ABOUT TEMPERATURE CALIBRATION
The Temperature Calibration is required to evaluate and adjust the precision and accuracy of the
Soberlive FRX's temperature measurements.
TEMPERATURE CALIBRATION PROCEDURE
To perform a Temperature Calibration, a calibration operator stands in
front of the Soberlive FRX and faces the camera.
The device will then take their temperature and use that as the
guideline for 'normal body temperature'.
Note: It is recommended that the calibration operator first check their
body temperature (e.g. with a separate accurate, reliable handheld
thermometer) to make sure that it is normal (within 36~36.5°C) before
performing the Temperature Calibration.
ADMIN TEMPERATURE CALIBRATION (ON DEVICE
STARTUP)
Each time the device is switched on, it takes 10 minutes to warm up
and will then require a Temperature Calibration.
This can only be performed by an Admin user or by any user (if enabled under Advanced Settings).
Note: The device will auto-restart daily at 4AM (device time) by default - this setting can be changed in
Device Settings. Any software updates are done automatically during restart as long as your device is
connected to the internet.
INTERVAL TEMPERATURE CALIBRATION
During normal operation, the device will also require a
Temperature Calibration at regular intervals (recommended /
default setting: 60 minutes)
When the device requires Temperature Calibration, the Thermal
Imaging Display will turn completely orange and the Notification
Box will show 'Please contact calibration operator' if anyone
attempts to take a test.
A normal test cannot be taken until a calibration operator
performs the Temperature Calibration.
MANUAL TEMPERATURE CALIBRATION
If a temperature calibration is needed outside of its regular
interval period (e.g. temperatures seem too high or too low, or is
measuring inconsistently), a manual temperature calibration can
be performed.
9