Vega VEGABAR 86 User manual

Operating Instructions
Suspension pressure transmitter with
ceramic measuring cell
VEGABAR 86
4 … 20 mA
Document ID: 45506

2
Contents
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbolism used............................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use................................................................................................................ 5
2.3 Warning about incorrect use............................................................................................. 5
2.4 General safety instructions............................................................................................... 5
2.5 CE conformity................................................................................................................... 5
2.6 Measuring range - permissible process pressure............................................................. 6
2.7 NAMUR recommendations .............................................................................................. 6
2.8 Environmental instructions ............................................................................................... 6
3 Product description
3.1 Conguration.................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 8
3.3 Packaging, transport and storage................................................................................... 10
3.4 Accessories and replacement parts ............................................................................... 10
4 Mounting
4.1 General instructions to use the instrument ..................................................................... 12
4.2 Ventilation and pressure compensation.......................................................................... 13
4.3 Level measurement........................................................................................................ 14
4.4 External housing ............................................................................................................ 15
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 17
5.2 Connecting..................................................................................................................... 18
5.3 Single chamber housing................................................................................................. 19
5.4 Housing IP 66/IP 68 (1 bar) ............................................................................................ 19
5.5 External housing with version IP 68 (25 bar)................................................................... 20
5.6 Switch-on phase............................................................................................................. 21
6 Set up with the display and adjustment module
6.1 Insert display and adjustment module............................................................................ 23
6.2 Adjustment system......................................................................................................... 24
6.3 Parameter adjustment - Quick setup .............................................................................. 25
6.4 Parameter adjustment - Extended adjustment................................................................ 26
6.5 Saving the parameter adjustment data........................................................................... 38
7 Setup with PACTware
7.1 Connect the PC.............................................................................................................. 40
7.2 Parameter adjustment .................................................................................................... 40
7.3 Saving the parameter adjustment data........................................................................... 41
8 Set up with other systems
8.1 DD adjustment programs ............................................................................................... 42
8.2 Field Communicator 375, 475 ........................................................................................ 42
9 Diagnosis, asset management and service
9.1 Maintenance .................................................................................................................. 43

3
Contents
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
9.2 Diagnosis memory ......................................................................................................... 43
9.3 Asset Management function........................................................................................... 43
9.4 Rectify faults................................................................................................................... 47
9.5 Calculation of total deviation (according to DIN 16086).................................................. 48
9.6 Exchange process assembly with version IP 68 (25 bar)................................................ 49
9.7 Exchanging the electronics module................................................................................ 50
9.8 Software update ............................................................................................................. 51
9.9 How to proceed in case of repair.................................................................................... 51
10 Dismounting
10.1 Dismounting steps.......................................................................................................... 52
10.2 Disposal ......................................................................................................................... 52
11 Supplement
11.1 Technical data ................................................................................................................ 53
11.2 Dimensions .................................................................................................................... 63
Safety instructions for Ex areas
PleasenotetheEx-specicsafetyinformationforinstallationandop-
eration in Ex areas. These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
Editing status: 2013-09-25

4
1 About this document
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instruc-
tionsformaintenanceandfaultrectication.Pleasereadthisinforma-
tion before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel.The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
• List
The dot set in front indicates a list with no implied sequence.
→ Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

5
2 For your safety
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
Model VEGABAR 86 is a pressure transmitter for level and gauge
measurement.
Youcannddetailedinformationontheapplicationrangeinchapter
"Product description".
Operational reliability is ensured only if the instrument is properly
usedaccordingtothespecicationsintheoperatinginstructions
manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specichazards,e.g.vesseloverllordamagetosystem
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of stand-
ard regulations and guidelines. The user must take note of the safety
instructionsinthisoperatinginstructionsmanual,thecountry-specic
installation standards as well as all prevailing safety regulations and
accident prevention rules.
Theinstrumentmustonlybeoperatedinatechnicallyawlessand
reliable condition.The operator is responsible for trouble-free opera-
tion of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
2.5 CE conformity
ThedevicefulllsthelegalrequirementsoftheapplicableECguide-
lines.ByaxingtheCEmarking,weconrmsuccessfultestingofthe
product.
YoucanndtheCECerticateofConformityinthedownloadsection
of our homepage.
Electromagnetic compatibility
Instruments with plastic housing as well as in four-wire or Ex-d-ia
version are designed for use in an industrial environment. Neverthe-
less, electromagnetic interference from electrical conductors and

