eXact Micro 7+ User manual

Index:
Parameter Page MENU
Free Chlorine (DPD-1) 4 CL1
Permanganate (DPD-1) 4 CL1
Total Chlorine (DPD-3) 5 CL1
Ozone (DPD-4) 6 CL1
Total Chlorine (DPD-4) 6 CL1
pH 6 PH2
Bromine 7 BR3
Total Alkalinity 7 AL4
Calcium Hardness (CaCO
3
) 8 CA5
Copper 8 CU6
High Range Free Chlorine (HRC) 9 HR8
(NOTE: The below tests are read as %Transmission and
require the use of a conversion table)
General TR7 Procedure 10 Tr7
Acid pH 11 Tr7
Alkali pH 12 Tr7
Ammonia 13 Tr7
Chloride (in Tap Water) 14 Tr7
Chloride (Pool Salt Systems) 15 Tr7
Chlorine Dioxide (DPD-1) 16 Tr7
Chromium (VI) 17 Tr7
Cyanuric Acid 18 Tr7
Fluoride 19 Tr7
Hydrogen Peroxide (Low Range) 20 Tr7
Hydrogen Peroxide (Mid Range) 21 Tr7
Total Iron, TPTZ 22 Tr7
Total Iron, Ferro (Fe
+2
/Fe
+3
) 23 Tr7
LR Total Hardness (up to 77ppm)
24 Tr7
Manganese 25 Tr7
Nitrate 26 Tr7
Nitrite 27 Tr7
Peracetic Acid (PAA) 28 Tr7
Sulfate (SO
4
) 29 Tr7
6XO¿GH+
2
S) 30 Tr7
Phosphate 31 Tr7
6SHFL¿FDWLRQV
About Your Photometer 3
Reference Standard Procedure 32
Tips for Best Accuracy 33
Battery Installation 34
Warranty (2 year) 34
Reorder Information 35
USEPA Compliance 2, 36
Kit Components 36
Visit us online at sensafe.com/micro7+ for up-to-date
product information & NEW tests available.
IDEAL FOR DRINKING WATER, POOLS AND SPAS,
ENVIRONMENTAL, & EDUCATIONAL TESTING
Revision 10/28/11
486691-K Standard Kit
486691-KP Pool/Spa Kit
486691-WD Well Driller Kit
The eXact®Micro 7+ Advanced
Photometer System has been
designed for use with the eXact®Strip
Micro reagent delivery system.
© 2011 Industrial THVW6\VWHPV,QF(;$&7
®
is a registered trademark of Industrial THVW6\VWHPV,QF5RFN+LOO6&86$
Micro 7+
Advanced Photometer System
863DWHQW1R863DWHQW1R6RXWK$IULFDQ3DWHQW1RDQGLQWHUQDWLRQDO
SDWHQWDSSOLFDWLRQVLQFOXGLQJ,QWHUQDWLRQDO3DWHQW$SSOQ1R3&786DQG(XU3DW$SS
USEPA, DIN, & ISO Compliant for Free & Total Chlorine Testing
(4500-CL G, DIN Standard 38 408 G4, ISO 7393/2)
Manufactured By: Industrial Test Systems, Inc.
/DQJVWRQ6WUHHW5RFN+LOO6&86$
Phone: 1-800-861-9712 - INSIDE THE U.S.
1-803-329-9712 - OUTSIDE THE U.S.
Fax: 1-803-329-9743
WWW.SENSAFE.COM
www.poolcheckonline.com
Micro 7+ is
Manufactured
and tested in
an ISO 9001
Facility
NOW WITH HIGH RANGE FREE CHLORINE (HR8 MENU)

2
eXact®Micro 7+ Photometer
F
E
D
C
B
A
G
H
+
–
ITEM PICTURED
LETTER
COMPONENT
NAME
AMixing Cap/Cell Cover
BCell (Built-in Plastic, 4 ml)
CLCD Display
DREAD Button
EMENU Button
FZERO/ON Button
GBattery Cover with loop for lanyard
HAAA Batteries (x4) (Not Included)
(Used for Cyanuric
Acid, Fluoride, Iron,
and Manganese tests,
and when testing
outdoors)
eXact®0LFUR0HWHU6SHFLÀFDWLRQV
Measurement Method: Photometric
Light Source: Light Emitting Diode (LED)
Wavelength: 525 nm
Transmission Range: 100 - 0.00 %T
Photometric Precision: +/- 0.1/0.01 %T
Automatic Range Selection: 6HH6SHFL¿FDWLRQVEHORZ
Display: 3-digit customized liquid crystal display
with annunciators
CELL Pathlength: 20mm
Cell Chamber: Custom-molded, proprietary, PET plastic
IXVHGLQWRFKDPEHUQRQUHPRYDEOH
Sample Required: 4 ml (0.13 oz)
Operating Temperature Range: 0 - 50°C (32° - 122°F)
Power Supply: $$$DONDOLQHEDWWHULHV1RW,QFOXGHG
Battery Life: !WHVWVZLWKDONDOLQHEDWWHULHV
Electromagnetic Compliance: (PLWWHG,QWHUIHUHQFH(1
(EMC) ,PPXQLW\WR,QWHUIHUHQFH(1
Waterproof Rating: ([FHHGV,3
Weight: ,QVWUXPHQWJR]
Dimensions: ,QVWUXPHQW:['[+FP
[[LQ
eXact® Strip Micro 7+ KDVEHHQGHVLJQHGWRRIIHUWKHXVHUDPRUH³*UHHQ´DQGFRVWHIIHFWLYHDOWHUQDWLYHWRWHVWLQJ
,QVWHDGRIXVLQJDPOZDWHUVDPSOHeXact® Strip Micro 7+ XVHVDPOZDWHUVDPSOHZKLFKXVHVXSWROHVV
FKHPLFDOSHUWHVW7KHDFFXUDF\RIWKHPHWHULVPDLQWDLQHGE\GHVLJQLQJWKHSKRWRFHOOZLWKDPPSDWKOHQJWK
We offer a “Green” Alternative
1
3HUIRUPDQFHYHUL¿HGZLWKYDULRXVVDOWV\VWHPVSRRODQGVSDZDWHU
VDPSOHVZLWKRSWLPDOZDWHUWHPSHUDWXUHDWÛ&Û)
2
0HDVXUHPHQWUHTXLUHVDFRQYHUVLRQFKDUWIRUYDOXH
eXact
®
0LFUR'LUHFW5HDG6SHFL¿FDWLRQV
R092211
CL1
PH2
BR3
AL4
CA5
CU6
TR7
HR8
)UHH&KORULQH
7RWDO&KORULQH
'3''3'
S+
Bromine
7RWDO$ONDOLQLW\
&DOFLXP+DUGQHVV
DV&D&2
&RSSHU&X
7UDQVPLVVLRQ
2
2WKHUWHVWVFRPLQJVRRQ
+5)UHH&KORULQH
SSP
SSP
SSP
SSP
SSP
SSP
7
SSP
±
SSP
±
SSP
±
SSP
S+
±
SSP
±
SSP
±
SSP
±
SSP
±
SSP
±
SSP
±
±
SSP
SSP
SSP
1
SSP
1
SSP
SSP
7
7
1
0HQX 7HVWVIRU15DQJH 5HVROXWLRQ $FFXUDF\

