not carry solid bodies. Magnetic particles accumulate on the
magnetic float and impair its function.
Check corrosion and antifreeze agents from the application for compatibility.
Notice:
The following requirements must be observed, otherwise the function of the flow monitor is
impaired or measurement results are falsified:
external magnetic fields will influence the switch contact. Keep sufficient distance to
magnetic fields (e.g. electric motors).
piping, process connections or supports made of ferromagnetic material influence
the magnetic field of the device. Keep a space of minimum 100 mm to those
materials (e.g. steel).
changes in cross-section, branch-offs or arcs in the pipe system impair measuring
accuracy. Ensure that the unimpeded flow sections are maintained (in front of
instrument 10 x nominal diameter, behind the instrument 5 x nominal diameter).
ever reduce the pipe diameter immediately before the device.
in the case of liquid media, take appropriate steps to ensure that the device is ven-
ted.
to ensure measuring accuracy, the device must always be completely filled with
medium.
vent the pipeline. If there are air pockets in the line during the measurement, this
could result in damage to the device caused by hydraulic shock. This may cause
malfunctions.
ensure that the plant is operating without cavitation. Cavitation may result in mal-
functions and damage to the device.
ensure that the medium is flowing continuously. Pulse-like staggered loads could
destroy the device. This may result in serious injury to the user.
if the mediums is contaminated by solids, a strainer must be installed before the
device (e.g. SF00 or SF01 from PKP)
DS02 Instruction manual 03/2020 page 5

Direction of flow:
Only install the flow monitor in one of the positions displayed in the drawing. The medium
must flow in the direction of the arrow (from a low to a high scale value).
DS02 Instruction manual 03/2020 page 6

onnecting devices equipped with Reed switches
Reed switches are basically designed for small contact ratings. To connect a load with
higher power consumption it is indispensable to use a contact protection relay (e.g. our
series MSR01)
If you connect directly a load to a Reed contact the following recommendations should be
considered.
one of the contact rating values printed on the switching unit must not to be exceeded,
even momentarily. This is valid for each of the given values individually: voltage, current,
power. The Reed contact integrated in the switching unit is very sensible to electrical over-
load
Danger of overload is given by the following applications:
•inductive load
•capacitive load
•lamp load
Inductive Load
Inductive loads consist e.g. of relay, contactors, solenoid valves, motors, electric engines,
etc.
WAR I G: Voltage spikes at shut down (up to 10 times of nominal voltage)
Protective measures: (examples)
(Flyback diode, e.g. type 1 4007)
apacitive Load
Capacitive loads consist e.g. of long connection cables or capacitive consumers.
WAR I G: High current spikes at switching on (this will exceed the nominal current)
Protective measures: (examples)
Limitation of current by a resistor
DS02 Instruction manual 03/2020 page 7

Lamp Load
Lamp loads consist e.g. by light bulbs, starting motors .
WAR I G: High current spikes at switching on, because the glowing spiral has low
resistance at low temperature.
Protective measures: (examples)
Limitation of current by a resistor
or preheating of the glowing
spiral.
onnecting to a PL
There is no need for protective measures by connecting the Reed switch to a PLC. The
Reed contacts are plated by Tunsten, Gold, and Rhodium located in a protective atmo-
sphere. They can be directly connected to the input terminals of a PLC without problems.
R - ircuits as protective measures (Boucherot cell, Snubber)
In practice the following values of resistor/capacitor cells give good results. evertheless,
the values given in the following tables are only recommendations for general purposes.
But it cannot be guaranteed that for specific applications more adequate Boucherot cells
may exist.
For Reed switches of 10 – 40 VA
Voltage [V] Resista ce [Ohm] Capacita ce [ F]
230 1500 330
115 470 330
48 220 330
24 100 330
For Reed switches of 40 – 100 VA
Voltage [V] Resista ce [Ohm] Capacita ce [ F]
230 1000 330
115 470 330
48 100 330
24 47 330
DS02 Instruction manual 03/2020 page 8

