Vetek VB2 Series User manual

Manual
MODEL MODEL CAPACITY/GRADUATION
VB2-30-EC Max 30 kg / 10 g
VB2-100-EC Max 100 kg / 10 (50) g
VB2-200-EC Max 200 kg / 20 (100) g
VB2-200-EC-TM Max 200 kg / 100 g
VB2-200-50-TM Max 200 kg / 50 g
VB2-30-05 Max 30 kg / 5 g
VB2-100-02 UNISCALE 100kg/2g Max 100 kg / 2 g
VB2-100-10 Max 100 kg / 10 g
WS-130 Max 130 kg / 10 g (special)
VB2-130-10 UNISCALE 130 kg /10 g 130kg/10g Max
VB2-200-10 Max 200 kg / 20 g
UNISCALE 200kg/5g Max 200 kg / 5 g
UNISCALE 200 kg /50 g 200kg/50g Max
OVERVIEW DIFFERENT RANGES
1
Scandinavia’s largest webshop for scales
Hantverksvägen 15
76493 Vaddo, Sweden
Tel. +46 (0)176-208920
WEIGHINGBLOCK VB2 SERIES AND UNISCALE
The Weighingblock is a robust scale. It is designed for portable use. When transported by car between
different working places we recommend using of a protected space or a case for the scale.
This will extend the interval for service.
Extern
matning
9VDC
Kommunika-
tionsport
Alkaliska
batterier
Kalibrerings-
omkopplare
Normal Kalibrering
ON
PRINT
NET
GROSS
OFF
ZERO
TARE
Manual_VB2_V1

user manual
www.vetek.com info@vetek.com 2
INTRODUCTION
Vetek’s new VB2 series of Weighingblocks are rugged digital scales cast in strong aluminum. The scales
come standard with a large (.75”) LCD screen for easy readout. All setup parameters may be entered via
the membrane panel keys. The scales have an ”intelligent” auto power off function and is equipped with
6 alkaline 1,5 V batteries LR6/AA. If the battery is not good enough to ensure a correct value the scale
will turn off. The scale is developed especially for the weighing of refrigerant cylinders, and to use as a
control when transferring refrigerants to other cylinders or equipment.
The scales setup parameters are altered through the Setup menu while a few other parameters are
altered through the User menu. The conguration section of the manual explains how to use the ve
front panel keys to maneuver and save settings in both menus.
There are several categories of scales:
Non - Automatic
weighing instruments and
Automatic
weighing
instruments. Weighingblocks and other scales used only for weighing for example, cylinders, are
Non -
Automatic
weighing instruments. When using a Non - Automatic scale in commercial purpose in any of
the EC countries it is prescribed by law that the scale must be EC-type approved. If the purpose of use
is not commercial, or if the scale’s intended use is as an automatic batching scale then there are no
legal demands.
If the Scale is used as a
Batching Scale
it is therefore an
Automatic
weighing instruments and there are
no legal demands. This however requires to be checked with the authorities in each country.
VB2 SERIES OF EC-TYPE APPROVED WEIGHINGBLOCKS: The parameters should not be changed because of
its approval. Only accepted authorities may do the verication. The user requires a good knowledge of
all concerning rules and laws. To get complete information we recommend vsiting the following website
http://www.sp.se ,or visiting other authorities’ web sites.
VB2 SERIES OF NOT APPROVED WEIGHINGBLOCKS: Cannot be veried.
The Weighingblocks are suitable for use in various weighing applications such as the weighing of:
refrigeration cylinders, re protection cylinders, CO2, cylinders, parcels, persons, etc.
TABLE OF CONTENTS
PAGE
Introduction, Keyboard Functions 2
Conguration and Setup Menu. Display 3-4
Parameters 5
Calibration 6
Specications and Displayed Error Codes 7
Fault localization and repair 8
Instruction for handling 9
Solenoid module VBB-10 9

user manual
www.vetek.com info@vetek.com 3
KEYBOARD FUNCTIONS
THE CALIBRATION SWITCH IS ONLY FOR CALIBRATION.
◄Indicator ON. In non EC-type approved version: Sends “Print” data to printer if scale is
stable and not in overload. Not active when “Continuous” option is selected in User Menu.
◄“Dot”. Extended graduation to 10 g for VB2-100-EC and 20 g for VB2-200-EC when
pushedand 5 sec after release.
◄Toggles between Gross and Net weight display only if a Tare has been established.
◄Sets scale to display “0” when in Gross mode, and within zero band range.
When pushed in for 3 seconds, the Scale switches off.
Used to establish a Tare◄(zero the Scale) while in either Gross mode or Net mode. This
operation cannot be performed at or below Gross zero.
CONFIGURATION
SETUP MENU.
To do this conguration you require a good acknowledge of how a digital electronic scale works.
Note: If the scale fails, it is not always certain the scale needs conguration.
Excluding the Normal Meny, the Weighingblock has two menus:
- The Setup menu; contains most of the indicator’s functional setup parameters,
consists of 15 separate menu selections, each with its own sub-menu of choices.
- The User menu; except A5 and A10 contains most of the indicator’s serial communication
parameters, which are not shown in this manual.
ON
PRINT
NET
GROSS
OFF
ZERO
TARE