3
About Your eXact®Micro 7+ Instrument
In order to save power, the meter is designed to turn off after 3 minutes (timed from the last button pressed).
Should the meter turn off in the middle of a test, the last stored zero in the meter will remain valid when the
meter is turned on again. Also, the test result is stored in memory for easy retrieval.
The eXact
®
Micro 7+ meter is controlled by three buttons:
1. ZERO/ON:KHQ¿UVWSUHVVHGWKLVEXWWRQWXUQVWKHPHWHURQ:KHQWKHPHWHULVRQDQGWKLVEXWWRQLV
pressed, it zeroes the sample in the cell. Once the meter is zeroed, this zero value applies to all parameters
and is stored and retained even when meter turns off. However, it is recommended that each new water
sample analyzed is zeroed before testing, to maximize sensitivity and accuracy.
2. MENU: With each press, the MENU button advances through the tests in the following sequence: CL1,
PH2, BR3, AL4, CA5, CU6, TR7, HR8. Each test menu can store up to 20 results. To retrieve the stored
results, go to the desired test using the MENU key. When the desired test is displayed, press and hold
down the MENU key. Continue holding down the MENU key to scroll the stored results for that test, starting
with the most recent result. The meter will display, from memory, the last 20 readings in sequence beginning
with -20, which is the latest result, followed by -19, which is the 2
nd
ODWHVWUHVXOWHWFDQG¿QDOO\ZKLFKLV
the oldest result retained. Only the last 20 readings are stored in each menu. This meter is able to store 160
results in memory (20 in each menu).
3. READ: When pressed once, this button starts the timer for the parameter being tested. When pressed a
second time the meter exits the timer and immediately prepares to colorimetrically measure the sample, and
simultaneously stores the measurement in memory.
If the parameter being measured is below or above the detection range, the display will show “LO” (Under
Range) or “HI´2YHU5DQJHUHVSHFWLYHO\7KLVIHDWXUHLVPHQXVSHFL¿FDQGGRHVQRWDSSO\WRDOOSDUDPHWHUV
AWWA
4500-Cl G
eXact®
DPD-1
Anhydrous DPD sulfate
Anhydrous Na2HPO4
Anhydrous KH2PO4Na2
EDTA
1.5%
33.4%
64.0%
1.1%
1.5%
33.4%
64.0%
1.1%
Component
(Free Chlorine)
About The Accuracy / Calibration Of The Micro 7+ System
$OOWHVWVKDYHEHHQFDOLEUDWHGXVLQJFHUWL¿HGUHIHUHQFHVWDQGDUGVDQGVWDQGDUGDQDO\WLFDO
VSHFWURSKRWRPHWULFPHWKRGV7KHDOJRULWKPVLQWKHVRIWZDUHUHÀHFWWKHEHVWFRUUHODWLRQRIWKHH;DFW
®
Micro 7+ Systems against the AWWA, US EPA , DIN, and ISO reference test methods for chlorine.
Studies show that the eXact
®
Micro 7+ System repeatedly agrees with an EPA Compliant reference
method greater than 99% (R2= 0.9989, 0 - 6.0 ppm - see page 36). The eXact
®
Micro 7+ Advanced
3KRWRPHWULF6\VWHPKDVEHHQIDFWRU\FDOLEUDWHGIRU\RXUFRQYHQLHQFH<RXFDQH[SHFWWKH¿[HG
calibrations in the meter to be valid for the life of the meter because of the quality, Long-Life LED, the
photo cell, and the software as written into the meter. This is why the meter comes with a 2-Year
Warranty.
NOTE: Test algorithms in the new version of photometer (serial numbers above 5000) give
accurate results in fresh and salt water.
Compliance VHULÀFDWLRQIRU)UHHDQG7RWDO&KORULQH7HVWLQJ
This DPD test system is accepted by most health departments because this test is USEPA (DIN
Standard 38 408 G4, ISO 7393/2) accepted for testing requirements for Free and Total Chlorine. The
Micro 7+ meter uses a wavelength of 525nm; and the compliance requirement is that the colorimeter
wavelength is between 490 and 530nm. The eXact
®
Strip Micro CL (DPD-1) uses the same reagents
and proportions, and the resulting solution pH is maintained between 6.2 and 6.5 as VSHFL¿HGby
AWWA (American Water Works Association) method 4500-Cl G. It should be understood that the
USEPA does not "approve" commercial DPD delivery systems such as reagent powder pillows, tablets,
dispensers, or eXact
®
Strip DPD delivery devices. The eXact
®
Strip Micro CL (DPD-1) for Free
Chlorine, and the eXact
®
Strip Micro CL (DPD-3) or the eXact
®
Strip Micro CL (DPD-4) for Total
Chlorine meet your reportable testing requirements because the eXact
®
Strip Micro CL delivers the
same chemicals in identical proportions (see table below); therefore, the system is compliant. Likewise,
AWWA proportions are followed as required for Total Chlorine measurements using Potassium Iodide.

4
DPD-1 Free Chlorine Test Procedure
1
6
5
2a
3
4
2b
Remove one (1) eXact
®
Strip Micro CL (DPD-1),
Part No. 486637 from the bottle before beginning
the test. Set the strip in a dry, convenient place and
recap the bottle immediately.
REMOVE STRIP
1
Press the ZERO/ON button to power the meter on;
the display will show all annunciators, then the
current MENU selection, followed by the last reading.
TURN METER ON
2
Dip the eXact
®
Strip Micro CL (DPD-1), Part No.
486637 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During
this time move the strip in a gentle back and forth
motion. Remove and discard the strip after “1” on
the display disappears.
DIP STRIP AND PRESS “READ”
6
Press the ZERO/ON button. The cursor will move
across the display followed by 0.00 PPM. Sample is
ready for testing.
ZERO METER
5
Rinse the CELL at least 3 times with the water sample
you will be testing - rinsing minimizes the potential for
cross-contamination from a previous test. Finally, fill
cell to capacity (4ml) with the water sample.
RINSE AND FILL CELL WITH SAMPLE
4
Press and re-press the MENU button until the
display shows the parameter CL1.
CL1 is also used for testing:
Total Chlorine (DPD-4), Ozone (DPD-4), Permanganate (DPD-1), and
Total Chlorine (DPD-3). (Contact ITS for specs and details if you are
planning on using CL1 for Permanganate or Ozone measurements)
SELECT TEST: CL1
3
CL
CL1
The cursor will move across the display
while the
meter prepares to measure the sample. Record result
displayed (this result is automatically stored in CL1).
RECORD RESULT DISPLAYED
7
DO NOT discard the sample from the Free Chlorine
test if you are planning to run eXact®Strip Micro DPD-3
(Total Chlorine) Procedure. Move directly to steps 8-10
on page 5. Otherwise, rinse the cell immediately.
For best results with this eXact®Micro 7+ Photometer and DPD-1 strips:
This version of the eXact®Micro 7+ meter has a range to 11ppm Chlorine. The algorithm for
Free Chlorine (CL1) with the use of DPD-1 strip gives most acurate results below 6ppm.
For Chlorine testing above 6ppm, you may use the eXact®Strip Micro HRC, Part No.
486672, product. This procedure can be found on page 9.

5
DPD-3 Total Chlorine Test Procedure
This procedure is only valid when run as a continuation of the
eXact
®
Strip Micro CL (DPD-1 Free Chlorine ) Test Procedure
located on the previous page.
8
Remove one (1) eXact
®
Strip Micro CL (DPD-3), Part No. 486638 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
10 PRESS READ AGAIN
Press READ again and the meter will count down and display the next reading. If this
reading matches the previous result, then record this as the Total Chlorine result.
This value
is automatically stored in CL1.
After testing is completed, rinse cell immediately. Record the
Total Chlorine as the highest value the meter displayed.
9
Dip the eXact
®
Strip Micro CL (DPD-3) into the CELL and immediately press READ. This
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip when “1” on the display disappears. The
cursor will move across the display while the meter prepares to measure the sample. This
result is automatically stored in CL1 (NOTE: The Iodide added with DPD-3 will, in the presence
of Combined Chlorine or Chloramines, convert into Iodine).
DIP STRIP AND PRESS “READ”
*NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires the reading to
be made after 2 minutes from the time the KI is added. For compliance testing, you must
time the two minutes and then make your measurement. NOTE: From testing in our lab,
water samples above 70°F (20°C), generally, reach a stabilized reading quicker then 2 minutes.
eXact
®
Strip Micro CL (DPD-1/DPD-3/DPD-4) Interferences (part nos. 486637/486638/486670)
Interfering Substance
Acidity
Alkalinity
Bromine & Bromamines, Br
2
Chlorine Dioxide, ClO
2
Copper, Cu
+2
Iodine, I
2
Manganese, oxidized (Mn
+4
, Mn
+7
)
or Chromium, oxidized (Cr
+6
)
Monochloramines (NH
2
Cl)
(applies to DPD-1 only)
Ozone, O
3
Peroxides
pH
Interfering Levels & Treatments
If sample has acidity above 150mg/L CaCO
3
test may not develop
full color. Neutralize to pH 6.0 to 7.0 with 0.5N Sodium hydroxide.
If sample has alkalinity above 200mg/L CaCO
3
test may not develop
full color. Neutralize to pH 6.0 to 7.0 with 0.5N Sulfuric acid.
Color similar to free chlorine reaction at all levels.
Color similar to free chlorine reaction at all levels.
Color development is reduced above 10 ppm (mg/L).
Color similar to free chlorine reaction at all levels.
See AWWA procedure 4500-CL F, 1(d) for removal of interferences.
Monochloramine interferences are known to occur in free chlorine
DPD methods. This interference is dependent on temperature and
monochloramine concentration.
Color similar to free chlorine reaction at all levels.
Interference is possible.
Typical pH samples of potable water with a pH of 6.0 to 9.0 are
OK. If outside this range adjust to pH 6.0 to 7.0 using acid
(0.5N Sulfuric acid) or base (0.5N Sodium hydroxide).
CL
CL1
CL1: Chlorine and Iodine react with N,N-diethyl-p-phenylenediamine as it is released from the strip to
form a magenta color, directly proportional to the Chlorine concentration. (Ozone, Bromine, and
Permanganate also form the color)

