
Low-Speed, Clinical Centrifuge
Quick Guide
Order & Inquiry : +82-2-3452-7736ㅣinfo@gyrozen.comㅣtechsupport@gyrozen.comㅣwww.gyrozen.com
Download corresponding user manual from www.gyrozen.com
Precautions and Maintenance Tips
EQG-1(Rev.1)
Interchangeable rotors
Clean rotor pivots and bucket grooves. Lubri-
cate them by coating the grease sometimes to
swing the buckets enough.
Lubricate the contact area
Remove a motor fixing cushion (polyethylene
(PE) foam) from the chamber of centrifuge be-
fore installation. The cushion is to prevent pos-
sible damages caused by motor shaking during
shipping or relocation.
Motor holding cushion
Clean the motor shaft. Remove the foreign
substances not to cause any vibration.
Motor shaft & rotor
Load the sample tubes
symmetrically.
Balance buckets and
samples
Check the rotor is installed correctly and is sit-
ting rmly on the motor shaft.
Rotor checking
Secure a safety zone or 30 cm
around centrifuge
Position the centrifuge with enough
space (30 cm) on each side of centrifuge
to ensure proper air circulation. Neither
hazardous materials nor persons should
be permitted within the area during op-
eration.
After washing or autoclaving, a cushion rub-
ber at the bottom of rotor sleeve (GLB-15A-
FA) may fall out of the sleeve and it can cause
imbalance. Also, the rubber has a correct di-
rection. The upside-down rubber cause poorly
tted tube from the sleeve. Check the rubber
in the sleeve and the direction of the rubber.
Cushion rubber of rotor sleeve
GRS-S-100-4 for
GLB-b100-100 / GLB-b50-100 /
GLB-10-8-100 / GLB-bd15-100 /
GLB-bdc15-100
GRS-S-50-4 for GLB-50-50 /
GLB-c50-50 / GLB-15-50/
GLB-c15-50 / GLB-d15-50 /
GLB-dc15-50
GRS-S-15-6 for GLB-15-15
GRA-S-15-16A incl.
GLB-15A-FA
GRS-S-mw-2
30 cm
30 cm
30 cm
Upside-down
Normal
Turn the power switch o after
using the centrifuge. Unplug
power cord before cleaning or
whenever the centrifuge is to
remain unused for a long period
of time.
Shut down power after
operation