AXE MMRD Series User manual

Rev.00
AL1
AL2
AL1
AL2
DIO
AXE MICROPROCESS DUAL INPUT RPM & LINE-SPEED CONTROLLER METER MMRD Series
■Features
◎Dual input measure and display sensor pulse or AC signal
◎16BIT DAC analog output can be modified,
◎Accuracy 0.03% F.S. ◎2 set independent alarm outputs with delay/hysteresis function
◎Input range(0~25KHz) ◎RS485 Communication interface, Protocol MODBUS RTU MODE
◎Readout range(0~99999) ◎BAUD RATE:38400/19200/9600/4800/2400
◎RPM or LINE-SPEED can be modified ◎Decimal point can be modified
◎Input pulse per revolution can be modified(1~99999) ◎0.4" LED highlight display
◎Diameter(LINE-SPEED)/Scale(RPM) can be modified ◎Man-machine interface, easy to operate
(0.0001~9.9999) ◎EEPROM Saving, data safekeeping about 10 years
◎Display average times can be modified(1~99) ◎Modified inside parameter, must have pass code
■Name of Parts
■Connect Diagram ■Input function jumper table
◎Contact input (PNP) ◎Sensor input (PNP 5V/12V)
◎Position 1 ON:IN(A)/IN(B) -> (NPN)
◎Contact input (NPN) ◎Sensor input (NPN 5V/12V)
◎Position 1 OFF:IN(A)/IN(B) -> (PNP)
◎Position 2 ON :IN(A)/IN(B) -> (0~50HZ)
◎Position 2 OFF:IN(A)/IN(B) -> (0~25KHZ)
■
Alarm Functio
n
description
◎When ACT=HI,DEL= 0 : Display value > Setting value(AL) + Hysteresis (HYS) à(Relay on)
Display value <= Setting value(AL) - Hysteresis (HYS) à(Relay off)
◎When ACT=LO,DEL= 0 : Display value >= Setting value(AL) + Hysteresis (HYS) à(Relay off)
Display value < Setting value(AL) - Hysteresis (HYS) à(Relay on)
◎When ACT=HI,DEL= 1 ~ 99 sec.: Display value > Setting value(AL) + Hysteresis (HYS) + Delay time(DEL) à(Relay on)
Display value <= Setting value(AL) - Hysteresis (HYS) à(Relay off)
◎When ACT=LO,DEL= 1 ~ 99 sec.: Display value >= Setting value(AL) + Hysteresis (HYS) à(Relay off)
Display value < Setting value(AL) - Hysteresis (HYS) + Delay time(DEL) à(Relay on)
◎When ACT=HI,DEL= -1 ~ -99 sec.:Display value > Setting value(AL) + Hysteresis (HYS) à(Relay one shoot(DEL) and then off)
Display value <= Setting value(AL) - Hysteresis (HYS) à(Relay restore normal after the procedure)
◎When ACT=LO,DEL= -1~ -99 sec.:Display value >= Setting value(AL) + Hysteresis (HYS)à(Relay restore normal after the procedure)
Display value < Setting value(AL) - Hysteresis (HYS) à(Relay one shoot(DEL) and then off)
IN2 Alarm Indicate
IN2 Display
Parameter Enter (Function call out)
Set-value down(Analog output adjustment call out)
Cursor shift
(
Alarm adjustment call out
)
ENT
FUNC
E
ALARM
A-ADJ
Communication Indicate
IN1 Display
IN1 Alarm Indicate
Set
-
value
up
COM
IN
12V
Inside jumper
SW1
SW2
1
1
2
2
ON
ON
O
FF
O
FF
COM
IN
12V
Sensor
Inside j
umper
SW1
SW2
1
1
2
2
O
FF
O
FF
O
FF
O
FF
4
3
2 1
SW1
4
3
2 1
SW
2
COM
IN
12V
COM
IN
12V
Sensor
Inside jumper
SW1
SW2
1
1
2
2
ON
ON
O
N
O
N
Inside jumper
SW1
SW2
1
1
2
2
ON
ON
O
FF
O
FF