6
CL
CL1
PH
PH2
6
Dip the eXact® Strip Micro PH, Part No. 486639 into the CELL and immediately press READ. This
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after “1” on the display
disappears*. The cursor will move across the display while the meter prepares to measure the
sample. Record result displayed (this result is automatically stored in PH2). After testing, rinse cell
immediately.
NOTE: For best results, Total Alkalinity of the sample should be 40-140 ppm.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. When the display shows 0.0 PH, the sample is ready for testing.
ZERO METER*
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\POZLWK
the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the MENU button until the display shows the parameter PH2.
SELECT TEST: PH2
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Remove one (1) eXact
®
Strip Micro PH, Part No. 486639 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
pH Test Procedure
NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires the reading to
be made after 2 minutes. 2 minute wait is not necessary for Ozone measurements.
6
Dip the eXact® Strip Micro CL (DPD-4), Part No. 486670 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion. Remove and discard the strip after “1” on the display disappears*. The cursor will move across
the display while the meter prepares to measure the sample. Record result displayed (this result is
automatically stored in CL1).
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button.
The cursor will move across the display,
followed by 0.00 PPM.
Sample is ready for testing.
ZERO METER*
7
Press READ again and the meter will count down and display the next Total Chlorine result. If this
reading matches the previous result, then record this as Total Chlorine value (this result is stored
in CL1). After testing is completed, rinse cell immediately. Record the Total Chlorine as the highest
value the meter displayed.
PRESS READ AGAIN
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
PLQLPL]HVWKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWR
capacity (4ml) with the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the MENU button until the display shows the parameter CL1.
SELECT TEST: CL1
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Remove one (1) eXact
®
Strip Micro CL (DPD-4 for Total Chlorine or Ozone), Part No.
486670 from the bottle before beginning the test. Set the strip in a dry, convenient place and
recap the bottle immediately.
REMOVE STRIP
DPD-4
(Total Chlorine or Ozone)
Test Procedure

7
Bromine DPD-1 Test Procedure
1
Remove one (1) eXact
®
Strip Micro CL (DPD-1), Part No. 486637 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact
®
Strip Micro CL (DPD-1), Part No. 486637 into the CELL and immediately press
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec). Remove and discard the strip after “1” on the
display disappears. The cursor will move across the display while the meter prepares to
measure the sample. Record result displayed (this result is stored in BR3 memory). After
testing is completed, rinse cell immediately and press MENU to select next test.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENU button until the display shows the parameter BR3.
SELECT TEST: BR3
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\POZLWK
the water sample.
FILL METER WITH SAMPLE
1
Remove one (1) eXact
®
Strip Micro AL, Part No. 486641 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact® Strip Micro AL, Part No. 486641 into the CELL and immediately press READ. This
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after “1” on the display disappears*.
NOTE: For water temperatures above 95°F/35°C (hot tubs), remove and discard the strip when the
timer displays “10”, countdown continues. For the hot water samples, a 10-second dip time is best.
The cursor will move across the display while the meter prepares to measure the sample. Record
UHVXOWGLVSOD\HGWKLVUHVXOWLVDXWRPDWLFDOO\VWRUHGLQ$/$IWHUWHVWLQJLVFRPSOHWHGULQVHFHOO
immediately.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER*
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENUEXWWRQXQWLOWKHGLVSOD\VKRZVWKHSDUDPHWHU$/
SELECT TEST: AL4
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\POZLWK
the water sample.
FILL METER WITH SAMPLE
Total Alkalinity Test Procedure
CL BR3
AL AL4

8
Calcium Hardness Test Procedure
1
Remove one (1) eXact
®
Strip Micro CA, Part No. 486629 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact
®
Strip Micro CA, Part No. 486629 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip after “1” on the display disappears.
The cursor
will move across the display while the meter prepares
to measure the sample. Record result
displayed (this result is automatically stored in CA5). After testing is completed, rinse cell
immediately and use brush to remove any residual from previous test.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENU button until the display shows the parameter CA5.
SELECT TEST: CA5
4
Rinse the CELL 2 or 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\PO
with the water sample.
FILL METER WITH SAMPLE
1
Remove one (1) eXact
®
Strip Micro CU, Part No. 486632 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact
®
Strip Micro CU, Part No. 486632 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip after “1” on the display disappears*. The
display will immediately start counting up from 1 to 20 (this extra time allows more thorough color
development).
At 20 sec, the cursor will move across the display while the meter prepares
to
measure the sample. Record result displayed (this result is automatically stored in CU6).
After
testing is completed, rinse cell immediately.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER*
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENU button until the display shows the parameter CU6.
SELECT TEST: CU6
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
PLQLPL]HVWKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWR
FDSDFLW\POZLWKWKHZDWHUVDPSOH
FILL METER WITH SAMPLE
Copper Test Procedure
CA5: Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate, directly
proportional to the Calcium concentration. Values are reported as Calcium Carbonate.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
CA
CA5
CU
CU6

