WIKA CPH7000 User manual

Operating instructions
Betriebsanleitung
EN
DE
Portable process calibrator, model CPH7000
Portabler Prozesskalibrator, Typ CPH7000
Portable process calibrator model CPH7000 with optional hand pump

2WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
EN
DE
Operating instructions model CPH7000 Page 3 - 64
Betriebsanleitung Typ CPH7000 Seite 65 - 125
Further languages can be found at www.wika.com.
© 02/2017 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!

EN
WIKA operating instructions portable process calibrator, model CPH7000 3
14204467.02 03/2019 EN/DE
Contents
Contents
1. General information 6
1.1 Software licence agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2. Short overview 7
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4 Product passport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3. Safety 8
3.1 Explanation of symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.5 Personnel qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.6 Labelling, safety marks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.6.1 Product label for process calibrator model CPH7000 . . . . . . . . . . . . . . . . . . . 11
3.6.2 Product label for reference pressure sensor model CPT7000 . . . . . . . . . . . . . . 11
4. Design and function 12
4.1 Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2 Functions and data transfer with the WIKA-Cal calibration software . . . . . . . . . . . . . 12
4.2.1 WIKA-Cal configuration (also possible with the demo version) . . . . . . . . . . . . . . 13
4.2.2 WIKA-Cal - Cal-Template. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2.3 WIKA-Cal - Log-Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.4 WIKA-Cal - switch test template. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3 Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.1 Rechargeable battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3.2 Using the power supply unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3.3 During charging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.4 Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.5 Mechanical connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.6 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.7 Barometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.8 Reference pressure sensor model CPT7000. . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.8.1 Connection of the model CPT7000 reference pressure sensor. . . . . . . . . . . . . . 21
4.8.2 Electrical connection of the CPT7000 reference pressure sensor to the CPH7000. . . 21
4.9 Carrying system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.10 Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.10.1 Applications (apps). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.10.2 Symbols of the status bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.10.3 Battery symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5. Transport, packaging and storage 24
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 Packaging and storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.3 Rechargeable battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

EN
4WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
Contents
6. Commissioning, operation 26
6.1 Electrical mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.1 ON/OFF key. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.1.1 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.1.2 Locking the screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.1.3 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.2.2 Application [Settings]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.2.3 Application [Info] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.2.4 Application [Remote] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2.5 Application [Service] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2.6 Further settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2.6.1 Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2.6.2 Rechargeable battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2.6.3 Changing the application task . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.3 Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.3.1 Configuration of the “Measure” function . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.3.1.1 Setting test item 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.3.1.2 Setting test item 2 and 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.3.2 Functions with symbol and meaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.3.3 Measuring the external reference pressure sensor on channel 1 or on channel 2 . . . 34
6.3.4 Measuring the external temperature probe . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.3.5 Measuring current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.6 Measuring current with simultaneous DC 24 V voltage supply . . . . . . . . . . . . . . 36
6.3.7 Measuring voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.3.8 Measuring voltage with simultaneous DC 24 V voltage supply . . . . . . . . . . . . . . 37
6.3.9 Simulating current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.10 Simulating current with simultaneous DC 24 V voltage supply . . . . . . . . . . . . . . 38
6.3.11 Manual pressure value acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.4 Calibrate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.4.1 Preparing the calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.4.1.1 Selecting the test item . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.4.1.2 Selecting the reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.4.1.3 Determining calibration points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.4.2 Carrying out the calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.4.3 Completing the calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.4.4 Representation of the calibration results . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.4.5 Repeating a calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.4.6 Deleting a calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5 Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5.1 Preparing a logger process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5.1.1 Setting test item 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5.1.2 Setting test item 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.5.1.3 Setting test item 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.5.1.4 Setting logger parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.5.2 Representation of logger results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.5.3 Analysing or repeating a logger process . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.5.4 Deleting logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

