CAS NC-1 User manual

NC-1 Series
Crane Scale
OWNER’S MANUAL

CONTENTS
PRECAUTIONS 4
INSTRUCTION 6
FEATURES & MAIN FUNCTIONS 6
TECHNICAL SPECIFICATION 7
DESCRITION OF OVERALL 8
THE OPERATIONS 10
TEST MODE 12
SET MODE 15
BATTERY OPERATION 19
REMOTE CONTROL 20
ASSEMBLE AND DISASSEMBLY 21
ERROR MESSAGE & TROUBLE SHOOTING 22

45
PRECAUTIONS
Do not disassemble the scale.
When any damage or
defect occurs, contact
your CAS authorized
dealer immediately for
proper repair.
Do not overload beyond the maximum
weight limit.
Do not pull the plug by its cord when
unplugging. Damaged cord could cause
electric shock or fire.
Avoid placing the scale near heater
or in direct sunlight.
Scale must be grounded to minimize
electricity static. This will minimize
defect or electric shock.
Use proper Adapter. Incorrect adapter
could damage the scale. Insert plug firmly to wall outlet to
prevent electric shock.
To reduce electric shock or incorrect
reading, Do not spill water on the scale
or place it in humid condition.
Warning
To prevent from fire occurring, Do not
place or use the scale near
flammable or corrosive
gas.
For consistent and accurate reading,
maintain periodical check by your
CAS authorized dealer.
Avoid sudden shock to the scale.
Internal mechanism could be damaged.
Place the scale on firm and temperature
consistent environment.
Attention
Keep the scale away from other
electromagnetic generating devices.
This may interfere with accurate
reading.
Take the battery out when scale is not in
use for long time. Leakage from the
batteries is hazardous.

7
6
INSTRUCTION
We greatly appreciate your purchase of the NC-I.
These goods perform excellently and exhibit splendid properties through strike tests.
Before using the NC-I, it is recommended that you read this manual carefully so you
may use this device to its full potential.
FEATURES
Appropriate hoist scale for industrial field
Useful Zero, Tare and Hold function
Weight Back-Up
Easy to move
Be able to control with remote controller
MAIN FUNCTIONS
User can change the stable condition and range of digital-filter at user's option
Self Hardware test (internal A/D value, Battery power, key and etc)
TECHNICAL SPECIFICATION
Kilogram Max Capacity
Division
U.S.A Max Capacity
version Division
Pound Max Capacity
Division
A/D Speed
Tare Subtraction
Display below zero
Excessive Weight
Weight Display Type
Character Display
Keys
Battery
Power and Battery
consumption
Operation Temperature
Rechargeable Current
Rechargeable Adapter
Rechargeable Time
WEIGHT
20kg 30kg 50kg 100kg 200kg 500kg
10g 10g 20g 50g 100g 200g
250lb(125kg) 500lb(250kg) 1000lb(400kg)
0.1lb(0.05kg) 0.2lb(0.1kg) 0.5lb(0.2kg)
250lb 500lb 1000lb
0.1lb 0.2lb 0.5lb
20 times/second
Full Capacity
minus sign
Maximum 9d (display OUEr )
LCD (5 digit Sign, 1 Inch Height)
ZERO, TARE, BAT, POWER, CHARGE
ON/OFF, ZERO, TARE, kg lb(HOLD)
7.2V , 2A (Li ion)
Use the Backlight Not use the Backlight
0.45 W 0.17 W
Approximate 22 Hours Approximate 58 Hours
10 … ~ 40 …
150 ~ 680
110/220VAC 12V, 1.5AH
4 ~ 5 Hours
5.5kg

KEYS FUNCTIONS
9
8
DESCRIPTION OF OVERALL
1. OVERALL VIEW
KEY BOARD
HOOK
JOINT RING ASSY
DISPLAY WINDOW
NAME
PLATE
FRONT VIEW REAR VIEW
SEALING
BOLT
ON/OFF
S/W
ON
OFF
JACK
RUBBER
2. DISPLAY
4. KEY
5. REAR PART
3.LAMP of DISPLAY PART(Front body)
SPEC PLATE
Power Switch
Main Power Switch (If use, you have to turn on this switch)
Charger Adaptor connector
If you want to recharge the battery, Connect to the adaptor.
POWER
Jack Cover
ON Being connected the Adapter
RED In charge GREEN Charge is completed
Receiving part from remote controller
Display the battery recharging time (rechargeable battery)-
about 6.7V remain
Below zero
ON when current weight is 0
ON when tare is activated
Weight Unit
LED
LCD
Power
Charge
Remocon
BAT
ZERO
TARE
OZ/lb/kg
TARE
ZERO
Power is ON while pressing for 1 second
Power is OFF while pressing for 2.0 seconds
Above function is possible while the power switch on the rear part is ON
Please turn off the power switch on rear part when not use for a long time.
It makes the current weight within the zero range to zero
(2% of maximum capacity)
Pressing ZERO key for 2 second, can enter the TEST Mode
It is used to weigh with a container
If press the tare key at unload state, Tare state is released
Pressing the ON/OFF key while tare key is pressed, can enter the SET Mode
It is used to weigh the moving things
If you press this key once more in the hold mode, hold state is released
(at AVERAGE HOLD, at MANUAL HOLD, at PEAK HOLD)
It is used to change the unit of weight(kg<->lgToggle U.S.A. version)
This key is not operated in the area which lb unit is used only

11
10
3. TARE OPERATION
Load tare on the hook.
Press "TARE" key.
"0" is displayed, and the tare is subtracted.
Remove tare on the hook.
Press "TARE" key.
The tare weight is cleared appers in the display window.
THE OPERATION
1. POWER ON
Press "ON/OFF" key.
2. ZERO OPERATION
Press "ZERO" key when the zero point is drifted.