6
2 For your safety
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
radiated emissions must be taken into account, as is usual with a
class A instrument according to EN 61326-1. If the instrument is used
inadierentenvironment,theelectromagneticcompatibilitytoother
instruments must be ensured by suitable measures.
2.6 Measuring range - permissible process
pressure
Thepermissibleprocesspressureisspeciedonthetypelabelwith
"process pressure", see chapter "Conguration". For safety reasons,
this range must not be exceeded. This applies also if a measuring
cell with higher measuring range (order-related) than the permissible
pressurerangeoftheprocessttingisinstalled.
2.7 NAMUR recommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
acceptedasthestandardineldinstrumentation.
ThedevicefulllstherequirementsofthefollowingNAMURrecom-
mendations:
• NE 21 – Electromagnetic compatibility of equipment
• NE 43 – Signal level for malfunction information from measuring
transducers
• NE53–Compatibilityofelddevicesanddisplay/adjustment
components
• NE107–Self-monitoringanddiagnosisofelddevices
For further information see www.namur.de.
2.8 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental pro-
tection.Theenvironmentmanagementsystemiscertiedaccording
to DIN EN ISO 14001.
Pleasehelpusfulllthisobligationbyobservingtheenvironmental
instructions in this manual:
• Chapter "Packaging, transport and storage"
• Chapter "Disposal"

7
3 Product description
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
3 Product description
3.1 Conguration
Thenameplatecontainsthemostimportantdataforidenticationand
use of the instrument:
2
1
13
14
15
16
12
11
5
3
6
4
7
8
9
10
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Field for approvals
4 Power supply and signal output, electronics
5 Protection rating
6 Measuring range
7 Permissible process pressure
8 Material, wetted parts
9 Hardware and software version
10 Order number
11 Serial number of the instrument
12 Symbol of the device protection class
13 ID numbers, instrument documentation
14 Reminder to observe the instrument documentation
15 Notied authority for CE marking
16 Approval directive
The type label contains the serial number of the instrument. With it
youcanndthefollowingdataonourhomepage:
• Product code of the instrument (HTML)
• Delivery date (HTML)
• Order-specicinstrumentfeatures(HTML)
• Operating instructions at the time of shipment (PDF)
• Order-specicsensordataforanelectronicsexchange(XML)
• Testcerticatepressuretransmitters(PDF)
Go to www.vega.com, "VEGA Tools" and "Serial number search".
Asanalternative,youcanndthedataviayourSmartphone:
• Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store"
• Scan the Data Matrix code on the type label of the instrument or
• Enter the serial number manually in the app
Type plate
Serial number

8
3 Product description
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
This operating instructions manual applies to the following instrument
versions:
• Hardware from 1.0.0
• Software version from 1.0.0
The scope of delivery encompasses:
• Pressure transmitter
• Documentation
– this operating instructions manual
– Testcerticatemeasuringaccuracy(optional)
– Operating instructions manual "Display and adjustment mod-
ule" (optional)
– Supplementary instructions manual "Heating for display and
adjustment module" (optional)
– Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
– Ex-specic"Safety instructions" (with Ex versions)
– ifnecessary,furthercerticates
3.2 Principle of operation
The VEGABAR 86 is suitable for the measurement of the following
process variables:
• Level
Fig. 2: Level measurement with VEGABAR 86
In combination with a slave sensor, VEGABAR 86 is also suitable for
electronicdierentialpressuremeasurement.
Youcannddetailedinformationintheoperatinginstructionsofthe
respective slave sensor.
The VEGABAR 86 is a suspension pressure transmitter for level
measurementinwells,basinsandopenvessels.Itsgreatexibility
Scope of this operating
instructions manual
Scope of delivery
Measured variables
Electronicdierential
pressure
Application area

