2Safety Summary Welch Allyn Spot Vital Signs LXi
WARNING To e n s u re pa t i e n t safe t y, u s e o nly a c c e ss o r i e s a n d s u p pli e s ( i . e .,
cuffs, hoses, temperature probes, SpO2sensors, etc.) recommended for or
supplied with Spot Vital Signs LXi. Using unapproved accessories with Spot Vital
Signs LXi can affect patient and/or operator safety.
WARNING Take care to prevent wate r or ot her fluid from entering any
connectors on the device. Should this occur, dr y the connectors with warm air.
Check the accuracy of all operating functions.
WARNING Every three months, inspect the blood pressure cuff, SpO2 cable,
and other accessories for fraying or other damage. Replace as necessary.
WARNING Do not use Spot Vital Signs LXi on patients who are on heart/lung
machines.
WARNING Electric shock hazard. There are no user-serviceable parts inside Spot
Vital Signs LXi other than battery replacement (see “Battery Replacement” on
page 46). An operator may only perform maintenance procedures specifically
described in this manual. For service, refer the device to an Authorized Service
Center.
WARNING This device is not intended for hand-held use during operation.
WARNING Do not autoclave.
WARNING This device complies with current required standards for
electromagnetic interference and should not present problems to other
equipment or be affected by other devices. As a precaution, avoid using this
device in close proximity to other equipment.
WARNING Welch Allyn is not responsible for the integrity of any mounting
installation. Welch Allyn recommends that the customer contact their Biomedical
Engineering Department or maintenance service to ensure professional
installation for safety and reliability of any mounting accessory.
WARNING The Spot Vital Signs LXi consists of high-quality precision parts.
Protect it from severe impact and shock. A qualified service technician must
check any Spot Vital Signs LXi that is dropped or damaged for proper operation
prior to further use. Do not use the Spot Vital Signs LXi if you notice any signs of
damage. Contact the Welch Allyn Customer Service Department for assistance.
WARNING Do not use an SpO2finger clip sensor and a blood pressure cuff
simultaneously on the same limb. Doing so may result in inaccurate pulse rate
and perfusion readings.
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC standards
(for example, IEC 60950) as appropriate to the device. Connecting additional
devices to the Spot LXi may increase leakage currents. To maintain operator and
patient safety, it is necessary to consider the requirements of IEC 60601-1-1.
WARNING For proper patient electrical isolation, use only a Welch Allyn power
supply (4500-101A) to charge Spot Vital Signs LXi and its attached peripheral
devices. Do not use an external charger while the printer or weight scale is
attached to Spot LXi.