Revision History | xi
ELx50 Operator's Manual
Revision Date Changes
J 4/07
Revised Chapters 1 through 5 primarily to support a redesign of the
aspiration carrier in the ELx50/8FMW washer: the vacuum grate is now
secured in place (under the plate seal gasket) by six screws around the
perimeter of the top of the carrier. Also revised vacuum levels for the
aspiration carrier: from –240 mmHg to –64 mmH for the 0.047”
diameter hole of the carrier’s vent port, and from –64 mmHg to –240
mmHg for the 0.020” diameter hole in the black vent plug.
K 10/07
General. Added information on the Liquid Level Alert™ option, which is
available for all ELx50 models. Added information on the new ‘First Strip’
run-time parameter. Additional cleanup and clarifications as described
below.
Chapter 1, Introduction. Software Variations: Clarified which features
and functions are supported for the different basecode part numbers.
Hardware Features: Simplified the description of the supply and waste
bottles provided with the instrument. Optional Accessories: Added
information on Liquid Level Alert option.
Chapter 2, Installation. Connecting Tubes and Bottles: Added
information on the Liquid Level Alert option components. Added Liquid
Alert port to drawings. Added instructions for connecting the sensor
cables to the select box, and the select box to the washer.
Chapter 3, Operation. Throughout: Added information on the ‘First Strip’
run-time parameter. Updated Menu Maps. General Usage Guidelines:
Added section for the Liquid Level Alert option. Linking Programs: Added
troubleshooting tip for resolving Link program creation issues. Setup:
Added information on new SENSOR option.
Chapter 4, Maintenance/Performance Verification. Recommended
Maintenance and Performance Verification Schedules: Divided the former
Maintenance AND Performance Verification Schedule table into two
separate tables. Decontamination: Added warning not to prepare a
stronger-than-recommended bleach solution.
Appendix A, Error Codes. General Errors: Added error codes 2800 and
2900, related to the Liquid Level Alert option.
L 1/08
General. Added information on the optional vacuum pump muffler for
ELx50 washers equipped with the new vacuum pump (washers
manufactured after January 2008). Additional changes are described
below.
Preface. Updated Directives in the CE Mark section and added
Unspecified Use Warning to Hazards and Precautions.
Chapter 1, Introduction. Added the optional vacuum pump muffler to the
Hardware Features and Accessories Included sections. Updated the
Performance Specifications section for the 96-well: changed residual
volume specs from ≤4.0 µl to ≤2.0 µl per well, and dispense precision
specs from ≤4.0% CV to ≤3.0% CV.
Chapter 2, Installation. Updated unpacking/packing artwork and
instructions to reflect new packaging materials, and added a graphic of
the vacuum pump muffler to illustrations of the washer’s rear panel.
Chapter 4, Maintenance and Performance Verification (IQ/OQ/PQ). In
the Evacuation Efficiency and Dispense Precision Tests and Worksheets
for the 8-, 8s-, and 12-channel manifolds, changed the residual volume
specs from ≤0.004 grams per well to ≤0.002 grams per well, and the
dispense precision specs from ≤4.0% CV to ≤3.0% CV.