Routine Care
1. After the depth to water has been recorded, the cable should
be carefully rewound onto the reel, the probe wiped dry and
placed into the probe holder.
2. The probe, weights, cable and reel can be wiped clean with
phosphate free (non-abrasive) detergent and warm water.
Do not submerge the reel.
3. Use of a Water Level Meter Carrying Bag adds to the service
life of the meter.
4. Use of the Tape Guide adds to the life of the cable.
Replacement Parts
The following parts can be provided should components
become lost or damaged.
1. Splice kits
2. Lights, dials, etc.
3. Reels and/or faceplates
4 Replacement cable with probe (complete)
5. Assembled probes on 10 ft or 3 m lengths of cable
6. Weights
Battery Replacement
• Battery type - alkaline, 9 volt.
1. The battery is housed in a convenient battery drawer located
in the faceplate of the Water Level Meter.
2. To replace the battery, simply press the drawer in, lift, then
pull.
3. The battery drawer should slide out of the faceplate enough
to pull it out.
4. Note the polarity. The positive (+) terminal should be
towards the small notch in the end of the drawer. Place new
battery in the drawer and slide it back into the faceplate.
Troubleshooting
SYMPTOM CAUSE REMEDY
No sound when probe
immersed in water.
Dead battery. Replace with 9V Alkaline.
Water Conductivity is very low. Increase sensitivity dial setting (turn clockwise) or
call Solinst for assistance.
Disconnected wires on circuit
board.
Check all connections inside hub of reel for
loose/disconnected wires - solder or reconnect.
Broken wire in cable. Locate break in cable - splice and seal, or replace.
(Contact Solinst)
Disconnected wire inside probe. Contact Solinst to obtain parts/repair instructions.
Instrument
continuously sounds
after being immersed
in water.
Water in probe. Probe may be
dirty which could interfere with
the circuit connection.
Contact Solinst for instructions to clean or replace
the probe.
Zero Measurement Point
Zero Measurement Point
P4 Probe (4 mm diameter)
P10 Probe (10 mm diameter)