EnglishDeutsch
 logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. WPI
  6. •
  7. Laboratory Equipment
  8. •
  9. WPI NANOLITER2010 User manual

WPI NANOLITER2010 User manual

This manual suits for next models

1

Other WPI Laboratory Equipment manuals

WPI LWCC-M-10 User manual

WPI

WPI LWCC-M-10 User manual

WPI Z-LITE User manual

WPI

WPI Z-LITE User manual

WPI NANOLITER2020 User manual

WPI

WPI NANOLITER2020 User manual

WPI Nanoliter 2000 User manual

WPI

WPI Nanoliter 2000 User manual

WPI MICRO-ePUMP User manual

WPI

WPI MICRO-ePUMP User manual

WPI SpectroPipetter SPT User manual

WPI

WPI SpectroPipetter SPT User manual

WPI PV850 User manual

WPI

WPI PV850 User manual

WPI MPS-4 User manual

WPI

WPI MPS-4 User manual

WPI MSI-150 User manual

WPI

WPI MSI-150 User manual

WPI M3301 User manual

WPI

WPI M3301 User manual

WPI CLEANRAY-1000 User manual

WPI

WPI CLEANRAY-1000 User manual

WPI Eagle E2 User manual

WPI

WPI Eagle E2 User manual

WPI CLEANRAY-2000 User manual

WPI

WPI CLEANRAY-2000 User manual

WPI UltraPath 2 Parts list manual

WPI

WPI UltraPath 2 Parts list manual

WPI LWCC-M-10 User manual

WPI

WPI LWCC-M-10 User manual

WPI REG2 User manual

WPI

WPI REG2 User manual

WPI ST5191 User manual

WPI

WPI ST5191 User manual

Popular Laboratory Equipment manuals by other brands

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Conviron GEN1000-GE Operator's manual

Conviron

Conviron GEN1000-GE Operator's manual

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

IKA IC control operating instructions

IKA

IKA IC control operating instructions

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

Agilent Technologies 5800 ICP-OES user guide

Agilent Technologies

Agilent Technologies 5800 ICP-OES user guide

Endress+Hauser Cleanfit CPA875 operating instructions

Endress+Hauser

Endress+Hauser Cleanfit CPA875 operating instructions

NI PXI-5422 CALIBRATION PROCEDURE

NI

NI PXI-5422 CALIBRATION PROCEDURE

Collomix Aqix operating instructions

Collomix

Collomix Aqix operating instructions

SPEX SamplePrep 6875 Freezer/Mill Series operating manual

SPEX SamplePrep

SPEX SamplePrep 6875 Freezer/Mill Series operating manual

Ocean Insight FLAME-NIR+ Installation and operation manual

Ocean Insight

Ocean Insight FLAME-NIR+ Installation and operation manual

Parker ALIGN-MG-NA Installation, operation and maintenance manual

Parker

Parker ALIGN-MG-NA Installation, operation and maintenance manual

BD 644787 user guide

BD

BD 644787 user guide

DENTAURUM Compact Megaplus Instructions for use

DENTAURUM

DENTAURUM Compact Megaplus Instructions for use

Biuged Laboratory Instruments BGD 626 instruction manual

Biuged Laboratory Instruments

Biuged Laboratory Instruments BGD 626 instruction manual

VWR SAS Super IAQ instruction manual

VWR

VWR SAS Super IAQ instruction manual

illumina MiSeqDx reference guide

illumina

illumina MiSeqDx reference guide

https://manuals.online logo
https://manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2026 Manuals.Online. All Rights Reserved.

Instrumenting scientific ideas
WORLD
PRECISION
INSTRUMENTS
NANOLITER2010
Nanoliter Injector with SMARTouch™ Controller
Serial No._____________________
www.wpiinc.com
QUICK START GUIDE
061217
Copyright © 2017 by World Precision Instruments. All rights reserved. No part of this publication may
be reproduced or translated into any language, in any form, without prior written permission of World
Precision Instruments, Inc.
Nanoliter 2010 with SMARTouch™ Controller
World Precision Instruments 1
Get started with your Nanoliter 2010 and SMARTouch™ Controller as quickly as
possible using these step by step instructions.
NOTE: This Quick Start Guide assumes that the operator back lled a micropipette
with mineral oil. For instructions on how to prepare the micropipette, see the
Nanoliter2010 Instruction Manual page 7.
1. Connect the AC power cord to the power supply unit and connect the power
supply unit to the MICRO2T controller. Plug the AC power cord into a wall outlet.
2 Connect the Nanoliter adapter cable to the output of channel 1 of the
SMARTouch™ controller and the RJ-11 connector of the Nanoliter2010 to the
RJ-11 female connector on the adapter. Make sure the embossed arrow on the
connector is facing upwards for proper pin alignment.
Adapter Cable connects to the
SMARTouch™ controller
Nanoliter 2010 Injector connection to
the adaper.
2 World Precision Instruments
3. Power on the unit. The following screen appears.
4. Touch the Initial Splash Screen anywhere. The MAIN MENU appears. Touch
DISPLAY ALL CHANNELS.
5. The syringe status screen appears. Press anywhere in the PUMP1 Display Area to
make Pump1 the active pump.
Nanoliter 2010 with SMARTouch™ Controller
World Precision Instruments 3
6. Touch the Congure button.
7. Press the Type box to bring up the Syringe Selection Screen.
8. In the Syringe Selection Screen, press the DWN key until you see the NL (Nanoliter
2010) type and press it. It will become highlighted. Then press Back to return to
the Pump Conguration Screen
4 World Precision Instruments
9. The CONFIGURATION screen appears. Touch the Volume Target display.
10. The Volume Target screen appears. Use the numerical keypad to enter 2500 nL.
Touch ENTER to set the volume and return to the CONFIGURATION screen.
11. Touch the DELIVERY RATE display.
Nanoliter 2010 with SMARTouch™ Controller
World Precision Instruments 5
12. The Volume Rate screen appears. Use the numerical keypad to enter a desired
injection rate in nL /sec. In this example, 500 nl/sec was entered. Touch ENTER to
set the delivery rate and return to the CONFIGURATION screen.
13. The Conguration screen shows that a target volume of 2500 nL was entered
and a delivery rate of 500 nL/s was entered. Touch BACK to return to the syringe
status {Home} screen.
14. Verify that the arrow inside the syringe is yellow and pointing in the injection
direction. If the arrow is not red and pointing to the left, touch DIRECTION to
change it. Touch the RUN button to start the injection.
6 World Precision Instruments
15. The injection will run until 2500 nL is injected and will then halt automatically.
Even though no injection actually happened, this will place the metal needle
plunger about 13.6 mm from the home (fully retracted) position. At this point
install the back lled micropipette on to the Nanoliter2010 injector as shown
below.
16. Press the DIRECTION button to change the direction to Withdraw. The arrow
turns green and points towards the back of the syringe depiction.
17. Press the CONFIGURE button.
Nanoliter 2010 with SMARTouch™ Controller
World Precision Instruments 7
18. Press the Volume Target area to bring up the Volume Target screen.
19. Use the keypad to enter the desired aliquot volume to be pulled into the
micropipette. We will enter 200 nL and press Enter.
20. Press the Delivery Rate area.
8 World Precision Instruments
21. Use the keypad to set the rate to 50 nL/sec and press Enter.
22. Press Back to go to the main Syringe information display.
23. Next place the micropipette tip into the aliquot to be loaded and press RUN. This
will pull 200 nL into the micropipette.