user manual
www.vetek.com info@vetek.com 4
LCD ENUNCIATOR MEANING
0Better known as the “Center of Zero” enunciator, this light is active whenever the
displayed weight is within ± 0.25 divisions of true zero.
NIndicates that the indicator is displaying net weight.
GIndicates that the indicator is displaying gross weight.
TIndicates that a tare weight has been established in the system.
lb, kg Indicates the unit of the displayed weight (normally disabled).
bAtt Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.
Indicates stable weighing.
SETUP MENU
1. Turn the power OFF.
2. Put Calibration Switch to the Calibration position.
3. Turn the indicator ON.
4. The display shows “F1” to indicate that the unit is in Setup menu mode.
5. To move to a new “F” heading, use TARE or ON to move up or down.
6. To move to the selection value, press ZERO once.
7. Increase and decrease the value with PRINT and TARE.
8. To save the value, press NET.
9. Press Dot to go back to parameter number.
10. To go back to normal menu put Calibration Switch to the Normal position.
NOTES ON THE SETUP MENU
There is an F21 sub-menu present that is for FACTORY USE ONLY.
DISPLAY
ON
PRINT
NET
GROSS
OFF
ZERO
TARE

user manual
www.vetek.com info@vetek.com 5
PARAMETER DESCRIPTION CODE/VALUE
F1
Graduations
Species number of the Weighingblocks graduations.
Value should be consistent with legalrequirements and
environmental limits on the useful system resolution.
500 1000 1500 2000
VB2-100-EC / 200-EC 2500 3000
VB2-30-EC 4000 5000 6000: VB2-
30-05 8000 10000: VB2-100-10 /
200-20 13000: WS-130
F2
Sampling rate
Sampling rate. 10 or 80Hz. Only choose 80hz for very fast
weighing applications. Normally always 10hz is chosen.
80
10: all versions
F3
Zero Track Band
Selects the range within which the scale will automatically
zero. Note that the scale must be in standstill to
automatically zero. Selections are in Display Divisions.
“d” = graduation
0d, 0.5d: , 1d: all EC-versions 3d: all
other versions 5d
F4
Zero Range
Selects the range within which the scale may be zeroed.
Note that the indicator must be in standstill to zero the scale.
100% (US) 1.9% (CE)
2%: all EC versions
20%: all other versions
F5
Motion Band
Sets the level at which motion is detected by comparing the
present display update with the previous one. If motion is
not detected for two seconds or more, scale is in standstill
and can process a Print or Zero command. Maximum value
varies depending on local regulations.
0,25 1d 3d all versions
5d 10d
F6
Digital Filter
Averages weight readings to produce higher accuracy. The
higher the lter number, the greater the accuracy but the
slower the response time. Choose 4 or 8 unless a very fast
response is needed.
1 2
4 8 all versions
F7
Overload Limit
Selects the desired formula which determines the point at
which the indicator shows overload. All selections are based
on the primary unit selected in F8.
“FS” = Full scale in primary units.
FS FS + 2% all none EC ver. FS +
1d FS + 9d all EC ver.
F8
Calib. Unit
Selects the primary base unit to be used in the calibration
process. Also the default unit for normal operation. “1” =
primary unit is lb. “2” = primary unit is in kg.
1
2 all versions
F9
Display Div.
Determines the desired weight increments. Value should be
consistent with legal requirements.
1: VB2-100-10/30-EC, WS-130
2: VB2-200-20
5: VB2-100-EC/30-05/100-50
F10
Decimal Pt. Determines location of the decimal point.
0 0,0
0,00 all versions except VB2-30-05
0,000 VB2-30-05 0,0000
F11
Batching funct. Activating Batching function (not be used with “EC”-Scales)
0: Approved Scales
1: Not Approved Scales
2: for scales with set points (F13,14)
F13
Set point 1 For special versions only Any value.
F14
Set point 2 For special versions only Any value.
F16
Zero Calibration
Places indicator into the zero calibration routine. Scrolling
down with the ZERO key one level begins the procedure. Press ZERO key to begin sequence
F17
Span Calibration
Places indicator into the span calibration routine. Scrolling
down with the ZERO key one level begins the procedure. Press ZERO key to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A”
menu to the default settings. USE WITH CAUTION! Press the ZERO key twice to execute.
A5
Disable the lb/kg
Key
Allows the lb/kg key to be disabled so that an operator
cannot accidentally press the key and change the displayed
units.
“0” = Disable the lb/kg key “1” = Enable the lb/kg key
”2” Activate the extended graduation for VB2-100-EC
0 all versions
1
2 (only for VB2-100-EC and for
batching Scales)
VB2 ENUNCIATOR DEFINITIONS

