Pyxis SP-350 User manual

Table of Contents
1 Introducon 2
1.1 MainFeatures......................................... 2
2 Specicaons 2
3 Unpacking Instrument 2
3.1 StandardAccessories ..................................... 2
3.2 OponalAccessories ..................................... 3
4 Installaon 4
4.1 BaeryInstallaon ...................................... 4
5 Instrument Overview 5
5.1 ControlKeys.......................................... 5
5.2 MainModuleOn/O ..................................... 5
6 Measurement 5
6.1 PTSAMeasurement...................................... 5
6.2 High Color and Turbidity Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
7 Calibraon 6
7.1 PTSA Calibraon (Two-Point with Zero) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
8 Device Informaon and Diagnosis 7
8.1 Sample Cell Cleanliness Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
8.2 Bluetooth Connecon to Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
8.3 FactoryReset ......................................... 12
9 Use with uPyxis® Mobile App 13
9.1 Download uPyxis® MobileApp................................ 13
9.2 Connecng to uPyxis® MobileApp.............................. 14
9.3 SengScreen......................................... 15
9.4 SystemScreen......................................... 16
10 Use with uPyxis® Desktop App 17
10.1 Install uPyxis® DesktopApp.................................. 17
10.2 Connecng to uPyxis® DesktopApp ............................. 18
10.3SystemScreen......................................... 19
10.4DatalogScreen ........................................ 20
10.5SengScreen......................................... 20
11 Device Maintenance and Precauon 21
11.1 Maintenance Best Pracces and Quick Tips . . . . . . . . . . . . . . . . . . . . . . . . . 21
11.2 Methods to Cleaning the SP-350 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
11.3Storage ............................................ 22
12 Contact Us 22

Warranty Informaon
Condenality
The informaon contained in this manual may be condenal and proprietary and is the property of Pyxis
Lab, Inc. Informaon disclosed herein shall not be used to manufacture, construct, or otherwise reproduce
the goods described. Informaon disclosed herein shall not be disclosed to others or made public in any
manner without the express wrien consent of Pyxis Lab, Inc.
Standard Limited Warranty
Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its opon, repair
or replace instrument components that prove to be defecve with new or remanufactured components
(i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether wrien or oral,
is expressed or implied.
Warranty Term
The Pyxis warranty term is thirteen (13) months ex-works. In no event shall the standard limited warranty
coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service
Damaged or dysfunconal instruments may be returned to Pyxis for repair or replacement. In some in-
stances, replacement instruments may be available for short duraon loan or lease.
Pyxis warrants that any labor services provided shall conform to the reasonable standards of technical com-
petency and performance eecve at the me of delivery. All service intervenons are to be reviewed and
authorized as correct and complete at the compleon of the service by a customer representave, or des-
ignate. Pyxis warrants these services for 30 days aer the authorizaon and will correct any qualifying
deciency in labor provided that the labor service deciency is exactly related to the originang event. No
other remedy, other than the provision of labor services, may be applicable.
Repair components (parts and materials), but not consumables, provided during a repair, or purchased
individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the in-
corporaon of a warranted repair component into an instrument extend the whole instrument’s warranty
beyond its original term.
Warranty Shipping
A Repair Authorizaon (RA) Number must be obtained from Pyxis Technical Support before any product can
be returned to the factory. Pyxis will pay freight charges to ship replacement or repaired products to the
customer. The customer shall pay freight charges for returning products to Pyxis. Any product returned to
the factory without an RA number will be returned to the customer. To receive an RMA you can generate a
request on our website at hps://pyxis-lab.com/request-tech-support/.
Pyxis Technical Support
Contact Pyxis Technical Support at +1 (866) 203-8397, [email protected], or by lling out a request for
support at hps://pyxis-lab.com/request-tech-support/.

1 Introducon
The Pyxis SP-350 is a uniquely designed handheld mulmeter that measures PTSA. It is a cuvee-less device.
Less than 5 mL water sample is needed to ll the sample cell for proper measurement.
1.1 Main Features
The SP-350 includes the following features:
• Measurement of the concentraon of uorescent tracer PTSA of a water sample.
• Pre-calibrated for measuring PTSA (pyrenetetrasulfonic acid) in the range of 0 to 300 ppb.
• The uorescence PTSA measurement is automacally compensated for sample color and turbidity
interference up to 60 NTU without preltraon.
• PTSA uorescence can be calibrated using a single standard through a user-friendly, menu-driven
procedure.
• Large color graphic screen that can be read in direct sunlight.
2 Specicaons
Table 1. SP-350 Specicaons
Item Specicaon*
P/N 50206
PTSA Range 0–300 ppb
PTSA Precision ±1% or ±1 ppb
Temperature Range 40–106 °F (4–41 °C)
Display 320×240 TFT-LCD, visible under direct sunlight
Power Supply 4 AA alkaline baeries
Typical Baery Life 10,000 readings
Dimension (L ×W×H) 6.30 ×2.91 ×1.30 inch (160 ×74 ×33 mm)
Weight†0.68 lbs (310 g)
Enclosure Rang IP67
*With Pyxis’s connuous improvement policy, these specicaons are subject to
change without noce.
†Baeries excluded
3 Unpacking Instrument
Remove the instrument and accessories from the shipping container and inspect each item for any damage
that may have occurred during shipment. Verify that all items listed on the packing slip are included. If any
3.1 Standard Accessories
• Four (4) AA alkaline baeries
• Bluetooth/USB Adapter for Desktop P/N: MA-NEB
• User Manual available online at hps://pyxis-lab.com/support/