Rev.00
Key Introduce Operation Manual
Key Function 1.In normal display,The key function is call out setting group
2.In parameter setting page,The key function is data Enter , and goto next page
Key Function 1.In normal display,The key function is call out alarm value setting page
2.Into parameter setting page,the parameter mark&data is alternate display,If need modify data can
press
key into setting procedure,The display is lock parameter data,this time must let off key
about 0.2 sec,press again,the cursor(twinkle express)is cycle moving left. (Key Response about 0.2
sec)
Key Function 1.Into parameter setting page,the parameter mark&data is
alternate display,If need modify data can
press
key into setting procedure,The display is lock parameter data,this time must let off key
about 0.2 sec,press again,the parameter data will increment. (Key Response about 0.2 sec)
Key Function 1.In normal display,The key function is call out adjustment analog output AZERO&ASPAN page
2.Into parameter setting page,the parameter mark&data is alternate display,If need modify data can
press
key into setting procedure,The display is lock parameter data,this time must let off key
about 0.2 sec,press again the parameter data will decrement. (Key Response about 0.2 sec)
&
Key Function 1.In setting group or setting page press
&
key return normal display,but if in setting page the
modify data will be lost
No Key in anything 1.In setting group or setting page no key in anything about 2 minutes,return normal display,but if in
Setting page the modify data will be lost
Step Parameter Mark Description
Parameter Mark
Operation Manual
1
Normal display
Press
/FUNC key into P.CODE setting page
2
P.CODE(Pass code input
page)
1.Key in 5 digit pass code with
or
or
key
2.Press
key,the pass code is right into setting group , otherwise
return normal display
SYS(System setting group)
ROP(Alarm setting group)
DOP
(Communication setting group)
3
AOP
(Analog output setting group)
1.Select setting group with
key
2.Press
key into setting page of selection setting group
4
SYS(System setting group)
Press key decide SYS setting group , press key into DP1 setting
page
4-1
DP1(Decimal Point IN-1)
Default = 0
1.Decide decimal point IN-1 position with
or
key (0 to 4)
2.Press
key enter data and into DP2 setting page
4-2
DP2(Decimal Point IN-2)
Default = 0
3.Decide decimal point IN-2 position with
or
key (0 to 4)
4.Press
key enter data and into TYPE setting page
4-3
TYPE(Type)
Default = RPM
1.Decide Type with
or
key(RPM/LINE)
2.Press
key enter data and into PPR-A setting page
4-4
PPR-A(Pulse Per Revolution
of input A)
Default = 1
1.Decide pulse per revolution of input A with
&
&
key(1~99999)
2.Press
key enter data and into PPR-B setting page
4-5
PPR-B(Pulse Per Revolution
of input B)
Default = 1
1.Decide pulse per revolution of input B with
&
&
key(1~99999)
2.Press
key enter data and into SCL-A setting page
4-6
SCL-A (Scale-A)
Default = 1.0000
1.Decide Scale-A with or or key (0.0001~9.9999)
2.Presskey enter data and into SCL-B setting page
Note:RPM(Scale = 0.0001~9.9999)
Line-Speed(rotation diameter = 0.0001~9.9999M)
4-7
SCL-B (Scale-B)
Default = 1.0000
1.Decide Scale-B with or or key (0.0001~9.9999)
2.Presskey enter data and into TBASE setting page
Note:RPM(Scale = 0.0001~9.9999)
Line-Speed(rotation diameter = 0.0001~9.9999M)
4-8
TBASE (Sampling Time
Base)
Default = 0.1
1.Decide sampling time base with
&
&
key(0.1~99.9sec)
2.Press
key enter data and into AVG setting page
4-9
AVG (Display Average
1.Decide display average times with
&
&
key(1~99)

