INTRODUCTION
Thank you for purchasing the Andatech® Prodigy 3.
The Andatech® Prodigy 3 is a state-of-the-art device used to detect Ethyl Alcohol from a valid
breath sample. It uses an electrochemical fuel cell to measure the concentration of alcohol from
the breath sample expelled from a human subject.
If alcohol is present, a corresponding voltage is generated from the fuel cell, which is proportional
to the alcohol content of the sample provided by a human subject. This voltage is then sent to the
CPU and converted to a subject’s BAC (blood alcohol concentration), which is then displayed.
The operator of this device should read the operation manual and be familiar with the device
operation before using it. Retain the manual for future reference.
Prodigy 3 Features
1. Stores more than 100,000 test records along with full data and photographs. Test records can be
uploaded to a computer for record keeping.
2. Uses disposable mouthpieces or sampling cups to prevent foreign matter from entering.
3. A hard plastic case is included for storing the device and its accessories for easy transport. When
not in use, the device and its accessories should be stored in this hard case to prevent damage.
4. Ability to take consecutive tests as needed. Recovery time between tests on average is 30
seconds (between positive tests).
5. Powered by a High-capacity lithium battery. The battery is located behind the sliding cover
on the back of the device. The level of power left in the battery is indicated on the screen. A
cigarette plug adapter is also included for use by the operator.
6. Storage temperature range is -30°C to 70°C and storage humidity is 15% to 90% RH (relative
humidity). Inappropriate storage of the device will shorten the lifespan of the fuel cell sensor.
7. Saves the GPS location of the test.
8. Immediate data transmission by 3G/4G and WIFI.
CONTENTS
SPECIFICATIONS
PRODUCT DIAGRAM
FEATURES
SCREEN DISPLAY
CONDITIONS FOR TESTING
OPERATING INSTRUCTIONS
PRINTER
DISCLAIMERS
LIMITS
WARRANTY REGISTRATION
CALIBRATION AND SERVICE
WARRANTY DETAILS
.....................................................3
................................................3
................................................................4
....................................................5
....................................6
...................................7
................................................................10
........................................................10
....................................................................11
..................................11
................................11
.............................................12

3
PRODUCT DIAGRAM
SPECIFICATIONS
CONTENTS:
1 x Prodigy 3 Breathalyser
1 x Hard Case
1 x Soft Pouch
2 x Rechargeable Batteries
1 x 12V Car Charging Cable
1 x 12V Power Adapter (AU Plug)
1 x Battery Charging Dock
1 x Micro USB Cable
1 x Sampling Cup
5 x Mouthpieces
10 x Printer Rolls
1 x User Manual
BAC Range 0.000% - 0.400%BAC
Result readings can also be changed to the following
formats. Please specify this during purchase or
calibration:
BAC BrAC
0.000 ~ 4.000 ‰ 0.000 ~ 2.000 mg/l BrAC
0.000 ~ 0.400%BAC 0.000 ~ 4.000 g/l
0~40.0 mg/100ml 0~200µg/100ml
0~2000µg/l
Accuracy ±0.005%BAC at 0.100%BAC
Warm Up Time Within 5 seconds
Response Time Within 5 seconds
Recycle Time 3 seconds
Operating System Android 5.1
Camera 4.3” TFT colour touch screen
GPS ±5M
Connection USB, WIFI
Printer Built-in Thermal Printer
Storage
Temperature
-30 to 70°C, 15 to 90%RH
Working
Temperature
-10 to 50°C
Power Supply 7.4V 2,000mAh NI-MH
rechargeable battery pack
Mouthpieces Manufacturer Recommended:
Unlimited tests every 6
months
Australian Standards
Requirement: Every 6 months
Memory Capacity 100,000 test records including
all information
Calibration Manufacturer Recommended:
Unlimited tests every 6
months
Weight Australian Standards
Requirement: Every 6 months
Dimensions (mm) 20cm x 22cm x 8.8cm
Warranty 1 Year
On/Off Button
Camera
Up Button
Touch Screen
Down Button
Printer
Button to
Remove
Mouthpiece
Sample Port