3.2 Oponal Accessories
The following oponal accessories can be ordered from Pyxis Customer Service ([email protected]om) or
Pyxis E-Store at hps://pyxis-lab.com/shop/.
Table 2.
Accessory Name P/N
Pyxis Carrying Case for SP-350 50725
Pyxis 100 ppb PTSA Calibraon Standard — 500 mL 21001
Pyxis 200 ppb PTSA Calibraon Standard — 500 mL 21000
Pyxis 300 ppb PTSA Calibraon Standard — 500 mL 21003
Pyxis Handheld Cleaning Kit SER-02

4 Installaon
4.1 Baery Installaon
The SP-350 is powered by four alkaline baeries. Typical baery life lasts for 10,000 measurements or 10
months. When the baery capacity is crically low, the SP-350 will display a ”LOW BATTERY” warning for 5
seconds and then automacally turn o.
*NOTE*Do not use rechargeable nickel cadmium (NiCad) or lithium baeries.
Replace the baeries to resume operaon of the SP-350 aer the baery warning. The SP-350 will auto-
macally turn on in the measurement mode aer new baeries are installaon.
The SP-350 baery compartment, shown in Figure 1, is on the back side of the instrument. Baeries are
held in place by a cover secured with two Phillips-head screws.
Figure 1. The SP-350 baery compartment
Install baeries using the following steps:
1. Remove the baery compartment cover by loosening the two screws.
2. Remove old baeries and dispose of them properly.
3. Following the posive and negave terminal signs in the compartment boom, snap four new AA
alkaline baeries rmly into the baery holder.
4. Replace the baery compartment cover and ensure that the sealing O-ring is lying at on the baery
holder.
5. Fasten the two screws.
*NOTE*Failure to properly seat the O-ring may result in water damage to the SP-350.

5 Instrument Overview
5.1 Control Keys
The SP-350 has three control keys, as shown in Figure 2. The le ( ), right ( ), and ok ( ) keys
are used to launch acons indicated on the LCD display directly above the keys. The labels above the keys
indicate the funcon associated with each key and funcons can be changed in dierent operaon modes.
Figure 2.
5.2 Main Module On/O
To turn on the SP-350: Press momentarily and release.
To turn o the SP-350: Press and hold for about three seconds. Release when the LCD display
turns o. The SP-350 turns itself o aer 60 seconds without user interacon detected. This is done to
conserve baery life.
*NOTE*This auto-me o seng may be customized by the user as desired through the
uPyxis® Mobile or Desktop App.
6 Measurement
6.1 PTSA Measurement
When powered on, the SP-350 will default to the PTSA measurement mode. The water sample can be
transferred to the sample cell using a pipee or lled directly from a faucet, sample bole, or sample
valve.
*NOTE*Special care should be taken when pouring the sample into the cell to avoid air
bubble entrainment, which can interfere with reading accuracy.
Before beginning a measurement, use the sample water to rinse the sample cell at least three mes. Allow
5–10 seconds for the SP-350 to stabilize. The values will be displayed in white with a blue background if
a stable value is reached (Figure 2). For a sample containing 100 ppb PTSA, the measured PTSA should be
stabilized within the range of 98–102 ppb.