EN
WIKA operating instructions portable process calibrator, model CPH7000 5
14204467.02 03/2019 EN/DE
Contents
6.6 Switch test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.6.1 Switch test with external voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.6.2 Switch test with DC 24 V voltage supply by means of CPH7000 . . . . . . . . . . . . . 48
6.6.3 Preparing the switch test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.6.3.1 Selecting the test item . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.6.3.2 Selecting the reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.6.4 Carrying out and completing the switch test . . . . . . . . . . . . . . . . . . . . . . . . 51
6.6.5 Analysing or repeating a switch test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.6.6 Delete switch test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7. Faults 52
8. Maintenance, cleaning and recalibration 53
8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.3 Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
9. Return and disposal 54
9.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
10. Specifications 55
10.1 Digital process calibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
10.2 Temperature probe Pt100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10.3 Atmospheric module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10.4 WIKA-Wireless . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10.5 Reference pressure sensor model CPT7000. . . . . . . . . . . . . . . . . . . . . . . . . . 58
10.6 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.7 Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.8 Patents, property rights. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.9 Dimensions in mm (in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11. Accessories 62
Appendix: EU declaration of conformity model CPH7000 126
Appendix: EU declaration of conformity model CPT7000 127

EN
6WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
1. General information
■The process calibrator model CPH7000 described in the operating instructions has been designed and manufactured using
state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001 and ISO 14001.
■These operating instructions contain important information on handling the instrument.Working safely requires that all
safety instructions and work instructions are observed.
■Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
■The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and
readily accessible to skilled personnel at any time. Pass the operating instructions on to the next operator or owner of the
instrument.
■Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
■The general terms and conditions contained in the sales documentation shall apply.
■Subject to technical modifications.
■Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with international standards.
■Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: CT 15.51
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
1.1 Software licence agreement
This product contains intellectual property, i.e. software programs which are licensed to be used by the end user/client
(hereinafter referred to as “end user”).
The license does not include the sale of the intellectual property.
The end user shall neither copy nor disassemble or recompile the software program.
The software program will be provided to the end user “as it is” without any guarantees, either expressly or
implicitly, including but not limited to warranties of marketability and fitness for purpose.The end user has to bear
the entire quality and performance risk related to the software program.
Both WIKA and its suppliers are not liable for any damage to the end user (including but not limited to general,
special, consequential or incidental damage, including loss of business profits, business interruption, loss of
business information, and the like) resulting from or relating to the delivery, use or performance of the software
program.
The software included in this product contains copyrighted software that is licensed under the GPL/LGPL. A copy of the
license texts is included in the packaging of this product.You may obtain the complete Corresponding Source code from
us for a period of three years after our last shipment of this product and/or spare parts therefor, which will be no earlier than
01/01/2030, for a fee of 10 €. Please use our contact form at CTSer[email protected] and write „Corresponding Source for
CPH7000“ in the memo line.This offer is valid to anyone in receipt of this information.
1. General information

EN
WIKA operating instructions portable process calibrator, model CPH7000 7
14204467.02 03/2019 EN/DE
2. Short overview
2.1 Overview
12
1 2 3
5
14
13
15 4
6
8
7
9
11 10
1Connection for external pressure sensor - channel 1
(only CPT7000)
2Internal sensor for test item
(optional, available only in conjunction with the
mechanical hand pump)
3Connection for external pressure sensor - channel 2
(only CPT7000)
4Sockets for 4 mm plugs for measuring and simulating
current (DC)
5Connection for atmospheric module or temperature
probe Pt100
6Display
7LED charging indicator
8Home button
9Mechanical hand pump (optional)
10 Fine adjustment valve
(optional, available only in conjunction with the
mechanical hand pump)
11 Vent valve
(optional, available only in conjunction with the
mechanical hand pump)
12 Switch to switch between pressure build-up or vacuum
(optional, available only in conjunction with the
mechanical hand pump)
13 Connection for power supply unit
14 ON/OFF key
15 Sockets for 4 mm plugs for measuring and simulating
voltages (DC)
1. General information / 2. Short overview
WARNING!
Installing modified versions of open source software components on the product will result in the loss of warranty.
Also support service and software updates will be refused.
Make sure to follow the safety precautions in the operating instructions. Improper access to the instrument is likely
to result in its damaging.