Electrical connection
The switch contacts employed in these devices are potential free and do not require a
power source.
Note:
Switch contacts and flow monitor have been optimally harmonized. After replacement of a
switch contact, the switch point must be readjusted.
The contacts open/change when the flow rate falls below the set value.
Grou di g of the device:
When installing the device in a pipeline, make sure that the device is grounded via the
pipeline. This prevents dangerous potential differences from occurring
Switching capacity of contacts:
Switchi g capacity
Type Size Co tact fu ctio A gle plug IP65 M12x1 plug IP67** Cable co ectio (1 m) IP67
DS02.1 1/4“ 1 = /O 140 VAC / 0,7 A / 20 VA
200 VDC / 1 A / 20 VA
125 VAC / 0,7 A / 20 VA
125 VDC / 1 A / 20 VA
140 VAC / 0,7 A / 20 VA
200 VDC / 1 A / 20 VA
2 = SPDT 150 VAC/DC / 1 A / 20 VA 125 VAC/DC / 1 A / 20 VA -/-
3SM = Ex- /O* gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
3UM = Ex-SPDT* -/-
DS02.2 1/2“ 1 = /O 230 V / 3 A / 60 VA 125 V / 3 A / 60 VA 230 V / 3 A / 60 VA
2 = SPDT 250 V / 1,5 A / 50 VA,
min load: 3 VA
125 V / 1,5 A / 50 VA,
min load: 3 VA
-/-
2X = SPDT for SPS 250 V / 1 A / 60 VA -/- -/-
3SM = Ex- /O* gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
3UM = Ex SPDT* -/-
3ST5 = Ex- /O T5*
3ST6 = Ex- /O T6*
-/- -/- 250 V / 2 A / 60 VA (2 m cable)
3UT5 = Ex-SPDT T5*
3UT6 = Ex-SPDT T6*
-/- -/- 250 V / 1 A / 30 VA,
min load: 3 VA (2 m cable)
DS02.3
DS02.4
3/4“
1“
1 = /O 250 V / 3 A / 100 VA
2 = SPDT 250 V / 1,5 A / 50 VA, min load: 3 VA
2X = SPDT for SPS 250 V / 1 A / 60 VA -/- -/-
3ST5 = Ex- /O T5*
3ST6 = Ex- /O T6*
-/- -/- 250 V / 2 A / 60 VA (2 m cable)
3UT5 = Ex-SPDT T5*
3UT6 = Ex-SPDT T6*
-/- -/- 250 V / 1 A / 30 VA,
min load: 3 VA (2 m cable)
* Exact max. switching capacity: see ATEX documents
** Protection class M12x1 plug or DS02.1 and DS02.2: IP65
DS02 Instruction manual 03/2020 page 9

ontacts for DS02.1 / DS02.2:
Switching contact with connector acc. to EN 175301-803:
Normally Ope (NOC) Cha ge Over (COC)
Switch position under no-flow condition Switch position under no-flow condition
The ground-pin is not used The ground-pin is not used
DS02 Instruction manual 03/2020 page 10

Switching contacts with plug connector M12x1 (mating connector not included in deliv-
ery)
Normally Ope (NOC) Cha ge Over (COC)
Switch position under no-flow condition Switch position under no-flow condition
Switching contact with cable:
The individual cores of the cable are numbered according to the following connection dia -
gram:
Normally Ope (NOC)
Switch position under no-flow condition
DS02 Instruction manual 03/2020 page 11

ontacts for DS02.2 / DS02.3 / DS02.4
Switching contact with connector acc. to EN 175301-803:
Normally Ope (NOC) Cha ge Over (COC)
Switch position under no-flow condition Switch position under no-flow condition
The ground-pin is not used The ground-pin is not used
DS02 Instruction manual 03/2020 page 12