*NOTE*The me required to reach a stable reading may be slightly longer if the water
sample temperature is signicantly dierent than the environmental temperature at which
the SP-350 had been equilibrated (stored).
6.2 High Color and Turbidity Warning
The SP-350 has extra channels to measure sample turbidity and color to automacally compensate sam-
ple color and turbidity interference. If sample turbidity and color values determined are too high, a PTSA
measurement warning will be displayed. In such a case, the user should lter the sample for PTSA mea-
surement.
7 Calibraon
7.1 PTSA Calibraon (Two-Point with Zero)
1. Rinse the sample cell three mes with DI water. Fill the sample cell with DI water.
*NOTE*In emergency, “non-PTSA” water, such as city water, may be used, but re-
calibrate using DI water for the zero step as soon as it is available.
2. Power on the SP-350 by pressing . Allow 5–10 seconds for the SP-350 to stabilize.
3. The unit is acvely reading and displaying PTSA. The values will be very low if DI water is used; PTSA
value should be near zero. A low non-zero value (e.g. 0.2 or 0.4, etc.) is not problemac.
4. Press P-Cal ( or ) to launch the PTSA CALIBRATION screen (Figure 3).
5. Press Zero ( ) to start the zero (blank) calibraon.
6. If the calibraon succeeds, a checkmark ( ) and instrucons for the slope calibraon will appear
(Figure 4).
7. Rinse the sample cell three mes with the desired PTSA standard. Fill the sample cell with the desired
PTSA standard.
8. Press Cycle ( ) to cycle between the PTSA standards 100, 200, and 300 ppb (it repeats). Ensure
the value selected matches the desired PTSA standard in the sample cell.
9. Press Slope ( ) to start the slope calibraon.
10. If the calibraon succeeds, a checkmark ( ), a ”Calibraon Success” message will appear (Figure
5). Otherwise, a warning message is displayed.
11. Calibraon is now complete. Press Exit ( ) to return to measurement mode.
*NOTE*If Exit is pressed before the second checkmark appears, the calibraon will
not be completed and must be re-done.

Figure 3. Figure 4. Figure 5.
8 Device Informaon and Diagnosis
The DEVICE INFORMATION screen is launched when System ( ) is pressed in the measurement mode.
This screen contains the device serial number, soware version, and hardware version (Figure 6). The bat-
tery life as a percentage and the MAC addresses for main module also shown.
Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen where raw measurement data are displayed
(Figure 7). The informaon has no use for normal operaon, but instead is used for device troubleshoong.
Provide an image of both the DEVICE INFORMATION screen and the SYSTEM DIAGNOSIS screen when you
Figure 6. Figure 7.

8.1 Sample Cell Cleanliness Check
The SP-350 is designed to provide reliable and accurate measurement on PTSA. Heavy fouling will pre-
vent the light from reaching the sensor, resulng in inaccurate readings. It is suggested that the SP-350
be checked for fouling and cleaned on a monthly basis. Heavily contaminated waters may require more
frequent cleanings. Cleaner water sources with less contaminaon may not require cleaning for several
months. The SP-350 is designed to carry out a Cleanliness Check as described below:
1. Power on the SP-350 by pressing .
2. Press System ( ) to launch the DEVICE INFORMATION screen.
3. Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starng BTLE...
to BTLE Started
5. Press Cleanliness ( ). An instrucon prompt appears to ask the user to put DI water into the
sample cell (Figure 8).
6. Pour DI water into the sample cell.
7. Press Conrm ( , , or ). The instrucon prompt will disappear and the SP-350 displays a
countdown toward the boom of the display.
8. Once the Cleanliness Check is completed a Clean message (Figure 9) or Sample cell fouled message
(Figure 10) will appear towards the boom of the display.
9. Cleanliness check is now complete. Press Exit ( ) to return to measurement mode.
Figure 8. Figure 9. Figure 10.

8.2 Bluetooth Connecon to Devices
The SP-350 uses a built-in Bluetooth Low Energy Connecon (BTLE) to connect wirelessly to a smart phone
via the uPyxis® Mobile App or to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the
uPyxis® Desktop App. To allow the SP-350 to connect via Bluetooth with other devices follow the steps
below:
1. Power on the SP-350 by pressing .
2. Press System ( ) to launch the DEVICE INFORMATION screen.
3. Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starng BTLE...
to BTLE Started (Figure 7).
5. Choose to connect via one of two opons:
(a) The uPyxis® Mobile App (see the Use with uPyxis® Mobile App secon), or
(b) The uPyxis® Desktop App (see the Use with uPyxis® Desktop App secon).

