screw (2), then mount the objectives onto the nosepiece.
2. Mounting and changing the lamps:
Before using the microscope, please first mount the lamps. Gently press the finger-shape
protrusion on the protective baffle (12) and take out the protective baffle (12) from the
illuminating hole, insert the lamp straight into the socket as far as it will go. Only use halogen
lamps of the specified rating. Replace the protective baffle (12) , Do not touch the glass part of a
lamp with your bare hands as your fingerprints may stick to the lamp, possibly reducing its level of
light output or causing damage to the lamp.
When changing the lamps, please pull down the plug of the electrical wire before changing it.
Change the lamp after it is cool enough. Press the finger-shape protrusion on the protective baffle
(12) lightly and take out the protective baffle (12) from the illuminating hole. Replace the bulb with
the new one; then remount the protective baffle (12) into the illuminating hole. During changing
the bulb, note not to directly touch the glass portion of the bulb and the protective baffle by hand
but to wear glove or use sheath. After finishing, wipe off the fingerprint and smudge on the glass
with the soft clean cloth dampened with absolute alcohol. Otherwise the bulb is easy to get split
and weaken the brightness of the whole illumination system.
3. Checking the input voltage:
The rated input voltage is indicated in the electric nameplate attached to the back of the arm.
In the case of the voltage of the local power supply according with the rated value, connect up the
power supply. The electric socket must have earthing.
4. Adjusting the illumination:
The illumination consists of a LED lamp or two halogen lamps with cup which form an angle
tilting each other. Rotate the knob for adjusting the illumination (9) to change the angle tilting each
other so as to get the ideal condition of illuminating.
5. Selecting the objectives magnification:
The microscope is equipped with various objectives for 2×,4×,6×magnification. The objective
with the lower magnification takes on the larger field of view and depth of field. In order to find
the object easily, advise to use first the lower magnification objective; then rotate the nosepiece
for 120 degree to change the objective magnification.
6. Adjusting the interpupillary distance:
According to the interpupillary distance of viewers, rotate the knob for adjusting the interpupillary