manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Smiths Medical
  6. •
  7. Medical Equipment
  8. •
  9. Smiths Medical CADD-legacy 1 User manual

Smiths Medical CADD-legacy 1 User manual

ALARMS & TROUBLESHOOTING CONTINUED
The CADD-Legacy®1 pump signals with audible alarms and messages
that appear on the display when a condition requires your attention.
Upstream Occlusion
Two-tone alarm
Fluid is not flowing from the reservoir to
the pump. Check for a kink or air bubble in
the tubing between the reservoir and pump.
Remove the occlusion to resume operation.
Or press STOP/START or NEXT to stop the
pump and silence alarm for two minutes, then
remove the occlusion and restart the pump.
The cassette is not aligned with the pump,
or is damaged, or a malfunction of the
pump sensor(s) is occurring. Reposition the
pump to silence the alarm. If repositioning
the pump does not silence the alarm within
two minutes, the pump will display “No
Disposable, Clamp Tubing.”
Air in fluid path. Press NEXT or STOP/
START to silence alarm. Make sure tubing is
threaded in air detector properly. If the fluid
path contains air, close clamps, disconnect
from patient and prime fluid path.
Pump is stopped. Start pump, if appropriate.
A value was not saved by pressing ENTER/
CLEAR. Press NEXT to resume programming.
An error has occurred. Remove pump
from service.
The pump was running when batteries
were removed or the batteries were
removed within 15 seconds after stopping
the pump. Batteries must be reinstalled.
IMPORTANT: Always stop the pump
before removing batteries.
The disposable was removed, or the
cassette is not aligned with the pump or
is damaged, or a malfunction of the pump
sensor(s) is occurring. Clamp the tubing
immediately. Press STOP/START or NEXT
to silence alarm.
The pump is due for service.
Air In Line Detected
Two-tone alarm
Stopped
Three beeps every
five minutes
Value Not Saved
Error
Two-tone alarm
Screen Displays
Current Pump Status
Two-beep
(long-short) alarm
No Message Displayed
Two-tone alarm
No Disposable,
Clamp Tubing
Two-beep
(long-short) alarm
Service Due
Two-tone alarm
CADD, CADD-Legacy, the Medication Cassette Reservoir design and the Smiths Medical
and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates
the trademark is registered in the U.S. Patent and Trademark Office and certain other
countries. ©2010 Smiths Medical. All rights reserved. 11/10 IN19872
Smiths Medical ASD, Inc.
St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com
Smiths Medical is part of the global technology business Smiths Group plc.
EC Authorized Representative
Smiths Medical International Ltd.
TN25 4BF, UK
Phone: +44 (0) 1233 722100
Product(s) described may not be licensed or available for sale in Canada or other
countries outside of the United States.
Smiths Medical Canada Ltd.
Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346
Warning: This Quick Reference Card should be used by clinicians
only. Do not permit patients to have access to this card, as the
information would allow access to all programming and operating
functions. Assistance with the CADD-Legacy®1 pump is available
to clinicians 24-hours-a-day by calling 1-800-258-5361.
Level of fluid in the reservoir is low.
Prepare to install new reservoir.
Reservoir volume has reached 0.0 ml. Press
NEXT or STOP/START to silence alarm.
Install a new reservoir, if appropriate.
Batteries are low, but pump is operable.
Change batteries soon.
Batteries are depleted and cannot operate
pump; install new batteries immediately.
With AC adapter attached, batteries have
been removed while pump was running,
or you tried to start pump with depleted
batteries. Install new batteries.
High pressure is being caused by an
occlusion in the fluid path between
the pump and the patient. Remove the
occlusion to resume operation. Or press
STOP/START or NEXT to stop the pump and
silence the alarm for two minutes, then
remove occlusion and restart pump.
ALARMS & TROUBLESHOOTING
The CADD-Legacy®1 pump signals with audible alarms and message
that appear on the display when a condition requires your attention.
CHANGING THE CASSETTE
To Remove Used
Cassette
