Water Level Meter Faceplate with Electronics Replacement
High Quality Groundwater and Surface Water Monitoring Instrumentation
(Page 1 of 2)
Model 101/102
Instructions
1. Place the reel on a flat surface, with the faceplate up.
Remove the battery from the Meter.
Tools and Materials Needed
1. One of the following Replacement Mk2 Faceplates
with Electronics
• #112765 - 101 P7 SC1000 Faceplate w Electronics
• #112759 - 101 P7 SC2000 Faceplate w Electronics
• #112670 - 101 P2 SC1000 Faceplate w Electronics
• #112656 - 101 P2 SC2000 Faceplate w Electronics
• #112020 - 102 SC1000 Faceplate w Electronics
• #111921 - 102M SC100 Faceplate w Electronics
2. Phillips or Robertson Screwdriver
3. Wire Cutters (only required for Mk1 Water Level Meters –
see photo below)
4. Pliers (only required for Mk1 Water Level Meters –
see photo below))
Back of Mk2 101 Water Level Meter Faceplate
Push-Release Fittings
Back of Mk1 101 Water Level Meter Faceplate
Molex
Connector
4. To connect the new faceplate with electronics, connect the
tape/cable to the new circuit board assembly:
a) If this is a Mk2 Water Level Meter, press down
on the white terminals on the circuit board and insert
the tape/cable leads. Release the terminals and the
leads should be secured.
For 101 Water Level Meter, the lead on the bottom of
the tape (numbers facing up) is inserted into the
terminal with a white square below it on the circuit
board (see next page for images).
For 102 Water Level Meters, the positive lead is
inserted into the terminal with a square etched
below it on the circuit board. The positive lead has
the pin connected to insulated wire, the negative pin is
connected to the braided wire.
b) If this is a Mk1 Water Level Meter, with a Molex
connector on the tape/cable leads, remove the Molex
connector by pushing out the two pins. Cut each pin
in half (see image on next page for correct location).
Use pliers to flatten the remainder of the pins against
the tape/cable lead, so it fits easily into the terminals
on the circuit board. Connect each tape/cable lead to
the new circuit board assembly, as described above in
step a.
Remove the Pins from the Tape Molex Connector (Mk1)
Continued overleaf...
Molex Connector
Note: The battery is located inside the reel hub of a 102M Water Level
Meter. Remove it in step 2.
2. Use the screwdriver to undo the three screws holding the
faceplate to the hub. Remove the faceplate.
3. If this is a Mk1 Water Level Meter, disconnect the Molex
connector that connects the circuit board to the tape/cable.
If this is a Mk2 Water Level Meter, press down on the white
terminals of the push-release fittings on the circuit board
and pull out to remove the tape/cable leads.
Printed in Canada
June 5, 2018
(#113261)
(Page 2 of 2)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
Square
Cut the Pins on Each Tape/Cable Lead in Half (as shown)
Press Down on the White Terminals on the Circuit Board and
Insert the Tape/Cable Leads
Release the White Terminals and the Leads Should be
Secured.
5. Install the battery.
6. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If the buzzer or light do
not activate, check all connections and the polarity of the
battery.
7. Reattach the faceplate to the reel using the three screws.

This manual suits for next models

7

Other Solinst Measuring Instrument manuals

Solinst Levelogger 3001 Use and care manual

Solinst

Solinst Levelogger 3001 Use and care manual

Solinst LevelSender 5 User manual

Solinst

Solinst LevelSender 5 User manual

Solinst Levelogger 3001 User manual

Solinst

Solinst Levelogger 3001 User manual

Solinst 101 P2 User manual

Solinst

Solinst 101 P2 User manual

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 101B User manual

Solinst

Solinst 101B User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 201 User manual

Solinst

Solinst 201 User manual

Solinst 101 P7 User manual

Solinst

Solinst 101 P7 User manual

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 101D User manual

Solinst

Solinst 101D User manual

Solinst 122 User manual

Solinst

Solinst 122 User manual

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 9700 SolSat 5 User manual

Solinst

Solinst 9700 SolSat 5 User manual

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 101 P2 Operating and safety instructions

Solinst

Solinst 101 P2 Operating and safety instructions

Solinst 615 ML User manual

Solinst

Solinst 615 ML User manual

Solinst 401 User manual

Solinst

Solinst 401 User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst Levelogger Series User manual

Solinst

Solinst Levelogger Series User manual

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 101 User manual

Solinst

Solinst 101 User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI 96813 instruction manual

Hanna Instruments

Hanna Instruments HI 96813 instruction manual

KALAS AK-1097 user manual

KALAS

KALAS AK-1097 user manual

PEAK INSTRUMENTS E-1000 Series User's manual and maintenance guide

PEAK INSTRUMENTS

PEAK INSTRUMENTS E-1000 Series User's manual and maintenance guide

Unipro UniGo One kit 1 user guide

Unipro

Unipro UniGo One kit 1 user guide

Danfoss SonoMeter 30 installation guide

Danfoss

Danfoss SonoMeter 30 installation guide

Amprobe VPC-10 instruction manual

Amprobe

Amprobe VPC-10 instruction manual

Amprobe MAN02-A user manual

Amprobe

Amprobe MAN02-A user manual

TQC POWDERSAVE II user manual

TQC

TQC POWDERSAVE II user manual

Rohde & Schwarz SMB100A operating manual

Rohde & Schwarz

Rohde & Schwarz SMB100A operating manual

V&A M266 Series user manual

V&A

V&A M266 Series user manual

Macherey-Nagel PF-3 manual

Macherey-Nagel

Macherey-Nagel PF-3 manual

ST DELTA SOLAR user manual

ST

ST DELTA SOLAR user manual

CSX P301 instruction manual

CSX

CSX P301 instruction manual

Thermo Scientific Scientific AquaSensors DataStick user guide

Thermo Scientific

Thermo Scientific Scientific AquaSensors DataStick user guide

Gima SP-10 user manual

Gima

Gima SP-10 user manual

ETS-Lindgren EMCenter user manual

ETS-Lindgren

ETS-Lindgren EMCenter user manual

ENMET ISA-M Operation & maintenance manual

ENMET

ENMET ISA-M Operation & maintenance manual

Vortex Ranger 1000 manual

Vortex

Vortex Ranger 1000 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.