9
3 Product description
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
throughdierentcableandtubeversionsallowstheinstrumenttobe
usedinmanydierentapplications.
Measured products are liquids.
Depending on the instrument version and measurement setuo, meas-
ured products can be also viscous or contain abrasive substances.
Sensor element is the CERTEC®measuring cell with robust ceramic
diaphragm.Theprocesspressuredeectstheceramicdiaphragm
and causes a capacitance change in the measuring cell. This capaci-
tance change is converted into an electrical signal and outputted as
measured value via the output signal.
The CERTEC®measuring cell has also an integrated temperature
sensor. The temperature value is outputted through:
• The display and adjustment module
• The current output
• The digital signal output
The measuring cell design depends on the selected pressure type.
Relative pressure: the measuring cell is open to atmosphere.The
ambient pressure is detected in the measuring cell and compensated.
Itthushasnoinuenceonthemeasuredvalue.
Absolute pressure: the measuring cell is evacuated and encapsu-
lated. The ambient pressure is not compensated and does hence
inuencethemeasuredvalue.
Relative pressure, climate-compensated: the measuring cell is
evacuated and encapsulated. The ambient pressure is detected
through a reference sensor in the electronics and compensated. It
thushasnoinuenceonthemeasuredvalue.
The following illustration shows the installation of the CERTEC®
measuring cell in the transmitter as well as the sealing concept.
1
2
3
4
5
Fig. 3: Front-ush installation of the CERTEC® measuring cell with double seal
1 Housing, transmitter
2 Measuring cell
3 Lateral seal for measuring cell
4 Additional, front seal for measuring cell
5 Diaphragm
Measured products
Measuring system
Pressure types
Seal concept

10
3 Product description
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
• Not in the open
• Dry and dust free
• Not exposed to corrosive media
• Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
• Relative humidity 20 … 85 %
3.4 Accessories and replacement parts
The display and adjustment module PLICSCOM is used for measured
value indication, adjustment and diagnosis. It can be inserted into the
sensor or the external display and adjustment unit and removed at
any time.
Youcanndfurtherinformationintheoperatinginstructions"Display
and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of
communication-capable instruments to the USB interface of a PC. For
parameter adjustment of these instruments, the adjustment software
PACTware with VEGA-DTM is required.
Youcanndfurtherinformationintheoperatinginstructions"Interface
adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 62 is suitable for measured value indication and adjustment
of sensors with HART protocol. It is looped into the 4 … 20 mA/HART
signal cable.
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
PLICSCOM
VEGACONNECT
VEGADIS 62

11
3 Product description
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
Youcanndfurtherinformationintheoperatinginstructions"VE-
GADIS 62" (Document-ID 36469).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
Youwillndadditionalinformationinthesupplementaryinstructions
manual "Protective cover" (Document-ID 34296).
Screwedangesareavailableindierentversionsaccordingtothe
following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5,
JIS B 2210-1984, GOST 12821-80.
Youcanndadditionalinformationinthesupplementaryinstructions
manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
Welded sockets are used to connect the sensors to the process.
Youcanndfurtherinformationinthesupplementaryinstructions
"Welded socket VEGABAR series 80" (Document-ID 45082).
The electronics module VEGABAR series 80 is a replacement part
forpressuretransmittersofVEGABARseries80.Thereisadierent
version available for each type of signal output.
Youcanndfurtherinformationintheoperatinginstructions"Elec-
tronics module VEGABAR series 80" (Document-ID 45054).
Protective cap
Flanges
Welded socket
Electronics module

12
4 Mounting
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
4 Mounting
4.1 General instructions to use the instrument
Make sure that all parts of the instrument exposed to the process are
suitable for the existing process conditions.
These are mainly:
• Active measuring component
• Processtting
• Process seal
Process conditions are particularly:
• Process pressure
• Process temperature
• Chemical properties of the medium
• Abrasionandmechanicalinuences
Youcanndthespecicationsoftheprocessconditionsinchapter
"Technical data" as well as on the nameplate.
Protect your instrument further through the following measures
against moisture penetration:
• Use the recommended cable (see chapter "Connecting to power
supply")
• Tighten the cable gland
• Loop the connection cable downward in front of the cable gland
This applies particularly to:
• Outdoor mounting
• Installations in areas where high humidity is expected (e.g. through
cleaning processes)
• Installations on cooled or heated vessels
Oninstrumentswithprocessttingthread,thehexagonmustbetight-
enedwithasuitablescrewdriver.Wrenchsizeseechapter"Dimen-
sions".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
In case of strong vibrations at the mounting location, the instrument
version with external housing should be used. See chapter "External
housing".
Higher process temperatures often mean also higher ambient
temperatures. Make sure that the upper temperature limits stated in
chapter "Technical data" for the environment of the electronics hous-
ing and connection cable are not exceeded.
Suitability for the process
conditions
Protection against mois-
ture
Screwing in
Vibrations
Temperature limits

