
8 OPAL • OPAL Plus SATEL
Pressing buttons and simultaneously restores default settings for the configured
sensor. Keeping the buttons pressed for 3 seconds restores default settings for all
sensors.
Pressing MODE button briefly allows starting configuring of another sensor. Blinking of the
proper LED will indicate which sensor is being configured (see: “Signaling in the configuring
mode”).
Configuring by means of OPT-1 keyfob [OPAL Plus]
Point the keyfob towards the detector and by pressing buttons (value decreasing) and
(value increasing) set the sensitivity / detection threshold of the sensor.
Pressing the button restores default settings for the configured sensor.
By pressing (the next sensor) or (the previous sensor) go to another sensor configuring.
Blinking of the proper LED will indicate which sensor is being configured (see: “Signaling in
the configuring mode”).
End of configuring mode
Press MODE button on the electronics board for 2 seconds or remove the common ground
from the SVCE terminal.
Note: If the configuring mode was activated by means of the MODE button, it is
automatically exited after 20 minutes since the last operation performed by the user.
6. Start-up and walk test
1. Power-up the detector. All LEDs will begin alternately blinking for 40 seconds, which
indicates the detector warm-up.
2. When the LEDs stop blinking, check that moving within the coverage area will activate the
alarm relay and make the LED light up red.
Separate testing of sensors
Testing of the sensors is done in the detector configuring mode. Starting the mode, choosing
the sensor and the way of changing sensitivity are described in detail in the chapter “Detector
configuring”.
1. Start the configuring mode.
2. Choose the sensor to be tested.
3. Check that moving in the coverage area will activate the alarm relay and make the
violation signaling LED light up.
4. If it is needed, readjust sensitivity and check the sensor operating.
7. OPT-1 keyfob
The keyfob is CR2032 3V lithium battery powered. Battery life depends entirely on the way
the keyfob is used. The battery lasts for at least 8 thousand button pressings.
The keyfob LED is ON if any button is pressed. If the battery is low, the LED blinks instead of
being ON. You should replace the battery then. The way of battery replacing is shown in
Figure 11. If the battery is low, the keyfob range decreases.