Switching contacts with plug connector M12x1 (mating connector not included in
delivery)
Normally Ope (NOC) Cha ge Over (COC)
Switch position under no-flow condition Switch position under no-flow condition
Switching contact with cable:
The individual cores of the cable are numbered according to the following connection dia -
gram:
Normally Ope (NOC) Cha ge Over (COC)
Switch position under no-flow condition Switch position under no-flow condition
DS02 Instruction manual 03/2020 page 13

Setting the switch point
The following instructions describe the procedure for a ormally Open Contact ( OC).
The actual state (open or closed), can be determined using a continuity meter.
1. Loosen the switch contact set screws (1) using a
screwdriver
2. Slide the switch contact to the flow value to be
monitored. Make sure that the arrow on the switch
contact label is in exact alignment with the desired
flow value on the body scale.
3. Re-tighten the switch contact housing set screws (1)
using a screwdriver. Observe the correct tightening
torque of 0,4 m.
The set switch point corresponds to the switch-off point of the switch contact by decreas-
ing flow.
DS02 Instruction manual 03/2020 page 14

Troubleshooting guide
Fault description Cause Remedy
The switch contact does not
switch
o medium flowing through
flow monitor
Check that medium is
flowing through the pipeline
Flow is too low or the switch
contact is set too high
Adjust the switch contact to
a lower flow rate
Use the device at another
measuring range
Increase the flow rate
Incorrect reduction fitting or
pipe diameter is too small
Correct pipe diameter
Float is stuck Disassemble and clean the
device
Switch contact is defective Remedy the cause of the
defect (short-circuit,
overload)
Replace the switch contact
Switch contact is
permanently switched
Flow is too high or the switch
contact is set too low
Reduce the flow
Adjust the switch contact to
a higher flow rate
Use the device at another
measuring range
Switch contact is
permanently switched
Float is stuck Disassemble and clean the
device
Switch contact is defective Remedy the cause of the
defect (short-circuit,
overload)
Replace the switch contact
The switch point is not the
same as the actual flow rate
Improper scale installed for
media used
Request proper conversion
table or scale for media
used
Incorrect reduction fitting or
pipe diameter is too small
Correct pipe diameter
Device is dirty Disassemble and clean the
device
Device is defective Remove device from system
and contact PKP
DS02 Instruction manual 03/2020 page 15

Maintenance / Maintenance plan
I tervals for replaci g wear parts
DS02 type flow monitors require very little maintenance due to the small number of mov-
ing parts. The intervals for the replacement of wear parts depend significantly on the oper -
ating conditions as well as on the composition of the medium flowing through the device.
For this reason, no intervals have been set by the manufacturer. The operator must
determine suitable intervals based on the local conditions and circumstances.
Mai te a ce work:
Visual inspection for dirt/soiling
Visual inspection for free movement of float
Visual inspection for leaks from the device
Check function of switch contact
For detailed i formatio o mai te a ce a d clea i g of the device, please refer to
a separate ma ual. Please ask for them if required.
Degree of protection (IP- ode)
Process co ectio Specificatio of
co ectio material
Degree of protectio
E 175301-803 with gland Ø of connection cable:
6-8 mm
IP65
M12x1 (DS02.1/2) Plug connector M12x1 IP65
M12x1 (DS02.3/4) Plug connector M12x1 IP67
Cable -- IP67
Returns
For returns please contact us:
PKP Prozessmesstechnik GmbH
Info@pkp.de
+49 (0) 6122-7055-0
DS02 Instruction manual 03/2020 page 16

