Logger operation Sensor settings
DS 300-P includes a data logger that can record up to 1,000,000 measurement values.
In order to configure the logger function and to activate/deactivate it, there are several
functions available which are explained below:
Set logging rate and
averaging option The logging rate defines the interval data should be recorded in
the memory. It will record from every activated channel a sam-
ple. The averaging option can be used to calculate an average
value. I.e. DS 300-P is measuring every 1 second and if the
logging rate is 10 seconds, it will calculate an average value
out of the last 10 measurement values and store it as a re-
corded sample.
Key start/stop log-
ging Starts or stops the data recording. Whenever a new recording
is started a new file is created in the memory.
Select logging
channels Those channels which should be recorded with the data logger
must be selected first. For example, if Flow is selected to be
recorded the flow value of all connected flow sensors will be
recorded. Attention: sensors connectable on port 3...6 are
shown as they have been selected in the sensor setting menu.
Set logger memory
mode It can be selected to stop logging when memory is full, or the
“wrap around mode” can be chosen, which then will overwrite
the oldest values as soon the memory is full.
Show logger memory Informs about the current protocol:
- number of channels programmed for logging
- number of values stored per channel
- how many values still can be recorded
- the estimated logging time
- memory status. Start/stop conditions
Set time start
condition DS 300-P can be programmed to start at a certain time. Pro-
grammed time must be 10 seconds later than the system time.
View protocol Single protocols (files) can be viewed. DS 300-P will show the
available protocols with date, number of channels, number of
recorded values per channel and min, max and mean values.
Delete protocol Single protocols (files) can be deleted. DS 300-P will show the
available protocols with date, number of channels and number
of recorded values per channel.
Format logger This function is used to free all memory of DS 300-P. It should
be used with care, as recorded protocols will be erased per-
manently.
The following chapter describes the settings related to the sensor itself, which can be
changed on the DS 300-P menu. For that purpose select in the menu the function
“Sensor settings”.
After the selecting “Sensor settings”, the next screen will show all detected flow/dew
point sensors of terminal 1 and 2 and the selected analogue sensor of terminal 3 to
6. Changes can be done individually for each sensor, by selecting the terminal where
the sensor is connected to.
The next chapters describe the available settings of the different sensors.
For CS flows sensors following settings can be made:
Set tube diameter: for flow calculation
Set total consumption: counter can be set to any value
Set flow unit: Selection of the desired flow unit
Set consumption unit: selection of the desired consumption unit
Set reference pressure: in order to calculate the standard flow
Set reference temperature: in order to calculate the standard flow
Set gas type: Select the gas which is measured with the flow sensor.
Note:
Reference pressure and reference temperature are not related to the actual process
pressure or temperature. They are used to calculate the standard flow at standard
conditions, for example: 1000 hPa, 20 ºC.
Changes on the sensor settings are downloaded immediately into the sensor as soon
the changes are confirmed by pressing <Enter>.