Hameg HM 2005 User manual

Oscilloscope
HM 2005
ENGLISH
MANUAL • HANDBUCH • MANUEL
AUTOSET
CT
A/ALT. DEL.TRIG
TR
B
RM
TRS
READ
OUT
PUSH
BOTH
PUSH
LONG
INV.
VAR.
B
ON
OFF
∆V
∆t1
∆t
1kHz
1MHz
INPUT CH I (HOR. INP.(X))
AC/DC
CAL.
0.2 Vpp
CURSOR
CH I/II
I/II
MENU max.
250 Vp
1MΩII
15pF
INPUT CH II
max.
250 Vp
1MΩII
15pF
TRIG. EXT. INP
max.
100 Vp
(Z)
x1/x10 INV.
x1/x10
200 MHz
ANALOG OSCILLOSCOPE
HM2005
Instruments
TRK
!
CAT
I
!!!
GD AC/DC GD
CH I CH II
DEL.POS.
ADD
CHP. VAR.VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
POWER INTENS
EXIT
SAVE
SET
1
FOCUS
9
RECALL
A
RO
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
Tsm005:BD 5RSsn001:t
CH1:=100mV CH2 :~100mV CHP

Subject to change without notice
2
MANUAL•HANDBUCH•MANUEL

3
Subject to change without notice
St.300802-Hüb/tke/hz
General information regarding the CE marking ............. 4
Specifications HM2005 ..................................................... 5
Important hints .................................................................. 6
Useof tilt handle ............................................................. 6
Safety ............................................................................. 6
Intended purpose and operating conditions .................. 6
EMC ............................................................................... 7
Warranty ......................................................................... 7
Maintenance .................................................................. 7
Protective Switch-Off ..................................................... 7
Power supply .................................................................. 7
Basics of signal voltage .................................................... 8
Type of signal voltage .................................................... 8
Amplitude Measurements ............................................. 8
Total value of input voltage ............................................ 9
Time Measurements...................................................... 9
Connection of Test Signal ............................................... 10
Controls and Readout ....................................................... 11
Menu ................................................................................... 21
FirstTime Operation .......................................................... 22
Trace Rotation TR ........................................................... 22
Probe compensation and use ........................................ 22
Adjustment at 1kHz ....................................................... 22
Adjustment at 1MHz ...................................................... 23
Operating modes of the
vertical amplifiers in Yt mode ......................................... 23
X-Y Operation ................................................................. 24
Phase comparison with Lissajous figures ..................... 24
Phase difference measurement
in DUAL mode (Yt) ......................................................... 24
Measurement of amplitude modulation ........................ 25
Triggering and Time Base .................................................. 25
Automatic Peak (value) -Triggering ................................. 26
Normal Triggering ........................................................... 26
- Slope ...................................................................... 26
Trigger coupling .............................................................. 26
Triggering of video signals ............................................. 27
Line triggering (~) ........................................................... 27
Alternate triggering ........................................................ 27
External triggering .......................................................... 28
Trigger indicator “TR” .................................................... 28
HOLD OFF-time adjustment .......................................... 28
B-Time Base (2nd Time Base)/
Triggering after Delay ..................................................... 29
AUTOSET ............................................................................ 29
Main Value Display ............................................................ 30
ComponentTester .............................................................. 30
General ........................................................................... 30
Using the Component Tester ......................................... 30
Test Pattern Displays ...................................................... 30
Testing Resistors ............................................................ 30
Testing Capacitors and Inductors .................................. 30
Testing Semiconductors ................................................ 31
Testing Diodes ............................................................... 31
Testing Transistors .......................................................... 31
In-Circuit Tests ................................................................ 31
Adjustments ....................................................................... 32
RS-232 Interface ................................................................. 32
Operation ....................................................................... 32
Baud-Rate Setting .......................................................... 32
Front panel HM2005 .......................................................... 34
Oscilloscope
HM 2005
Table of contents

Subject to change without notice
4
General information regarding the CE marking
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product
standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for residential, commercial
and light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the acceptance limits.
For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding
emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened cables must be
used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3 meters and
not be used outside buildings. If an interface has several connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG are
suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for
a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened
cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an influence of such
signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments
specifications may result from such conditions in individual cases.
4. RF immunity of oscilloscopes.
4.1 Electromagnetic RF field
The influence of electric and magnetic RF fields may become visible (e.g. RF superimposed), if the field intensity is high. In most cases the coupling
into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/or radiation. The device under test as well
as the oscilloscope may be effected by such fields.
Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each amplifier
stage is higher than the total –3dB bandwidth of the oscilloscope, the influence RF fields of even higher frequencies may be noticeable.
4.2 Electrical fast transients / electrostatic discharge
Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via test leads and/or control
cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals may effect the trigger unit and/or may become
visible on the CRT, which is unavoidable. These effects can also be caused by direct or indirect electrostatic discharge.
KONFORMITÄTSERKLÄRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
Herstellers HAMEG GmbH
Manufacturer Industriestraße 6
Fabricant D-63533 Mainhausen
Die HAMEG GmbH bescheinigt die Konformität für das Produkt
The HAMEG GmbH herewith declares conformity of the product
HAMEG GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Oszilloskop/Oscilloscope/Oscilloscope
Typ / Type / Type: HM2005
mit / with / avec: —
Optionen / Options / Options: —
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
15.01.2001
Angewendete harmonisierte Normen / Harmonized standards applied / Normes
harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994
EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility /
Compatibilité électromagnétique
EN 61326-1/A1
Störaussendung / Radiation / Emission:Tabelle / table / tableau 4; Klasse / Class /
Classe B.
Störfestigkeit / Immunity / Imunitee:Tabelle / table / tableau A1.
EN 61000-3-2/A14
Oberschwingungsströme / Harmonic current emissions / Émissions de courant
harmonique: Klasse / Class / Classe D.