8.2.1 Calibrate an ST-500 Series Sensor with the SP-350 via Bluetooth
The SP-350 can be used to verify the result of an inline Pyxis ST-500 Series sensor by measuring the sample
water taken from the inline sensor sample line. The SP-350 can then be used to calibrate the inline sensors
over the Bluetooth connecon. To calibrate an inline sensor, follow the steps below:
1. Power on the SP-350 by pressing .
2. Press System ( ) to launch the DEVICE INFORMATION screen.
3. Press Comm ( ) to launch the COMMUNICATION screen (Figure 11).
4. Press Scan ( ) to begin scanning for Bluetooth devices.
5. Discoverable devices will begin to populate on the display with their name and MAC-Address (Figure
12).
6. If more than one device appears in the Device list, press »( ) to cycle through the devices.
7. If no devices or the incorrect device appear in the Device list, press Scan ( ) to re-scan for discov-
erable devices.
8. Press Connect ( ) to begin pairing to the selected sensor.
9. When the connecon is established, the SP-350 displays the latest PTSA measurement from the con-
nected sensor (Figure 13).
10. Fill the main module sample cell with the same sample water that the sensor is measuring.
11. Press Read ( ) to see PTSA measurement from the SP-350along with the sensor measurement
(Figure 14).
12. Press Calib ( ) to begin sensor PTSA calibraon.
13. The SP-350 will take the sensor PTSA measurement three mes to verify the calibraon (Figure 15).
*NOTE*It takes about one minute for the sensor to approach the calibrated reading
and the three verifying readings may not be exactly the same as the value measured
by the SP-350Press Read ( ) again to take more readings from the sensor, if nec-
essary.
14. If the calibraon is successful, a ”Verify calibraon - OK” message will appear on the top of the display
(Figure 16).
15. Calibraon is now complete. Long press Calib ( ) to return to measurement mode.

8.3 Factory Reset
Use the following steps to restore all device parameters to factory default:
1. Power on the SP-350 by pressing .
2. Press System ( ) to launch the DEVICE INFORMATION screen.
3. Press Diagnosis ( or ) to launch the SYSTEM DIAGNOSIS screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starng BTLE...
to BTLE Started .
5. Press Help ( ) to launch the HELP screen (Figure 17).
6. Press Factory Reset ( or ). The display updates as shown in Figure 18 appear and the user
can choose one of three opons:
(a) Press to start the factory reset, or
(b) Press Cancel ( ) to return to the HELP screen, or
(c) Press Exit ( ) to abandon the factory reset enrely.
7. Aer a successful factory reset, the message “Factory reset done.” will appear on the display.
8. Press Exit ( ) to return to measurement mode.
Figure 17. Figure 18.

9.2 Connecng to uPyxis® Mobile App
Connect the SP-350 sensor to a mobile smart phone according to the following steps:
1. Follow the steps in the Bluetooth Connecon to Devices secon to make the SP-350 discoverable.
2. Open uPyxis® Mobile App.
3. On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices.
4. If the connecon is successful, the SP-350 and its Serial Number (SN) will be displayed (Figure 20).
5. Press on the SP-350 image.
Figure 20.

9.4 System Screen
From the System screen, users can change the Device Name, nd the Serial Number,Hardware Version,
and Firmware Version, as well as update the rmware of the SP-350 by pressing Check Update. If a rmware
update is available, press Get Firmware. Once the new rmware is downloaded, press Upgrade Firmware.
*NOTE*The rmware update process takes some me and will require the SP-350 to stay
within range (approximately 10 without obstrucons) for the enre duraon of the up-
date.
Once the update is complete, the SP-350 will reboot which will disconnect the SP-350 from the uPyxis®
Mobile App.
Figure 22.

10 Use with uPyxis® Desktop App
[uPyxis® Desktop App installaon]
10.1 Install uPyxis® Desktop App
Download the latest version of uPyxis® Desktop soware package from: hps://pyxis-lab.com/upyxis/ this
setup package will download and install the Microso.Net Framework 4.5 (if not previously installed on the
PC), the USB driver for the USB-Bluetooth adapter (MA-NEB), the USB-RS485 adapter (MA-485), and the
main uPyxis® Desktop applicaon. Double click the uPyxis.Setup.exe le to install.
Figure 23. uPyxis® Desktop App installaon
Click Install to start the installaon process. Follow the screen instrucons to complete the USB driver and
uPyxis installaon.
Other manuals for SP-350
1
This manual suits for next models
1
Table of contents
Other Pyxis Measuring Instrument manuals

Pyxis
Pyxis ST-500 User manual

Pyxis
Pyxis PRISM RT-200 User manual

Pyxis
Pyxis SP-910 User manual

Pyxis
Pyxis SP-350RO User manual

Pyxis
Pyxis ST-590 Series User manual

Pyxis
Pyxis SP-350P User manual

Pyxis
Pyxis ST-525SS User manual

Pyxis
Pyxis SP-900 User manual

Pyxis
Pyxis SP-380P User manual

Pyxis
Pyxis SP-910 User manual

Pyxis
Pyxis IK-765SS-O3 User manual

Pyxis
Pyxis ST-774 User manual

Pyxis
Pyxis SP-380P User manual

Pyxis
Pyxis SP-350 User manual

Pyxis
Pyxis ST-540SS-HR User manual

Pyxis
Pyxis ST-601 User manual

Pyxis
Pyxis SP-400 User manual

Pyxis
Pyxis CR-301 User manual

Pyxis
Pyxis SP-200 OXIPOCKET User manual

Pyxis
Pyxis SP-395 User manual