13
4 Mounting
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
2
1
Fig. 4: Temperature ranges
1 Process temperature
2 Ambient temperature
4.2 Ventilation and pressure compensation
Ventilation and pressure compensation are carried out with VE-
GABAR86viaalterelement.Itisairpermeableandmoisture-
blocking.
Caution:
Thelterelementcausesatime-delayedpressurecompensation.
When quickly opening/closing the housing cover, the measured value
can change for approx. 5 s by up to 15 mbar.
Foreectiveventilation,thelterelementmustalwaysbefreeof
buildup.
Caution:
Donotuseahigh-pressurecleaner.Thelterelementcouldbedam-
aged, which would allow moisture into the housing.
Thefollowingparagraphsdescribehowthelterelementisarranged
inthedierentinstrumentversions.
Thelterelementismountedintotheelectronicshousing.Ithasthe
following functions:
• Ventilation electronics housing
• Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
Filter elements
Instruments in non-Ex
and Ex-ia version

14
4 Mounting
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
1 2 3
4 4 4
Fig. 5: Position of the lter element - non-Ex, Ex-ia version
1 Plastic, stainless steel precision casting
2 Aluminium
3 Stainless steel, electro-polished
4 Filter element
With the following instruments a blind plug is installed instead of the
lterelement:
• Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-
lariesinxconnectedcable
• Instruments with absolut pressure
Thelterelementisintegratedintheprocessassembly.Itislocatedin
a rotatable metal ring and has the following function:
• Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
Turnthemetalringinsuchawaythatthelterelementpointsdown-
ward after installtion of the instrument. This provides better protection
against buildup.
21
3
4
3
4
Fig. 6: Position of the lter element - Ex-d version
1 Single chamber housing, aluminium, stainless steel precision casting
2 Double chamber housing, aluminium, stainless steel precision casting
3 Rotatable metal ring
4 Filter element
Instruments with absolute pressure have a blind plug mounted
insteadofthelterelement.
4.3 Level measurement
Keep the following instructions in mind for the measurement setup:
Instruments in Ex-d ver-
sion
Measurement setup

15
4 Mounting
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
• Donotmounttheinstrumentclosetothellingstreamoremptying
area
• Mount the instrument so that it is protected against pressure
shocks from the stirrer
Fig. 7: Measurement setup with level measurement
4.4 External housing
1
2
3
4
Fig. 8: Arrangement measurement loop, external housing
1 Vessel
2 Sensor
3 Connection cable sensor - external housing
4 External housing
5 Signal cable
1. Mark the holes according to the following drilling template
2. Fasten wall mounting plate with 4 screws
Conguration
Mounting

16
4 Mounting
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
90 mm
(3.54")
R 3,5 mm
(0.14")
3mm
(0.12")
70 mm
(2.76")
8 mm
(0.32")
93 mm
(3.66")
110 mm
(4.33")
Fig. 9: Drilling template - wall mounting plate