EN 61000-3-3
Spannungsschwankungen u. Flicker /Voltage fluctuations and flicker /
Fluctuations de tension et du flicker.
Datum /Date /Date Unterschrift / Signature /Signatur
E. Baumgartner
Technical Manager /Directeur Technique
General Information

5
Subject to change without notice
Specifications
Vertical Deflection
Operating modes: Channel I or II separate,
Channel I and II: alternate or chopped
Chopper Frequency: approx. 0.5MHz
Sum or difference: from CH I and CH II
Invert: both channels
XY-Mode: via channel I (X) and channel II (Y)
Frequency range: 2x DC to 200MHz (-3dB)
incl. Bandwidth Limiter:
2x DC to approx. 50MHz (-3dB)
Min Rise Time: <1.75ns
Overshoot: ≤1%
Deflection coefficients: 12 calibrated steps
1mV to 2mV/div.: ±5% (DC – 10MHz (-3dB))
5mV/div. to 5V/div.: ±3% in 1-2-5 sequence
with variable 2.5:1 up to 12,5V/div.
Input impedance: 1MΩII 15pF
Input coupling: DC-AC-GD (ground)
Input voltage: max. 250V (DC + peak AC)
Delay line: approx. 70ns
Tr i g ge r i n g
Automatic (pp): ≤≤
≤≤
≤20Hz-300MHz (≥0.5div.)
Normal with lev. contr.: DC-300MHz (≥0.5div.)
Indicator for trigger action: LED
Slope: positive or negative
Sources: Channel I or II, line and external
ALT. Triggering: CH I/CH II (≥0.8div.)
Coupling: AC (10Hz – 300MHz), DC (0 – 300MHz),
HF (50kHz – 300MHz), LF (0 –1.5kHz),
NR (noise reject): 0-50MHz (≥0.8div.)
Triggering time base B: normal with level control
and slope selection (0 – 300 MHz)
Active TV Sync. Separator: field and line, + / –
External: ≥0.3Vpp (0 – 200MHz)
Horizontal Deflection
Time base A: 23 calibrated steps (±3%)
from 0.5s/div. – 20ns/div. in 1-2-5 sequence
variable 2.5:1 up to 1.25s/div.
X-Mag. x10: 2ns/div. (±5%)
Holdoff time: variable to approx. 10:1
Time base B: 19 calibrated steps (±3%)
from 20ms/div. to 20ns/div. in 1-2-5 sequence
Operating modes: A or B, alternate A/B
Bandwidth X-amplifier: 0 – 5MHz (-3dB)
Input X-amplifier: via Channel I
Sensitivity: see CH I
X-Y phase shift: <3° below 220kHz
Operation / Control
Manual (front panel switches)
Auto Set (automatic parameter selection)
Save/Recall: 9 user-defined parameter settings
Readout: Display of parameter settings
Cursor measurement: ∆V, ∆t or ∆1/t (frequ.)
Remote control: with built in RS-232 interface
ComponentTester
Test voltage: approx. 7Vrms (o/c) approx. 50Hz
Test current: approx. 7mArms (s/c) approx. 50Hz
One test lead is grounded (Safety Earth)
General Information
CRT: D14-375GH, 8x10cm, internal graticule
Acceleration voltage: approx. 14kV
Trace rotation: adjustable on front panel
Calibrator: 0,2V ±1%, ≈ 1kHz/1MHz (tr<4ns)
Z-Input (Intensity modulation):max. +5V (TTL)
Line Voltage / Power consumption: 100-240V
AC ±10% / approx.43Watt. 50/60Hz
Min./Max. ambient temperature: +10°C ...+ 40°C
Protective system: Safety class I (IEC1010-1)
Weight: approx. 5.9kg.
Size (W x H x D): 285 x 125 x 380 mm
Color: techno-brown
HM 2005 –
Analog Oscilloscope (200 MHz)
Autoset, Save/Recall, Readout/Cursor and RS-232 Interface
2 x DC to 100 MHz, 2 x 1mV/div to 5V/div
Time Base A: 0.5s/div to 2ns/div,
Time Base B: 20ms/div to 2ns/div
Triggering DC to 300 MHz, Peak Triggering; Component Tester
1 kHz/1 MHz Calibrator, TV Sync Separator,
Built in Adjustment Menu
This microprocessor controlled oscilloscope have been designed for a wide
multitude of applications in service and industry. For ease of operation, the
AUTOSET function allows for signal related automatic setup of measuring
parameters. On screen alphanumeric readout and cursor functions for voltage,
time and frequency measurement provide extraordinary operational convenience.
Nine different user defined instrument settings can be saved and recalled without
restriction.
The built in RS-232 serial interface permits remote controlled operation by a
PC. The outstanding feature of the HM 2005 include a second time base with the
ability to magnify, over 1000 times, extremely small portions of the input signal.
The second time base has its own triggering controls, to allow a stable and
precisely referenced display of asynchronous or jittery signal segments. The
trigger circuit is designed to provide reliable triggering to over 300 MHz at signal
levels as low as 0.5 div. An active TV Sync Separator for TV signal tracing
ensures accurate triggering even with noisy signals.
The HM2005have a built in switchable calibrator, which checks the instrument’s
transient response characteristics. The essential high frequency compensation of
wide band probes can be performed with this calibrator. The built in adjustment
menu allows closed case adjustment procedures. The instruments offer the right
combination of triggering control, frequency response, and time base versatility to
facilitate measurements in a wide range of applications - in laboratory as well as in
field service use. It is another example of HAMEGs dedication to engineering
excellence.
Accessories supplied: Operators Manual and PC-Software on CD-ROM, 2
Probes 1:1/10:1 and Line Cord.