Flow Measurement and Monitoring
DS02
Miniature Variable Area
Flowmeter and Switch-
All Metal Version
•for low viscosity liquids and gases
•small mounting dimensions
•brass or stainless steel version
•robust design without a measuring glass tube
•universal installation osition
•high switching accuracy
• o tional Ex- version acc. to ATEX
•analogue transmitter
4...20 mA available
Description:
The flow switch model DS02 works according to a modified
variable area principle. The float is guided in a cylindrical
measuring tube by means of a spring. The flowing medium
moves the float in the flow direction.
A Reed contact is mounted outside the meter in a sealed
housing. This reed contact is encapsulated in a continuously
ad ustable housing and thus protected from external
influences.
When the float reaches the position of the Reed contact the
switch will close. With higher flows the float moves further
upward until it reaches a built-in float stop, still keeping the
switch closed. This ensures a bistable switch function at any
time. An external pointing instrument is magnetically coupled
to the float and indicates the flow rate.
Typical application:
The variable area flowmeters and monitors DS02 are used to
measure and monitor continuous flow rates of low-viscosity
liquids or gaseous media.
Areas of applications are:
•cooling systems
•engineering
•medical technology
•pharmaceutical and chemical industries
•research and development
PKP Prozessmesstechnik GmbH
Borsigstr. 24 • D-65205 Wiesbaden
S +49 (0) 6122-7055-0 • T +49 (0) 6122 7055-50
[email protected] • www.pkp.de
20201026