To Attach New
Cassette
Res Vol Low
Three beeps
Reservoir Volume
Empty
Two-tone alarm
Low Bat
Three two-tone beeps
every five minutes
Battery Depleted
Two-tone alarm
Battery Removed
Pump Won’t Run
Two-tone alarm
High Pressure
Two-tone alarm
1. Stop the pump.
2. Close all tubing clamps.
3. Disconnect tubing from patient.
4. Use coin to unlatch used cassette.
5. Remove and discard used cassette.
1. Close all tubing clamps.
2. Insert new cassette hooks into pump’s hinge pins.
3. Place pump with cassette on firm surface and
push down on top of pump so cassette fits
tightly against the pump.
4. Insert a coin into the lock, push in and turn
until the line on the lock lines up with the
arrow on the side of the pump.
5. Gently twist and pull on cassette to make sure
it is attached properly.
6. Press ENTER/CLEAR to reset reservoir volume,
if appropriate.
7. If the Air Detector is in use, make a small loop
of tubing underneath the air detector and hold
it with your thumb. Place the tubing over the
groove in the air detector and tuck it under the
catch. Seat the tubing into the groove by gently
pulling the tube upward until it is under the
retention nubs and flat in the groove.
WARNING: Attach the cassette (the part of the
Medication Cassette reservoir or CADD®
administration set that attaches to the pump)
properly. An improperly attached or detached
cassette could result in unregulated gravity
infusion of medication from the fluid container
or a reflux of blood. If you are using a CADD®
administration set or Medication Cassette
reservoir that does not have the flow stop
feature (catalog number does not start with
21-73xx): You must use a CADD®extension set
with an integral anti-siphon valve or a CADD®
administration set with either and integral
or add on anti-siphon valve to protect against
unregulated gravity infusion that can result
from an improperly attached cassette.
®
®
A
B
C
D
E
F
G
H
I
Rate is in
ml/24 hrs
CADD-Legacy®1
Infusion Pump
Model 6400
Quick Reference Card for Clinicians
Lock Level 0 (LL0)
´
Î
ADisplay shows programmed values
and messages
BAC Adapter Jack used to plug in
AC Adapter
CAccessory Jack (not currently in use)
DIndicator Light indicates AC power in use
EAir Detector
FCassette (part of reservoir or administration set
that attaches to pump)
GThreaded Mounting Hole for use with Polemount
Bracket Adapter (back of pump)
HBattery Compartment (back of pump)
ICassette Latch attaches cassette to pump (side of pump)
KEYPAD
STOP/START Stops and starts the infusion
ENTER/CLEAR Enters or clears displayed value
PRIME Fills tubing with fluid
LOCK Displays or changes Lock Level (security level)
NEXT Advances to next programming screen
Increases or decreases displayed values
or scrolls through menu items
ON/OFF Turns the pump on or off (low power)
Read the entire Operator’s Manual before operating the CADD-Legacy®
ambulatory infusion pump. Failure to properly follow warnings, cautions
and instructions could result in death or serious injury to the patient.
IN19872.indd 1 10/27/11 2:53 PM
PUMP OPERATIONS
1. Stop the pump.
2. Press LOCK.
3. Press ´orÎuntil desired lock level appears.
4. Press LOCK or ENTER/CLEAR.
5. Press ´orÎuntil the lock level code appears.
6. Press LOCK or ENTER/CLEAR.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) appears on the display.
2. Release STOP/START key. STOPPED
will appear on the display when the
pump is stopped.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) disappears from
the display.
2. Release STOP/START key. RUN will appear on the
display when the pump is running.
Pump must be stopped and in LL0 or LL1.
WARNING: Do not prime the fluid path with the
tubing connected to a patient as this could result
in over delivery of medication or air embolism.
1. Press and hold PRIME until the word PRIME
appears on the display, along with
(– – – – – – – – – – – –).
2. Release the PRIME key.
3. Press and hold PRIME. Priming appears on the
display. Continue priming until the fluid path is
free of air.
4. Press NEXT to return to the main screen.
1. Stop the pump.
2. Press NEXT to display the Reservoir Volume screen.