HM 2005 · Specifications

Subject to change without notice
6
20°
10°
A
E
D
C
B
Important hints
This oscilloscope is easy to operate. The logical arrangement of
the controls allows anyone to quickly become familiar with the
operation of the instrument, however, experienced users are also
advised to read through these instructions so that all functions
are understood.
Immediately after unpacking, the instrument should be checked
for mechanical damage and loose parts in the interior. If there is
transport damage, the supplier must be informed immediately.
The instrument must then not be put into operation.
Symbols
ATTENTION - refer to manual
Danger - High voltage
Protective ground (earth) terminal
Use of tilt handle
To view the screen from the best angle, there are three different
positions (C, D, E) for setting up the instrument. If the instrument
is set down on the floor after being carried, the handle
automatically remains in the upright carrying position (A). In order
to place the instrument onto a horizontal surface, the handle
should be turned to the upper side of the oscilloscope (C). For
the D position (10° inclination), the handle should be turned to
the opposite direction of the carrying position until it locks in
place automatically underneath the instrument.
For the E position (20° inclination), the handle should be pulled
to release it from the D position and swing backwards until it
locks once more. The handle may also be set to a position for
horizontal carrying by turning it to the upper side to lock in the B
position. At the same time, the instrument must be lifted, because
otherwise the handle will jump back.
Safety
This instrument has been designed and tested in accordance
with IEC Publication 1010-1, Safety requirements for electrical
equipment for measurement, control, and laboratory use. The
CENELEC regulations EN 61010-1 correspond to this standard. It
hasleft the factory in a safe condition. This instruction manual
contains important information and warnings which have to be
followed by the user to ensure safe operation and to retain the
oscilloscope in a safe condition.
The case, chassis and all measuring terminals are connected to
the protective earth contact of the appliance inlet. The instrument
operates according to Safety Class I (three conductor power cord
with protective earthing conductor and a plug with earthing
contact). The accessible metal parts (case, sockets, jacks) and
the mains/line supply contacts (live, neutral) have been tested
against insulation breakdown with 2200V DC. The mains/line plug
shall only be inserted in a socket outlet provided with a protective
earth contact. The protective action must not be negated by the
use of an extension cord without a protective conductor.
Most cathode-ray tubes develop X-rays. However, the dose
equivalent rate falls far below the maximum permissible value of
36pA/kg (0.5mR/h).
Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured against any
unintended operation. The protection is likely to be impaired if, for
example, the instrument
shows visible damage,
fails to perform the intended measurements,
has been subjected to prolonged storage under unfavorable
conditions (e.g. in the open or in moist environments),
has been subject to severe transport stress (e.g. in poor
packaging).
Intended purpose and operating conditions
Attention!
This instrument must be used only by qualified experts who
are aware of the risks of electrical measurement.
Due to safety reasons the instrument must only be connected to
a properly installed power outlet, containing a protective earth
conductor. The protective earth connection must not be broken.
The mains/line plug must be inserted before connections are
made to measuring circuits.
CAT I
This instrument is intended for measurement of circuits either
not at all or not directly connected to mains/line (CAT I). Direct
measuring (without galvanic separation) within measuring
category II, III and IV circuits is inadmissible.
Circuits of a device under test are not connected directly with
mains/line if the device under test is operated via an isolation
transformer (Safety Class II). It is also possible to use suitable
transducers (e.g. Current Probe) that meet the demands of
protective class II, to measure indirect at mains/line. The
transducer’s measuring category must be observed.
Measuring Categories
The measuring category relates to transients on mains/line.
Transients are short and fast voltage and current changes that
may appear periodically or aperiodically. The amplitude of
transients increases with decreasing distance to the low voltage
installation source.
Category IV: has the highest demands on isolation etc. and is
required for measurement at the low voltage installation source
(e.g. at the supply meter).
Category III: is for measurement within the building installation
at e.g. terminal block, power switch, fixed motor, wall outlet etc.
Category II: measurement at circuits connected to the low
voltage installation such as home appliances, portable tools etc.
Important hints

7
Subject to change without notice
Field of application
The instrument is specified for operation in industry, light industry,
commercial and residential environments.
The instrument has been designed for indoor use. The permissible
ambient temperature range during operation is +10°C (+50°F) ...
+40°C (+104°F). It may occasionally be subjected to temperatures
between +10°C (+50°F) and -10°C (+14°F) without degrading its
safety. The permissible ambient temperature range for storage or
transportation is -40°C (-40°F) ... +70°C (+158°F). The maximum
operating altitude is up to 2200m (non-operating 15000m). The
maximum relative humidity is up to 80%.
If condensed water exists in the instrument it should be
acclimatized before switching on. In some cases (e.g. extremely
cold oscilloscope) two hours should be allowed before the
instrument is put into operation. The instrument should be kept
in a clean and dry room and must not be operated in explosive,
corrosive, dusty, or moist environments. The oscilloscope can be
operated in any position, but the convection cooling must not be
impaired. The ventilation holes may not be covered. For continuous
operation the instrument should be used in the horizontal position,
preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if the instrument
has warmed up for 30minutes at an ambient temperature between
+15°C (+59°F) and +30°C (+86°F). Values without tolerances are
typical for an average instrument.
EMC
This instrument conforms to the European standards regarding
the electromagnetic compatibility. The applied standards are:
Generic immunity standard EN50082-2:1995 (for industrial
environment) Generic emission standard EN50081-1:1992 ( for
residential, commercial and light industry environment).
This means that the instrument has been tested to the highest
standards. Please note that under the influence of strong
electromagnetic fields, such signals may be superimposed
on the measured signals.
Under certain conditions this is
unavoidable due to the instrument’s high input sensitivity, high
input impedance and bandwidth. Shielded measuring cables,
shielding and earthing of the device under test may reduce or
eliminate those effects.