EN
8WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
2.2 Description
The model CPH7000 process calibrator is a battery-powered, precise, portable calibrator for the on-site calibration and
checking of analogue pressure measuring instruments, pressure transmitters and process transmitters. Furthermore, pressure
switches can be checked and the switch point determined.With the CPH7000, not only can transmitters be checked, but also
simulated and tested.
The integrated hand pump (optional) enables you to generate a pressure of up to 25 bar (360 psi).This pressure can be
measured by means of an integrated pressure sensor (optional, available only in conjunction with the pump).
For pressure ranges of -1 ... 10,000 bar (-14.5 ... 145,000 psi), the model CPT7000 external precision pressure transmitter is
available. In conjunction with the data logger, it is also possible to carry out a leakage test in a very simple manner.
2.3 Scope of delivery
■Process calibrator model CPH7000
■Power supply unit
■Operating instructions
■Service case with 2 connection cables (4 mm plugs)
■Calibration certificate
Cross-check scope of delivery with delivery note.
2.4 Product passport
The product passport can be retrieved from the product page or directly from the corresponding web application (https://portal.
wika.com/serial/).
3. Safety
3.1 Explanation of symbols
DANGER!
... indicates a directly dangerous situation resulting in serious injury or death, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to property or the
environment, if not avoided.
DANGER!
... identifies hazards caused by electrical power. Should the safety instructions not be observed, there is a risk of
serious or fatal injury.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
2. Short overview / 3. Safety

EN
WIKA operating instructions portable process calibrator, model CPH7000 9
14204467.02 03/2019 EN/DE
3. Safety
3.2 Intended use
The model CPH7000 process calibrator is a portable multi-function calibrator for the calibration of a variety of measuring
instruments. An integrated mechanical hand pump (optional) enables you to generate pressures up to 25 bar (360 psi); an
integrated electrical module also enables you to supply transmitters or sensors with a current of max. 30 mA (voltage (idling) =
DC 24 V), in addition to the typical measurement of current and voltage signals.
The process calibrator can be used to carry out and document a complete calibration process. In addition to the functions
of displaying, measuring and calibrating, the functions logging and switch test are also available.The use of the WIKA-Cal
software also allows a complete documentation.
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed. Improper handling or operation of the
instrument outside of its technical specifications requires the instrument to be taken out of service immediately and inspected
by an authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from humidity, impacts, strong magnetic
fields, static electricity and extreme temperatures, do not insert any objects into the instrument or its openings). Plugs and
sockets must be protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
3.3 Improper use
DANGER!
Danger to life from explosion!
Improper use of the process calibrator causes a risk of explosion that can result in death.
▶The process calibrator should not be thrown into fire, since the built-in battery can explode.
WARNING!
Injuries, material and environmental damage due to an improper use of the process calibrator!
Improper use of the process calibrator causes a direct danger to life.
▶Do not throw the process calibrator into water → this can result in the destruction of the safety circuit, in heat
generation, in inflammation, in the formation of oxyhydrogen or corrosion and in the generation of electrolytes.
▶Overcharging, reverse charging and too high charging currents can result in fire or excessive gas formation.
▶The use of incorrect, unsuitable power supply units can lead to overheating, fire and to the destruction of the
rechargeable battery.
▶Crushing due to mechanical damage may result in the escape of electrolytes, in an internal short circuit, in
heating or fire.
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
▶Refrain from unauthorised modifications to the instrument.
▶Do not use the instrument within hazardous areas.
▶There must be no external pressure acting on the CPH7000.
▶Do not use the CPT7000 with abrasive or viscous media.
▶Only dry, clean air can be used as medium for the CPH7000.
▶It is forbidden to open the instrument.
▶Do not open the service flap on the rear. Only the manufacturer is allowed to open it.
▶Do not apply a voltage greater than the specified voltage to the instrument. See chapter 10 “Specifications”.
▶Make sure that the test probes never contact a voltage source while the test leads are connected to the
terminals.
▶Do not use the CPH7000 if it is damaged. Before using the process calibrator, check that there are no cracks
or missing plastic parts on the case. Pay particular attention to the insulation of the connectors.
▶Select the proper function and correct measuring range for the measurement.

