manual
5
How to calibrate
**you must have an accurate the full load kilogram weight or combination of weights in order to
calibrate**
1. The scale must be powered OFF. Press and hold [ZERO] rst, then Press [ON/OFF] while keeping
the [ZERO] button depressed, then release both keys. The display will show the A/D value (a series
of random numbers).
2. Wait for the stable indicator to be displayed, Press [UNIT] the display will show “0SAVE”, then the
display will return to the random A/D value.
3. Place the full load kilograms weight(s) on platform. Wait for the stable indicator to display, then
press [UNIT], the display will show “0SAVE”. Calibration is complete, remove the weight(s).
4. Turn the scale OFF (Press the ON/OFF key) , then turn it back ON and check some weight readings.
If calibration is still incorrect, repeat calibration but try it more slowly.
If calibration is repeatedly incorrect then: 1) Replace the batteries, 2) Try a more stable surface or
vibration and interference free location, 3) Be sure there is no static electricity on the scale (use an
anti- static spray to remove static electricity)
FEATURES
Auto shut o (optional) Auto calibration
Auto zero tracking AC Power (optional)
Removable display w/3ft cord kg /lb conversion
Large LCD (6digits 15 mm high) hold function
␣Largesquareweighingtray.
SPECIFICATIONS
**SPECIAL WARNING**
Cell-Phones,Cordless-Phones,andanyradio-frequencydevicecancausetemporaryinterference
and cause the scale to temporarily not work properly. Please do not use any electronic device near
the shipping scale. Just like in an Airplane, do not use your cell-phone near the scale when it is in
use. While there is NO risk of permanent damage to the scale, interference can cause an incorrect
calibration or incorrect weight readings.
Model number Capacity Division
FCS-30 30kg 10g
FCS-60 60kg 20g
FCS-100 100kg 20g/50g
FCS-150 150kg 50g/100g
Platter / Tray 330×320mm
Net/gross weight 4.1kg/4.5kg
Package Standard carton: 44.5×34.6×7(cm3)
5 Units in one box: 46×36×39(cm3)
Operating
Temperature
Optimum: 15-30°C
Power source 6×AA Batteries or AC/DC
Adapter 9~12V/100mA (optional)
www.vetek.com info@vetek.com