Models:
Measuring ranges:
water: 5...60 ml/min – 60...150 l/min
air: 0,6...2,2 Nl/min – 200...650 Nl/min
(referenced to 1 bar abs, 20°C)
Materials: Brass or stainless steel versions
Technical Data:
Max. Pressure:
DS 2.1 DS 2.2 DS 2.3/4
Liquids: 200 bar (MS)
300 bar (VA)
300 bar (MS)
350 bar (VA)
300 bar (MS)
350 bar (VA)
Gases: 300 bar (MS)
350 bar (VA)
300 bar (MS)
350 bar (VA)
250 bar (MS)
300 bar (VA)
Pressure loss: DS02.1: 0,02–0,2 bar
DS02.2: 0,02–0,3 bar
DS02./4: 0,02–0,4 bar
Max. media-
temperature: 100 °C (optional 160 °C) for liquids
120 °C (optional 160 °C) for gases
Ex-devices acc. to. ATEX-marking
Operating temp.: 70 °C with analogue transmitter AZ06
Electr. Connection: DS 2.1 and DS 2.2:
angle plug acc. to EN 175301-803,
form C (DIN 43650)
DS 2.3 and DS 2.4
angle plug nach EN 155301-803,
form A (DIN 43650),
Ex-contact 3S and 3U with 2 m cable
optional: cable connection
round plug M12 x 1 acc. to EN 50044,
angle plug with LED or glow lamp
Accuracy: ± 10 % of full scale
(for vertical installation)
Materials:
Brass version:
Wetted parts:
Meas. tube: brass (outside nickel plated)
Spring: st. steel 1.4571
Gaskets*: NBR, optional: EPDM, FKM
Magnet: ferrite
all other wetted parts: brass
Stainless steel version 1.4571:
Wetted parts:
Gaskets*: FKM, optional EPDM, NBR
Magnet: ferrite
all other wetted parts: stainless steel1.4571
*only for reduced connection
Order Code:
Order Number: DS 2.
Miniature variable area flowmeter-
and switch
1. 1. 1. W13. 1. 1. .
Connection female thread:
1 = G 1/4 1N = 1/4“ NPT*
2 = G 1/2 2N = 1/2“ NPT*
3 = G 3/4* 3N = 3/4“ NPT*
4 = G 1 4N = 1“ NPT*
*) reduced
Material:
1 = brass
2 = stainless steel 1.4571
Scale:
1 = for water
2 = for air (at 1 bar abs., 20 °C)
Measuring ranges:
Water Air:
only DS 2.1:
W101 = 5–60 ml/min L1002 = 0,6–2,2 Nl/min
W102A = 40–130 ml/min L1006 = 1,7–6,0 Nl/min
W106 = 0,1–0,6 l/min L1008 = 2,5–8,0 Nl/min
W11 = 0,2–1,2 l/min L1012 = 3–12 Nl/min
W12 = 0,4–2 l/min L1022 = 3–22 Nl/min
W13 = 0,5–3 l/min L1024 = 7–24 Nl/min
W15 = 1,0–5 l/min L1034 = 12–34 Nl/min
L1056 = 16–56 Nl/min
L1080 = 20–80 Nl/min
only DS 2.2:
W202 = 0,02–0,2 l/min L2010 = 2,5–10 Nl/min
W206 = 0,2–0,6 l/min L2020 = 5,5–20 Nl/min
W21 = 0,4–1,8 l/min L2030 = 8–30 Nl/min
W23 = 0,8–3,2 l/min L2035 = 10–35 Nl/min
W27 = 2–7 l/min L2090 = 45–90 Nl/min
W213 = 3–13 l/min L2220 = 55–220 Nl/min
W220 = 4–20 l/min L2240 = 65–240 Nl/min
W230 = 8–30 l/min L2300 = 80–300 Nl/min
L2525 = 140–525 Nl/min
DS 2.3 or DS 2.4
W3030 = 10–30 l/min L30180 = 60–180 Nl/min
W3045 = 15–45 l/min L30300 = 100–300 Nl/min
W3060 = 20–60 l/min L30650 = 200–650 Nl/min
W3090 = 30–90 l/min
only DS 2.4
W4150 = 60–150 l/min
Addition S...= special scale
Number of contacts:
0 = without contact (only for devices with indication and/or AZ06)
1 = 1 contact
2 = 2 contacts
Contact function / Analogue output:
(contact or analogue transmitter available)
0 = without
1 = N/O
2 = SPDT
2X = SPDT for SPS application (for devices from 1/2")
3ST5 = Ex-N/O, T5 (100 °C), with 2 m cabel, not for DS02.1
3ST6 = Ex-N/O, T6 (80 °C), with 2 m cabel, not for DS02.1
3UT5 = Ex-SPDT, T5 (100 °C), with 2 m cabel, not for DS02.1
3UT6 = Ex-SPDT, T6 (80 °C), with 2 m cabel, not for DS02.1
3SM = Ex-N/O, only DS02.1 and DS02.2
3UM = Ex-SPDT, only DS02.1 and DS02.2
AZ06.1.I = analogue transmitter 4...20 mA
AZ06.1.V = analogue transmitter 0...10 V
Options:
0 = without
1 = please specify in plain text
AZ = with pointer indication (not for DS02.1)
HT = high temperature version 160 °C
M12 = round plug M12 x 1 acc. to EN 50044 (Tmax. 85 °C)
Kx = cable version 1 m, 2 m, 5 m or 10 m
PKP Prozessmesstechnik GmbH
Borsigstr. 24 • D-65205 Wiesbaden
S +49 (0) 6122-7055-0 • T +49 (0) 6122 7055-50
[email protected] • www.pkp.de
Flow