4
PLEASE REFER TO DIAGRAM ON PREVIOUS PAGE
1. Disposable Mouthpiece
Each mouthpiece is individually packaged and sealed for hygiene reasons. Before each new
Standard Test, attach a new mouthpiece to the sample port to prevent contamination from a
previous test.
2. Sampling Cup
The sampling cup can be used for Passive Tests and is reusable as long as no alcohol is detected
in the previous test. Each sampling cup is also individually sealed. Attach a sampling cup to the
sample port to take a passive test. Do not allow the mouth to touch the sampling cup during
testing. After use, remove the sampling cup and clean it with alcohol or boiling water. Allow it to
dry completely before re-using it.
3. Button to Remove Mouthpiece
The mouthpiece can be removed from the breathalyser hygienically by pushing this button up.
This will eject the mouthpiece for easy disposal without touching the mouthpiece.
4. Touch Screen Display
This displays the instructions and the alcohol concentration reading of breath tests taken on the
unit. Select the icons and buttons on the screen using the stylus pen provided.
5. Camera
Takes a picture of the subject providing the breath sample.
6. On/Off Button
Press and hold the On/Off button to turn the device on or turn off.
7. Battery Charging
The battery can be charged while inside the device or using the included battery charging dock.
The battery charging indicator turns red when the battery is charging, and green when the
battery is fully charged.
8. External Power Connectors
Connectors for 12V car (cigarette) adapter or optional 12V power transformer. The battery can
be charged while inside the device.
9. USB Cable
Use the USB cable to upload test records stored in the internal memory and upgrade the
software.
FEATURES

5
1. GPS Indicator
2. WIFI Indicator
3. 3G Indicator
4. Battery Indicator
5. Date & Time Indicator
6. Standard Test
7. Passive Test
8. Test Records
9. Calibration
10.Setting
11.GPS
12.3G/4G Setting (Optional)
13.WIFI Setting
14.Last Calibration Date
15.Next Calibration Date
16.Test Record Counter
SCREEN DISPLAY
Confirm / Enter
Next Screen
Restore
Previous Screen
Print
Edit
Cancel
Save
Search

6
MEASUREMENTS & CONDITIONS FOR TEST
• Use a clean mouthpiece for each test.
• The optimum temperature range for operating the unit is -10°C to 50°C. Use within this
temperature range will limit condensation from a subject’s breath and ensure an accurate
reading is taken.
• Do not allow samples containing smoke,tobacco or other chemical products to enter the unit, as
damage to the fuel cell sensor can occur.
• It is recommended that the operator observe the subject for approximately 20 minutes before
administering the breath test. This is to ensure the subject does not consume anything.
Following this procedure will ensure the subject’s saliva is clear of alcohol so that it does not
alter the outcome of the breath sample test.
• There must be an interval of at least 15 minutes after alcohol consumption as alcohol residue
in the mouth may give a false reading. Such residue may also be left by aromatic drinks,
mouthwash, medicines and drops, as well as by belching and vomiting. Mouth rinsing with water
or non-alcoholic drinks does not substitute for an interval.
• A person’s blood alcohol concentration (BAC) can continue to rise for up to 2 hours after
drinking alcohol. If a result is close to the designated BAC, it is recommended to wait and retest
after 20 minutes.
• It can take 10 hours or more for a person’s BAC to return to zero after a high BAC has been
reached. A further test should be carried out later in the day or the following morning.
CONDITIONS FOR TESTING
LIABILITY FOR PROPER FUNCTION OR DAMAGE
The liability for proper functioning of the instrument is irrevocably transferred to the owner
or operator in the event the instrument is serviced or repaired by personnel not employed
or authorised by Andatech. In such an event, the warranty will be void.
OPERATING CONDITIONS
Working temperature of the Andatech® Prodigy 3: -10 to 50°C
Storage temperature range of the Andatech® Prodigy 3: -30 to 70°C, 15 to 90%RH
Note: Inappropriate storage of the device will shorten the lifespan of the fuel cell sensor.