9
TR TR7
75'LIIHUHQWLRQVUHDFWZLWKDVSHFL¿FLQGLFDWRUWRIRUPDFRORURUDSUHFLSLWDWHWKDWSURSRUWLRQDOO\LQGLFDWHV
WKHFRQFHQWUDWLRQRIWKHLRQSUHVHQWE\WKHWUDQVPLVVLRQYDOXHIRXQG$FRQYHUVLRQWDEOHLVWKHQXVHGWR
GHWHUPLQHWKHLRQFRQFHQWUDWLRQXVLQJWKHWUDQVPLVVLRQYDOXH7KHDGYDQWDJHRIXVLQJWUDQVPLVVLRQ
PHDVXUHPHQWLVWKDWPDQ\GLIIHUHQWLRQVFDQEHGHWHUPLQHGLQRQH0(18ZKLFKH[SDQGVWKHÀH[LELOLW\RI
WKLVPHWHU2QFH\RXKDYHGHWHUPLQHGWKH7UHVXOWIRUWKHWHVW\RXUDQ¿QGWKLV7UHVXOWLQWKHFRQYHUVLRQ
FKDUWDQGUHDGWKHFRQFHQWUDWLRQFRUUHVSRQGLQJIRUWKLV77KH0LFURJLYHVWKH7YDOXHVDVGLJLWV
H[DPSOHEXWRQO\WKH¿UVWWZRGLJLWVDUHXVHGLQWKHFKDUWVURXQGRII\RXU7YDOXHWRWZRGLJLWV7HVWV
WKDWFDQEHXVHGZLWKWKH0LFURLQ750(18DUHOLVWHGRQSDJH7\SLFDOO\WKHPRVWFRPPRQWHVWVDUH
VXSSOLHGZLWKWKHFRQYHUVLRQFKDUWLQWKLVERRNOHW)RUFRQYHUVLRQFKDUWVDQGLQIRUPDWLRQDERXWWKHWHVWVQRW
JLYHQLQWKLVERRNOHWYLVLWRXUZHEVLWH<RXFDQDOVRXVHWKH750(18WRGHYHORS\RXURZQFXVWRP
FRQYHUVLRQFKDUWIRUXQXVXDOVDPSOHVWRJHWPRUHDFFXUDWHUHVXOWV
6
Dip the eXact
®
Strip Micro (or add Reagent) into the CELLLPPHGLDWHO\SUHVVREAD7KLVVWDUWV
the 20 SECONDFRXQWGRZQWLPHU'XULQJWKLVWLPHPRYHWKHVWULSLQDJHQWOHEDFNDQGIRUWK
PRWLRQRemove and discard the after “1” on the display disappears*7KHFXUVRUZLOOPRYH
DFURVVWKHGLVSOD\ZKLOHWKHPHWHUSUHSDUHVWRPHDVXUHWKHVDPSOH5HFRUGUHVXOWGLVSOD\HGWKLV
UHVXOWLVDXWRPDWLFDOO\VWRUHGLQ75
$IWHUWHVWLQJLVFRPSOHWHGULQVHFHOOLPPHGLDWHO\
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ONEXWWRQ7KHFXUVRUZLOOPRYHDFURVVWKHGLVSOD\IROORZHGE\100 %T
7KHVDPSOHLVUHDG\IRUWHVWLQJ
ZERO METER*
4
Rinse the CELLDWOHDVWWLPHVZLWKWKHZDWHUVDPSOH\RXZLOOEHWHVWLQJULQVLQJ
PLQLPL]HVWKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWR
FDSDFLW\POZLWKWKHZDWHUVDPSOH
FILL METER WITH SAMPLE
3
3UHVVDQGUHSUHVVWKHMENUEXWWRQXQWLOWKHGLVSOD\VKRZVWKHSDUDPHWHUTR7
7HVWVOLVWHGRQSDJHVWKURXJKZLOOUHTXLUHWKH750(18DQGWKHFRQYHUVLRQFKDUWV
LQFOXGHG2WKHUWHVWVOLVWHGRQSDJHZLOOUHTXLUHWKHXVHRIDFRQYHUVLRQFKDUWDYDLODEOHDW
ZZZVHQVDIHFRP
SELECT TEST: TR7
2
Press the ZERO/ONEXWWRQWRSRZHUWKHPHWHURQWKHGLVSOD\ZLOOVKRZDOODQQXQFLDWRUV
WKHQWKHFXUUHQW0(18VHOHFWLRQIROORZHGE\WKHODVWUHDGLQJ
TURN METER ON
1
5HPRYHRQHeXact
®
Strip MicroSDUWQXPEHULVGHSHQGHQWXSRQWKHWHVWEHLQJUXQ
IURPWKHERWWOHEHIRUHEHJLQQLQJWKHWHVW6HWWKHVWULSLQDGU\FRQYHQLHQWSODFHDQG
UHFDSWKHERWWOHLPPHGLDWHO\
REMOVE STRIP
Transmission Test Procedure
*127(:KHQWHVWLQJRXWGRRUVVXQOLJKWIRUEHVWDFFXUDF\XVHWKH0L[LQJ&DS&HOO&RYHUZKHQ=HURLQJDQG5HDGLQJWKHVDPSOH