3. Press ENTER/CLEAR to reset the value to
previously programmed amount.
1. Press and hold ON/OFF until pump beeps and
powers up.
1. PressandholdON/OFFuntil(••••••••••••)
appears on the display.
2. Release ON/OFF key.
3. The screen will go blank as the pump goes into a
lower power state.
1. Stop the pump.
2. Push down and hold the arrow button on the
battery door while sliding the door off. Remove
and discard old batteries.
IMPORTANT: Always stop pump before
removing batteries.
3. Install new batteries, matching polarities shown
on pump. Replace battery door and close.
4. Start the pump.
WARNING: If a gap is present anywhere between
the battery door and the pump housing, the
door is not properly latched. If the battery door
becomes detached or loose, the batteries will not
be properly secured which could result in loss of
power or non-delivery of medication.
WARNING: Do not use rechargeable NiCad or
nickel metal hydride (NiMH) batteries. Do not use
carbon zinc (“heavy duty”) batteries.
Change the
Lock Level
Stop the
Pump
Start the
Pump
Prime the
Fluid Path
Reset the
Reservoir
Volume
Turn the
Pump On
Turn the
Pump Off
Change the
Batteries
PUMP OPERATIONS CONTINUED
Pump must be stopped and in LL0.
1. Press LOCK.
2. Press LOCK or ENTER/CLEAR.
3. Press
´orÎuntil the biomed code appears.
4. Press LOCK or ENTER/CLEAR.
5. Press NEXT to go through settings.
6. Press NEXT until AIR DETECTOR appears.
7. Press
´orÎuntil desired setting appears.
8. Press ENTER/CLEAR.
9. Press NEXT.
WARNING: When the air detector is turned off,
the pump will not detect air in the fluid path.
Periodically inspect the fluid path and remove
any air to prevent air embolism.
10. Press NEXT until UPSTREAM SENSOR appears.
11. Press ´orÎuntil desired setting appears.
12. Press ENTER/CLEAR.
13. Press NEXT.
WARNING: When the upstream occlusion sensor
is turned off, the pump will not detect occlusions
upstream (between pump and reservoir).
Periodically inspect the reservoir for decreasing
reservoir volume and inspect the fluid path
for kinks, a closed clamp, or other upstream
occlusions. Upstream occlusions could result in
under- or non-delivery of medications.
14. Press NEXT until NEXT FOR BIOMED,
ENTER FOR MAIN appears.
15.Press ENTER/CLEAR to return to main
screen. Or press NEXT to continue in Biomed
settings loop.
Main Screen Displays status of pump (STOPPED or RUN)
Reservoir Volume Volume of fluid in reservoir
Continuous Rate Continuous rate of infusion in ml/24 hrs
Given Total amount of medication delivered since
display was last cleared
Air Detector Indicates if air detector is turned on or off
(status only) and high or low sensitivity level
Upstream Sensor Indicates if turned on or off
(status only)
PROGRAMMING SCREENS
To Access
To Turn Air
Detector
ON-HIGH,
ON-LOW
or OFF
To Turn
Upstream
Sensor
ON/OFF
To Exit Biomed
Functions
BIOMED
FUNCTIONS
PROGRAMMING / SET-UP SEQUENCE
Batteries must be installed; pump must be stopped and in LL0.
NOTE: Value Not Saved is displayed if a value is scrolled and ENTER/
CLEAR is not pressed. Press NEXT to continue programming.
WARNING: If the pump is dropped or hit, inspect it for damage. Do
not use a pump that is damaged or is not functioning properly.
Enter Reservoir
Volume
Enter Continuous
Rate in ml/24 hrs
Clear Milliliters
Given
Verify Air Detector
Status
Verify Upstream
Sensor Status
Verify Programming
Main Screen Press NEXT.
1. Press ´orÎto select desired reservoir
volume.
2. Press ENTER/CLEAR.
3. Press NEXT.
4. Press ´orÎto select desired
Continuous Rate in ml/24 hours.
5. Press ENTER/CLEAR.
6. Press NEXT.
WARNING: This pump delivers medication
at a continuous rate in milliliters per
24 hours (ml/24 hrs). Programming
the pump at a delivery rate other than
what is prescribed will cause over- or
under-delivery of medication.
7. Press ENTER/CLEAR to clear the given
value. The display will show 0.00.
8. Press NEXT.
9. Verify the setting is correct. (To change
setting, see Biomed Functions section.)
10. Press NEXT.
11. Verify the setting is correct. (To change
setting, see Biomed Functions section.)
12. Press NEXT.
13. Press NEXT repeatedly to review program.
IN19872.indd 2 10/27/11 2:53 PM

