Vega VEGABAR 81 User manual

Process pressure
Pressure transmitter
VEGABAR 81,82,83
Product Information

Contents
2Process pressure –Pressure transmitter
Contents
1Measuring principle .......................................................................................3
2Type overview ...........................................................................................4
3Instrument selection .......................................................................................5
4Selection criteria .........................................................................................6
5Housing overview.........................................................................................7
6Mounting ...............................................................................................8
7Electronics -4…20 mA-two-wire ............................................................................9
8Electronics -4…20 mA/HART -two-wire ......................................................................10
9Electronics -Profibus PA ..................................................................................11
10 Electronics -Foundation Fieldbus ............................................................................12
11 Electronics -Slave sensor for electronic differential pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
12 Operation .............................................................................................14
13 Dimensions ............................................................................................15
Take note of safety instructions for Ex applications
Please note the Ex specific safety information which you can find on our homepage www.vega.com/downloads under "Approvals"and which
comes with every instrument.In hazardous areas you should take note of the corresponding regulations,conformity and type approval
certificates of the sensors and power supply units.The sensors must only be operated on intrinsically safe circuits.The permissible electrical
values are stated in the certificate.
45078-EN-140625

1Measuring principle
1.1Overview
The VEGABAR 81,82 and 83 can be used universally for the measure-
ment of gases,vapours and liquids.The instrument series is designed for
all applications in industrial measurement engineering.The pressure
transmitters are the ideal solution for special as well as standard applica-
tions.Setup and commissioning is always the same.The VEGABAR 81,
82 and 83 offer maximum reliability and operational safety.
1.2Measuring cell technology
VEGABAR 81
The VEGABAR 81 is equipped with a chemical seal system.It consists of
a process diaphragm as well as an isolating liquid.The process dia-
phragm is available in different materials.
The process pressure acts on the sensor element via the chemical seal
system and an internal stainless steel diaphragm.The process pressure
causes a resistance change which is converted into a corresponding
output signal and outputted as measured value.
1
3 4
2
Fig.1:Configuration of the piezoresistive measuring cell in VEGABAR 81
1Sensor element
2Base element
3Silicone oil filling
4Process diaphragm
For measuring ranges up to 40 bar,a piezoresistive sensor element with
internal transmission liquid is used,and for measuring ranges 100 bar and
higher,a strain gauge (DMS)sensor element on the back side of the
stainless steel diaphragm (dry)is used.
VEGABAR 82
The sensor element is the CERTEC
®
measuring cell with front-flush,
abrasion resistant ceramic diaphragm.The hydrostatic pressure of the
medium or the process pressure causes a capacitance change in the
measuring cell via the diaphragm.This change is converted into an ap-
propriate output signal.
16 µm
1
2
3
Fig.2:Configuration of the CERTEC
®
measuring cell in VEGABAR 82
1Diaphragm
2Soldered glass bond
3Base element
The CERTEC
®
measuring cell is also equipped with a temperature sen-
sor.The temperature value can be displayed via the display and adjust-
ment module or processed via the signal output.
VEGABAR 83
The process pressure acts via the process diaphragm to the sensor ele-
ment.The process pressure causes a resistance change which is con-
verted into a corresponding output signal and outputted as measured
value.The process diaphragm is available in different materials.
For measuring ranges up to 40 bar,a piezoresistive sensor element with
internal transmission liquid is used,and for measuring ranges 100 bar and
higher,a strain gauge (DMS)sensor element on the back side of the
stainless steel diaphragm (dry)is used.
With small measuring ranges or higher temperatures,the METEC
®
measuring cell is used.It consists of the ceramic-capacitive CERTEC
®
measuring cell and a special,temperature-compensated chemical seal
system.
4321
Fig.3:Configuration of the METEC
®
measuring cell in VEGABAR 83
1Process diaphragm
2Isolating liquid
3FeNi adapter
4CERTEC
®
measuring cell
Measuring principle
Process pressure –Pressure transmitter 3
45078-EN-140625