EN
10 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
▶When using the electrical module, set the measurement type first and then connect the connection cables.
▶Only tighten the vent valve by hand, as it is a needle valve.Tightening with excessive force can result in a
damage of the needle valve or the sealing.
▶Inspect the test leads for damaged insulation or exposed metal. Check the continuity of the leads. Damaged
test leads should be replaced before using the process calibrator.
▶When using test probes, keep fingers away from the test probe contacts. Keep your fingers behind the test
probes' finger guards.
▶First connect the common lead, and then the live lead.When disconnecting, remove the live test lead first.
▶Disconnect test leads before changing to another measurement or source function.
▶When the battery indicator lights up red, load the rechargeable battery of the CPH7000, in order to avoid an
incorrect display or a data loss.
▶To avoid any damage to the process calibrator or to the inspection equipment, always use the proper
connection cable, function and area for the respective measuring application.
▶The switch valve can be switched from positive pressure to vacuum or from vacuum to positive pressure only
in depressurized state.
▶Only use the accessories specified and authorized by WIKA.
▶If pressure is applied over a long period of time, the pump can be damaged.
▶Only use pressure sensors model CPT7000! The use of other pressure sensors can result in the destruction of
the measuring instrument and of the pressure sensor.
▶Make sure that the internal sensor is not overloaded by the pump.
The electrical module can generate max. 30 mA and DC 24 V and measure max. 30 mA and DC 30 V.
With the integrated pump, pressures from -0.85 ... +25 bar (-12.3 ... +360 psi) can be generated.
Any use beyond or different to the intended use is considered as improper use.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for legal obligations regarding safety at
work.
The safety instructions within these operating instructions, as well as the safety, accident prevention and environmental
protection regulations for the application area must be maintained.
The operator is obliged to maintain the product label in a legible condition.
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to equipment.
▶The activities described in these operating instructions may only be carried out by skilled personnel who have
the qualifications described below.
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowledge of country-specific regulations,
current standards and directives, are capable of carrying out the work described and independently recognising potential
hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
3. Safety

EN
WIKA operating instructions portable process calibrator, model CPH7000 11
14204467.02 03/2019 EN/DE
3.6 Labelling, safety marks
3.6.1 Product label for process calibrator model CPH7000
The product label is located on the rear, in the upper part.
1Approval-related data for radio approval 6Accuracy
2Safety instructions for radio approval 7Input signals
3Permissible ambient temperature 8Pressure measuring range
4Serial number 9Model
5Date of manufacture (month-year)
3.6.2 Product label for reference pressure sensor model CPT7000
1Serial number
2Serial number external sensor
3Date of manufacture (month-year)
4Accuracy
5Pressure measuring range
6Model
Symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.
9
8
7
256 4 3
1
6
5
4
3
1
2
3. Safety