6
Dip the eXact
®
Strip Micro HR, Part No. 486672 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion.
Remove and discard the strip after “1” on the display disappears*. The
display will immediately start counting up from 1 to 120 (this extra time allows more thorough
color development).
At 120 sec, the cursor will move across the display
while the meter
prepares to measure the sample. Record result displayed (this result is automatically stored in
HR8).
After testing is completed, rinse cell immediately with brush and water.
DIP STRIP - (read carefully and follow procedure closely)
5Press the ZERO/ON button. The cursor will move across the display, followed by 0.0 PPM.
The sample is ready for testing.
ZERO METER*
2Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3Press and re-press the MENU button until the display shows the parameter HR8.
SELECT TEST: HR8
4Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\PO
with the water sample.
FILL METER WITH SAMPLE
High Range Free Chlorine Test Procedure
10
TR7
HR
HR8
1Remove one (1) eXact®Strip Micro HR, Part No. 486672 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Cyanuric Acid Table
Cyanuric Acid results require the table below. Follow eXact
®
Micro 7+ Cyanuric Acid Test Procedure (above) using
eXact®Reagent CY, Part No. 481652-II.
NOTE: For levels above 95ppm Cyanuric Acid, dilute the sample ½ or ¼ with distilled water and retest.
eXact
®
Reagent CY, Part No. 481652-II - for 4mL Samples
Rev. 040811-BT
%T
90
80
70
60
50
40
30
20
10
0
987654 3 2 1 0
11
5
Shake the bottle of eXact
®
Reagent CY, Part No. 481652-II to mix the chemical in the bottle. Then,
DGG¿YHGURSVRIH;DFW
®
Reagent CY to the cell and cap meter cell with mixing cap.
Press
READ to start timer, SODFHWKXPERU¿QJHURYHUFDSDQGPL[WKHVDPSOHE\WXUQLQJWKHPHWHU
upside-down repetitively during the 20 SECOND countdown. NOTE: Cover the cap completely
DQGKROG¿UPO\
When timer displays 1SODFHPHWHURQÀDWVXUIDFH7LPHWKHUHDFWLRQLQWKH
CELL for 40 seconds (timer not included). After the 40 seconds, press READ again to start
another
20 SECOND
countdown. At the end of the 20 seconds, the cursor will move across the
display while the meter prepares to measure the sample. Record result displayed (this result is
automatically stored in CY7). After testing is completed, rinse cell immediately and use brush to
remove any residual from previous test.
ADD REAGENT, CAP, PRESS “READ”, AND MIX
6
Find the “TR7” result in the table below to determine the Cyanuric Acid concentration in ppm
(parts per million). (Example: a “TR7” result of 75.3 (use only the 75 for the chart) equals a
Cyanuric Acid value of 6.8 ppm). Record result.
After testing is completed, rinse cell immediately.
USE TABLE
4
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
ZERO METER*
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the MENU button until the display shows the parameter TR7.
SELECT TEST: TR7
3
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\PO
with the water sample.
FILL METER WITH SAMPLE
Cyanuric Acid Test Procedure
CY TR7
0 0.5 1 1.5 1.8 2 2.2 2.4 2.7 2.9
3.2 3.4 3.7 3.9 4.2 4.4 4.9 5.2 5.4 5.6
5.9 6.1 6.3 6.5 6.8 7.3 7.6 7.8 8 8.3
8.8 9 9.3 9.5 9.8 10.3 10.5 10.7 11.2 11.5
11.7 11.9 12.2 12.7 13.2 13.5 13.7 14.2 14.4 14.6
15.1 15.6 16.1 16.3 16.6 17.1 17.6 18.1 18.6 19
19.5 20 20.5 21 21.5 22 22.5 22.9 23.4 23.9
24.4 24.9 25.9 26.9 27.3 27.8 28.8 29.8 30.8 31.7
32.2 33.2 34.2 35.6 36.6 38.1 39.6 41 43 44.9
46.9 49.3 52.2 55.2 59.1 64 70 80 95 >95
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Total Iron, TPTZ (Fe+2/Fe+3) Table
Iron results require the table below.
eXact
®
Reagent EZ Open REDUCER, Part No. 486601 - for 4mL Samples
Rev. 020311 TPTZThis table was calibrated using Fe+2 Iron Standards
%T
90
80
70
60
50
40
30
20
10
0
987654 3 2 1 0
7
Find the “TR7” result in the table below to determine the Iron concentration in ppm (parts per
million). (Example: a “TR7” result of 85.3 (use only the 85 for the chart) equals an Iron value of
0.08 ppm). Record result.
After testing is completed, rinse cell immediately.
USE TABLE
5
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
ZERO METER*
4
Tilt meter to discard about 0.2mL water in order to leave room for powder reagent. Add the
contents of one eXact®Reagent EZ Open REDUCER, Part No. 486601 to the CELL and cap
meter cell with mixing cap. Press READ to start the 20 SECOND countdown timer, place thumb
over cap, and mix the sample by turning the meter upside-down repetitively. When time displays
1, hold the meter upright and the cursor will move across the display, informing you that it is
about to measure the sample (ignore this result). Wait 40 seconds (timer not included).
ADD REAGENT, CAP, AND MIX
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the MENU button until the display shows the parameter TR7.
SELECT TEST: TR7
3
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Fill cell to capacity (4ml) with the
water sample.
FILL METER WITH SAMPLE
12
6
Dip the eXact
®
Strip Micro FE (TPTZ), Part No. 486631 into the CELL and immediately press
READ. This starts the 20 SECOND countdown timer. During this time, move the strip in a gentle
back and forth motion. Remove and discard the strip after “1” on the display disappears*. The
cursor will move across the display while the meter prepares to measure the sample (ignore this
result). Wait 20 seconds (timer not included) and press READ to start an additional 20 SECOND
countdown. When time displays 1, the cursor will move across the display, informing you that
it is about to measure the sample. Record result displayed (this result is automatically stored in TR7).
DIP STRIP AND PRESS “READ”
Total Iron, TPTZ (Fe+2/Fe+3) Test Procedure
TR7
FE
<0.02 <0.02 <0.02 <0.02 <0.02 <0.02 <0.02 <0.02 <0.02 0.02
0.02 0.04 0.05 0.06 0.08 0.09 0.10 0.11 0.12 0.14
0.15 0.16 0.17 0.18 0.20 0.21 0.22 0.24 0.25 0.26
0.28 0.30 0.31 0.32 0.34 0.36 0.37 0.38 0.40 0.42
0.43 0.44 0.46 0.48 0.50 0.52 0.54 0.56 0.57 0.58
0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74 0.76 0.80
0.82 0.84 0.86 0.90 0.92 0.94 0.98 1.00 1.02 1.06
1.10 1.12 1.16 1.20 1.24 1.28 1.32 1.36 1.40 1.46
1.50 1.56 1.62 1.68 1.76 1.84 1.92 2.00 2.10 2.20
2.30 2.44 2.58 2.74 2.92 3.14 3.42 4.00 5.00 >5.00
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Nitrate (NO3-) Table
Nitrate results require the table below. Follow eXact
®
Micro 7+ Nitrate (NO
3-
)
Test Procedure (above) using
eXact®Strip Micro NO3, Part No. 486655
NOTE: For levels above 45ppm Nitrate, dilute the sample ½ or ¼ with distilled water and retest.
eXact
®
Strip Micro NO
3
, Part No. 486655 - for 4mL Samples
Rev. BT 040711
%T
90
80
70
60
50
40
30
20
10
0
987654 3 2 1 0
5
Dip the eXact
®
Strip Micro NO
3
, Part No. 486655 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a very gentle
back and forth motion. Remove and discard the strip after “1” on the display disappears*.
The cursor will move across the display
, while the meter prepares to measure the sample.
Time the reaction in the cell for 580 seconds (timer not included). During this time, the meter
will shut off. When 580 seconds have elapsed, turn meter on and wait for the display to show last
reading. Then, press READZKLFKZLOOVWDUWD¿QDO20 SECOND countdown.
The cursor will
move across the display
, while the meter prepares to measure the sample. Record result
displayed (this result is automatically stored in TR7).
DIP STRIP -
(read carefully and follow procedure closely)
6
Find the “TR7” result in the table below to determine the Nitrate concentration in ppm
(parts per million). (Example: a “TR7” result of 66.3 (use only the 66 for the chart) equals a
Nitrate value of 3.40 ppm). Record result.
After testing is completed, rinse cell immediately
to remove zinc dust.
USE TABLE
4
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
ZERO METER*
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the MENU button until the display shows the parameter TR7.
SELECT TEST: TR7
3
Add sample water to the cell. For accurate results, brush the cell to remove any excess zinc
from previous tests. Rinse the CELL at least 3 times with the water sample you will be
testing - rinsing minimizes the potential for cross-contamination from a previous test. Finally,
¿OOFHOOWRFDSDFLW\POZLWKWKHZDWHUVDPSOH
FILL METER WITH SAMPLE
NOTE: Divide the above Nitrate result by 4.4 to determine Nitrate value as Nitrogen (NO
3
as N)
13
NO3
Nitrate (NO3-) Test Procedure
TR7
This table was calibrated using NO3-Nitrate Standards
0 0 0 0 0 0.1 0.2 0.3 0.4 0.5
0.6 0.7 0.8 0.9 1 1.1 1.2 1.4 1.5 1.6
1.8 1.9 2 2.1 2.2 2.3 2.5 2.6 2.7 2.9
3 3.1 3.3 3.4 3.5 3.7 3.8 3.9 4.1 4.2
4.4 4.5 4.7 4.8 5 5.2 5.3 5.5 5.6 5.8
6 6.2 6.4 6.6 6.7 6.9 7.1 7.3 7.5 7.8
8 8.2 8.4 8.7 8.9 9.2 9.5 9.8 10.1 10.4
10.7 11.1 11.4 11.8 12.2 12.6 13.1 13.6 14.1 14.7
15.3 16 16.8 17.6 18.5 19.6 20.9 22.3 24 25.6
27.6 30 33 37.3 43 45 >45 >45 >45 >45
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Rev. 041311-BT
%T
90
80
70
60
50
40
30
20
10
0
987654 3 2 1 0
14
CH
TR7
eXact
®
Strip Micro CH, Part No. 481657-II - for 4mL Samples
Sodium Chloride results require the table below. Follow eXact
®
Micro 7+ Chloride (as NaCl) Test Procedure (above)
using
eXact®Strip Micro CH, Part No. 481657-II. NOTE: To convert the NaCl value to Chloride (Cl–),
multiply the value from the chart below by 0.6. (Example: 1380ppm NaCl = 828ppm Cl–)
Chloride
(as NaCl)
Table
7
Find the “TR7” result in the table below to determine the Sodium Chloride concentration in ppm
(parts per million). (Example: a “TR7” result of 65.3 (use only the 65 for the chart) equals a Sodium
Chloride value of 1400 ppm). Record result.
After testing is completed, rinse cell immediately.
USE TABLE
6
Dip the eXact
®
Strip Micro CH (as NaCl), Part No. 481657-II into the CELL and immediately
press READ. This starts the 20 SECOND countdown timer. During this time move the strip
in a gentle back and forth motion. Remove and discard the strip after “1” on the display
disappears*. The cursor will move across the display while the meter prepares to measure
the sample. Record result displayed (this result is automatically stored in TR7).
DIP STRIP
5
Press the ZERO/ON button.
The cursor will move across the display,
followed by 100 %T.
Meter is ready for testing.
ZERO METER*
4
Use the small pipet provided to add 0.2ml of the sample to the water in the cell.
This is a
1 to 20 dilution of the pool water that is required to perform this test. NOTE: If you are using the
more accurate dilution procedure with the Dilution Kit (487200) or Mini Dilution Kit II (487202),
follow the procedure as provided with those kits.
ADD THE SAMPLE
3
Using deionized or distilled water, rinse the CELL at least 3 times to minimize potential cross
FRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOWKHFHOOWRWKHWRSZLWKGHLRQL]HGRUGLVWLOOHGZDWHU
PREPARE THE CELL
2
Press and re-press the MENU button until the display shows the parameter TR7.
SELECT TEST: TR7
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
Salt/Chloride (as NaCl) Test Procedure
For Pool and Spa Salt Systems
0 40 80 120 160 200 240 280 320 360
400 440 480 520 560 600 640 680 720 760
800 840 880 960 1000 1040 1080 1120 1160 1200
1240 1280 1320 1360 1400 1440 1480 1520 1560 1600
1640 1680 1720 1760 1800 1840 1880 1920 1960 2000
2040 2060 2100 2140 2180 2200 2240 2280 2320 2360
2400 2460 2500 2540 2580 2620 2660 2700 2740 2780
2820 2860 2900 2940 2980 3040 3080 3140 3200 3240
3280 3340 3400 3460 3520 3580 3640 3700 3760 3840
3940 4040 4200 4280 4440 4660 5000 5840 7000 >7000
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