13
12
TEST MODE
How to enter TEST Mode
Pressing the key for 2 seconds in the normal mode, enter the test
mode with display the message Press the key or key
to execute the test mode.(test2~5 is same sequence.)
I f press the key in the test mode, test mode is ended and then
enter the normal mode.
(Ref. You can't enter the test mode while in the state of Hold or Tare.)
Test menu ( tESt 1~5)
: Display Test
: Load Cell and A/D Conversion Test
: The IC(CPU) Temperature Test
: Battery Test
: Key Test
4. UNIT CONVERSIONS
Place an item to be weight on the hook.(GRAM MODE)
lb
Remove weight on the hook.
Press "kg/lb" key.(POUND MODE)
Place an item to be weight on the hook.
ZERO
TEST 2
FUNCTION : Load Cell and A/D Conversion TEST
KEYS
: execute test 2
LCD SCREEN DESCRIPTION
Display TEST 2 state
Display the digital value
Corresponding to current weight
Available keys
KEYS FUNCTIONS
ZERO
Exit Test mode and then enter the Normal mode
Use the backlight(only TEST 1)
Move the next menu and then execute TEST Mode
or
or
TEST 1
FUNCTION : LCD TEST
KEYS
Backlight
execute
test 1
LCD SCREEN DESCRIPTION
Display TEST 1 state
Display all digits on the screen
ZERO

15
14
TEST 3
FUNCTION : The IC (CPU) Temperature TEST
KEYS
: execute test 3
LCD SCREEN DESCRIPTION
Display TEST 3 state
Display the temperature of IC [CPU]
or
TEST 4
FUNCTION : Battery TEST
KEYS
: execute test 4
LCD SCREEN DESCRIPTION
Display TEST 4 state
Display the current value of battery
or
SET Menu (F0 F6)
TEST 5
FUNCTION : Key Test
KEYS
: execute test 5
LCD SCREEN DESCRIPTION
Display TEST 5 state
Pressing the key to test shows
(Power, Zero, Tare, HoLd)
or
SET MODE
How to enter SET Mode
Turn on the power while pressing key , after displaying the message
and then enter the SET Mode.
Return to the normal mode if you press the key or key long.
TARE
Menu
F 0
F 1
F 2
F 3
Menu
F 4
F 5
F 6
Contents
Setting the value of Digital-Filter
Setting the stable time of weight
Setting the stable range of weight
Setting the Back Light
Contents
Weight Backup
Setting the hold type
Setting the time of Auto Display Off
Available keys
TARE
ZERO
or
or
Menu Cycle ( F0 ~F6F0)
Increase the setting value
Decrease the setting value
Store the current setting value (Not exit)
After storing all setting value and then exit SET Mode
KEYS DESCRIPTION

17
16
KEYS
LCD SCREEN Setting value
DESCRIPTION
00 : Not use Digital-Filter
(Speed of display conversion is fast)
~
09 : Speed of display conversion is slow
TARE
F 0
FUNCTION : Setting the value of Digital-Filter
KEY
LCD SCREEN Setting value
DESCRIPTION
000 : Setting the time within 0.5 second
09 : Setting the time within 1.4 second
TARE
F 1
FUNCTION : Setting the stable time of weight (step : 0.1 sec.)
KEY
LCD SCREEN Setting value
DESCRIPTION
00 : Setting the range within 0.5 division
09 : Setting the range within 4.0 division
TARE
F 2
FUNCTION : Setting the stable range of weight (step : 0.5 division)
KEY
LCD SCREEN Setting value
DESCRIPTION
Always Off
Always ON
Back Light ON whenever weight is loaded
Back Light ON Whenever the weight is stable
TARE
F 3
FUNCTION : Setting the stable range of weight (step : 0.5 division)
KEY
LCD SCREEN Setting value
DESCRIPTION
Not use
Use (Zero, Tare, Status)
TARE
F 4
FUNCTIN : Weight Backup
KEY
LCD SCREEN Setting value
DESCRIPTION
TARE
F 5
FUNCTION : Hold Type
Calculate the average of
moving weight
Calculate the peak of
moving weight
Average
Hold
Peak
Hold
Description of HOLD Function
1. When the F5 is set to 00
Case 1 ( When the hold key is pressed in unloaded state) : Auto-Average Hold Function
If you press the Hold key in unloaded state, after displaying the and
the Hold function Is started.
Just When the weight is changed(above the 1divison) , being displayed
and the changed weight is hold
If you want to release the Hold function, have to press Hold key in unloaded
state certainly
When the Hold function is released, being displayed the message
Case 2 (When the hold key is pressed in loaded state) : Manual Hold Function
When you want to hold the current weight, just press Hold key.
After displaying the and the Hold function Is started.