Warranty
HAMEG warrants to its Customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years. This warranty shall
not apply to any defect, failure or damage caused by improper
use or inadequate maintenance and care. HAMEG shall not be
obliged to provide service under this warranty to repair damage
resulting from attempts by personnel other than HAMEG
representatives to install, repair, service or modify these products.
In order to obtain service under this warranty, Customers must
contact and notify the distributor who has sold the product. Each
instrument is subjected to a quality test with 10 hour burn-in
before leaving the production. Practically all early failures are
detected by this method. In the case of shipments by post, rail
or carrier the original packing must be used. Transport damages
and damage due to gross negligence are not covered by the
guarantee.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone number
(dialing code and telephone or direct number or department
designation) is stated for possible queries, this helps towards
speeding up the processing of guarantee claims.
Maintenance
Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it largely
certain that all signals are displayed with the accuracy on which
the technical data are based. Purchase of the HAMEG scope
tester HZ 60, which despite its low price is highly suitable for
tasks of this type, is very much recommended. The exterior of
the oscilloscope should be cleaned regularly with a dusting brush.
Dirt which is difficult to remove on the casing and handle, the
plastic and aluminum parts, can be removed with a moistened
cloth (99% water +1% mild detergent). Spirit or washing benzene
(petroleum ether) can be used to remove greasy dirt. The screen
may be cleaned with water or washing benzene (but not with
spirit (alcohol) or solvents), it must then be wiped with a dry
clean lint-free cloth. Under no circumstances may the cleaning
fluid get into the instrument. The use of other cleaning agents
can attack the plastic and paint surfaces.
Protective Switch Off
This instrument is equipped with a switch mode power supply. It
has both over voltage and overload protection, which will cause
the switch mode supply to limit power consumption to a
minimum. In this case a ticking noise may be heard.
Power supply
The instrument operates on mains/line voltages between 100VAC
and 240VAC. No means of switching to different input voltages
has therefore been provided.
The power input fuse is externally accessible. The fuse holder
and the 3 pole power connector is an integrated unit. The power
input fuse can be exchanged after the rubber connector is
removed. The fuse holder can be released by lever action with
the aid of a screwdriver. The starting point is a slot located on
contact pin side. The fuse can then be pushed out of the mounting
and replaced.
The fuse holder must be pushed in against the spring pressure
and locked. Use of patched fuses or short circuiting of the fuse
holder is not permissible; HAMEG assumes no liability whats-
oever for any damage caused as a result, and all warranty claims
become null and void.
Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: time lag.
Attention!
There is a fuse located inside the instrument within
the switch mode power supply:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).
The operator must not replace this fuse!
Important hints

Subject to change without notice
8
Basics of signal voltage
Type of signal voltage
The oscilloscope HM2005 allows examination of DC voltages and
most repetitive signals in the frequency range up to at least
200MHz (-3dB). The vertical amplifiers have been designed for
minimum overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when measuring
high frequency signals.These errors become noticeable at approx.
100MHz. At approx. 120MHz the reduction is approx. 10% and
the real voltage value is 11% higher. The gain reduction error can
not be defined exactly as the -3dB bandwidth of the amplifiers
differ between 200MHz and 220MHz.
For sine wave signals the -6dB limits are approx. 280MHz.
When examining square or pulse type waveforms, attention must
be paid to the harmonic content of such signals. The repetition
frequency (fundamental frequency) of the signal must therefore
be significantly smaller than the upper limit frequency of the
vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be used
for triggering.This is the case with bursts, for instance.To obtain
a well triggered display in this case, the assistance of the variable
holdoff function or the second time base may be required.
Television video signals are relatively easy to trigger using the
built in TV-Sync-Separator (TV).
For optional operation as a DC or AC voltage amplifier, each vertical
amplifier input is provided with a DC/AC switch. DC coupling
should only be used with a series connected attenuator probe or
at very low frequencies or if the measurement of the DC voltage
content of the signal is absolutely necessary.
When displaying very low frequency pulses, the flat tops may be
sloping with AC coupling of the vertical amplifier (AC limit frequency
approx. 1.6 Hz for 3dB). In this case, DC operation is preferred,
provided the signal voltage is not superimposed on a too high DC
level. Otherwise a capacitor of adequate capacitance must be
connected to the input of the vertical amplifier with DC coupling.
This capacitor must have a sufficiently high breakdown voltage rating.
DC coupling is also recommended for the display of logic and pulse
signals, especially if the pulse duty factor changes constantly.
Otherwise the display will move upwards or downwards at each
change. Pure direct voltages can only be measured with DC-coupling.
The input coupling is selectable by the AC/DC pushbutton. The
actual setting is displayed in the readout with the “ = “ symbol
for DC- and the “ ~ “ symbol for AC coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data normally
refers to effective values (rms = root-mean-square value).
However, for signal magnitudes and voltage designations in
oscilloscope measurements, the peak-to-peak voltage (Vpp) value
is applied. The latter corresponds to the real potential difference
between the most positive and most negative points of a signal
waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting peak-
to-peak value must be divided by 2x√2 = 2.83. Conversely, it
should be observed that sinusoidal voltages indicated in Vrms
(Veff) have 2.83 times the potential difference in Vpp. The
relationship between the different voltage magnitudes can be
seen from the following figure.
Voltage values of a sine curve
Vrms = effective value; Vp= simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.
The minimum signal voltage which must be applied to the Y input
for a trace of 1div height is 1mVpp (± 5%) when this deflection
coefficient is displayed on the screen (readout) and the vernier is
switched off (VAR-LED dark). However, smaller signals than this
may also be displayed. The deflection coefficients are indicated
in mV/div or V/div (peak-to-peak value).
The magnitude of the applied voltage is ascertained by multiplying
the selected deflection coefficient by the vertical display height
in div. If an attenuator probe x10 is used, a further multiplication
by a factor of 10 is required to ascertain the correct voltage value.