Dimensions:
DS 2.1 DS 2.2 – DS 2.4
Contacts:
The contact opens/changes, if the flow level has fallen under the
ad usted value
* Exact max. switching capacity: see ATEX documents
** Protection class M12x1 plug for DS02.1 and DS02.2: IP65
ATEX-designations:
Contacts 3SM and 3UM for DS 2.1/2.:
ATEX II 2 G Ex ib IIC and ATEX II 2 D Ex ib IIIC
for connection to a certified intrinsically safe circuit,
temperature range -5 °C < TService < 45 °C, Li=0, Ci=0
Dimensions:
Type Dimensions [mm] Weight
SW D B G T L [g]
DS 2.1
DS 2.2
DS 2.3
DS 2.4
17
27
41
41
17
31
47
47
47
52
76
76
G 1/4
G 1/2
G 3/4
G 1
10
14
21
17
65
90
152
130
140
350
1200
1050
Note: DS02.2...AZ and DS02.3(4)...AZ: square housing
Contacts 3ST5, 3ST6, 3UT5, 3UT6 for DS 2.2/3/4.:
ATEX II 2 G Ex mb IIC T6 Gb, ATEX II 2 D Ex tb IIIC T80 °C Db
ATEX II 2 G Ex mb IIC T5 Gb, ATEX II 2 D Ex tb IIIC T100 °C Db
(with cable connection, Standard 2 m only)
PKP Prozessmesstechnik GmbH
Borsigstr. 24 • D-65205 Wiesbaden
S +49 (0) 6122-7055-0 • T +49 (0) 6122 7055-50
[email protected] • www.pkp.de
Switching capacity
Type Size Contact function Angle plug IP65 M12x1 plug IP67** Cable connection (1 m) IP67
DS02.1 1/4“ 1 = N/O 140 VAC / 0,7 A / 20 VA
200 VDC / 1 A / 20 VA
125 VAC / 0,7 A / 20 VA
125 VDC / 1 A / 20 VA
140 VAC / 0,7 A / 20 VA
200 VDC / 1 A / 20 VA
2 = SPDT 150 VAC/DC / 1 A / 20 VA 125 VAC/DC / 1 A / 20 VA -/-
3SM = Ex-N/O* gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
3UM = Ex-SPDT* -/-
DS02.2 1/2“ 1 = N/O 230 V / 3 A / 60 VA 125 V / 3 A / 60 VA 230 V / 3 A / 60 VA
2 = SPDT 250 V / 1,5 A / 50 VA,
min load: 3 VA
125 V / 1,5 A / 50 VA,
min load: 3 VA
-/-
2X = SPDT for SPS 250 V / 1 A / 60 VA -/- -/-
3SM = Ex-N/O* gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
gas: < 30 V / 0,101 A / 0,76 W
dust: < 30 V / 0,25 A / 0,75 W
3UM = Ex SPDT* -/-
3ST5 = Ex-N/O T5*
3ST6 = Ex-N/O T6*
-/- -/- 250 V / 2 A / 60 VA (2 m cable)
3UT5 = Ex-SPDT T5*
3UT6 = Ex-SPDT T6*
-/- -/- 250 V / 1 A / 30 VA,
min load: 3 VA (2 m cable)
DS02.3
DS02.4
3/4“
1“
1 = N/O 250 V / 3 A / 100 VA
2 = SPDT 250 V / 1,5 A / 50 VA, min load: 3 VA
2X = SPDT for SPS 250 V / 1 A / 60 VA -/- -/-
3ST5 = Ex-N/O T5*
3ST6 = Ex-N/O T6*
-/- -/- 250 V / 2 A / 60 VA (2 m cable)
3UT5 = Ex-SPDT T5*
3UT6 = Ex-SPDT T6*
-/- -/- 250 V / 1 A / 30 VA,
min load: 3 VA (2 m cable)

Accessories (see se arate data sheets):
• Needle valves SNV 1, SNV 2
• Ball valves SKG 1, SKG 2
• Dirt traps SF , SF 1
• Protection relay MSR 1
• M12 Plug connector PVC-cable SM12
Notes:
The specified measuring/switching ranges apply when the
instrument is installed vertically and the flow rate is from bottom
to top.
Other installation positions or operating densities deviating from
the specified specifications increase the specified measuring
error.
Special scales for different media and operating conditions are
available on request.
The specified switching points are shut-off points at falling flow
rates. Please note that the switch-on points are higher due to
the hysteresis.
For applications where pressure surges are to be expected,
please contact PKP!
PKP Prozessmesstechnik GmbH
Borsigstr. 24 • D-65205 Wiesbaden
S +49 (0) 6122-7055-0 • T +49 (0) 6122 7055-50
[email protected] • www.pkp.de
This manual suits for next models
3