Patient Cable Connections
The accessory kit of the electrocardiograph
includes a 10-lead patient cable. This cable is
pluggedintothe patient cablesocket onthe right-
hand side of the unit and secured with the two
screws.
The AT-1 is CF rated. The patient connection
is fully isolated and defibrillation protected.
Protection against defibrillation voltage is
howeveronly ensured, ifthe originalSCHILLER
patient cable (Part-no. 2.400070 / USA:
2.400071) is used. Make sure that during ECG
recording neither the patient nor the conducting
parts of the patient connection or the electrodes
(including the neutral electrode) come into
contactwith otherpersonsor conductingobjects
(even if these are earthed).
The quality of the ECG is dependent on the
preparation and the resistance between the skin
andthe electrode.To ensure a good quality ECG
and minimise the skin/electrode resistance,
remember the following points:
1. Ensure that the patient is warm and
relaxed.
2. Shave electrode area before cleaning.
3. Thoroughlyclean theareawith alcohol.
4. Place the C4 electrode first - in the fifth
intercostal space on midclavicular line.
Thenplace:
•C1 in fourth intercostal space at the right
sternalborder
•C2 in fourth intercostal space at the left
sternalborder
•C3between, andequidistantto, C4and C2
•C6 onleftmidaxillarylineonthesamelevel
as C4
•C5between, andequidistantto, C4and C6
The electrode placements shown on the following page are labelled
withthe colorsaccording toIEC requirements. TheequivalentAHA
colors are given on the table opposite.
Patient Cable Connections
The accessory kit of the electrocardiograph
includes a 10-lead patient cable. This cable is
pluggedintothe patient cablesocket onthe right-
hand side of the unit and secured with the two
screws.
The AT-1 is CF rated. The patient connection
is fully isolated and defibrillation protected.
Protection against defibrillation voltage is
howeveronly ensured, ifthe originalSCHILLER
patient cable (Part-no. 2.400070 / USA:
2.400071) is used. Make sure that during ECG
recording neither the patient nor the conducting
parts of the patient connection or the electrodes
(including the neutral electrode) come into
contactwith otherpersonsor conductingobjects
(even if these are earthed).
The quality of the ECG is dependent on the
preparation and the resistance between the skin
andthe electrode.To ensure a good quality ECG
and minimise the skin/electrode resistance,
remember the following points:
1. Ensure that the patient is warm and
relaxed.
2. Shave electrode area before cleaning.
3. Thoroughlyclean theareawith alcohol.
4. Place the C4 electrode first - in the fifth
intercostal space on midclavicular line.
Thenplace:
•C1 in fourth intercostal space at the right
sternalborder
•C2 in fourth intercostal space at the left
sternalborder
•C3between, andequidistantto, C4and C2
•C6 onleftmidaxillarylineonthesamelevel
as C4
•C5between, andequidistantto, C4and C6
The electrode placements shown on the following page are labelled
withthe colorsaccording toIEC requirements. TheequivalentAHA
colors are given on the table opposite.