For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity can
be reduced up to a ratio of 2.5 to 1 (
please note “Controls and
Readout”
). Therefore any intermediate value is possible within
the 1-2-5 sequence of the attenuator(s).
With direct connection to the vertical input, signals up to
100Vpp may be displayed (attenuator set to 5V/div,
variable control to 2.5:1).
With the designations
H= display height in div,
U= signal voltage in Vpp at the vertical input,
D= deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given quantities:
However, these three values are not freely selectable.They have
to be within the following limits (trigger threshold, accuracy of
reading):
Hbetween 0.5 and 8div, if possible 3.2 to 8div,
Ubetween 1mVpp and 40Vpp,
Dbetween 1mV/div and 5V/div in 1-2-5 sequence.
Basics of signal voltage

9
Subject to change without notice
Examples:
Set deflection coefficient D= 50mV/div (0.05V/div),
observed display height H= 4.6div,
required voltage U= 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp,
set deflection coefficient D= 1V/div,
required display height H= 5:1 = 5div.
Signal voltage U= 230Vrmsx2
√
2 = 651Vpp
(voltage > 400Vpp, with probe 100:1: U= 65.1Vpp),
desired display height H= min. 3.2div, max. 8div,
max. deflection coefficient D= 6.51:3.2 = 2.03V/div,
min. deflection coefficient D= 6.51:8 = 0.81V/div,
adjusted deflection coefficient D= 1V/div.
The previous examples are related to the crt graticule reading.
The results can also be determined with the aid of the ∆V cursor
measurement (
please note “Controls and Readout”
).
The input voltage must not exceed 250V, independent from the
polarity.
If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
exceed + or –250V. So for AC voltages with a mean value of zero
volt the maximum peak to peak value is 500Vpp.
If attenuator probes with higher limits are used, the probes
limits are valid only if the oscilloscope is set to DC input
coupling.
If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 250V.
The attenuator consists of a resistor in the probe and the 1MΩ
input resistor of the oscilloscope, which are disabled by the AC
input coupling capacity when AC coupling is selected. This also
applies to DC voltages with superimposed AC voltages.
It also must be noted that due to the capacitive resistance of the
AC input coupling capacitor, the attenuation ratio depends on
the signal frequency. For sine wave signals with frequencies
higher than 40Hz this influence is negligible.
With the above listed exceptions HAMEG 10:1 probes can be
used for DC measurements up to 600V or AC voltages (with a
mean value of zero volt) of 1200Vpp. The 100:1 probe HZ53
allows for 1200V DC or 2400Vpp for AC.
It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer bridging
the attenuator series resistor will break down causing damage
to the input of the oscilloscope.
However, if for example only the residual ripple of a high voltage
is to be displayed on the oscilloscope, a normal x10 probe is
sufficient. In this case, an appropriate high voltage capacitor
(approx. 22-68nF) must be connected in series with the input tip
of the probe.
With Y-POS. control (input coupling to GD) it is possible to use a
horizontal graticule line as reference line for ground potential
before the measurement. It can lie below or above the horizontal
central line according to whether positive and/or negative
deviations from the ground potential are to be measured.
Total value of input voltage
The dotted line shows a voltage alternating at zero volt level. If
superimposed on a DC voltage, the addition of the positive peak
and the DC voltage results in the max. voltage (DC + ACpeak).
Time Measurements
As a rule, most signals to be displayed are periodically repeating
processes, also called periods.The number of periods per second
is the repetition frequency. Depending on the time base setting
(TIME/DIV.-knob) indicated by the readout, one or several signal
periods or only a part of a period can be displayed. The time
coefficients are stated in ms/div, µs/div or ns/div.The following
examples are related to the crt graticule reading. The results can
also be determined with the aid of the ∆T and 1/∆T (frequency)
cursor measurement (
please note “ Controls and Readout”
).
The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout .
Uncalibrated, the time base speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermediate
value is possible within the 1-2-5 sequence.
With the designations
L= displayed wave length in div of one period,
T= time in seconds for one period,
F= recurrence frequency in Hz of the signal,
Tc = time coefficient in ms, µs or ns/div and the relation
F= 1/T, the following equations can be stated:
However, these four values are not freely selectable. They have
to be within the following limits:
Lbetween 0.2 and 10div, if possible 4 to 10div,
Tbetween 2ns and 5s,
Fbetween 0.5Hz and 300MHz,
Tc between 20ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 2ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).
Examples:
Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
to be continued see next page
Basics of signal voltage

Subject to change without notice
10
required period T = 7x100x10-9 = 0.7µs
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.
TV-Line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.
When investigating pulse or square waveforms, the critical feature
is the rise time of the voltage step. To ensure that transients,
ramp-offs, and bandwidth limits do not unduly influence the
measuring accuracy, the rise time is generally measured between
10% and 90% of the vertical pulse height. For measurement,
adjust the Y deflection coefficient using its variable function
(uncalibrated) together with the Y-POS. control so that the pulse
height is precisely aligned with the 0% and 100% lines of the
internal graticule.The 10% and 90% points of the signal will now
coincide with the 10% and 90% graticule lines. The rise time is
given by the product of the horizontal distance in div between
these two coincident points and the calibrated time coefficient
setting. The fall time of a pulse can also be measured by using
this method.
With a time coefficient of 2ns/div (X x10 magnification active),
the example shown in the above figure results in a total measured
rise time of
ttot = 1.6div x 2ns/div = 3.2ns
When very fast rise times are being measured, the rise times of
the oscilloscope amplifier and of the attenuator probe has to be
deducted from the measured time value. The rise time of the
signal can be calculated using the following formula.
tr= √ttot2- tosc2- tp2
In this ttot is the total measured rise time, tosc is the rise time of
the oscilloscope amplifier (HM2005 approx. 1.75ns), and tpthe
rise time of the probe (e.g. = 1.4ns). If ttot is greater than 16ns,
then ttot can be taken as the rise time of the pulse, and calculation
is unnecessary.