Rev.00
times)
Default = 5
2.Press
key enter data and into CODE setting page
4-10
CODE(Pass Code)
Default = 0
1.Decide pass code with
&
&
key(0~99999)
2.Press
key enter data and into LOCK setting page
4-11
LOCK(Panel Lock)
Default = NO
1.Decide panel lock with
&
key(NO or YES)
2.Press
key enter data and return SYS setting group
5
ROP(Alarm setting group)
Press
key decide ROP setting group,press
key into ACT1.1
setting page
5-1
ACT1.1(Alarm Active 1-1)
Default = HI
1.Decide Alarm Active 1-1 with
or
key(HI or LO)
2.Press
key enter data and into ACT1.2 setting page
5-2
ACT1.2(Alarm Active 1-2)
Default = HI
1.Decide Alarm Active 1-2 with
or
key(HI or LO)
2.Press
key enter data and into ACT2.1 setting page
5-3
ACT2.1(Alarm Active 2-1)
Default = HI
1.Decide Alarm Active 2-1 with
or
key(HI or LO)
2.Press
key enter data and into ACT2.2 setting page
5-4
ACT2.2(Alarm Active 2-2)
Default = HI
1.Decide Alarm Active 2-2 with
or
key(HI or LO)
2.Press
key enter data and into HYS1.1 setting page
5-5
HYS1.1(Alarm Hysteresis
1-1)
Default = 0
1.Decide Hysteresis 1-1 with or or key(0~999)
2.Press key enter data and into HYS1.2 setting page
5-6
HYS1.2(Alarm Hysteresis
1-2 )
Default = 0
1.Decide Hysteresis 1-2 with or or key(0~999)
2.Press key enter data and into HYS2.1 setting page
5-7
HYS2.1(Alarm Hysteresis
2-1 )
Default = 0
1.Decide Hysteresis 2-1 with or or key(0~999)
2.Press key enter data and into HYS2.2 setting page
5-8
HYS2.2(Alarm Hysteresis
2-2)
Default = 0
1.Decide Hysteresis 2-2 with or or key(0~999)
2.Press key enter data and into DEL1.1 setting page
5-9
DEL1.1(Alarm Delay 1-1)
Default = 0
1.Decide Alarm Delay 1-1 with or or key(-99~99 sec)
2.Press key enter data and into DEL1.2 setting page
Note:-1 ~ -99
sec = Alarm active time
1 ~ 99 sec = Alarm delay time
5-10
DEL1.2(Alarm Delay 1-2)
Default = 0
1.Decide Alarm Delay 1-2 with or or key(-99~99 sec)
2.Press key enter data and into DEL2.1 setting page
Note:-1 ~ -99
sec = Alarm active time
1 ~ 99 sec = Alarm delay time
5-11
DEL2.1(Alarm Delay 2-1)
Default = 0
1.Decide Alarm Delay 2-1 with or or key(-99~99 sec)
2.Press key enter data and into DEL2.2 setting page
Note:-1 ~ -99
sec = Alarm active time
1 ~ 99 sec = Alarm delay time
5-12
DEL2.2(Alarm Delay 2-2)
Default = 0
1.Decide Alarm Delay 2-2 with or or key(-99~99sec)
2.Press key enter data and return ROP setting group
Note:-1 ~ -99
sec = Alarm active time
1 ~ 99 sec = Alarm delay time
6
DOP(Communication setting
group)
press
key decide DOP setting group,press
key into ADDR setting
page
6-1
ADDR(Communication –
Address setting page )
Default = 0
1.Decide address with
or
or
key(0~255)
2.Press
key enter data and into BAUD setting page
6-2
BAUD(Communication Baud
Rate setting page)
Default = 19200
1.Decide baud rate with
or
key(38400,19200,9600,4800,2400)
2.Press
key enter data and into PARI setting page