19
BATTERY OPERATION
1. CHARGE
A. The rechargeable battery inside the NC scale must be charged using the
voltage which don't exceed 10% of the voltage indicated on the adapter.
B. If the voltage for charging exceed the above limit, the battery would be
damage.
C. Check if an output voltage is DC 12V 1.5A.
D. Remove ADAPTOR JACK COVER of REAR PANEL and connect
adaptor with an scale like the following figure.
2. Discharge
If you use the battery continuously, discharge time is as fallows.
A. with the backlight -> approximate 22 hours
B. without the backlight -> approximate 50 hours
E. If adaptor is connected with the scale, Red lamp is on(Power) and Charge
lamp is on(Red : charge, Green : complete charge)
F. Time of charge is approximate 4~5 hours.
G. If main power switch is off, charge is not operate.(Red lamp is blinking.)
Do check main power switch(Main power switch is located Rear panel)
18
In Hold state
- if the weight is not changed, the current weight is not changed
- if the weight is changed(above the 10% of Max capa), the current weight is changed and
the Hold function is released
- Whenever unloaded states, the Hold function is released automatically
- Also you want to release the Hold function, have to press Hold key in unloaded state
certainly
- If Hold function is released, you can see the massage .
2. When the F5 is set to 01
The maximum value of weight is showed when the Hold key is pressed
If the Hold key is pressed once, the peak Hold function is started
If Hold function is released, you can see the massage .
If the Hold key is pressed again, the Hold function is released
If Hold function is released, you can see the massage .
(Ref> If the weight is 0 lb or 0 kg, this function is not operated.)
KEY
LCD SCREEN Setting value
DESCRIPTION
Not use
Display all off in 3 minutes
Display all off in 6 minutes
Display all off in 9 minutes
TARE
F 6
FUNCTION : Setting the time of Display off automatically
Ref. 1. Operates the function of F6 which anything is set when the weight is in zero state.
2. The Display is ON when the weight is not in zero state or when the any key
In the remote controller and front cover is pressed

21
20
3. LOW BATTERY("BAT.") SIGNAL
A. To protect over-discharge of Battery, A. To protect over-discharge of Battery,
BAT LAMP is on when battery has to be charge. (80% of full charged voltage)
B. A scale turns power off automatically below the fixed voltage
(75% of full charged voltage). You can show LobAt message before turn power off.
BAT LAMP is on when battery has to be charge.
(80% of full charged voltage)
B. A scale turns power off automatically below the fixed voltage
(75% of full charged voltage).
You can show LobAt message before turn power off.
REMOTE CONTROL
ASSEMBLE AND DISASSEMBLY
DEVICE PIN
HOOK
LOAD PIN
Blinking whenever key is pressed
Display ON/OFF(Toggle)
It makes the current weight within the zero range to zero
(2% of maximum capacity)
It is used to weigh with a container
It is used to weigh the moving things It is used to change the unit of
weight (kg lb Toggle U.S.A. version)
This key is not operated in the area which lb unit is used only
RED LED
ON/OFF KEY
ZERO KEY
TARE KEY
HOLD KEY/
Kg -> lb KEY
1.5V
AA Size(14.5 X 50.5 m/m) - Battery
Place batteries in correct polarity(+)(-)direction.
Do not use new batteries with the used
Voltage
Size
Note on the
replacing battery
* Remote controller is working within 60 and distance of 7~10m
ON
OFF
Red LED
ON/OFF Key
TARE Key ZERO Key
HOLD Key
(kg lb)

23
22
ERROR MESSAGES & TROUBLE SHOOTING
Err 01
Cause
The Indicator can't be initialized due to platform shaking
Management
Put the platform on a flat place without shaking and power on
If you can see the same massage continuously, call your CAS dealer please.
Err 02
Cause
Initial A/D value of zero is too high or low, when the power is ON
Or in the case of NC-I is lying on the flat
Management
Check that the load cell is well connected with main unit
Or calibration is done wrongly
Don't lie the NC-1 on the flat
If you can see the same massage continuously, call your CAS dealer please.
Err 09 (Zero Key)
Cause
The current weight exceeds zero range (2 % of Maximum Capacity)
Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.
Err 13 (Zero Range when the Power is ON)
Cause
The initial zero range exceeds above 10% of maximum capacity
Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.
BAT or LobAt
Cause
Need to recharge of battery
(After display LobAt , the power is off automatically in 2 second)
Management
Connect the adaptor with outlet of charge battery of rear part
(Charging time is about 4~5 hours)
If you can see the same massage continuously, call your CAS dealer please.
oUEr
Cause
Too heavy object is put on platform so that it may exceed the maximum capacity
Management
Never put the weight exceeding the maximum capacity
If you can see the same massage continuously, call your CAS dealer please.
If the load cell is damaged, replace the load cell
* Ref. The function of this manual can be changed without previous warning !!