2Type overview
VEGABAR 81 VEGABAR 82 VEGABAR 83
Measuring cell Piezoresistive/DMS CERTEC
®
Piezoresistive/strain gauge,METEC
®
Diaphragm Metal Ceramic Metal
Media gases,vapours and liquids,also
aggressive ones,at high temper-
atures
gases,vapours and liquids,also with
abrasive substances
gas,vapours and liquids,also vis-
cous
Process fitting Thread from ½" or ½NPT
Flanges from DN 20
Boltings,tube isolating diaphragm
each from DN 25
Thread from 1"or ½NPT
Flanges from DN 25
Extension fittings from 1"
Thread from 1"or ½NPT
Flanges from DN 20
Boltings,tube isolating diaphragm
each from DN 25
Material
Process fitting
316L316L,PVDF,Alloy C22,Alloy C276 316L
Diaphragm material 316L,Alloy C276,Tantalum,gold on
316L
Al
2
O
3
ceramic 316Ti,Elgiloy 2.4711,
Alloy C276,gold-coated,gold/rhodi-
um-coated
Measuring cell seal -FKM,EPDM,FFKM -
Isolating liquid Silicone oil,high temperature oil,
halocarbon oil,med.white oil
Dry measuring system Silicone oil,Halocarbon oil
Med.white oil
Measuring range -1…+1000 bar/-100 …+100 MPa
(-14.5…+14500 psig)
-1…+100 bar/-100 …+10 MPa(-
-14.5…+1450 psig)
-1…+1000 bar/-100 …+100 MPa
(-14.5…+14500 psig)
Smallest measuring
range
0.4bar/40 kPa(5.802 psig)0.025 bar/2.5kPa(1.45 psig)0.1bar/10 kPa(1.45 psig)
Process temperature -40 …+400 °C(-40 …+752 °F) -40 …+150 °C(-40 …+302 °F) -40 …+200 °C(-40 …+392 °F)
Smallest deviation <0.2% < 0.05 % < 0.075 %
Signal output 4…20 mA,4…20 mA/HART,Profibus PA,Foundation Fieldbus,Modbus
Interface Digital interface for Slave-Master combination
Indication/Adjustment PLICSCOM,PACTware,VEGADIS 81,VEGADIS 62
Approvals SIL,shipbuilding,ATEX,IEC,overfill protection,FM,CSA,GOST
Type overview
4Process pressure –Pressure transmitter
45078-EN-140625

3Instrument selection
Application area
VEGABAR 81
The VEGABAR 81 is a pressure transmitter with chemical seal for pres-
sure and level measurement.The process-adapted chemical seal system
of VEGABAR 81 ensures reliable measurement even in highly corrosive
and hot media.
VEGABAR 82
The VEGABAR 82 pressure transmitter can be used universally for meas-
urement of gases,vapours and liquids.Even materials like sand are no
problem for the abrasion-resistant ceramic measuring cell.The VEGA-
BAR 82 is an economical solution for a multitude of applications in all
areas of industry.
VEGABAR 83
The VEGABAR 83 is a pressure transmitter for pressure measurement of
gases,vapours and liquids in all industries.VEGABAR 83 offers special
advantages for applications with high pressures.
Configuration and housing protection classes
1
2
3
Fig.4:Example of a VEGABAR 82 with connection G1½and plastic housing with
protection IP 66/IP 67
1Housing cover with integrated display and adjustment module (optional)
2Housing with electronics
3Process fitting with measuring cell
1
2
3
4
5
Fig.6:Example of a VEGABAR 82 with flange and aluminium housing in protection
IP 66/IP 68,1bar
1Housing cover with integrated display and adjustment module (optional)
2Housing with electronics
3Process fitting with measuring cell
4Cable gland
5Connection cable
1
2
3
5
4
Fig.8:Example of a VEGABAR 82 with flange and stainless steel housing with
protection rating IP 68 and external electronics
1Housing cover with integrated display and adjustment module (optional)
2Housing with electronics
3Housing base
4Connection cable
5Process fitting with measuring cell
Measured variables
The pressure transmitters VEGABAR 81,82 and 83 are suitable for meas-
urement of the following process variables:
lProcess pressure
lLevel
Fig.10:Process pressure measurement
In conjunction with a Slave sensor for electronic differential pressure
measurement,the instruments are also suitable for the measurement of
the following process variables:
lDifferential pressure
lLevel
lFlow
lDensity
lInterface
Fig.11:Differential pressure measurement via Master-Slave combination
Instrument selection
Process pressure –Pressure transmitter 5
45078-EN-140625

