What you see
or hear
FLASHING “1”
“Low Pressure/Leak”
Single beep.
Device is making
more noise.
• Clamp the tubing.
•
audible alarm resets, there is a leak below
the clamp – often in the dressing. Reopen
the clamp before addressing the leak.
Gently press around drape to check for
leaks. If leak is found, patch with extra
drape material.
•
audible alarm continue, there is a leak
above the clamp. Check tubing connection
at the canister. Check to ensure the canister
is fully seated and locked. Check for cracks
in the canister or lid separation.
If found, replace the canister.
• Open the clamp.
The alarm will reset, the pressure light will
Leaks often occur over areas of moist skin,
creases or folds in skin, and wrinkles in the
drape.They can occur if the drape snags on
clothes or bed sheets.
There is an air leak in
either the dressing or the
tubing connections.
Problem What to do More Information
FLASHING “2”
“Canister Full”
Two-tone beep.
• Clamp the tubing.
•
OFF button.
• Press the canister release button above the
canister and slide the full canister out.
Cap and dispose of properly. Slide new
canister in, align the short ports and click
into place.
• Open the clamp and press
the ON button to resume therapy.
The canister full alarm begins when
the canister is 90% full, but the device
will continue to work until the canister
If the Cardinal Health™ PRO family of
entering the canister will cause a false
canister full alarm and the canister must
be changed.
The canister is full.
Plug in the device. Use only the power cord that came with the
Cardinal Health™ PROfamily of devices.
Whenthe device is getting power, a green
light will illuminate on the front of the device.
A yellow light below the greenlight will
show that the battery is charging. It will
The battery is low and
will run out in about
30 minutes.
FLASHING “3”
“Low Battery”
Three-tone beep.
Dispose as per facility/region/provincial
guidelines.
This alarm cannot be Muted or manually
reset by cycling power.
Device is at end of life
cycle.
FLASHING “4”
“Service Timer”
Four beeps every 10 seconds.
Unlock the device.
Pressure lock-out
is engaged.
Pressure Setting will
not change.
No action needed.
No action needed.
Intermittent mode maintains target
25 mmHg for two minutes.
may be removed from the wound and
stay in the tubing. The foam will be
compressed normally and the device will
be quiet.
You can use intermittent mode to move
This is NOT a problem.
The device is operating in
intermittent mode.
This is NOT a problem.
An amber light is showing on
the front of the device below
the pressure numbers.
not moving in the tube.
To lock therapy setting, press and hold the
ON button for three seconds.
The device will beep three times indicating
therapeutic setting is locked.
Repeat process to unlock.
07/15
Cat. 6708888, 47-0010, 47-1230
© 2015 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO
and SPEEDCONNECT are trademarks or registered trademarks of Cardinal Health.
US Pat. 7,532,953, 7,608,066, 8,066,243.
Cardinal Health
1500 Waukegan Road
Waukegan, IL 60085 USA
cardinalhealth.com
Please contact your sales representative
for the electronic instructions for use.