Rev.00
6-3
PARI(Communication Parity
Check setting page)
Default = n82
1.Decide parity check with
or
key(n82,n81,even,odd)
2.Press
key enter data and return DOP setting group
7
AOP(Analog output setting
group)
Press
key decide AOP setting group , press
key into AO.SEL
setting page
7-1
AO.SEL(Analog Output
Select)
Default = IN1
1.Decide Analog Output Select with or key(IN1 or IN2)
2.Press key enter data and into ANLO setting page
7-2
ANLO(A/O Zero According
to Display setting page)
Default = 0
1.Decide A/O Zero According to Display with
or
or
key
(0~99999)
2.Press
key enter data and into ANHI setting page
7-3
ANHI(A/ O Span According
to Display setting page)
Default = 99999
1.Decide A/ O Span According to Display with
or
or
key
(0~99999)
2.Press
key enter data and return AOP setting group
Step Parameter mark description Parameter mark
Operation manual
8
Normal display
Press
/ALARM key about 3 sec,into AL1-1 setting page
8-1
AL1-1 (IN1 Alarm value 1-1)
Default = 0
1.Decide IN1 alarm value 1-1 with or or key(0~99999)
2.Press key enter data and into AL1-2 setting page
8-2
AL1-2 (IN1 Alarm value 1-2)
Default = 0
1.Decide IN1 alarm value 1-2 with or or key(0~99999)
2.Press key enter data and into AL2-1 setting page
8-3
AL2-1 (IN2 Alarm value 2-1)
Default = 0
1.Decide IN2 alarm value 2-1 with or or key(0~99999)
2.Press key enter data and into AL2-2 setting page
8-4
AL2-2 (IN2 Alarm value 2-2)
Default = 0
1.Decide IN2 alarm value 2-2 with or or key(0~99999)
2.Press key enter data and return normal display
Step Parameter mark description Parameter mark
Operation manual
9
Normal display
Press
/A-ADJkey about 3 sec,into AZERO adjustment page
9-1
AZERO(Analog Output
Zero Adjustment page)
Default = 0
1.Adjustment analog output zero with
or
or
key(±6000)
2.Press
key enter data and into ASPAN adjustment page
9-2
ASPAN(Analog Output
Span Adjustment page)
Default = 0
1.Adjustment analog output span with
or
or
key(±6000)
2.Press
key enter data and return normal display
Appendix
Error Mark description Error Mark Analyze & Description
1
Input over range error
detect
Input signal over range(0~25KHz)
2
Display over range error
detect
Display over range(99999)
3
EEPROM error detect
1.External interference when EEPROM read/write
2.EEPROM write over 1 million times(guarantee 10 years)
Please power reset,if still display E-00,doing following step:
1.E-00 & No alternate display for inquire reset EEPROM
2.Decide Yes with or key,press key return normal display
3.EEPROM was reset,Please follow step 1~9 set again