4Selection criteria
VEGABAR
81
VEGABAR
82
VEGABAR
83
Wear through process Aggressive products ●–●
Abrasive products –●–
Process temperature up to 150 °C(302 °F)● ● ●
200 °C(302 °F)● ● ●
400 °C(752 °F)●– –
Measuring system Dry –●–
Oil-filled ●–●
Version process fittings Front-flush ● ● ●
Hygienic ● ● ●
Largest measuring range 100 bar (10 MPa)● ● ●
400 bar (40 MPa)●–●
1000 bar (100 MPa)●–●
Smallest measuring range 400 mbar (40 kPa)● ● ●
100 mbar (10 kPa) – ● ●
25 mbar (2,5kPa) – ●–
Vacuum applications up to 1mbar
abs
(100 Pa) – ●–
Suitability for industry-specific ap-
plications
Aggregates and min-
ing industry
–● ●
Chemical ● ● –
Power generation ● ● –
Foodstuffs● ● ●
Metal production –● ●
Offshore ● ● –
Paper ● ● ●
Petrochemical ● ● –
Pharmaceutical ● ● –
Shipbuilding –● ●
Environment and recy-
cling industry
–●–
Water,waste water –●–
Cement industry –● ●
Selection criteria
6Process pressure –Pressure transmitter
45078-EN-140625

5Housing overview
Plastic PBT
Protection rating IP 66/IP 67 IP 66/IP 67
Version Single chamber Double chamber
Application area Industrial environment Industrial environment
Aluminium
Protection rating IP 66/IP 67,IP 66/IP 68 (1bar)IP 66/IP 67,IP 66/IP 68 (1bar)
Version Single chamber Double chamber
Application area Industrial environment with increased
mechanical stress
Industrial environment with increased
mechanical stress
Stainless steel 316L
Protection rating IP 66/IP 67
IP 69K
IP 66/IP 67
IP 66/IP 68 (1bar)
Version Single chamber,electropolished Single chamber,precision casting Double chamber,precision casting
Application area Aggressive environment,food pro-
cessing,pharmaceutical
Aggressive environment,extreme mechanical stress
Separate version
Material Stainless steel 316Lplastic PBT
Stainless steel 316L
Protection rating IP 68 (25 bar)IP 65
IP 66/IP 67
Function Transmitter External electronics
Application area Extremely moist environment Industrial environment
Housing overview
Process pressure –Pressure transmitter 7
45078-EN-140625

6Mounting
Installation position
The instruments function in any installation position.But the installation
position can influence the measurement,depending on the measuring
system.This can be compensated by a position correction.
It is useful to select an installation position you can easily reach for mount-
ing and connecting as well as later retrofitting of an display and adjustment
module.For this purpose,the housing can be rotated by 330°without the
use of any tools.You can also install the indicating and adjustment module
in four different positions (each displaced by 90°).
Mounting examples and measurement setups
The following illustrations show mounting examples and measurement
setups.
Process pressure measurement
The VEGABAR measures the pressure in a pipeline.
Fig.12:Process pressure measurement on a pipeline with VEGABAR
Level measurement
The VEGABAR measures the level in a vessel.
Fig.14:Level measurement in a vessel with VEGABAR
Differential pressure measurement
The VEGABAR measures the pressure difference between input and
output of a filter.
Fig.16:Differential pressure measurement on a filter with VEGABAR and Slave
sensor
Flow measurement
The VEGABAR measures the flow via the pressure difference between
the front and back sides of an orifice plate.
Fig.18:Flow measurement through an orifice with VEGABAR and a Slave sensor
Density measurement
VEGABAR measures the density of the medium in a vessel.
h
Fig.20:Density measurement on a vessel with VEGABAR and Slave sensor
Interface measurement
VEGABAR measures the interface of two liquids in a vessel.
h
Fig.22:Interface measurement on a vessel with VEGABAR and Slave sensor
Mounting
8Process pressure –Pressure transmitter
45078-EN-140625