Calculation of the example in the figure above results in a signal
rise time
tr= √3.22– 1.752– 1.42 = 2.28ns
The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to measure
in any display position and at any signal amplitude. It is only
important that the full height of the signal edge of interest is
visible in its full length at not too great steepness and that the
horizontal distance at 10% and 90% of the amplitude is measured.
If the edge shows rounding or overshooting, the 100% should
not be related to the peak values but to the mean pulse heights.
Breaks or peaks (glitches) next to the edge are also not taken
into account. With very severe transient distortions, the rise and
fall time measurement has little meaning. For amplifiers with
approximately constant group delay (therefore good pulse
transmission performance) the following numerical relationship
between rise time tr (in ns) and bandwidth B (in MHz) applies:
Connection ofTest Signal
In most cases briefly depressing the AUTOSET causes a useful
signal related instrument setting.The following explanations refer
to special applications and/or signals, demanding a manual
instrument setting.
The description of the controls is explained
in the section “Controls and Readout”
.
Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the input
coupling switch to AC (readout). The attenuator should
initially be set to 5V/div.
Sometimes the trace will disappear after an input signal has been
applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is only
3-8div. With a signal amplitude greater than 40Vpp and the
deflection coefficient (VOLTS/DIV.) in calibrated condition, an
attenuator probe must be inserted before the vertical input. If,
after applying the signal, the trace is nearly blanked, the period
of the signal is probably substantially longer than the set time
deflection coefficient (TIME/DIV.). It should be switched to an
adequately larger time coefficient.
The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as HZ32
or HZ34, or reduced through a x10 or x100 attenuator probe. The
use of test cables with high impedance circuits is only
recommended for relatively low frequencies (up to approx. 50kHz).
Basics of signal voltage

11
Subject to change without notice
For higher frequencies, the signal source must be of low impedance,
i.e. matched to the characteristic resistance of the cable (as a rule
50Ω). Especially when transmitting square and pulse signals, a
resistor equal to the characteristic impedance of the cable must
also be connected across the cable directly at the Y-input of the
oscilloscope. When using a 50Ωcable such as the HZ34, a 50Ω
through termination type HZ22 is available from HAMEG. When
transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become visible
if the correct termination is not used. A terminating resistance is
sometimes recommended with sine signals as well. Certain
amplifiers, generators or their attenuators maintain the nominal
output voltage independent of frequency only if their connection
cable is terminated with the prescribed resistance. Here it must be
noted that the terminating resistor HZ22 will only dissipate a
maximum of 2 Watts. This power is reached with 10Vrms or at
28.3Vpp with sine signal. If a x10 or x100 attenuator probe is used,
no termination is necessary. In this case, the connecting cable is
matched directly to the high impedance input of the oscilloscope.
When using attenuators probes, even high internal impedance
sources are only slightly loaded (approx. 10MΩII 12pF or 100MΩII
5pF with HZ53).Therefore, if the voltage loss due to the attenuation
of the probe can be compensated by a higher amplitude setting, the
probe should always be used. The series impedance of the probe
provides a certain amount of protection for the input of the vertical
amplifier. Because of their separate manufacture, all attenuator
probes are only partially compensated, therefore accurate
compensation must be performed on the oscilloscope (see Probe
compensation ).
Standard attenuator probes on the oscilloscope normally reduce
its bandwidth and increase the rise time. In all cases where the
oscilloscope bandwidth must be fully utilized (e.g. for pulses with
steep edges) we strongly advise using the probe HZ52 (x10 HF).
This can save the purchase of an oscilloscope with larger
bandwidth.
The probe mentioned has a HF-calibration in addition to low
frequency calibration adjustment. Thus a group delay correction
to the upper limit frequency of the oscilloscope is possible with
the aid of an 1MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are not
noticeably changed with these probe types and the waveform
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscopes individual pulse response.
If a x10 or x100 attenuator probe is used, DC input coupling
must always be used at voltages above 250V. With AC
coupling of low frequency signals, the attenuation is no
longer independent of frequency, pulses can show pulse
tilts. Direct voltages are suppressed but load the
oscilloscope input coupling capacitor concerned. Its
voltage rating is max. 250 V (DC + peak AC). DC input
coupling is therefore of quite special importance with a
x100 attenuation probe which usually has a voltage rating
of max. 1200 V (DC + peak AC). A capacitor of
corresponding capacitance and voltage rating may be
connected in series with the attenuator probe input for
blocking DC voltage (e.g. for hum voltage measurement).
With all attenuator probes, the maximumAC input voltage must
be derated with frequency usually above 20kHz. Therefore the
derating curve of the attenuator probe type concerned must be
taken into account.
The selection of the ground point on the test object is important
when displaying small signal voltages. It should always be as
close as possible to the measuring point. If this is not done,
serious signal distortion may result from spurious currents
through the ground leads or chassis parts. The ground leads on
attenuator probes are also particularly critical. They should be as
short and thick as possible. When the attenuator probe is
connected to a BNC-socket, a BNC-adapter, should be used. In
this way ground and matching problems are eliminated. Hum or
interference appearing in the measuring circuit (especially when
a small deflection coefficient is used) is possibly caused by
multiple grounding because equalizing currents can flow in the
shielding of the test cables (voltage drop between the protective
conductor connections, caused by external equipment connected
to the mains/line, e.g. signal generators with interference
protection capacitors).
Controls and Readout
The following description assumes that the instrument is
not set to “COMPONENTTESTER” mode.
If the instrument is switched on, all important settings are
displayed in the readout. The LED´s located on the front panel
assist operation and indicate additional information. Incorrect
operation and the electrical end positions of control knobs are
indicated by a warning beep.