Rev.00
MMRD Modbus RTU Mode Protocol Address Map
Data format 16Bit/32Bit, sign bit 8000~7FFF(-32768~32767),80000000~7FFFFFFF(-2147483648~2147483647)
Address
Name Description Accept
0000 DP1 IN1 Decimal Point,Input Range 0000~0004(0~4)(0:100,1:10-1,2:10-2,3:10-3,4:10-4) R/W
0001 DP2 IN2 Decimal Point,Input Range 0000~0004(0~4)(0:100,1:10-1,2:10-2,3:10-3,4:10-4) R/W
0002 TYPE Input Type,Input Range 0000~0001(0~1)(0:RPM,1:LINE) R/W
0003 LOCK Panel Lock,Input Range 0000~0001(0~1)(0:NO,1:YES) R/W
0004 ACT1.1 Alarm Active 1-1,Input Range 0000~0001(0~1)(0:HI,1:LO) R/W
0005 ACT1.2 Alarm Active 1-2,Input Range 0000~0001(0~1)(0:HI,1:LO) R/W
0006 ACT2.1 Alarm Active 2-1,Input Range 0000~0001(0~1)(0:HI,1:LO) R/W
0007 ACT2.2 Alarm Active 2-1,Input Range 0000~0001(0~1)(0:HI,1:LO) R/W
0008 HYS1.1 Alarm Hysteresis 1-1,Input Range 0000~03E7(0~999) R/W
0009 HYS1.2 Alarm Hysteresis 1-2,Input Range 0000~03E7(0~999) R/W
000A
HYS2.1 Alarm Hysteresis 2-1,Input Range 0000~03E7(0~999) R/W
000B
HYS2.2 Alarm Hysteresis 2-2,Input Range 0000~03E7(0~999) R/W
000C
DEL1.1 Alarm Delay 1-1,Input Range FF9D~0063(-99~99) R/W
000D
DEL1.2 Alarm Delay 1-2,Input Range FF9D~0063(-99~99) R/W
000E
DEL2.1 Alarm Delay 2-1,Input Range FF9D~0063(-99~99) R/W
000F
DEL2.2 Alarm Delay 2-2,Input Range FF9D~0063(-99~99) R/W
0010 ADDR Communication Address,Input Range 0000~00FF(0~255) R/W
0011 BAUD Communication Baud Rate,Input Range 0000~0004(0~4)(0:38400,1:19200,2:9600,3:4800,4:2400) R/W
0012 PARI Communication Parity Check,Input Range 0000~0003(0~3)(0:N82,1:N81,2:EVEN,3:ODD) R/W
0013 AO.SEL
Analog Output Select,Input Range 0000~0001(0~1)(0:IN1,1:IN2) R/W
0014 TBASE Sampling Time Base, Input Range 0001~03E7(1~999) R/W
0015 AVG Display Average times, Input Range 0001~0063(1~99) R/W
0016 A_ZERO
Analog Output Zero Adjust,Input Range E890~1770(-6000~6000) R/W
0017 A_SPAN
Analog Output Span Adjust,Input Range E890~1770(-6000~6000) R/W
0018 CODE Pass Code,Input Range 00000000~0001869F(0~99999)high word R/W
0019 Pass Code,Input Range 00000000~0001869F(0~99999)low word R/W
001A
SCL-A Scale-A,Input Range 00000001~0001869F(1~99999)high word R/W
001B
Scale-A,Input Range 00000001~0001869F(1~99999)low word R/W
001C
SCL-B Scale-B,Input Range 00000001~0001869F(1~99999)high word R/W
001D
Scale-B,Input Range 00000001~0001869F(1~99999)low word R/W
001E
PPR-A Pulse Per Revolution of input A, Input Range 00000001~0001869F(1~99999) high word R/W
001F
Pulse Per Revolution of input A, Input Range 00000001~0001869F(1~99999) low word R/W
0020 PPR-B Pulse Per Revolution of input B, Input Range 00000001~0001869F(1~99999) high word R/W
0021 Pulse Per Revolution of input B, Input Range 00000001~0001869F(1~99999) low word R/W
0022 AL1-1 IN1 Alarm value 1-1,Input Range 00000000~0001869F(0~99999)high word R/W
0023 IN1 Alarm value 1-1,Input Range 00000000~0001869F(0~99999)low word R/W
0024 AL1-2 IN1 Alarm value 1-2,Input Range 00000000~0001869F(0~99999)high word R/W
0025 IN1 Alarm value 1-2,Input Range 00000000~0001869F(0~99999)low word R/W
0026 AL2-1 IN2 Alarm value 2-1,Input Range 00000000~0001869F(0~99999)high word R/W
0027 IN2 Alarm value 2-1,Input Range 00000000~0001869F(0~99999)low word R/W
0028 AL2-2 IN2 Alarm value 2-2,Input Range 00000000~0001869F(0~99999)high word R/W
0029 IN2 Alarm value 2-2,Input Range 00000000~0001869F(0~99999)low word R/W
002A
ANLO Analog Output Zero According to Display,Input Range 00000000~0001869F(0~99999)high word R/W
002B
Analog Output Zero According to Display,Input Range 00000000~0001869F(0~99999)low word R/W
002C
ANHI Analog Output Span According to Display,Input Range 00000000~0001869F(0~99999)high word R/W
002D
Analog Output Span According to Display,Input Range 00000000~0001869F(0~99999)low word R/W
002E
DISP1 IN1 Display Value, Display Range 00000000~0001869F(0~99999)high word R
002F
IN1 Display Value, Display Range 00000000~0001869F(0~99999)low word R
0030 DISP2 IN2 Display Value, Display Range 00000000~0001869F(0~99999)high word R
0031 IN2 Display Value, Display Range 00000000~0001869F(0~99999)low word R
0032 STATUS
Display & Alarm Status, Display Range 0000~00FF(0~255) Bit0:AL1-1,Bit1:AL1-2,Bit2:AL2-1,
Bit3:AL2-2,Bit4:DISP1 DOFL,Bit5:DISP1 IOFL,Bit6:DISP2 DOFL,Bit7:DISP2 IOFL R
Other AXE Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Dimetix
Dimetix FLS-C Application note

Hanna Instruments
Hanna Instruments HI 93703 instruction manual

CONDTROL
CONDTROL NEO X1-360 user manual

BIFFI
BIFFI ELBS-20 Installation, operation and maintenance manual

sauter
sauter FK 10 instruction manual

Hanna Instruments
Hanna Instruments HI 83225 instruction manual