Solinst 101 User manual

12V Power Reel Operating Instructions
High Quality Groundwater and Surface Water Monitoring Instrumentation
(Page 1 of 2)
Model 101
® Solinst is a registered trademark of Solinst Canada Ltd
Operation of the Controller
1. Set the speed dial on zero.
2. Connect to a 12V DC power supply using the clips provided.
3. Slowly turn the speed dial clockwise or counterclockwise
until the motor just starts to turn.
Controller
Input Ratings
10 - 16V DC @ 25A
Front Panel Controls
Speed dial: starts and stops motor, adjusts motor speed and
direction of rotation
Reset: 25A rating
Ambient Temperature Conditions
Temperature: 0 - 40˚C
Humidity: 90% RH or less, non-condensing
Motor
Lubrication
The shaft bearings contain grease fittings. After one year,
insert multipurpose chassis grease, and re-apply once per
year. The gear box is lubricated for life, and should never
need to be lubed.
Servicing
Disconnect power source from motor and any accessory
devices and allow motor to come to a complete stand still
Note: The Power Reel can use heat embossed polyethylene flat
tape or Solinst laser-marked PVDF flat tape
Caution: Motor must be grounded in accordance with the
national electrical code and local codes by trained
personnel to prevent serious electrical shocks.
Ensure the air in-take fan at the end of the motor is
clear.
If down-hole component gets stuck, do not use power
motor. To avoid injury, turn motor off immediately,
and attempt retrieval manually.
Water Level Meter Operation
Equipment Check
1. The level of sensitivity can be adjusted higher or lower by
turning the switch clockwise or counter clockwise.
2. Depress the Battery Test button to test the battery and
circuitry.
3. Submerse the probe in tap water. This completes the circuit
and activates the buzzer and light.
Water Level Measurement
The zero measurement point is at the tip of the sensor pin,
located at the centre of the P2 Probes, and where the black
Delrin tip meets the stainless steel body of the P7 Probes.
The light and buzzer activate when the zero point enters water.
To ensure accuracy, lower and raise the probe a few times and
then record the depth measurement from the tape at the top
of the well.
Using the Tape Guide
Feed the tape into and out of the well using the groove in the
top of the Tape Guide.
The tape guide has been designed to:
• improve accuracy when reading water levels,
• easily obtain repeatable measurements,
• prevent tape being cut by well casing,
• allow the tape and probe to hang straight from the side of
the well.
When using the tape guide, the measuring point is offset from
the top of casing. To adjust your measurements to the top of the
casing, simply subtract the amount indicated on the front of the
tape guide (i.e. 6 cm or 2/10 ft.).
Tape Guide
Fit onto well
casing
Measurement Point
Continued overleaf...
Lowering the Tape
When lowering the tape, the speed will be about 2 ft/s,
increasing by 2 ft/s for each additional 1000 ft of tape down-
hole. To reduce the speed increase during lowering, turn the
speed dial down gradually. If the speed dial is at its lowest
setting (1) and the speed is still too fast, decrease the speed
by reversing the reel direction (i.e. rewind). This serves as a
dynamic brake, and enables you to stop at the desired depth.
Retrieving the Tape
When rewinding the tape, the speed also depends on the length
of tape down-hole. At maximum depth (for longer tapes with
5000 ft down-hole), the speed will be less than 0.2 ft/s, even
at the maximum speed setting (8). The speed will gradually
increase to 2 ft/s as the tape winds closer to the top of the well.
Printed in Canada
November 16, 2012
(#110370)
(Page 2 of 2)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
®Solinst is a registered trademark of Solinst Canada Ltd.
Routine Care
1. After the depth to water has been recorded, the tape should
be carefully rewound onto the reel, the probe wiped dry and
placed into the probe holder.
2. The probe, tape and reel can be cleaned with phosphate
free (non-abrasive) detergent and warm water.
3. Remove any dirt or water from around the central sensor
pin. If the central sensor pin is corroded or coated, use
emery cloth to polish it.
4. Check the P2 probe seal/strain relief and replace the black
heat shrink if there is any cracking or other damage.
5. Use of the Tape Guide adds to the life of the tape.
Battery Replacement (battery type - alkaline, 9 volt.)
1. The battery is housed conveniently
on the front of the grey box with
the sensitivity switch.
2. To replace the battery, open the
compartment with a screw driver.
3. Note the polarity and place a new
battery in the compartment. Snap
the compartment cover back into
place.
Replacement Parts
The following parts can be provided should they become lost
or damaged:
Probes and seal kits, splice kits, lights, switches, etc., reels, and
replacement tape with probes (complete) or probe seals.
SYMPTOM CAUSE REMEDY
No sound
when
probe
immersed
in water
Dead battery. Replace with 9V Alkaline.
Water
conductivity is
very low.
Increase sensitivity switch setting (turn
clockwise) or call Solinst for assistance.
Disconnected
wires on circuit
board.
Check all connections inside hub of reel
for loose/disconnected wires - solder or
reconnect.
Broken wire in
tape.
Locate break in tape - splice and seal
(contact Solinst).
Disconnected
wire inside
probe.
Contact Solinst to obtain parts/repair
instructions.