7Electronics -4…20 mA-two-wire
Configuration of the electronics
The pluggable electronics is mounted in the electronics compartment of
the instrument and can be exchanged by the user when servicing is re-
quired.The electronics is completely encapsulated to protect against
vibration and moisture.
The terminals for voltage supply as well as the contact pins with I²Cinter-
face for parameter adjustment are located on the upper side of the elec-
tronics.In the double-chamber housing,the terminals are located in the
separate terminal compartment.
Voltage supply
Power supply and current signal are carried on the same two-wire cable.
The operating voltage can differ depending on the instrument version.
You can find the data of the voltage supply in chapter "Technical data"in
the operating instructions manual of the respective instrument.
Provide a reliable separation between the supply circuit and the mains
circuits according to DIN EN 61140 VDE 0140-1.
Data of the voltage supply:
lOperating voltage
-9.6…35 V DC
lPermissible residual ripple -Non-Ex,Ex-ia instrument
-for U
N
12 V DC:≤0.7V
eff
(16 …400 Hz)
-for U
N
24 V DC:≤1.0V
eff
(16 …400 Hz)
lPermissible residual ripple -Ex-d-ia instrument
-for U
N
24 V DC:≤1.0V
eff
(16 …400 Hz)
Keep in mind the following additional factors that influence the operating
voltage:
lLower output voltage of the power supply unit under nominal load (e.g.
with a sensor current of 20.5mAor 22 mAin case of fault)
lInfluence of additional instruments in the circuit (see load values in
chapter "Technical data"of the operating instructions of the respective
instrument)
Connection cable
The instrument is connected with standard two-wire cable without screen.
If electromagnetic interference is expected which is above the test values
of EN 61326-1for industrial areas,screened cable should be used.
In the product configurator of VEGABAR,different cable glands are avail-
able.This selection covers all cable diameters in the range of 4…12 mm
(0.16 …0.47 in).
Cable screening and grounding
If screened cable is required,we recommend connecting the cable
screen on both ends to ground potential.In the sensor,the screen must
be connected directly to the internal ground terminal.The ground terminal
on the outside of the housing must be connected to the ground potential
(low impedance).
Connection
Single chamber housing
5
12
+
( ) (-)
6 7 8
4...20mA
2
3
4
1
Fig.24:Electronics and terminal compartment,single chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3For external display and adjustment unit
4Ground terminal for connection of the cable screen
Electronics -4…20 mA-two-wire
Process pressure –Pressure transmitter 9
45078-EN-140625

8Electronics -4…20 mA/HART -two-wire
Configuration of the electronics
The pluggable electronics is mounted in the electronics compartment of
the instrument and can be exchanged by the user when servicing is re-
quired.The electronics is completely encapsulated to protect against
vibration and moisture.
The terminals for voltage supply as well as the contact pins with I²Cinter-
face for parameter adjustment are located on the upper side of the elec-
tronics.In the double-chamber housing,the terminals are located in the
separate terminal compartment.
Voltage supply
Power supply and current signal are carried on the same two-wire cable.
The operating voltage can differ depending on the instrument version.
You can find the data of the voltage supply in chapter "Technical data"in
the operating instructions manual of the respective instrument.
Provide a reliable separation between the supply circuit and the mains
circuits according to DIN EN 61140 VDE 0140-1.
Data of the voltage supply:
lOperating voltage
-9.6…35 V DC
lPermissible residual ripple -Non-Ex,Ex-ia instrument
-for U
N
12 V DC:≤0.7V
eff
(16 …400 Hz)
-for U
N
24 V DC:≤1.0V
eff
(16 …400 Hz)
lPermissible residual ripple -Ex-d-ia instrument
-for U
N
24 V DC:≤1.0V
eff
(16 …400 Hz)
Keep in mind the following additional factors that influence the operating
voltage:
lLower output voltage of the power supply unit under nominal load (e.g.
with a sensor current of 20.5mAor 22 mAin case of fault)
lInfluence of additional instruments in the circuit (see load values in
chapter "Technical data"of the operating instructions of the respective
instrument)
Connection cable
The instrument is connected with standard two-wire cable without screen.
If electromagnetic interference is expected which is above the test values
of EN 61326-1for industrial areas,screened cable should be used.
In the product configurator of VEGABAR,different cable glands are avail-
able.This selection covers all cable diameters in the range of 4…12 mm
(0.16 …0.47 in).
We generally recommend the use of screened cable for HART multidrop
mode.
Cable screening and grounding
If screened cable is required,we recommend connecting the cable
screen on both ends to ground potential.In the sensor,the screen must
be connected directly to the internal ground terminal.The ground terminal
on the outside of the housing must be connected to the ground potential
(low impedance).
Connection
Single chamber housing
5
12
+
( ) (-)
6 7 8
4...20mA
2
3
4
1
Fig.25:Electronics and terminal compartment,single chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3For external display and adjustment unit
4Ground terminal for connection of the cable screen
Double chamber housing
4...20mA
2
3
12
+
( ) (-)
1
Fig.26:Terminal compartment,double chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3Ground terminal for connection of the cable screen
Electronics -4…20 mA/HART -two-wire
10 Process pressure –Pressure transmitter
45078-EN-140625

