manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ADC
  6. •
  7. Thermometer
  8. •
  9. ADC Adtemp Temple Touch 427 User manual

ADC Adtemp Temple Touch 427 User manual

Other ADC Thermometer manuals

ADC Adtemp ULTRA 417 User manual

ADC

ADC Adtemp ULTRA 417 User manual

ADC Adtemp V 418 Owner's manual

ADC

ADC Adtemp V 418 Owner's manual

ADC Adtemp II 412 Maintenance manual

ADC

ADC Adtemp II 412 Maintenance manual

ADC Adtemp 424 User manual

ADC

ADC Adtemp 424 User manual

ADC Adtemp 429 User manual

ADC

ADC Adtemp 429 User manual

ADC Adimals 426 Manual

ADC

ADC Adimals 426 Manual

ADC Adtemp 413 User manual

ADC

ADC Adtemp 413 User manual

ADC Adtemp 413 User manual

ADC

ADC Adtemp 413 User manual

ADC Adtemp 433 User manual

ADC

ADC Adtemp 433 User manual

ADC Adtemp 427 TempleTouch User manual

ADC

ADC Adtemp 427 TempleTouch User manual

ADC Adtemp TempleTouch 427 User manual

ADC

ADC Adtemp TempleTouch 427 User manual

ADC Adtemp 422 Maintenance manual

ADC

ADC Adtemp 422 Maintenance manual

ADC Adtemp 422 Maintenance manual

ADC

ADC Adtemp 422 Maintenance manual

ADC Adimals 426 User manual

ADC

ADC Adimals 426 User manual

ADC Adtemp ULTRA 417 Maintenance manual

ADC

ADC Adtemp ULTRA 417 Maintenance manual

ADC Adtemp II 412 Maintenance manual

ADC

ADC Adtemp II 412 Maintenance manual

ADC Adtemp V 418 Maintenance manual

ADC

ADC Adtemp V 418 Maintenance manual

ADC Adtemp 415FL User manual

ADC

ADC Adtemp 415FL User manual

ADC Adtemp 423 User guide

ADC

ADC Adtemp 423 User guide

ADC Adtemp V 418 Maintenance manual

ADC

ADC Adtemp V 418 Maintenance manual

ADC Adtemp II 413 Maintenance manual

ADC

ADC Adtemp II 413 Maintenance manual

ADC Adtemp 412 User manual

ADC

ADC Adtemp 412 User manual

ADC Adtemp 427 Temple Touch Maintenance manual

ADC

ADC Adtemp 427 Temple Touch Maintenance manual

ADC Adtemp Mini 432 User manual

ADC

ADC Adtemp Mini 432 User manual

Popular Thermometer manuals by other brands

Benetech GM900 instruction manual

Benetech

Benetech GM900 instruction manual

Parton PTD-400 manual

Parton

Parton PTD-400 manual

GAME 15900 owner's manual

GAME

GAME 15900 owner's manual

scigiene FP-180 instruction manual

scigiene

scigiene FP-180 instruction manual

Ametek Land UNO U1 600/1600C-L user guide

Ametek Land

Ametek Land UNO U1 600/1600C-L user guide

Delton 981A instruction manual

Delton

Delton 981A instruction manual

Taylor 5286498 instruction manual

Taylor

Taylor 5286498 instruction manual

Baker MM2P instruction manual

Baker

Baker MM2P instruction manual

Chino IR-FA Series instructions

Chino

Chino IR-FA Series instructions

Vive DMD1010 user manual

Vive

Vive DMD1010 user manual

Tarmo 320281 manual

Tarmo

Tarmo 320281 manual

Citizen CTD 504 instruction manual

Citizen

Citizen CTD 504 instruction manual

TFA 14.1500 instruction manual

TFA

TFA 14.1500 instruction manual

Gotemp THB4002 user guide

Gotemp

Gotemp THB4002 user guide

GREISINGER G 1202 Quick reference guide

GREISINGER

GREISINGER G 1202 Quick reference guide

Amprobe TMD-53W user manual

Amprobe

Amprobe TMD-53W user manual

Trotec BP10 operating manual

Trotec

Trotec BP10 operating manual

Radiance instruments RT616 Series product manual

Radiance instruments

Radiance instruments RT616 Series product manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