user manual
www.vetek.com info@vetek.com 6
CALIBRATION
We recommend the scale to be calibrated at yearly intervals.
The minimum test weight that can be used is 70% of full-scale capacity.
TO ADJUST THE ZERO POINT USING THE F16 ZERO ADJUST PROCEDURE:.
1. Turn the power OFF.
2. Put Calibration Switch to the Calibration position.
3. Turn the indicator ON.
4. Scroll to “F 16”. Use TARE or ON to move up or down.
5. Press ZERO. The display will momentarily show “C 0” followed by a value. (If you
want, press Dot to go back to parameter number.)
6. After making sure that there are no weights on the platform, press ZERO to zero out
the displayed value.
7. Press NET to save the zero point value. The display will show “EndC0”momentarily,
and then revert back up to F16. At this time, proceed to the F17 span calibration to
complete indicator calibration.
TO ADJUST THE MAX POINT USING THE F17 SPAN ADJUSTS PROCEDURE:
1. Scroll to “F 17”, then press ZERO to enter span calibration menu.
2. The display will momentarily show “C 1” for the span calibration, followed by a value with one ashing digit.
This value will be zero. Place the test weight on the Scale.
3. Pressing TARE or ON will change the position of the ashing digit.
4. Increase the ashing digit by pressing Dot. Decrease the ashing digit by pressing ZERO.
5. After setting the exact value, press NET to save the value.
6. If the adjustment was successful, the display will show “EndC1” momentarily, and then revert back up to F17.
“Err0” – The calibration test weight or the adjusted keyed-in weight is larger than the full capacity of the
scale. Change the calibration test weight or check the input data.
“Err1” – Change the calibration test weight or check the input data.
“Err2” – The internal resolution of the scale is not high enough to accept the calibration value.
Select a larger parameter for the Span Gain (F2).
7. Put the Calibration Switch to the normal mode.
8. Check the Scale
ON
PRINT
NET
GROSS
OFF
ZERO
TARE

user manual
www.vetek.com info@vetek.com 7
SPECIFICATIONS AND DISPLAYED ERROR CODES
GENERAL SPECIFICATIONS
MODEL CAPACITY /
GRADUATION
WEIGHT
(KG) DIMENSION MM BATTERY
VB2-30-EC Max 30 kg / 50 g 3,6 210 x 260x 65 6 of alk. 1,5 V batteries LR6/AA for 15 hrs operation
VB2-100-EC Max 100 kg /10 (50) g 4,5 260 x 345 x 55 6 st alk. 1,5 V batteries LR6/AA for 24 hrs. operation
WS-100-EC Max 100 kg / 50 g 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs operation
VB2-200-EC Max 200 kg / 100 g 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-30-05 Max 30 kg / 5 g
Not approved version 3,6 210 x 260x 65 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-100-02 Max 100 kg / 2 g 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-100-10 Max 100 kg / 10 g
Not approved version 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-130-10
WS-130
Max 130 kg / 10 g
Not approved version 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-100-50 Max 100 kg / 50 g
Not approved version 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
VB2-200-10 Max 200 kg / 20 g
Not approved version 4,5 260 x 345 x 55 6 of alk. 1,5 V batteries LR6/AA for 15 hrs. operation
- Alternative power is a 9 VDC adaptor (centre minus).
OPERATOR INTERFACE
Display 0.75” (19 mm) 7-segment, Liquid Crystal, 6-Digit
Additional Symbols Net, Gross, Stable, Tare, lb, kg, Zero
Keyboard 5-key at membrane panel
ENVIRONMENTAL
Operating Temperature 0 to +30 grad C (-10 to +40 grad C not approved version)
Storage Temperature -25 to +70 grad C
DISPLAYED ERROR CODES
CODE MODE MEANING / POSSIBLE SOLUTION
¤¤¤¤ Normal Operating M. Gross Overload. A weight greater than the rated capacity has been applied to the scale. Remove the weight from
the platter or try re-calibrating the scale. Otherwise, check for a bad load cell connection or possible load cell
damage due to overloading.
bAtt Normal Operating M. Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.
Err 0 Span Calibration
Mode (F17)
Keyed-in weight value is larger than full-scale capacity. Use a smaller test weight or check keyed-in value.
Err 1 Span Calibration
Mode (F17)
Keyed-in weight value is less than 1% of full-scale capacity. Use a larger test weight or check keyed-in value.
Err 2 Span Calibration
Mode (F17)
There is not enough load cell signal to produce the internal counts necessary to properly calibrate the scale. First check
all load connections. Use F16 mode to view internal counts. See Appendix C for more info.
Err 3 All Modes Non-volatile memory read error. One or more setup parameters have been lost.
Err 4 All Modes Non-volatile memory writes error. Indicator needs service.
Err 5 Key-in Span Calibration
Mode (F20)
You have attempted to enter a zero value for C1. Enter a known calibration value greater than zero.
Err 7 Initialization No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up.
Err 9 Normal Operating Span calibration value has been lost. Re-calibrate the Scale.