Other Solinst Measuring Instrument manuals

Solinst 107 Guide

Solinst

Solinst 107 Guide

Solinst 101 SC2000 Operating and safety instructions

Solinst

Solinst 101 SC2000 Operating and safety instructions

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 401 User manual

Solinst

Solinst 401 User manual

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 101 Parts list manual

Solinst

Solinst 101 Parts list manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 101 P7 User manual

Solinst

Solinst 101 P7 User manual

Solinst LevelSender 5 User manual

Solinst

Solinst LevelSender 5 User manual

Solinst 201 User manual

Solinst

Solinst 201 User manual

Solinst 101M User manual

Solinst

Solinst 101M User manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst Levelogger Series User manual

Solinst

Solinst Levelogger Series User manual

Solinst 101 P7 SC1000 User manual

Solinst

Solinst 101 P7 SC1000 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 101 P2 User manual

Solinst

Solinst 101 P2 User manual

Solinst 101D User manual

Solinst

Solinst 101D User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 101 P2 Operating and safety instructions

Solinst

Solinst 101 P2 Operating and safety instructions

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 122 User manual

Solinst

Solinst 122 User manual

Popular Measuring Instrument manuals by other brands

PCE Americas PCE-SMM 1 manual

PCE Americas

PCE Americas PCE-SMM 1 manual

Dostmann Electronic GD 383 user manual

Dostmann Electronic

Dostmann Electronic GD 383 user manual

Laserworld Ecoline Series manual

Laserworld

Laserworld Ecoline Series manual

ADVANTEST R3131A manual

ADVANTEST

ADVANTEST R3131A manual

Noyes M series user guide

Noyes

Noyes M series user guide

sauter SAUTER TE SERIES instruction manual

sauter

sauter SAUTER TE SERIES instruction manual

NSFOCUS NTA NX3-2000E installation guide

NSFOCUS

NSFOCUS NTA NX3-2000E installation guide

Keysight Technologies U5309A Startup guide

Keysight Technologies

Keysight Technologies U5309A Startup guide

GREISINGER GMH 3750 operating manual

GREISINGER

GREISINGER GMH 3750 operating manual

Sony FDDR-7000 installation manual

Sony

Sony FDDR-7000 installation manual

Mountz TorqueMate Plus Operation manual

Mountz

Mountz TorqueMate Plus Operation manual

Apera Instruments PC60 instruction manual

Apera Instruments

Apera Instruments PC60 instruction manual

Instantel Blastmate III Operator's manual

Instantel

Instantel Blastmate III Operator's manual

Agilent Technologies U2722A datasheet

Agilent Technologies

Agilent Technologies U2722A datasheet

Hilti PD-I manual

Hilti

Hilti PD-I manual

SkylinkHome Monitor GM-434RTL manual

SkylinkHome

SkylinkHome Monitor GM-434RTL manual

CORNET ED-88T user manual

CORNET

CORNET ED-88T user manual

Ossila T2006A1 user manual

Ossila

Ossila T2006A1 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.