17
5 Connecting to power supply
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
• Connect only in the complete absence of line voltage
• If overvoltage surges are expected, overvoltage arresters should
be installed
Power supply and current signal are carried on the same two-wire
cable.Theoperatingvoltagecandierdependingontheinstrument
version.
Thedataforpowersupplyarespeciedinchapter"Technical data".
Provide a reliable separation between the supply circuit and the
mains circuits according to DIN EN 61140 VDE 0140-1.
Keepinmindthefollowingadditionalfactorsthatinuencetheoperat-
ing voltage:
• Lower output voltage of the power supply unit under nominal load
(e.g. with a sensor current of 20.5 mA or 22 mA in case of fault
message)
• Inuenceofadditionalinstrumentsinthecircuit(seeloadvaluesin
chapter "Technical data")
The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326-1 for industrial areas, screened cable should
be used.
Use cable with round cross section for instruments with housing and
cablegland.Toensurethesealeectofthecablegland(IPprotection
rating),ndoutwhichcableouterdiameterthecableglandissuitable
for.
• 5 … 9 mm (0.20 … 0.35 in)
• 6 … 12 mm (0.24 … 0.47 in)
• 10 … 14 mm (0.40 … 0.55 in)
Useacableglandttingthecablediameter.
With plastic housing, the NPT cable gland or the Conduit steel tube
must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data".
If screened cable is required, we recommend connecting the cable
screen on both ends to ground potential. In the sensor, the screen
must be connected directly to the internal ground terminal.The
ground terminal on the outside of the housing must be connected to
the ground potential (low impedance).
With Ex systems it must be ensured that the grounding corresponds
to the installation regulations.
In electroplating and CCP systems (cathodic corrosion protection) it
mustbetakenintoaccountthatsignicantpotentialdierencesexist.
Safety instructions
Voltage supply
Connection cable
Cable gland ½ NPT
Cable screening and
grounding

18
5 Connecting to power supply
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
This can lead to unacceptably high shield currents if the cable shield
is grounded at both ends.
Information:
Themetallicpartsoftheinstrument(processtting,transmitter,con-
centric tube, etc.) are conductively connected with the inner and outer
ground terminal on the housing.This connection exists either directly
via connecting metallic parts or, in case of instruments with external
electronics, via the screen of the special connection cable.
Youcanndspecicationsonthepotentialconnectionsinsidethe
instrument in chapter "Technical data".
5.2 Connecting
The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
The connection to the display and adjustment module or to the inter-
face adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the
electronics.To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn-
ing it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Fig. 10: Connection steps 5 and 6 - Single chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Connection technology
Connection procedure

19
5 Connecting to power supply
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
Information:
Solidcoresaswellasexiblecoreswithwireendsleevesareinsert-
eddirectlyintotheterminalopenings.Incaseofexiblecoreswithout
end sleeves, press the terminal from above with a small screwdriver;
the terminal opening is freed. When the screwdriver is released, the
terminal closes again.
Youcanndfurtherinformationonthemax.wirecross-sectionunder
"Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry.The seal ring must
completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
Theelectricalconnectionishencenished.
5.3 Single chamber housing
The following illustration applies to the non-Ex as well as to the Ex-ia
version.
5
12
+
( ) (-)
678
4...20mA
2
3
4
1
Fig. 11: Electronics and connection compartment, single chamber housing
1 Voltage supply/Signal output
2 For display and adjustment module or interface adapter
3 For external display and adjustment unit or Slave sensor
4 Ground terminal for connection of the cable screen
5.4 Housing IP 66/IP 68 (1 bar)
1
2
Fig. 12: Wire assignment x-connected connection cable
1 brown (+) and blue (-) to power supply or to the processing system
2 Shielding
Electronics and connec-
tion compartment
Wire assignment, con-
nection cable

20
5 Connecting to power supply
VEGABAR 86 • 4 … 20 mA
45506-EN-131010
5.5 External housing with version IP 68 (25 bar)
Fig. 13: VEGABAR 86 in IP 68 version 25 bar, non-Ex and axial cable outlet,
external housing
1
2
3
5
1
2
( )
+
(
-
)
6
7
8
4...20m
A
Fig. 14: Electronics and connection compartment
1 Electronics module
2 Cable gland for voltage supply
3 Cable gland for connection cable, transmitter
Overview
Electronics and connec-
tion compartment for
power supply
Other manuals for VEGABAR 86
14
Table of contents
Other Vega Transmitter manuals

Vega
Vega VEGADIF 65 User manual

Vega
Vega VEGABAR 54 User manual

Vega
Vega VEGABAR 86 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 66 User manual

Vega
Vega VEGADIF 85 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 86 User manual

Vega
Vega VEGABAR 83 User manual

Vega
Vega T-677 Series User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 81 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGABAR 64 User manual

Vega
Vega VEGABAR 83 User manual

Vega
Vega VEGABAR 54 User manual

Vega
Vega VEGAFLEX 80 series User guide

Vega
Vega VEGABAR 14 User manual

Vega
Vega VEGABAR 87 User manual

Vega
Vega VEGADIF 65 User manual