Other manuals for CADD-legacy 1

2

This manual suits for next models

1

Other Smiths Medical Medical Equipment manuals

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical CADD-legacy Plus 6500 Reference guide

Smiths Medical

Smiths Medical CADD-legacy Plus 6500 Reference guide

Smiths Medical cadd-solis VIP 2120 Technical manual

Smiths Medical

Smiths Medical cadd-solis VIP 2120 Technical manual

Smiths Medical CADD-Solis VIP User manual

Smiths Medical

Smiths Medical CADD-Solis VIP User manual

Smiths Medical CADD-Legacy PLUS Pump User manual

Smiths Medical

Smiths Medical CADD-Legacy PLUS Pump User manual

Smiths Medical 1884SS User manual

Smiths Medical

Smiths Medical 1884SS User manual

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Solis VIP User manual

Smiths Medical

Smiths Medical CADD-Solis VIP User manual

Smiths Medical CADD-Solis VIP User manual

Smiths Medical

Smiths Medical CADD-Solis VIP User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Solis Reference guide

Smiths Medical

Smiths Medical CADD-Solis Reference guide

Smiths Medical Graseby 1200 User manual

Smiths Medical

Smiths Medical Graseby 1200 User manual

Smiths Medical SurgiVet AES User manual

Smiths Medical

Smiths Medical SurgiVet AES User manual

Smiths Medical Portex Blue Line Ultra User manual

Smiths Medical

Smiths Medical Portex Blue Line Ultra User manual

Smiths Medical CADD User manual

Smiths Medical

Smiths Medical CADD User manual

Smiths Medical SurgiVet V3404 Plus User manual

Smiths Medical

Smiths Medical SurgiVet V3404 Plus User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical level 1 HOTLINE HL-90 User manual

Smiths Medical

Smiths Medical level 1 HOTLINE HL-90 User manual

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical CADD-MS 3 User manual

Smiths Medical

Smiths Medical CADD-MS 3 User manual

Popular Medical Equipment manuals by other brands

Hospira Symbiq System Operating Manual

Hospira

Hospira Symbiq System Operating Manual

Chattanooga LIGHTFORCE XLi Instructions for use

Chattanooga

Chattanooga LIGHTFORCE XLi Instructions for use

Gima DENVER user manual

Gima

Gima DENVER user manual

Omron CFX-WR17 Instructions for use

Omron

Omron CFX-WR17 Instructions for use

Dermlite DL3N instructions

Dermlite

Dermlite DL3N instructions

Braemar Fusion MCT user manual

Braemar

Braemar Fusion MCT user manual

Levabo Seat All Flexi quick start guide

Levabo

Levabo Seat All Flexi quick start guide

Cantel SCOPE BUDDY PLUS manual

Cantel

Cantel SCOPE BUDDY PLUS manual

Quasar Biotech MD BLUE instruction manual

Quasar Biotech

Quasar Biotech MD BLUE instruction manual

Datex-Ohmeda S/5 Avance Technical reference manual

Datex-Ohmeda

Datex-Ohmeda S/5 Avance Technical reference manual

Stryker LUCAS 3 Standard operating procedure

Stryker

Stryker LUCAS 3 Standard operating procedure

Dräger Infinity C700 Quick reference guide

Dräger

Dräger Infinity C700 Quick reference guide

human care SHP600 user manual

human care

human care SHP600 user manual

Propolair A4 manual

Propolair

Propolair A4 manual

Hanger ACP Synchrony user manual

Hanger

Hanger ACP Synchrony user manual

Medela Thopaz Instructions for use

Medela

Medela Thopaz Instructions for use

Gima 34056 Use and maintenance book

Gima

Gima 34056 Use and maintenance book

Lightmed LightLink-CXL Service manual

Lightmed

Lightmed LightLink-CXL Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.