15
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
Become familiar with the meter and the different tests by reading the instructions carefully.
The Free Chlorine, Combined Chlorine, and Total Chlorine reagents are compliant for meeting
USEPA (4500-Cl G); ISO 7393/2; and German DIN 38408 G4-2 requirements.
Observe the dip time (as required for the test) for accurate results.
7HVWLPPHGLDWHO\DIWHU¿OOLQJWKHCELL with water sample when testing for oxidizers such as
Chlorine and Bromine (Ozone can be measured in CL3 MENU).
Be sure the CELL LV¿OOHGWRFDSDFLW\POHVSHFLDOO\IRUS+DQG7RWDO$ONDOLQLW\
Rinse the CELL with clean water immediately after completing each test. Some reagents may
stain the CELL if not rinsed shortly after use. Other reagents including Cyanuric Acid, Chloride,
DQG&DOFLXP+DUGQHVVPD\FRDWWKH&(//ZDOO,WLVUHFRPPHQGHGDIWHUWKHVHWHVWVWRXVHWKH
Cell Cleaning Brush with water to clean the CELL.
Just before testing, rinse the sample CELL with the sample water several times to get a
representative sample. (Use deionized or distilled water for rinsing if you have a limited amount of
sample).
Store the meter and all test materials out of direct sunlight and away from chemical storage areas.
Minimize exposure of meter and test reagents to heat above 38°C (100°F).
Dry the outside of the meter when testing is complete or before storage of the meter.
When running a DPD-1 Free Chlorine test AFTER a Total Chlorine DPD-3, a Total Chlorine
'3'RUD+5&KORULQHWHVWULQVLQJLVYHU\LPSRUWDQWWRUHPRYHUHVLGXDO.,ZKLFKPD\LQWHUIHUH
Each eXact
®
Strip Micro is valid for ONLY one test. Discard strip after single use in regular refuse
that is inaccessible to children and pets.
Each bottle of eXact
®
Strip Micro contains the quantity of strips notated on the bottle. Due to the
VWULSVOLWWLQJSURFHVV\RXPD\¿QGRQHRUWZRVWULSVWKDWDUHQRWLFHDEO\VPDOOHURUODUJHULQZLGWK
than the normal strips in the bottle. These should be discarded. Using these strips may give
unreliable results.
Each table supplied has a unique revision number located in the bottom right corner of the table. It
LVUHFRPPHQGHGWKDW\RXYLVLWZZZVHQVDIHFRPDWOHDVWHYHU\PRQWKVWRFKHFNIRUDQ\XSGDWHG
revisions.
The eXact
®
Micro 7+ Meter is not compatible for use with DPD-1, DPD-3, and DPD-4 powder
pillows, tablets, and liquids available from other manufacturers. Accurate results can only be
guaranteed by using genuine eXact
®
Micro strips or reagents (reorder information on page 19).
Our lab testing with the Micro 7+ meter has shown that zeroing and measuring of the sample
normally does not require any cell cover for accurate results, except in sunlight. To obtain optimal
accuracy when testing with the meter outdoors (sunlight), use the Mixing Cap/Cell Cover when
zeroing and reading the sample.
Remove batteries when meter is not used for more than a month (Warranty Requirement).
,WLVUHFRPPHQGHGWKDW3RRODQG6SDVDPSOHVIRUR[LGL]HUVVXFKDV&KORULQHEHWDNHQLQFKHV
below the surface as follows: submerge meter with open cell facing down 18 inches, and then turn
PHWHUXSULJKWDWWKDWGHSWKWR¿OOWKHFHOO5HPRYHPHWHUIURPZDWHUZLWKWKHVDPSOHIRUWHVWLQJ
eXact
®Micro 7+ Tips For Best Accuracy

16
The following are some common messages that may be displayed, including error messages. If an
error message other than those listed below is displayed, please contact technical support in the USA
at (803) 329-0162 (ext. 0).
eXact
®Micro 7+ Meter Messages
LCD Message
HI
LO
LO
ER
Description
In READ mode: test sample concentration is above
WKHPHDVXUHPHQWUDQJHWHVWVSHFL¿F
In READ mode: test sample concentration is below
WKHPHDVXUHPHQWUDQJHWHVWVSHFL¿F
In ZERO mode: sample absorbance (due to a
cloudy or colored sample or a dirty cell) is too high
to zero, the meter will read “LO”.
Excessive stray light detected. Normally this
does not occur, even when testing in sunlight.
Low battery indication.
Corrective Action
Dilute and retest. Dilution Kit available
(Part Number 487200).
Sample value is below measurement
range.
'LOXWHVDPSOH¿OWHUVDPSOHRUFOHDQFHOO2QHRI
these options should remedy the problem.
Place the LIGHT BLOCKING CAP over the CELL
for zeroing and for reading result. Moving
WRDVKDGHGDUHDFDQDOVR¿[WKLVSUREOHP
Replace the batteries.
About The Built-In Cell
The built-in CELLLVWUDQVSDUHQWSODVWLFDQGZKHQ¿OOHGWRWKHWRSFRQWDLQVPO7KHVWXUG\CELL design
will last for over 20,000 readings. Scratches on the CELL will not interfere or compromise the accuracy of
WKHUHDGLQJVEHFDXVHRILWV¿[HGSRVLWLRQ)RUEHVWDFFXUDF\ULQVHFHOOZLWKFOHDQZDWHULPPHGLDWHO\DIWHU
a test is completed. Do not use solvents, such as acetone, to clean the cell. When the CELL becomes
stained or cloudy from repeated testing, or when the meter does not blank when you press the ZERO/ON
button, the cell needs to be cleaned. Clean as follows)LOOFHOOZLWKFOHDQZDWHUDQGPRYHWKHCell
cleaning brush up-and-down and back-and-forth along the walls of the cell. Afterwards, rinse the cell and
the meter is ready for use again. Cleaning the cell regularly is especially recommended after you run a
test that is using turbidity or precipitation chemistry for analysis (Calcium Hardness and Cyanuric Acid).
eXact®Photometer 2-Year Limited Warranty
Registration of your eXact®photometer must be received within 30 days from date of purchase to
activate the warranty. The eXact®photometer is warranted to be free from defects in materials and
workmanship for a period of two (2) years from the date of purchase by the customer. ITS will repair or
replace any part of the product which is deemed to be faulty or otherwise defective. The non-transferable
warranty does not cover product damage caused by abuse (such as crushing a tablet in the cell) or
improper use. If the meter is faulty or otherwise defective contact ITS by phone (+1-803-329-9712 Ext. 0)
returning the photometer to ITS. Damage caused by improper packing of the photometer for return
shipment to ITS will not be covered by the warranty. Customer is responsible for shipping charges to
ITS. ITS pays postage when photometer is returned to customer. A maximum processing fee of $75 will
be charged for repair or replacement of non-registered photometers and damages not covered by this
warranty. Registration is available over the phone (+1-803-329-9712 Ext. 0) or online at
KWWSZZZVHQVDIHFRPPLFURZDUUDQW\3HUVRQDOGDWDLVNHSWFRQ¿GHQWLDO
1. Unscrew the O-ring sealed battery cover counter-clockwise. Use proper sized pliers if necessary.
Do not disturb the sealing O-ring. Batteries are not included.
2. Remove the used batteries and install 4 new AAA batteries following the diagram for
correct polarity (see diagram). We recommend high quality AAA alkaline batteries be used.
4. Replace the battery cover. Be sure to tighten the cover securely. This is necessary for meter to
be waterproof.
5. Dispose of the used batteries in accordance with your
local regulations.
3UHVV=(5221EXWWRQWRFRQ¿UPWKHPHWHUWXUQVRQ
The meter is now ready for operation.
7. Meter will not work if battery orientation is incorrect.
To Install/Replace “AAA” Batteries:
TOP VIEW