7
OPERATING INSTRUCTIONS
Refer to “Conditions for Testing” before conducting a Standard or Passive test.
On/Off
Press and hold the power button for 1-2 seconds to switch the device on or off.
Taking a standard test
1. Attach a mouthpiece onto the sample port and select the Standard Test icon.
2. The breathalyser will take a Blank Test to ensure the mouthpiece and sensor are clear of alcohol
from previous tests.
3. If the Blank Test result shows 0.000, the unit will proceed to test mode (skip to step 5).
4. If alcohol is detected, verify that a clean mouthpiece is being used and select retest.
5. Use the stylus pen to input information.
6. When the screen displays “Please Blow”, take a deep breath and blow into the mouthpiece with
a slow and steady breath until the BEEP tone stops and the Time and Volume bars are full.
• A photo will be taken while the subject is blowing.
• A passive test can be taken by pressing the icon while the subject is still blowing.
7. The test result will display on the screen. Press to go to the next screen (review results).
8. If the subject did not blow within 30 seconds, the administrator can press to confirm that
the subject refused to take a test, or press to allow the subject to take a test again,
or to exit.
9. If the subject failed to give a valid breath sample, the administrator can press to confirm or
to allow the subject to take the test again.
10.Select to print the test result.
11.Push the button underneath the mouthpiece to remove the mouthpiece and take another test.
Taking a passive test
1. Attach a sampling cup onto the sample port and select the Passive Test icon.
2. When the screen displays “Please Blow”, take a deep breath and blow into the sampling cup
until the BEEP tone stops.
3. Press to take a manual sample while the subject is still blowing into the sampling cup.
4. An “Alcohol Detected” or “No Alcohol” result will display on the screen.
5. The next test can be taken immediately after a “No Alcohol” result.
6. If “Alcohol Detected” result was displayed, use a mouthpiece to perform a Standard Test to
confirm the test result.

8
7. If alcohol was detected during a passive test, clean the sampling cup or use a new sampling cup
to prevent residual alcohol from affecting the next test result.
8. Select to print the test result.
Note: A passive test only checks for the presence of alcohol in a breath sample. An accurate test
result can only be obtained via a Standard Test using a mouthpiece. Passive test results will not be
stored in the device but can be printed immediately after a test is taken.
Record
1. The device can store over 100,000 test records along with the subject’s photo.
2. Select the Record icon to view details of past test records.
3. Select to print out individual test records.
4. Select to search for test records based on their record number, date, time, mode or result.
Then press again to confirm.
Calibration
1. When the screen displays “Calibration Needed”, it means the device has passed its calibration
due date.
2. Contact an authorised calibration centre to get the device calibrated.
GPS
1. Shows the longitude and latitude of the device location.
2. The longitude and latitude of the test is stored and printed on the printout.
3. Uncheck the “Print GPS” tickbox under the “Print” Settings to disable printing the GPS location
on test records.
Note: The GPS function requires 3-8 minutes after startup to begin locating correctly. Ensure the
device is in an open area in order to receive a signal. If no GPS signal is detected, the longitude and
latitude will not be shown on the printout.
3G/4G
1. Applicable for devices with 3G/4G module only.
2. Insert SIM card.
3. Apply your preferred cellular network settings.

9
WIFI
1. Enter the WI-FI information to connect to the WIFI.
Setting
1. Basic Settings
Adjust the screen brightness level and sleep time.
2. Time Setting
Set the device date and time.
3. Print
• Print GPS: Tick the checkbox to print the GPS location
of the test on the printout.
• Print Photo: Tick the checkbox to print the photo of
the subject on the printout.
4. Device Info
Displays the device information such as the software
version number, next and last calibration date, and total
number of tests.
5. Upload
If a backend server is used for data upload, the IP
address of the server is shown (IP address can only
be changed in Advanced Settings). Set Auto Upload if
needed.
6. Advanced
Only authorised calibration technicians can enter the
device’s Advanced Settings.
7. Skip Data Entry
Tick this checkbox to skip entering data before each
Standard Test. Untick to re-enable entering data.

10
PRINTER
DISCLAIMERS
The Prodigy 3 features a built-in thermal printer.
• This product must be used strictly in accordance with the instructions in this manual.
• This product is intended to provide the user with an indication of alcohol on the breath. However,
neither the manufacturer nor the supplier gives any warranty (express or implied) that the
product can give any indication of blood alcohol level.
• A person’s BAC (Blood Alcohol Concentration) can only be accurately measured by the
administration of a blood test. This product only measures breath alcohol content which, in turn,
only gives an approximate indication of blood alcohol concentration.
INSERT/REPLACE PAPER:
Pull the handles on the side to open the printer cassette and inset thermal paper. When inserting
a new roll of paper, make sure the glossy/outer side of the paper is facing the serrated tear strip
once the door is closed.
NOTE: If nothing can be printed, change the orientation of the paper roll and try again.
STANDARD THERMAL PAPER WIDTH
57mm x 24mm