AdtempTM 427
TempleTouchTM
Thermometer
USE, CARE &
MAINTENANCE
Intended Use:
A Temple Touch thermometer is a device which enables accurate body tem-
perature measurements obtained by using a unique probe, which measures
the heat flow from the blood vessels to the skin and converts it to body temper-
ature from the temple area in approximately 6-8 seconds.
Safety Information:
• High, prolonged fever requires immediate medical attention, especially for young children.
• Do not allow children to take their temperatures unattended.
• Read the enclosed instructions carefully to ensure accurate temperature readings. Temperature readings are
affected by many factors including exercise, drinking hot or cold beverages prior to taking a reading as well
as technique.
• WARNING: Keep batteries out of children’s reach. Swallowing a battery could be harmful. Batteries
should not be recharged or placed into extreme heat as they may explode.
• WARNING: It is recommended to remove the batteries from the thermometer when it is not used for a
long period. Do not dispose of batteries in fire, recharge, put in backwards, disassemble, mix with used
or other battery types – they may explode or leak and cause injury. Waste batteries should be deposited in a
batteries recycling trash only.
• WARNING: This product may contain a chemical known to the state of california to cause cancer, birth
defects, or other reproductive harm.
• CAUTION: Do not drop the thermometer or expose to heavy shock.
• CAUTION: Do not immerse thermometer in water or any other liquid.
• CAUTION: Do not store the unit in direct sunlight, at a high temperature, in high humidity or dust.
• CAUTION: Stop using the thermometer if it operates erratically or if display malfunctions.
• CAUTION: Clean the thermometer probe before storing.
• CAUTION: Do not attempt to disassemble the unit except to replace the batteries.
• CAUTION:
Do not autoclave. Please note the cleaning procedures described in this manual
Specifications
Feature Performance / Data: Measurement time 6-8 seconds
Measurement location: Temple area of the forehead
Measurement range: 95°F to 107.6°F (35°C to 42°C)
Operation environment: 60.8°F-104°F (16° C - 40° C)
Humidity: 15-95% (non condensing)
Storage environment: -4°F to 122°F (-20° C - 50° C)
Humidity: 15-95% (non condensing)
Measurement readout: Celsius or Fahrenheit
Accuracy: Complies with ASTM E 1112-00 standard requirements
Display: TN LCD 0.63" X 0.44" (16 mm X 11 mm)
Battery: 2 AAA 1.5volt batteries
Battery life: Minimum 1000 full duty cycles over 24 month period
Weight: 1.3oz (37gr) including batteries
Dimensions: 4.7" x 1.17" x 1" (12 cm x 3.0 cm x 2.4 cm )
Memory: Last reading display when turned on
Auto shut off: After 60 seconds
Case material: Medical grade ABS plastic
Probe material: Medical grade Stainless steel ,medical grade ABS plastic
Warranty: 2 year
IB p/n 93-427-00 rev 3 Printed in China
Display
Window
Multi Function
Button
Parts and Assembly
Non invasive
sensor
WARRANTY
This Thermometer is guaranteed for a period of 24
months from date of purchase. This warranty does not
cover batteries, or damage resulting from improper
handling or use. Specific legal rights stipulated in your
state may guide this warranty. If you have a legitimate
claim under this warranty, send item(s) postage paid
to Service Center, Attn: Repair Dept., 55 Commerce
Dr., Hauppauge, NY 11788. Please include your
name and address, phone no., proof of purchase, and
a brief note explaining the problem, your return ad-
dress and 2.00 to cover shipping and handling.
This manual is available online in a
variety of lan ua es,
follow the links for lan ua e options.
www.adctoday.com/care
To re ister your product visit us at
www.adctoday.com/register
Da ao Electronics Factory
Xia Nan Industrial Zone
Yuan Zhou Town
Huizhou City, Guan don
CHINA
Medisim Limited
G.G. Communication Centre,
Neve Ilan, 90850 Israel
Distributed by: ADC®
55 Commerce Drive
Hauppau e NY 11788
Made in China
www.adctoday.com
Patent #7597668
EC REP
0473
93-427-00 rev 3.qxp_Layout 1 8/12/16 4:55 PM Page 1
Quick Start:
1. Press and release the MULTI-FUNCTION button. 3. Hold firmly to temple, wait for beep.
2. Wait 2 seconds. 4. Read results.
5. Before re-measuring: wait at least 1 minute.
Measuring Guidelines:
1. Measurement body location is the temple, in proximity to the temporal artery, between the end of the
eyebrow and hairline.
2. Do not place the thermometer on scarred tissue compromised by skin disorder, open sores or abrasions
and patients in trauma.
3. Temperature readings are affected by many factors including exercise and drinking hot or cold beverages
prior to taking a reading.
4. Consult your physician when using drug therapies that may raise the local skin temperature in the
forehead; this may lead to an incorrect reading.
5. Do not place the thermometer on skin exposed to direct sunlight, fireplace heat, air conditioner flow and
cold/warm compress therapies.
6. Wait at least one inute between consecutive easure ents for the unit to readjust to
roo te perature.
7. If the thermometer has been stored at a cold temperature, allow the unit to warm naturally to room