17
eXact®Strip Micro 7+ Reagent Reorder Information
* Results utilize the Tr-7 (Transmission) meter function and require the use of a conversion table. See respective test procedures for more information and tables.
7RWDO&KORULQH'3'7HVWUHTXLUHV)UHH&KORULQH'3'WREHUXQ¿UVW
N/A
N/A
N/A
12 - 180 ppm
0 - 14 ppm
10 - 500 ppm
0 - 11 ppm
0 - 11 ppm
0 - 300 ppm
0 - 11 ppm
0 - 11 ppm
0 - 11 ppm
0.04 - 8 ppm
0 - 11 ppm
0.01 - 4.5 ppm
6.2 - 8.4 pH
0 - 5 ppm
0 - 350 ppm
0 - 12 ppm
0 - 1.5 ppm
0 - 95 ppm
0.1 - 1.1 ppm
0 - 2.5 ppm
0 - 3 ppm
0 - 33 ppm
0 - 1400 ppm
0 - 13 ppm
0.02 - 5 ppm
0.15 - 3.5 ppm
0 - 300 ppm
2 - 77 ppm
0.03 - 1.5 ppm
0.04 - 2.3 ppm
0 - 45 ppm
0 - 4.5 ppm
0 - 7.5 ppm
3 - 6.2 pH
8 - 9.6 pH
5.9 - 9 pH
0.03 - 4 ppm
0 - 150 ppm
0 - 13 ppm
3 - 580 NTU
PARAMETER
No. PART NO. # OF TESTS DETECTION RANGE CHEMISTRY
eXact
®
Strip Micro (4mL ) Reagent SpeciÀcations - For use with eXact
®
Micro 7+, Part no. 486691
eXact
®
Micro Carrying Case w/ foam
Dilution Kit
Reference Standard
Alkalinity, Total
Bromine (DPD-1)
Calcium (as CaCO
3
)
Chlorine, Free (DPD-1)
Chlorine, Free (DPD-1)
Chlorine, HR Free
Chlorine, Total (DPD-3)**
Chlorine, Total (DPD-4)
Chlorine, Total (DPD-4)
Copper (Cu
+2
)
Ozone (DPD-4)
Permanganate (DPD-1)
pH
Ammonia (as NH
3
/NH
4+
)*
Chloride (as NaCl)*
Chlorine Dioxide (DPD-1)*
Chromium (VI)*
Cyanuric Acid*
Fluoride (as F
-
)*
Hydrazine*
Hydrogen Peroxide LR*
Hydrogen Peroxide MR*
Hydrogen Peroxide HR (DPD-4)*
Iodine (DPD-1)*
Total Iron, TPTZ (Fe
+2
/Fe
+3
)*
Total Iron, Ferro (Fe
+2
/Fe
+3
)*
Total Hardness (as CaCO
3
)*
LR Total Hardness (as CaCO3)*
Manganese (as Mn
+2
)*
Molybdate (as MoO
4
)*
Nitrate (as NO
3
-
)*
Nitrite (as NO
2
-2
)*
Peracetic Acid (PAA) (DPD-4)*
pH, Acid*
pH, Alkali*
pH, BT*
Phosphate*
Sulfate (as SO
4
-2
)*
6XO¿GHDV+
2
S)*
Turbidity* (as transmission)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
486001
487200
486602
486641
486636
486629
486637
484051
486672
486638
486670
484054
486632
486634
486676
486639
483343-MK
481657-II
486633
486614
481652-II
486643
486649
486616
486648
486670
486637
486650
481623
486673
486630
486606
486653
486655
486623
486674
486624
486609
486652
486814
486608
486646
None
N/A
N/A
15
100
100
50
100
100 Foils
50
100
100
100 Foils
50
100
100
100
25
25
100
50
60
25
50
50
50
100
100
50
50
50
100
24
50
50
50
100
50
50
100
50
50
50
No Reagent
N/A
N/A
N/A
Alizarin Red S + Citrate
DPD
Oxalic Acid
DPD
DPD
KI + Buffer
Kl
DPD + Kl
DPD + KI
Biquinoline
DPD + Kl
DPD
Phenol Red
Salicylate Method
Silver (ppt)
DPD
Diphenylcarbazide
Melamine (ppt)
SPADNS
4-Dimethylaminocinnamaldehyde
DPD + PO
4
+ MoO
4
+ KI
DPD + MoO
4
+ KI + acid
DPD + KI
DPD
TPTZ + PP
1,10 Phenanthroline
Phthalein Purple
Phthalein Purple
PAN + Cyanide
Alizarin Red S + Buffer
Zinc Reduction
Chromotropic Acid
DPD + KI
Alizarin Red S
Thymol Blue
Bromothymol Blue and Thymol Blue
Molybdate Method
Barium (ppt)
NPS
Turbidity values as NTU
If your required procedure is not listed in this manual, please see the back page for our contact information.
NOTE: Because most of our products are test strips or use reagents that have little or no hazard in the quantity sold, MSDS sheets are not
supplied with the test. The exceptions are the Manganese (486606) test, which comes with 2 strips and one liquid reagent (PAN); Fluoride
(486643) test, which is a liquid reagent (SPADNS), and Iron (481623) test, which is a powder reagent. Hydrazine (486649) uses a liquid and strip.
To ensure optimal performance, store your eXact®kit in a cool, dry place away from excess heat (below
100°F / 38°C), moisture, and oxidizers such as Chlorine and Bromine.

18
For US Inquiries and Re-Orders:
Industrial Test Systems, Inc.
1875 Langston Street,
Rock Hill, SC 29730 USA
Phone: 1-800-861-9712 - INSIDE THE U.S.
1-803-329-9712 - OUTSIDE THE U.S.
Fax: 1-803-329-9743
WWW.SENSAFE.COM
www.poolcheckonline.com
For European & Middle East Inquiries and Re-Orders:
ITS Europe, LTD
The UK Centre for Homeland Security
Building 7, Chilmark
Salisbury, Wiltshire SP3 5DU, United Kingdom
Tel: +44 (0)1722 717911 Fax: +44 (0) 1722 717941
WWW.ITSEUROPE.CO.UK
www.poolcheckonline.com
Free Chlorine results are compared using the eXact®Strip Micro CL (DPD-1) with the eXact®Micro 7+ Meter in
Menu CL1 and Hach®DR890 Colorimeter in Program 9 and Program 12 using Hach®powder pillows.
(486691-K)
Standard Kit
Includes:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
(486691-KP)
Pool/Spa Kit
Includes:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Reagent Cyanuric Acid (481652-II)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
(486691-WD)
Well Driller Kit
Includes:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Micro TPTZ Total Iron (486650)
eXact
®
Micro Manganese (486606)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
The eXact®Micro 7+ Line of Kits
Contact Information
eXact®Strip Micro DPD-1 Accuracy
y = 1.0402x
R
2
= 0.9989
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0
Hach Free Chlorine, ppm
ITS Free Chlorine, ppm
Hach vs ITS
Linear (Hach vs ITS)
DPD Free Chlorine study January 21, 2011
DR890
Micro 7+
0.00 0.00
0.41 0.39
0.92 0.96
0.79 0.73
1.28 1.32
2.70 2.73
3.20 3.22
4.45 4.75
5.30 5.53
Meter Menu Range (PPM) Resolution
Micro 7+ CL1 0.00 to 5.99 0.01
6.0 to 11.0 0.1
DR890
Program 9
0.00 to 2.20 0.01
Program 12
0.0 to 11.0 0.1
Hach
®
is a registered trademark of Danaher Corporation