Except for the power pushbutton (POWER), the calibrator
frequency pushbutton (CAL. 1kHz/1MHz), the focus control
(FOCUS) and the trace rotation control (TR) all other controls
are electronically selected. All other functions and their settings
can therefore be remote controlled and stored.
The front panel is subdivided into sections. – On the top,
immediately to the right of the CRT screen, the following controls
and LED indicators are placed.
(1) POWER
Pushbutton and symbols for ON (I) and OFF (O).
After the oscilloscope is switched on, all LEDs are lit and an
automated instrument test is performed. During this time
the HAMEG logo and the software version are displayed on
the screen. After the internal test is completed successfully,
the overlay is switched off and the normal operation mode is
present. Then the last used settings become activated and
one LED indicates the ON condition.
Some mode functions can be modified (SETUP) and/or
automated adjustment procedures (CALIBRATE) can be
called if the “MAIN MENU” is present.
For further infor-
mation please note“MENU”
.
(2) AUTOSET
Briefly depressing this pushbutton (
please note
“AUTOSET”
) automatically selectsYt mode.The instrument
is set to the last usedYt-mode setting (CH I, CH II or DUAL).
Controls and Readout
AUTOSET
TR
B
RM READ
OUT
200 MHz
ANALOG OSCILLOSCOPE
HM2005
Instruments
!
POWER INTENS
EXIT
SAVE
SET
1
FOCUS
9
RECALL
A
RO
21 3 4 5 6 7

Subject to change without notice
12
Even if alternating time base mode or B time base mode
was active before, the instrument is switched automatically
to A time base mode.
Please note “AUTOSET”
.
Automatic CURSOR supported voltage measurement. If
CURSOR voltage measurement is present, the CURSOR
lines are automatically set to the positive and negative peak
value of the signal. The accuracy of this function decreases
with higher frequencies and is also influenced by the signal‘s
pulse duty factor.
In DUAL mode the CURSOR lines are related to the signal
which is used for internal triggering. If the signal height is
insufficient, the CURSOR lines do not change.
(3) RM
The remote control mode can be switched on or off via the
RS232 interface. In the latter case the “RM” LED is lit and
the electronically selectable controls on front panel are
inactive. This state can be left by depressing the AUTOSET
pushbutton provided it was not deactivated via the interface.
(4) INTENS - READOUT
Knob with associated pushbutton and LEDs.
This control knob is for adjusting the A and B traces and
readout intensity. Turning this knob clockwise increases and
turning it counterclockwise decreases the intensity.
The READOUT pushbutton below is for selecting the function
in two ways.
Depending on the actual time base mode and with the
readout (RO) not switched off, briefly pressing the READOUT
pushbutton switches over the INTENS knob function
indicated by a LED in the sequences:
A - RO - A in condition A time base,
A - RO - B - A if alternate time base mode is present and
B - RO - B in condition B time base.
XY mode: A - RO - A.
Component Test: A - RO - A.
Pressing and holding the READOUT pushbutton switches
the readout on or off. In readout off condition the INTENS
knob function can consequently not be set to RO. Briefly
pressing the pushbutton causes the following sequences:
condition sequence
A time base A - A
Alternate A/B A - B - A
B time base B - B
XY mode A - A
Component Test A - A
Switching the readout off, may be required if interference is
visible on the signal(s). Such interference may also originate
from the chopper generator if the instrument is operated in
chopped DUAL mode. All INTENS-settings are stored after
the instrument is switched off.
The AUTOSET function switches the readout on and selects
A time base mode (A-LED lit). The INTENS setting for each
function is automatically set to the mean value, if less
intensity was previously selected.
(5) TR
The trace rotation control can be adjusted with a small
screwdriver (
please note “trace rotationTR”
)
(6) FOCUS
This control knob effects both the trace and the readout
sharpness.
(7) SAVE / RECALL
The instrument contains 9 non volatile memories.These can
be used by the operator to save instrument settings and to
recall them.This relates to all controls which are electronically
selected.
Press the SAVE pushbutton briefly to start the save
procedure.The readout then indicates the letter “S” followed
by a cipher between 1 and 9, indicating the memory location.
If the instrument settings stored in this memory location
must not be overwritten, briefly press the SAVE or the
RECALL pushbutton to select another memory location. Each
time the SAVE pushbutton is briefly pressed the memory
location cipher increases until the location number 9 is
reached. The RECALL pushbutton function is similar but
decreases the memory location cipher until 1 is reached.
Press and hold SAVE for approx. 3 seconds to write the
instruments settings in the memory and to switch the
associated readout information (e.g. “S8”) off.
To recall a front panel setup, start that procedure by briefly
pressing the RECALL pushbutton.The readout then indicates
the letter “R” and the memory location number. If required,
select a different memory location as described above. Recall
the settings by pressing and holding the RECALL pushbutton
for approx. 3 seconds.
If the SAVE or the RECALL pushbutton was depressed
inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.
Switching the instrument off results in an automatic SAVE
procedure of the present settings in memory location 9 and
overwrites the data in that location. If the instrument settings
in memory location 9 are of importance, RECALL 9 before
switching the instrument off.
Attention:
Make sure that the signal to be displayed is similar to the
one that was present when the settings were stored. If
the signal is different (frequency, amplitude) to the one
during storage then a distorted display may result.
The setting controls and LED’s for the Y amplifiers, modes,
triggering and time bases are located underneath the sector of
the front panel described before (Responding picture see next
page).
(8)TRS – Pushbutton wit associated LED.
The instrument contains a trace separation function which
is required in the alternate time base mode to separate the
B time base trace from the A time base in Y direction.
Controls and Readout
AUTOSET
TR
B
RM
READ
OUT
200 MHz
ANALOG OSCILLOSCOPE
HM2005
Instruments
!