9Electronics -Profibus PA
Configuration of the electronics
The pluggable electronics is mounted in the electronics compartment of
the instrument and can be exchanged by the user when servicing is re-
quired.The electronics is completely encapsulated to protect against
vibration and moisture.
The terminals for voltage supply as well as the plug with I²Cinterface for
parameter adjustment are located on the upper side of the electronics.In
the double-chamber housing,these connection elements are located in
the separate terminal compartment.
Voltage supply
The voltage supply is provided by a Profibus DP /PA segment coupler.
Data of the voltage supply:
lOperating voltage
-9…32 V DC
lMax.number of sensors per DP/PA segment coupler
-32
Connection cable
Connection is carried out with screened cable according to Profibus spec-
ification.
In the product configurator of VEGABAR,different cable glands are avail-
able.This selection covers all cable diameters in the range of 4…12 mm
(0.16 …0.47 in).
Make sure that the entire installation is carried out according to the Pro-
fibus specification.In particular,make sure that the bus is terminated with
suitable terminating resistors.
Cable screening and grounding
In systems with potential equalisation,connect the cable screen directly to
ground potential at the power supply unit,in the connection box and at the
sensor.The screen in the sensor must be connected directly to the inter-
nal ground terminal.The ground terminal outside on the housing must be
connected to the potential equalisation (low impedance).
In systems without potential equalisation,connect the cable screen di-
rectly to ground potential on the power supply unit and the sensor.In the
connection box or T-distributor,the screen of the short stub to the sensor
must not be connected to ground potential or to another cable screen.
Connection
Single chamber housing
5
0 0
5
1
6
2
7
3
8
4
9
0
5
1
6
2
7
3
8
4
9
1
0
1
6 7 8
Bus
2
3
4
5
12
+
( ) (-)
1
Fig.27:Electronics and terminal compartment,single chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3Selection switch for bus address
4For external display and adjustment unit
5Ground terminal for connection of the cable screen
Double chamber housing
Bus
5
12
+
( ) (-)
6 7 8
2
3
4
1
Fig.28:Terminal compartment,double chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3Ground terminal for connection of the cable screen
Electronics -Profibus PA
Process pressure –Pressure transmitter 11
45078-EN-140625

10 Electronics -Foundation Fieldbus
Configuration of the electronics
The pluggable electronics is mounted in the electronics compartment of
the instrument and can be exchanged by the user when servicing is re-
quired.The electronics is completely encapsulated to protect against
vibration and moisture.
The terminals for voltage supply as well as the plug with I²Cinterface for
parameter adjustment are located on the upper side of the electronics.In
the double-chamber housing,these connection elements are located in
the separate terminal compartment.
Voltage supply
Power supply via the H1Fieldbus cable.
Data of the voltage supply:
lOperating voltage
-9…32 V DC
lmax.number of sensors
-32
Connection cable
Connection is carried out with screened cable according to Fieldbus
specification.
In the product configurator of VEGABAR,different cable glands are avail-
able.This selection covers all cable diameters in the range of 4…12 mm
(0.16 …0.47 in).
Make sure that the entire installation is carried out according to the Field-
bus specification.In particular,make sure that the bus is terminated with
suitable terminating resistors.
Cable screening and grounding
In systems with potential equalisation,connect the cable screen directly to
ground potential at the power supply unit,in the connection box and at the
sensor.The screen in the sensor must be connected directly to the inter-
nal ground terminal.The ground terminal outside on the housing must be
connected to the potential equalisation (low impedance).
In systems without potential equalisation,connect the cable screen di-
rectly to ground potential on the power supply unit and the sensor.In the
connection box or T-distributor,the screen of the short stub to the sensor
must not be connected to ground potential or to another cable screen.
Connection
Single chamber housing
12
( ) (-)
1
5
0
1
0
1
+6 7 8
Bus
2
3
4
5
Fig.29:Electronics and terminal compartment,single chamber housing
1Voltage supply/Signal output
2Contact pins for the display and adjustment module or interface adapter
3Selection switch for bus address
4For external display and adjustment unit
5Ground terminal for connection of the cable screen
Double chamber housing
Bus
5
12
+
( ) (-)
6 7 8
2
3
4
1
Fig.30:Terminal compartment,double chamber housing
1Voltage supply/Signal output
2For display and adjustment module or interface adapter
3Ground terminal for connection of the cable screen
Electronics -Foundation Fieldbus
12 Process pressure –Pressure transmitter
45078-EN-140625