18
(486691-K)
Standard Kit
Inclut:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
(486691-KP)
Pool/Spa Kit
Inclut:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Reagent Cyanuric Acid (481652-II)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
(486691-WD)
Well Driller Kit
Inclut:
1 eXact
®
Micro 7+ Meter (486691)
eXact
®
Strip Micro DPD-1 (486637-25)
eXact
®
Strip Micro DPD-3 (486638-25)
eXact
®
Strip Micro pH (486639-25)
eXact
®
Strip Micro Total Alkalinity (486641-25)
eXact
®
Strip Micro Calcium Hardness
(486629-25)
eXact
®
Strip Micro Copper (486632-25)
eXact
®
Micro TPTZ Total Iron (486650)
eXact
®
Micro Manganese (486606)
eXact
®
Strip Micro High Range Chlorine
(486672)
1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
eXact®Micro 7+ Kits
y = 1.0402x
R
2
= 0.9989
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0
Hach Chlore Libre, ppm
ITS Chlore Libre, ppm
Hach vs ITS
Linéaire (Hach vs ITS)
Étude DPD Chlore Libre Janvier 21, 2011
DR890
Micro 7+
0.00 0.00
0.41 0.39
0.92 0.96
0.79 0.73
1.28 1.32
2.70 2.73
3.20 3.22
4.45 4.75
5.30 5.53
Meter Menu Range (PPM) Resolution
Micro 7+ CL1 0.00 to 5.99 0.01
6.0 to 11.0 0.1
DR890
Program 9
0.00 to 2.20 0.01
Program 12
0.0 to 11.0 0.1
Hach
®
est une marque déposée de Danaher Corporation
Pour commander ou demander des
informations aux Etats-Unis:
Industrial Test Systems, Inc.
1875 Langston Street,
Rock Hill, SC 29730 USA
Téléphone: États-Unis : 1-800-861-9712
Extérieur des États-Unis : 1-803-329-9712
Télécopieur : 1-803-329-9743
WWW.SENSAFE.COM
www.poolcheckonline.com
Pour commander ou demander des
informations en Europe et au Moyen-Orient:
ITS Europe, LTD
UK Centre for Homeland Security
Building 7, Chilmark
Salisbury, Wiltshire SP3 5DU, United Kingdom
Téléphone: -44 (0) 1722 717911
Télécopieur: +44 (0) 1722 717941
WWW.ITSEUROPE.CO.UK
www.poolcheckonline.com
Précision des bandes eXact
®
Strip Micro DPD-1
Les résultats de chlore libre sont comparés en utilisant eXact®Strip Micro CL (DPD-1) avec le photomètre
eXact®Micro 7+ dans le menu CL1 et le colorimètre DR890 et les sachets de Hach®, programmes 9 et 12.
Pour nous joindre :

17
Informations concernant vos nouvelles commandes de bandes de réactifs eXact®Strip Micro 7+
*Les résultats s’appuient sur la fonction Tr-7 (Transmission) du photomètre et demandent l’utilisation d’une table de conversion.
Voir les procédures d’analyses respectives pour les tables de conversion et pour plus d’information.
**L’analyse de chlore total DPD-3 exige qu’une analyse de chlore libre DPD-1 (486637) soit d’abord exécutée.
N/A
N/A
N/A
12 - 180 ppm
0 - 14 ppm
10 - 500 ppm
0 - 11 ppm
0 - 11 ppm
0 - 300 ppm
0 - 11 ppm
0 - 11 ppm
0 - 11 ppm
0.04 - 8 ppm
0 - 11 ppm
0.01 - 4.5 ppm
6.2 - 8.4 pH
0 - 5 ppm
0 - 350 ppm
0.03 - 6.5 ppm
0 - 1.5 ppm
0 - 95 ppm
0.1 - 1.1 ppm
0 - 2.5 ppm
0 - 3 ppm
0 - 33 ppm
0 - 1400 ppm
0 - 11.9 ppm
0.02 - 5 ppm
0.15 - 3.5 ppm
0 - 300 ppm
2 - 77 ppm
0.03 - 1.5 ppm
0.04 - 2.3 ppm
0 - 45 ppm
0 - 4.5 ppm
0 - 4 ppm
3 - 6.2 pH
8 - 9.6 pH
5.9 - 9 pH
0.03 - 4 ppm
0 - 150 ppm
0 - 13 ppm
3 - 580 NTU
Paramètres
Nu. Pièce
Quantité d’analyses
Portée Produit de traitement
Caractéristiques concernant les réactifs eXact
®
Strip Micro (4 ml) – Utiliser exclusivement avec eXact
®
Micro 7+, pièce numéro 486691
Le boitier de transport avec mousse eXact
®
Micro
Kit de Dilution
Référence Standard
Alcalinité Totale
Brome (DPD-1)
Calcium (CaCO
3
)
Chlore Libre (DPD-1)
Chlore Libre (DPD-1)
Chlore Libre, HR
Chlore Total (DPD-3)**
Chlore Total (DPD-4)
Chlore Total (DPD-4)
Cuivre (Cu
+2
)
Ozone (DPD-4)
Permanganate (DPD-1)
pH
Ammoniac (NH
3
/NH
4+
)*
Chlorure (sous forme NaCI)*
Dioxyde de Chlore (DPD-1)*
Chrome (VI)*
Acide Cyanurique*
Fluorure*
Hydrazine*
Peroxyde d'Hydrogène LR*
Peroxyde d'Hydrogène MR*
Peroxyde d'Hydrogène HR*
Iode (DPD-1)*
Fer Total , TPTZ (Fe
+2
/Fe
+3
)*
Fer Total, Ferro (Fe
+2
/Fe
+3
)*
Dureté Totale (CaCO
3
)*
Dureté Totale LR (CaCO3)*
Manganèse*
Molybdate (MoO
4
)
Nitrate (NO
3
)*
Nitrite (as NO
2
)*
Acide Peracétique (PAA) (DPD-4)
pH, Acide*
pH, Alcalin*
pH, BT*
Phosphate*
Sulfate*
Sulfide (sous forme H2S)*
Turbidité* (TR)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
486001
487200
486602
486641
486637
486629
486637
484051
486672
486638
486670
484054
486632
486670
486637
486639
483343-MK
481657-II
486637
486614
481652-II
486643
486649
486616
486648
486670
486637
486650
481623
486673
486630
486606
486653
486655
486623
486670
486624
486609
486652
486814
486608
486646
Sans #
N/A
N/A
15
100
100
50
100
100 Foils
50
100
100
100 Foils
50
100
100
100
25
25
100
50
60
25
50
50
50
100
100
50
50
50
100
24
50
50
50
100
50
50
100
50
50
50
Aucun réactif
N/A
N/A
N/A
Rouge alizarine S et citrate
DPD
Acide oxalique
DPD
DPD
Kl + Buffer
KI
DPD + Kl
DPD + KI
Biquinoline
DPD + Kl
DPD
Rouge de phénol
Salicylate
Argent
DPD
Diphénylcarbazide
Mélamine
SPADNS
4-Dimethylaminocinnamaldehyde
DPD + PO
4
+ MO
4
+ KI
DPD + MO
4
+ KI + acide
DPD + KI
DPD
TPTZ + PP
1,10 Phenanthroline
Phtaléine pourpre
Phtaléine pourpre
PAN + cyanure
Rouge alizarine S et Buffer
Zinc réducteur
Acide chromotropique
DPD + KI
Rouge alizarine S
Thymosulfophtaléine
bleu de bromothymol et bleu de thymol
Méthode de molybdate
Baryum
NPS
Turbidité (valeurs en NTU)
Si vous désirez obtenir des informations sur les procédures et des instructions qui ne sont pas dans ce manuel,
référez vous à la liste de contacts et d'informations au verso.
NOTE : La plupart de nos produits sont des bandes d’analyses et des réactifs contenant peu ou pas de risque, vu la quantité
vendue. Les fiches techniques santé-sécurité (pour matières dangereuses) ne sont donc pas fournies avec. Les tests de manganèse
(486606), viennent avec 2 bandes et un réactif liquide (NPA) et le fluorure (486643) qui utilise un réactif liquide (SPADNS) et le test
du fer (481623) qui utilise un réactif en poudre et l'Hydrazine (486649) qui utilise une languette et un liquide, sont des exceptions.
Pour maximiser la performance, rangez votre trousse eXact® dans un endroit frais, sec, loin de toute
chaleur excessive (sous 100°F / 38°C), de moisissure ou d’oxydants, tels que le chlore et le brome.
Other manuals for Micro 7+
1
Table of contents
Other eXact Measuring Instrument manuals

eXact
eXact Micro 10 User manual

eXact
eXact Smart Photometer System iDip User manual

eXact
eXact Chlorine Plus User manual

eXact
eXact eXact Micro 20 User manual

eXact
eXact 486201 User manual

eXact
eXact Well EZ User manual

eXact
eXact idip User manual

eXact
eXact pH+ User manual

eXact
eXact Eco-Check User manual

eXact
eXact Micro 7+ User manual