PARAMETER DESCRIPTION CODE/VALUE
F1
Graduations
Species number of the Weighingblocks graduations.
Value should be consistent with legalrequirements and
environmental limits on the useful system resolution.
500 1000 1500 2000
VB2-100-EC / 200-EC 2500 3000
VB2-30-EC 4000 5000 6000: VB2-
30-05 8000 10000: VB2-100-10 /
200-20 13000: WS-130
F2
Sampling rate
Sampling rate. 10 or 80Hz. Only choose 80hz for very fast
weighing applications. Normally always 10hz is chosen.
80
10: all versions
F3
Zero Track Band
Selects the range within which the scale will automatically
zero. Note that the scale must be in standstill to
automatically zero. Selections are in Display Divisions.
“d” = graduation
0d, 0.5d: , 1d: all EC-versions 3d: all
other versions 5d
F4
Zero Range
Selects the range within which the scale may be zeroed.
Note that the indicator must be in standstill to zero the scale.
100% (US) 1.9% (CE)
2%: all EC versions
20%: all other versions
F5
Motion Band
Sets the level at which motion is detected by comparing the
present display update with the previous one. If motion is
not detected for two seconds or more, scale is in standstill
and can process a Print or Zero command. Maximum value
varies depending on local regulations.
0,25 1d 3d all versions
5d 10d
F6
Digital Filter
Averages weight readings to produce higher accuracy. The
higher the lter number, the greater the accuracy but the
slower the response time. Choose 4 or 8 unless a very fast
response is needed.
1 2
4 8 all versions
F7
Overload Limit
Selects the desired formula which determines the point at
which the indicator shows overload. All selections are based
on the primary unit selected in F8.
“FS” = Full scale in primary units.
FS FS + 2% all none EC ver. FS +
1d FS + 9d all EC ver.
F8
Calib. Unit
Selects the primary base unit to be used in the calibration
process. Also the default unit for normal operation. “1” =
primary unit is lb. “2” = primary unit is in kg.
1
2 all versions
F9
Display Div.
Determines the desired weight increments. Value should be
consistent with legal requirements.
1: VB2-100-10/30-EC, WS-130
2: VB2-200-20
5: VB2-100-EC/30-05/100-50
F10
Decimal Pt. Determines location of the decimal point.
0 0,0
0,00 all versions except VB2-30-05
0,000 VB2-30-05 0,0000
F11
Batching funct. Activating Batching function (not be used with “EC”-Scales)
0: Approved Scales
1: Not Approved Scales
2: for scales with set points (F13,14)
F13
Set point 1 For special versions only Any value.
F14
Set point 2 For special versions only Any value.
F16
Zero Calibration
Places indicator into the zero calibration routine. Scrolling
down with the ZERO key one level begins the procedure. Press ZERO key to begin sequence
F17
Span Calibration
Places indicator into the span calibration routine. Scrolling
down with the ZERO key one level begins the procedure. Press ZERO key to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A”
menu to the default settings. USE WITH CAUTION! Press the ZERO key twice to execute.
A5
Disable the lb/kg
Key
Allows the lb/kg key to be disabled so that an operator
cannot accidentally press the key and change the displayed
units.
“0” = Disable the lb/kg key “1” = Enable the lb/kg key
”2” Activate the extended graduation for VB2-100-EC
0 all versions
1
2 (only for VB2-100-EC and for
batching Scales)
VB2 ENUNCIATOR DEFINITIONS