11 Electronics -Slave sensor for electronic differential pressure
Configuration of the electronics
The pluggable electronics is mounted in the electronics compartment of
the instrument and can be exchanged by the user when servicing is re-
quired.The electronics is completely encapsulated to protect against
vibration and moisture.
The terminals for voltage supply as well as the contact pins with I²Cinter-
face for parameter adjustment are located on the upper side of the elec-
tronics.In the double-chamber housing,the terminals are located in the
separate terminal compartment.
Voltage supply
The Slave sensor can be mounted up to 25 m away from the Master
sensor,by which it is directly powered.Aseparate power supply is not
required.
Connection cable
VEGABAR is connected to the sensor with standard four-wire,screened
cable.
In the product configurator of VEGABAR,different cable glands are avail-
able.This selection covers all cable diameters in the range of 4…12 mm
(0.16 …0.47 in).
Cable screen and grounding
Connect the cable screen on both ends to ground potential.InVEGABAR
Slave and Master sensor,the screen must be connected directly to the
internal ground terminal.The ground terminal on the outside of the re-
spective housing must be connected to the potential equalisation (low
impedance).
Connection
Single chamber housing
5 6 7 8 4
2
1
Master
Fig.31:Electronics and connection compartment Slave sensor
1To the Master sensor
2Ground terminal for connection of the cable screen1)
Electronics -Slave sensor for electronic differential pressure
Process pressure –Pressure transmitter 13
1) Connect screen here. Connect ground terminal on the outside of the housing to ground as prescribed. The two terminals are galvanically connected.
45078-EN-140625

12 Operation
12.1Overview
The sensors can be adjusted with the following adjustment media:
lWith the display and adjustment module
lWith external display and adjustment unit
lan adjustment software according to FDT/DTM standard,e.g.
PACTware and PC
as well as via systems from other manufacturers,dependent on the signal
output:
lA HART handheld (4…20 mA/HART)
lThe adjustment program AMS (4…20 mA/HART and Foundation
Fieldbus)
lThe adjustment program PDM (Profibus PA)
lAconfiguration tool (Foundation Fieldbus)
The entered parameters are generally saved in the sensor,optionally also
in the display and adjustment module or in the adjustment program.
12.2Display and adjustment module
PLICSCOM
The pluggable display and adjustment module is used for measured value
indication,operation and diagnosis.It is equipped with an illuminated full
dot matrix as well as four keys for adjustment.
Fig.32:Display and adjustment module PLICSCOM
The display and adjustment module is integrated in the respective sensor
housing or in the external display and adjustment unit.After mounting,the
sensor as well as the display and adjustment module are splash-proof
even without housing cover.
12.3PACTware/DTM
As an alternative to the display and adjustment module,the sensor can
also be configured via a Windows PC.To do this,the configuration soft-
ware PACTware and a suitable instrument driver (DTM)according to the
FDT standard are required.The current PACTware version as well as all
available DTMs are compiled in a DTM Collection.The DTMs can also be
integrated into other frame applications according to the FDT standard.
All device DTMs are available as a free-of-charge standard version and as
a full version that must be purchased.In the standard version,all functions
for complete setup are already included.An assistant for simple project
configuration simplifies the adjustment considerably.Saving/printing the
project as well as import/export functions are also part of the standard
version.
In the full version there is also an extended print function for complete
project documentation as well as a save function for measured value and
echo curves.In addition,there is a tank calculation program as well as a
multiviewer for display and analysis of the saved measured value and
echo curves.
Connection of the PC via VEGACONNECT
The interface converter VEGACONNECT is required for connection of the
PC.On the computer side,the connectionis made via USB interface.The
VEGACONNECT is placed instead of the display and adjustment module
to the sensor,the connection to the sensor is made automatically.As an
alternative the connection via the HART signal can be carried out on any
position of the signal cable with 4…20 mA/HART sensors.
3
1
USB
2
OPEN
TWIST
LOCK
Fig.33:Connection via VEGACONNECT and USB
1VEGACONNECT
2plics
®
sensor
3USB cable to the PC
Necessary components:
lVEGABAR
lPC with PACTware and suitable DTM
lVEGACONNECT
lVoltage supply/Processing system
12.4Alternative adjustment programs
PDM
For HART and Profibus PA sensors,device descriptions are available as
EDDs for the adjustment program PDM.The device descriptions are al-
ready included in the current version of the PDM.Newer instrument driv-
ers that are not yet delivered with the PDM are available in the download
section.
AMS
For HART and Foundation Fieldbus sensors,device descriptions are
available as EDDs for the adjustment program AMS.The device descrip-
tions are already included in the current version of the AMS.Newer instru-
ment drivers that are not yet delivered with the AMS are available in the
download section.
Operation
14 Process pressure –Pressure transmitter
45078-EN-140625