11
WARRANTY
The manufacturer warrants for 1 year from
the date of purchase, the product to be free
from defects in workmanship or material
(excluding calibration) under normal use.
The manufacturer’s obligation under this
warranty is limited to replacing, adjusting
or repairing the unit if returned along
with its proof of purchase. This warranty
is void if the unit has been tampered with,
damaged or abused in any way.
To register your warranty online, go to
my.andatech.com.au
LIMITS
AUSTRALIAN BLOOD ALCOHOL LIMITS: Adult License 0.05%
Some drivers are restricted to lower levels of BAC. Please refer to your local transport authority
for your correct limit. If in any doubt do not drink and drive.
The safest level when driving is none at all. Alcohol impairs your judgement and can make you
drowsy and sleepy.
CALIBRATION AND SERVICE
Andatech recommends the AlcoSense®
Prodigy 3 is calibrated every 6 months
(unlimited tests) to maintain Australian
Standards.
For recalibration and technical support,
please go to my.andatech.com.au or
contact:
Andatech
Hotline: 1300 800 200 (within Australia)
+613 8899 6900 (International)
Email: [email protected]
Website: www.andatech.com.au
• In these circumstances the user of this product should not rely on the information provided
thereby as an accurate indication of his blood alcohol concentration or of his level of intoxication
or of his fitness to drive and in particular should not take any action in reliance on such
information as the product can provide no more than an indication of alcohol on the breath.
• If the user of this product makes any decision or takes any action based on the use of the product
such decision or such action shall be at the sole risk of the user and neither the manufacturer
nor the supplier assumes any responsibility for such decision or action. In particular, neither
the manufacturer nor the supplier shall be in any way liable if the user of the product asserts
that the product led him to believe that he was fit to operate a motor vehicle and is subsequently
found not to be so fit.
• In particular, neither the manufacturer nor the supplier shall be in any way liable if the user of
the product asserts that the product led them to believe they were fit to operate a motor vehicle
and are subsequently found not to be so fit.
THINGS TO REMEMBER:
1. Always wait a minimum of 15 minutes after your last drink before testing.
2. NEVER take a sip of alcohol and blow immediately into the device.
3. Ignoring these two points WILL cause damage to the unit that is NOT covered by warranty.

12
It is our aim to provide you with quality products that you can trust. Our products come with
guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to
a replacement or a refund for a major failure and for compensation for any other reasonably
foreseeable loss or damage.
You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable
quality and the failure does not amount to a major failure.
The Guarantee
The product is guaranteed to be free from defects in workmanship and parts for a period of 1 year
from the date of purchase. Defects that occur within this warranty period, under normal use and
care, will be repaired, replaced or refunded at our discretion, solely at our option with no charge for
parts and labour. The benefits conferred by this warranty are in addition to all rights and remedies in
respect of the product that the consumer has under the Trade Practices Act and similar state laws.
Proof of Purchase
This warranty is valid for the original purchase and is not transferable. Please keep your purchase
docket or receipt as proof of purchase and as proof of date on which the purchase was made. The
purchase docket (or a copy) or receipt must be presented with the warranty when making a claim
under this warranty.
Service during the Warranty Period
to get an RA# (Return Authorisation Number). Ensure that the RA# is clearly stated on the outside
of the packaging and that the product is properly packaged so that no damage occurs to the product
during transit. Shipping of the product back to us for warranty will be at your cost. A product return
without the RA# or proof of purchase will not be accepted.
Extent of Warranty
This warranty is limited to defects in workmanship or parts. All defective products or parts will be
repaired or replaced. This warranty does not cover manuals and packaging.
Normal Wear and Tear
This warranty does not cover normal wear and tear to the product or parts.
Exclusions
This warranty does not cover:
• Calibration costs.
• Any defect caused by an accident, misuse, abuse, improper operation, lack of reasonable care,
unauthorised modification, loss of parts, tampering or attempted repair by a person not authorised
by the distributor.
• Any product that has been damaged by a lightning strike either directly or indirectly or a main
power surge or liquid ingress.
• The product if it is located outside of Australia.
• Any damage caused by improper power input or improper cable connection.
WARRANTY DETAILS

13

Andatech Pty Ltd
www.andatech.com.au
1300 800 200
Table of contents
Other Andatech Test Equipment manuals

Andatech
Andatech SOBERLIVE FR User manual

Andatech
Andatech ALX3000 User manual

Andatech
Andatech AlcoSense SOBERPOINT 3 User manual

Andatech
Andatech AlcoSense SURETY User manual

Andatech
Andatech Alcosense Zenith Breathalyser User manual

Andatech
Andatech Sentry User manual

Andatech
Andatech AkcoSense User manual

Andatech
Andatech SOBERMATE S3 User manual

Andatech
Andatech DrugSense Orascan User manual

Andatech
Andatech AlcoSense AccessPoint User manual