EN
12 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
4. Design and function
4. Design and function
The model CPH7000 process calibrator is a battery-powered, precise, portable calibrator for the on-site calibration and
checking of analogue pressure measuring instruments, pressure transmitters and process transmitters. Furthermore, pressure
switches can be checked and the switch point determined.With the CPH7000, not only can transmitters be checked, but also
simulated and tested.
The CPH7000 can be operated by means of a clearly structured touchscreen. The integrated rechargeable battery allows the
CPH7000 to be used daily in the field.
4.1 Design
The CPH7000 can optionally be equipped with an integrated reference pressure sensor as well as an integrated manual
pressure generation (pressure sensor only available in conjunction with pressure generation), with which pressures from
-0.85 ... +25 bar (-12.3 ... +360 psi) can be generated.
A pressure connection enables a calibration item (test item) to be connected to the reference pressure sensor as well as to
the integrated pressure generation of the CPH7000.The integrated pump (optional) allows you to generate different pressure
ratings and thus to carry out a calibration.
An external separate pressure generation up to 25 bar (360 psi) is thus not necessary.
Transmitter output signals can simultaneously be measured via an electrical module, thus enabling the simultaneous display
of the reference and test item value on the screen.This allows a complete transmitter calibration to be carried out with just the
CPH7000.
In addition to the integrated sensor, it is also possible to connect up to two external reference pressure sensors to the
CPH7000, thus enabling the measurement of pressure also in other pressure ranges or differential pressure. Communication
between CPH7000 and pressure sensors is digital.
The ambient atmospheric pressure is measured via an integrated barometric reference which can optionally be integrated
in the case.This also enables gauge pressure sensors to be converted into absolute pressure.The ambient conditions
(temperature and humidity) can be measured by means of an additional external atmospheric module.To the connection for
the atmospheric module, it is also possible to connect a temperature probe Pt100.
4.2 Functions and data transfer with the WIKA-Cal calibration software
The calibrator offers the possibility to create calibration routines quickly and easily, but also to carry out preconfigured
calibration routines. An integrated memory allows the recording of logger cycles as well as calibrations on site.
The completed calibration processes can be transmitted to a PC via WIKA-Wireless.
As soon as a connection exists over WIKA-Wireless, it can communicate with the WIKA-Cal calibration software from version
1.0.66. Data can also be transferred using Windows Explorer.
A complete paperless transmitter calibration can be carried out using the CPH7000 and WIKA-Cal.
As the CPH7000 features an electrical module, the hand-held also has to be set up as a multimeter.
Compatibility between the CPH7000 process calibrator and the WIKA-Cal evaluation software is only possible
with the following firmware versions and higher.
CPH7000: 1.1.4 or higher
WIKA-Cal: 1.0.82 or higher

EN
WIKA operating instructions portable process calibrator, model CPH7000 13
14204467.02 03/2019 EN/DE
4. Design and function
4.2.1 WIKA-Cal configuration (also possible with the demo version)
1. In WIKA-Cal, in the “Object manager”, open the menu item “Workplace”.
⇒The CPH7000 must first be defined as a standard and assigned to the workplace.
2. Call up the Wireless function.
⇒The Wireless monitor will open.
3. Click into the address field.
⇒The address will be displayed automatically. If required, correct this.
⇒The communication is working properly if the pressure value displayed on the instrument is shown after pressing the
[Test] button.
4. Via [Config], the configuration dialogue for the instrument can be called up.
In the configuration window, the four functions “General information”, “Sensors”, “Signal output” and “Ambient
conditions” are available.
General information
Here are all the general parameters of the CPH7000 used.
Specific pressure units with the associated factor can be set or also specific temperature coefficients entered for the Pt100
probe.

EN
14 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
4. Design and function
Sensors
The following sensors can be selected:
INT = Internal reference pressure sensor
EXT1 = External CPT7000 reference pressure sensor to channel 1
EXT2 = External CPT7000 reference pressure sensor to channel 2
RTD = External Pt100 temperature probe
IN = Integrated electrical module
EXT1, EXT2 and RTD are only displayed when these are also connected.
Signal output
Here, the current and voltage values can be read in and overwritten. The prerequisite for this is that under “Measure”, a
current or voltage channel has also been selected.

EN
WIKA operating instructions portable process calibrator, model CPH7000 15
14204467.02 03/2019 EN/DE
4. Design and function
Ambient conditions
Here, the measured values from the atmospheric module and barometer are displayed.
4.2.2 WIKA-Cal - Cal-Template
The following options are available with the “Cal-Template”:
“New calibration”, “Recalibration”, “Upload” and “Download”
New calibration
A new calibration certificate will be created. All parameters from the calibration item and reference instrument must be
entered, and then the calibration can start. By pressing on the “Measuring results” table, a new window opens in which the
communication parameters are requested. Here, the CPH7000 used must be selected, then the measurement starts.

EN
16 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
4. Design and function
Recalibration
Select the required certificate from the database. Only completed certificates (marked with a green tick) can be recalibrated. A
copy of the certificate will be created and opened, and when the [Measuring results] table is pressed, the measurement can
be restarted.
Upload
With “Upload”, a calibration routine can be defined and installed on the CPH7000.
All parameters must be entered.The measuring points and measurement series can be specified by default.With upload, the
calibration routine will be created and uploaded to the CPH7000.
Download
The calibrations stored on the CPH7000 can be downloaded and archived using “Download”.
▶Select reference instrument and click on “Download”.
⇒Then the available data sets are shown in the selection window.
4.2.3 WIKA-Cal - Log-Template
The following options are available with the “Log-Template”:
New log
The “New Log” function opens a new logger protocol.
Following the input of all parameters, the “Wireless address” line must be pressed in the communication window. Select the
CPH7000 used and confirm.
By pressing the [Measuring results] graphic, the logger process is started.

