
5
Holding the positioning rod firmly at its correct depth and orientation, slide the slide-hammer
back up the 1/4” rod, then side it quickly back to the anvil striking it a sharp firm blow. This will
shear the rivet holding the wedge to the platen and will pull the wedge into the platen thereby
expanding it against the wall of the borehole.
After the first blow, take another reading on the readout box and observe the change in reading.
The recommended preloads are as follows: For the EX size a reading change of 2000 digits on
channel F, for the BX size a reading change of 400 digits on channel B, for the NX size a reading
change of 200 digits on channel B.
Use as many blows of the hammer as is necessary to achieve this reading. Stop hammering
when successive blows produce little or no change of reading. Continued hammering can
break the wedge! When the target reading has been achieved, or if successive hammer blows
produce little or no change, disconnect the 1/4” rod from the wedge yoke by turning clockwise.
Remove the 1/4” rod from the hole, and then disengage the setting tool from the stressmeter by
pulling on it.
For multiple installations of gauges in a single hole, route the lead wires from deeper gauges
through the recess in the side of the setting tool head. Maintain tension on these wires as
subsequent gauges are pushed into the hole.
If necessary, after setting the gauges and obtaining the final readings, push the leads back into
the borehole and seal the borehole using an expandable rockbolt anchor or a short bolt. This will
discourage vandalism if this is a problem.
Note that the stressmeter initial readings will probably diminish slightly over the first day or two
as the instrument beds firmly into place.
2.5 Recovering the Stressmeter
After tests, the stressmeter can be removed from the borehole by using the setting tool.
Only the larger setting rods are required, along with the setting tool head, which is used to strike
the outer tip of the wedge. This will drive the wedge out from under the platen and allow the
stressmeter to be pulled from the hole using the electrical cable. Make sure that the setting head
is oriented so that the flat part of the front face lies opposite the wedge. The entire stressmeter
can sometimes be recovered in this way i.e., the wedge, platen, and stressmeter body.
In order to reuse the stressmeter it will require a new nylon screw. (A few spare nylon screws are
included in each shipment). However, there is a good chance that the wedge and platen may
dislodge in the borehole and be lost; therefore, it is advised to order spares of these also.
(Note: The BX size stressmeter and the NX size stress meter both use the same wedge. However,
there are two holes in the wedge. The one nearest the tip is for the BX size; the one farthest from
the tip is for the NX size).