user manual
www.vetek.com info@vetek.com 8
FAULT LOCALIZATION AND REPAIR
*PLEASE NOTE: IF A VERIFIED SCALE IS REPAIRED IT PROBABLY NEEDS A NEW VERIFICATION, CHECK WITH
AUTHORITIES IN YOUR COUNTRY.
FAULT LOCALIZATION
IF THE WEIGHINGBLOCK IS “DEAD”
Change to new alkaline batteries.1.
Change the membrane key buttons if you have reason to believe it’s broken.2.
IF THE SCALE STARTS BUT THE DISPLAY SHOWS WRONG.
1. Adjust and Calibrate the Scale.
2. If it not works, the Scale must be demounted.
DEMOUNTING
1. Demount the bottom plate of the Weighingblock trough the two screws covered by a piece of tape. Be careful
with the keyboard connection.
2. Check the inside of the Scale visually. Look for bad cables and connections.
3. If needed, clean the PC-board.
IF THE SCALE SHOWS F1
1. Probably the Calibration Switch has been changed to calibration mode
2. Reset to normal mode.
CHECK THE LOADCELL
Check the Loadcell with a universal volt instrument, see table. The left pin is number 1.
Terminal J1 Description Colours Loadcell
AG
Colours Loadcell
1042
Colours
Loadcell LOC o
FAS
Colours
Lastcell SP4
E+ (1) Out +5 VDC* Brown Green Red White
S+ (2) In + Yellow Red Green Green
E- (3) Out 0 VDC Green Black Black Black
S- (4) In - White White White Red
IF IT IS PROBLEM TO GET A READING ON THE DISPLAY, OR IT IS IMPORTANT TO CALIBRATE:
1. Check the excitation voltage (E+ and E-) from the Indicator. Should be 5,0* (+-0,4) VDC.
2. Check the input to the Indicator (S+ and S-). The output from the Loadcell S+ and S- (pin 2 and 4) will
increase from aprox. 0 to 10 mV analogue to the capacity range (100 or 30 kg). Note S+ is plus and S- is
minus. The polarity is important.
3. If the input is wrong ( <0 mV or > +10 mV with unloaded scale) check the Loadcell. Have only E+ and E-
connected (or probably an external Power 5 VDC) and check the output from the Loadcells directly on the
wires. Output should be 0 mV to +10 mV depending on the load.