EN
WIKA operating instructions portable process calibrator, model CPH7000 17
14204467.02 03/2019 EN/DE
4. Design and function
Repeat logging
Similar to “Recalibration”, it is possible to repeat logger sequences.
Download
The logger sequences stored on the CPH7000 can be downloaded and archived using “Download”.
4.2.4 WIKA-Cal - switch test template
Here, only the option “Download” is available:
The switch test procedures stored on the CPH7000 can be downloaded and archived using “Download”.
The switch test function in the WIKA-Cal software has to be activated under “Object manager / Templates”.
When this box is checked, the template for the switch test will appear in the main menu window (see chapter
4.2.2 “WIKA-Cal - Cal-Template”).
4.3 Voltage supply
The internal Lithium-Ion battery, which can be easily charged with the battery charger supplied with the equipment, serves as
the voltage supply for the instrument.
To charge the CPH7000 rechargeable batteries, the mains plug of the power supply unit must always be plugged in to a mains
socket and accessible, so that one can always remove it from the mains socket without difficulty.
When using the electrical module, better results are achieved if the battery is not charged during the
measurement.
To avoid incorrect measurements, recharge the battery as soon as the battery indicator flashes. If the battery discharges too
much, the CPH7000 will automatically turn off.
The battery life is up to 8 hours in continuous operation (without backlighting, WIKA-Wireless is not active and the electrical
module does not supply voltage/current).
In the upper right corner of the display there is a symbol for the battery capacity. For instructions on batteries see chapter
4.10.3 “Battery symbol”).

EN
18 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
4. Design and function
4.3.1 Rechargeable battery
The integrated lithium-ion battery is subject to the requirements of the Dangerous Goods Directive. Special
requirements for packaging and labelling must be observed when shipping. A dangerous goods expert must be
consulted when preparing the package.
Do not ship the CPH7000 if the rechargeable battery is damaged or defective.
Observe the different dangerous goods requirements relative to the respective modes of transport and any other
national regulations.
The rechargeable battery is permanently installed in the model CPH7000 process calibrator.
The battery can be charged only using the power supply unit included in the scope of delivery.
4.3.2 Using the power supply unit
DANGER!
Danger to life caused by electric current
Upon contact with live parts, there is a direct danger to life.
▶Only use the power supply unit from WIKA supplied with the instrument!
▶If there is any visible damage to the case or the wiring, do not use the power supply unit!
▶Never install nor store the power supply unit in the following locations, as this can lead to a failure in operation:
■Places where there is strong humidity or condensation
■Outdoors
▶Disconnect the power supply unit from the mains supply when it won’t be used for a longer period of time.
▶The power supply unit is maintenance-free. It must not be opened (danger of electrical shock).
▶Before cleaning, disconnect the power supply unit from the mains supply. Do not clean with chemical cleaning
agents. Only clean with a dry cloth.
▶The power supply unit may only be used at an ambient temperature of 0 ... 40 °C (32 ... 104 °F) (humidity: up to
90 % relative humidity, non-condensing).
To avoid measurement uncertainties, use the CPH7000 only without a connected power supply unit.
The instrument is delivered with a charge level of 25 ... 50 % and should be fully charged once before being used.
The battery capacity status (charge state in %) is displayed shortly after the instrument is switched on.
When the power supply unit is connected to the CPH7000, the battery will be charged, even if the CPH7000 is
switched off.
The typical charging time of the rechargeable battery is < 5 h.
■When the power supply unit is no longer being used, the mains plug should be disconnected from the mains socket. Do not
leave the rechargeable battery connected to the power supply unit for longer than one day, since overcharging can shorten
its service life.
■Should the rechargeable battery still not be fully charged after 24 hours, contact the manufacturer. When not being used, a
fully charged battery will lose its charge over time.
■Extreme temperatures have an adverse effect on battery charging. As a result, the battery may first need to be either cooled
or warmed, as appropriate.
■When the battery is almost completely discharged, the message “low BAT” appears in the display.To avoid a data loss, the
instrument must be charged immediately.