13 Dimensions
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
112 mm
(4.41")
M20x1,5/
½ NPT
~ 84 mm
(3.31")
M16x1,5
112 mm
(4.41")
M20x1,5/
½ NPT
12
ø 79 mm
(3.11")
1Single chamber housing
2Double chamber housing
Aluminium housing
21
ø 86 mm
(3.39")
~ 116 mm
(4.57")
116 mm
(4.57")
M20x1,5
M20x1,5/
½ NPT
~ 87 mm
(3.43")
M16x1,5
ø 86 mm
(3.39")
120 mm
(4.72")
M20x1,5/
½ NPT
1Single chamber housing
2Double chamber housing
Stainless steel housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
117 mm
(4.61")
M20x1,5/
½ NPT
~ 59 mm
(2.32")
ø 80 mm
(3.15")
112 mm
(4.41")
M20x1,5/
½ NPT
~ 87 mm
(3.43") ø 86 mm
(3.39")
120 mm
(4.72")
M20x1,5/
½ NPT
M16x1,5
321
1Single chamber housing,electropolished
2Single chamber housing,precision casting
2Double chamber housing,precision casting
VEGABAR 81
109 mm
(4.29")
89 mm
(3.5")
D
1 2
1Flange version
2Version with tube isolating diaphragm
VEGABAR 82
2
3
1
20,8 mm
(0.82")
45 mm
(1.77")
SW 27 mm
(1.06")
G ½
ø 18,4 mm
(0.72")
ø 27 mm
(1.06")
G1 ½
50 mm
(1.97")
22 mm
(0.87")
ø 55 mm
(2.17")
SW 46 mm
(1.81")
50 mm
(1.97")
1Threaded version G1½A
2Threaded version G½A,inner G¼A
3Threaded version G½A,inner G¼A,PVDF
VEGABAR 83
1
2
3
118 mm
(4.65")
46 mm
(1.81")
78 mm
(3.07")
ø 64 mm
(2.52")
ø 84 mm
(3.31")
G ½
61 mm
(2.40")
SW 27 mm
(1.06")
32 mm
(1.26")
3 mm
(0.12")
ø 3 mm
(0.12") ø 6 mm
(0.24")
1Threaded version G1½A
2Threaded version G1½A,200 °C
3Tri-Clamp version 2"
The listed drawings are only an excerpt of the available process fittings.
You can find more drawings at www.vega.com/downloads under "Draw-
ings".
Dimensions
Process pressure –Pressure transmitter 15
45078-EN-140625

VEGA Grieshaber KG
AmHohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
Fax +49 7836 50-201
E-Mail:info.de@vega.com
www.vega.com
You can find at www.vega.com
loperating instructions manuals
lspecification sheet
lSoftware
ldrawings
lcertificates
lapprovals
and much,much more
Subject to change without prior notice 45078-EN-140625
Other manuals for VEGABAR 81
16
This manual suits for next models
2
Table of contents
Other Vega Transmitter manuals

Vega
Vega VEGABAR 83 User manual

Vega
Vega VEGABAR 81 User manual

Vega
Vega VEGABAR 55 User manual

Vega
Vega VEGABAR S 14 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGAPULS 62 User manual

Vega
Vega VEGABAR 83 User manual

Vega
Vega VEGABAR 17 User manual

Vega
Vega VEGABAR 86 User manual

Vega
Vega VEGABAR 51 User manual

Vega
Vega VEGABAR S 16 User manual

Vega
Vega VEGABAR 66 User manual

Vega
Vega VEGABAR 51 User manual

Vega
Vega VEGABAR 54 User manual

Vega
Vega VEGABAR 83 User manual

Vega
Vega VEGABAR 86 User manual

Vega
Vega VEGABAR 67 User manual

Vega
Vega VEGADIF 65 User manual