user manual
www.vetek.com info@vetek.com 9
LCD Indikator Betydelse
0Better known as the “Center of Zero” enunciator, this light is active whenever the
displayed weight is within ± 0.25 divisions of true zero.
NIndicates that the indicator is displaying net weight.
GIndicates that the indicator is displaying gross weight.
TIndicates that a tare weight has been established in the system.
lb, kg Indicates the unit of the displayed weight(normally disabled).
bAtt Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.
Indicates stable weighing.
INSTRUCTIONS FOR HANDLING WEIGHINGBLOCK VB2-SERIES AND UNISCALE
These instructions are valid for the new design with waist and rounded corners, shipped after
2002-09-18, and with a serial number higher than 74201.
The following instructions are for the scale
user
, and comprises this last section of the user manual. If
you require even further information, contact you wholesaler; Vetek AB or download information from
our web site: http://www.vetek.com
We highly recommend that the user of this scale gains good knowledge of all rules and laws pertaining to
its usage. To get complete information we recommend visiting the concerning Authorities’ web sites.
ENUNCIATOR DEFINITIONS
Pictured above is the LCD display. Before reading the value, the stable indicator must be ”on”.
If not, wait a few seconds until it appears.
HANDLING
1. Press ON to start the scale.
2. If necessary, press ZERO to obtain a weight reading of zero.
3. Press OFF to switch off the scale. Hold 3 seconds.
4. The scale switches off automatically after 10 minutes when not in use
WEIGHING (E.G. A CYLINDER WITH LIQUID)
1. Unload scale and press ZERO.
2. Place the object (cylinder, container, etc) on the scale’s platter and allow the weight indication to stabilize.
If the item weight exceeds the scale’s weight capacity, it displays “”. Press TARE to zero the
Scale. The display shows zero weight and turns the NET enunciator on.
3. NET/GROSS toggles between Gross and Net weight display only if a Tare has been established.
4. “Dot” extends graduation to 10 g for VB2-100-EC when pushed and 5 seconds after release.
GENERAL ADVICE
• When the Scale is not in use and is transported between different working places we recommend to use a
case or any other protection against vibrations and other hard environmental conditions.
• Protect the scale with a transparent plastic bag if there is any risk of leakage or when working in a wet
environment.
• Use onle 6 pcs alkaline 1,5 V batteries LR6/AA. Alternative power is a 9 VDC adaptor (centre minus).

user manual
www.vetek.com info@vetek.com 10
COMMUNICATION PORT (NOT VALID FOR UNISCALE)
THESE INSTRUCTIONS ARE ONLY VALID FOR THE NEW DESIGN, WITH WAIST AND ROUNDED CORNERS
(FOR UNITS SHIPPED AFTER 2003-03-01).
PRINTER FOR THE WEIGHT VALUE
Many printers in the market can be connected ( e.g. Martels MCP9800).
For more information, contact Vetek AB.
PC/NOTEBOOK
The Scale can be connected to a PC/Notebook. For more information, contact Vetek AB.
SOLENOID MODULE VBB-10
(NOTE: THIS FUNCTION IN VB2 IS STANDARD FOR SCALES DELIVERED AFTER MARCH 2003)
One time set up: F11 must be “1” and A5 must be “2” (normally done when delivered).
HANDLING
1. Mount the Solenoid module to the cylinder.
2. Connect the ”zero modem cable” and the Mains cable. (For the points 3 – 10 below it is no need for the
Scale to be on the oor, you can keep it in your hands)
Start the Scale with ON and wait a few seconds.3.
To set the Batching Weight:4.
Press, ”Dot”, display shows 0,0 (ashing).5.
Press ON until the rst relevant digit ash.6.
Change to the wished value with ”Dot”.7.
Press ON until the next relevant digit ash.8.
Change to the wished value with ”Dot”.9.
When the Scale shows wished value, save with NET10.
Press ZERO it the Scale not shows zero and put the object on the Scale. The Scale shows the gross weight.11.
Batching: Press TARE. The Scale shows zero and the valve is opened.12.
When the weight difference on the Scale is same as Batching Weight the valve will close.13.
The Scale shows the Batching Weight. Press NET/GROSS for gross weight.14.
GENERAL ADVICE
• Batching too quickly gives less accuracy.
Do not touch the tube to the cylinder during batching.•
VBB-10 CONSIST OF:
VBB-10-10 BOX WITH SOLENOID
VBB-10-11 “ZERO MODEM CABLE” 9-PIN FEMALE/FEMALE
VBB-10-12 MAINS CABLE.
Other manuals for VB2 Series
1
This manual suits for next models
16
Table of contents
Other Vetek Scale manuals

Vetek
Vetek WLC 1/A2 User manual

Vetek
Vetek JKD 250 User manual

Vetek
Vetek FCS Series User manual

Vetek
Vetek VRB-20 User manual

Vetek
Vetek VC Series User manual

Vetek
Vetek AS Series User manual

Vetek
Vetek XA Series User manual

Vetek
Vetek CW263BT User manual

Vetek
Vetek OCS-XZ-003 User manual

Vetek
Vetek VB2 Series User manual

Vetek
Vetek ERT DYNALINK User manual

Vetek
Vetek WBA-300 User manual

Vetek
Vetek VRH-300 User manual

Vetek
Vetek JWL-1 User manual

Vetek
Vetek SELL User manual

Vetek
Vetek BABY 630 User manual

Vetek
Vetek OCS-P Series User manual

Vetek
Vetek ACS 20AYE User manual

Vetek
Vetek LPWN-1530 User manual

Vetek
Vetek FR-H-100 User manual