EN
WIKA operating instructions portable process calibrator, model CPH7000 19
14204467.02 03/2019 EN/DE
4. Design and function
4.3.3 During charging
WARNING!
Physical injuries and damage to property and the environment
The temperature range over which the Lithium-Ion battery can be charged is 0 ... 40 °C (32 ... 104 °F).
▶Do not charge the lithium-ion battery outside this temperature range.This may result in heating or demolition.
In addition, the performance of the Lithium-Ion battery can be affected and the service life reduced.
4.4 Electrical connections
1Connection for external pressure sensor model CPT7000 to channel 1 (EXT 1)
2Connection for external pressure sensor model CPT7000 to channel 2 (EXT 2)
3Connection for atmospheric module (AMB) or temperature probe Pt100 (RTD)
4Connection for power supply unit
5Vout - Sockets for 4 mm plugs: Connection for DC 24 V supply
6Vin - Sockets for 4 mm plugs: Connection for measuring voltages (DC 30 V) (V_IN)
7mA - Sockets for 4 mm plugs: Connection for measuring and simulating current (30 mA) (mA)
8GND - Ground connection (GND)
Any external circuit connected to this instrument must be protected against electrical shock by additional
insulation or insulation enhanced against potentially dangerous active voltages.
3
4
5 6 7 8
2
1

EN
20 WIKA operating instructions portable process calibrator, model CPH7000
14204467.02 03/2019 EN/DE
4. Design and function
4.5 Mechanical connection
9G 1/8 female thread
Optionally, it can be used only in conjunction with the hand pump (INT).
10 Pumping unit
In order not to put a strain on the case, it is recommended to hold the pressure connection on the CPH7000 by
means of an open-ended spanner.
The connections of the CPH7000 without pumping unit are provided with plugs. In order to maintain the IP
protection class, these plugs should not be removed.
4.6 Pump
CAUTION!
Physical injuries and damage to property and the environment
Improper handling of the pump can damage the process calibrator.
▶Do not use the pump when switched off.
The pump is a consumable. Regular maintenance is recommended after 100,000 pump cycles at the latest. Under normal use
conditions, this corresponds to a service life of 2 years.
Recommended procedure in vacuum range
1. Always pull out the piston rod until it stops.
2. Wait 5 seconds.
3. Carry out another pumping stroke until the desired pressure is reached.
4.7 Barometer
The barometric sensor is a very stable, high-accuracy absolute pressure sensor designed to measure the daily atmospheric
pressure.
It can be used to emulate gauge pressure using an absolute pressure sensor or to emulate absolute pressure using a gauge
pressure sensor. For a seamless absolute pressure emulation, a bi-directional measuring range with a start of measuring
range of -1 bar (-14.5 psi) is recommended.
4.8 Reference pressure sensor model CPT7000
For the CPH7000 process calibrator, a variety of reference pressure sensors with measuring ranges from 250 mbar (4 psi) up
to 10,000 bar (145,000 psi) and vacuum pressure sensors with an accuracy of 0.025 % FS are available. These sensors can
be replaced on the instrument very quickly and without using any tools.
When the process calibrator is switched on, the connected reference pressure sensor will be automatically detected.
9
10
Other manuals for CPH7000
1
Table of contents
Languages:
Other WIKA Test Equipment manuals

WIKA
WIKA mensor CPG2500 User manual

WIKA
WIKA CPB 5800 User manual

WIKA
WIKA CPH7000 User manual

WIKA
WIKA CalibratorUnit CPU6000 Series User manual

WIKA
WIKA Wally Box III CPH7600 User manual

WIKA
WIKA CPB 5000 DP User manual

WIKA
WIKA CPB 5000 User manual

WIKA
WIKA CEP6500 User manual

WIKA
WIKA DELTA-comb DPGS40TA User manual

WIKA
WIKA CPB3800 User manual