temperature before using.
8. Performance may be degraded if operated or stored outside stated temperature/humidity ranges.
Taking Temperature:
Use this thermometer on a clean dry forehead. Wipe away any perspiration,
move hair aside from the temporal area and apply the probe to bare skin.
1. Press and release the Multi-Function button, all segments will be displayed along with a
single audible beep (Figure 1).
2. The previous result will be displayed for 2 seconds (Figure 2).
3. Wait for the hourglass icon to disappear (Figure 3). Do not try to take a measurement
before the hourglass icon disappears, as this may compromise the results.
4. Place the thermometer's probe in the center of the temple area between the end of
the eyebrow and the hairline. Gently press and hold it against the temple to ensure
good contact with the skin is maintained (Figures 4a + 4b). Correct placement and
good contact with the skin ensure reliable readings.
5. As soon as the thermometer is in place, the measurement will start automatically.
A running circle will appear indicating that the measurement is progressing successfully
(Figure 5). Make sure that the thermometer and the patient do not move while
measurement is in progress.
6. A temperature reading will be displayed within 6-8 seconds. One long beep will be
heard to indicate that measurement is complete and temperature will be displayed for about 1 minute.
7. Press and release the Multi-Function button to turn the thermometer off, or it will shut off automatically
after 1 minute.
If you need to take another reading – wait at least a inute after turning the ther o eter
off and begin again. If you take a easure ent before 1 inute has elapsed, the device
will perfor a 10 to 0 countdown to allow the device to re-adjust.
DO NOT TRY TO MEASURE DURING THE COUNTDOWN. If you atte pt to do so,
the device will display the previous result, beep 8 ti es and turn itself off auto atically.
Setting the Operating Scale °F or °C
The device setting is by default fahrenheit (°F). To change the scale to °C, or back to °F:
1. While off, press and hold the Multi-function button for 5 seconds, on the screen you
will see the icon (°F, °C) changes.
2. Release the button, the hourglass icon will disappear in 5 seconds indicating that
the device is ready for use.
Changing the battery:
To install/replace, first switch off the device. Remove the battery cover by sliding it out of its
housing. Firmly attach the batteries to their place, making sure polarity is correct. (see engraving on the
bottom of battery compartment). Slide the battery cover back into place.
Troubleshooting and special indications:
The thermometer has been designed to display feedback-warnings to ensure proper usage
and accurate temperature measurement. When an error appears, the device will "beep" 5
times, and then the thermometer will turn off automatically. In the event of a malfunction or in-
correct temperature measurement, the following warnings will be displayed:
HI: If the measured temperature is above 107.6°F (42°C) the device will beep and display HI”.
Remedy: Turn the unit OFF, wait 1 minute and then ON again. Retake temperature using proper technique.
If still reads HI, this indicates that the patient has a high fever. Seek immediate medical attention. NOTE: This is
not a device malfunction.
LO: If the measured temperature is below 95°F (35°C) the device will beep and display “LO”.
Remedy: Turn the unit OFF, wait 1 minute and then ON again. Retake temperature using
proper technique. If still reads LO, this indicates that the patient has a low body temperature.
NOTE: This is not a device malfunction.
Er.1: If the thermometer is too cold or too hot to operate. The device will beep 5 times and
display “Er.1”
Remedy: 1. Verify room temperature within product range 60.8°F-104°F (16° C - 40° C).
2. Wait 1 minute and then retake temperature using proper technique.
Er.2: If the temperature did not register because of poor contact, patient movement or poor
placement, the device will beep 5 times and display “Er.2”
Remedy: Turn the unit OFF, wait 1 minute, turn it ON again and retake temperature using
proper technique.
Er.3: If the unit malfunctioned, the device will beep 5 times and display “Er.3”
Remedy: Do not use the thermometer – Call 1-800-232-2670 for service.
Low Battery: Battery icon is on continuously.
Remedy: The battery has enough remaining life for approximately 10 more measurements
Drained Battery: Battery icon appears as unit is turned on and the device turns itself OFF.
Remedy: Replace battery.
Standards:
Complies with EN 12470-3 and ASTEM E 1112-00
How to care for this product:
• Wipe the tip and probe with a cotton swab moistened with 70% isopropyl alcohol before & after each
use. After cleaning, wait at least 2 minutes before taking a measurement for the unit to re-adjust to
room temperature.
• Use only a soft dry cloth to clean the body of the thermometer (if necessary). Never use an abrasive
cleanser or submerge the thermometer in water or other liquids.
93-427-00 rev 3.qxp_Layout 1 8/12/16 4:55 PM Page 2