POWER INTENS
EXIT
SAVE
SET
1
FOCUS
9
RECALL
A
RO
21 3 4 5 6 7

13
Subject to change without notice
Consequently this function is only available in alternate time
base mode. After theTRS pushbutton was pressed once the
LED related to that pushbutton is lit.
The Y-POS. I (9) control knob is then operative as vertical
position control for the trace of the B time base. The
maximum position shift is approx. +/- 4 div. Without a change
of the Y-POS. I (9) control the trace separation function is
switched off automatically after approx. 10 seconds.The trace
separation function can also be left by pressing the TRS
pushbutton.
(9)Y-POS. I – Control knob with a double function.
Y-Position channel I:
The vertical trace position of channel I can be set with this
control knob. In ADD (addition) mode both (Y-POS. I and Y-
POS. II) control knobs are active.
Y-Position B-trace in alternate time base mode:
In alternate time base mode, this control knob can be used
to separate the B time base trace from the A time base trace.
Please noteTRS (8)
.
DC voltage measurement:
If no signal is applied at the INPUT CHI (28), the vertical trace
position represents 0 Volt. This is the case if INPUT CHI (28)
or in addition (ADD) mode, both INPUT CHI (28) and INPUT
CHII (32), are set to GD (ground) and automatic triggering
(AT (12)) is present to make the trace visible. The trace then
can be set to vertical position which is suited for the following
DC voltage measurement.
After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the trace
position in vertical direction. The DC voltage then can be
determined by taking the deflection coefficient, the probe
factor and the trace position change in respect to the previous
0 Volt position into account.
”0 Volt” Symbol:
The determination of the ”0 Volt” position is not necessary
if the readout is switched on and the software setting ”DC
REFERENCE = ON” is selected in the ”SETUP” submenu
”Miscellaneous”. Then the ”⊥⊥
⊥⊥
⊥” symbol to the left of the
screen‘s vertical center line always indicates the ”0 Volt”
trace position in CHI and DUAL mode.
The ”0 Volt” position symbol (⊥⊥
⊥⊥
⊥) will not be displayed in XY
and ADD (addition) mode.
(10) BW Limit – Pushbutton with associated BWL-LED.
Pressing this pushbutton switches the BWL–LED and readout
display on or off. The amplifier(s) bandwidth will be reduced
if deflection coefficients from 5mV/div. to 5V/div. are present
and BWL (Bandwidth Limit) is active.The bandwidth reduction
causes less amplifier noise and therefore increases the trace
sharpness.
On 1mV/div. and 2mV/div. deflection coefficient settings, the
bandwidth is already reduced for the benefit of higher input
sensitivity. The BWL function is practically ineffective in this
settings.
(11) Y-POS. II – Control knob.
The vertical trace position of channel II can be set with this
control knob. In ADD (addition) mode both (Y-POS. I and Y-
POS. II) control knobs are active. If the instrument is set to
XY mode this control knob is inactive and the X-POS. knob
must be used for a horizontal position shift.
DC voltage measurement:
If no signal is applied at the INPUT CHII (32), the vertical
trace position represents 0 Volt. This is the case if INPUT
CHII (32) or in addition (ADD) mode, both INPUT CHI (28)
and INPUT CHII (32), are set to GD (ground) and automatic
triggering (AT (12)) is present to make the trace visible. The
trace then can be set to vertical position which is suited for
the following DC voltage measurement.
After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the trace
position in vertical direction. The DC voltage then can be
determined by taking the deflection coefficient, the probe
factor and the trace position change in respect to the previous
0 Volt position into account.
”0 Volt” Symbol:
The determination of the ”0 Volt” position is not necessary
if the readout is switched on and the software setting ”DC
REFERENCE. = ON” is selected in the ”SETUP” submenu
”Miscellaneous”. Then the ”⊥⊥
⊥⊥
⊥” symbol to the left of the
screen‘s vertical center line always indicates the ”0 Volt”
trace position in CHI and DUAL mode.
The ”0 Volt” position symbol (⊥⊥
⊥⊥
⊥) will not be displayed in XY
and ADD (addition) mode.
(12)NM - AT - (SLOPE)
Pushbutton with a double function and associated NM-LED.
NM - AT selection:
Press and hold the pushbutton to switch over from automatic
(peak value) to normal triggering (NM LED above the
pushbutton lit) and vice versa. If the LED is dark, automatic
(peak value) triggering is selected. Whether the peak value
detection in automatic trigger mode is automatically activated
or not, depends on the trigger coupling setting (TRIG. MODE).
The way the trigger point symbol in the readout responds on
different LEVEL control knob settings indicates the situation:
1. If the trigger symbol can not be shifted in the vertical
direction when a signal is not applied or the signal height
is not sufficient, the peak value detection is active.
2. Under the condition that the trigger point symbol cannot
be shifted in such a way that it leaves the signal display
on the screen, the peak value detection is active.
Controls and Readout
MADE IN GERMANY
A/ALT. DEL.TRIG
TRS
PUSH
BOTH
PUSH
LONG
VAR.
B
CH I CH II
DEL.POS.
ADD
CHP. VA R .VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
17 18 19 20 22 23
24 25
26 27
21
910 12
11 14 161513
8

Subject to change without notice
14
3. The peak value detection is switched off if the trigger point
can be set outside the maximum peak values of the signal,
thus causing an untriggered signal display.
Slope selection:
Briefly pressing this pushbutton selects which slope of the
signal is used for triggering the time base generator. Each
time this pushbutton is briefly pressed, the slope direction
switches from falling edge to rising edge and vice versa.
The current setting is displayed in the readout under item
“TR: source, SLOPE, coupling”. The last setting in A time
base mode is stored and still active if the alternate (A and B)
or B time base are selected. This allows for a different slope
setting regarding the B time base if the DEL. TRIG. function
is active. The slope direction chosen for the B time base is
indicated in the readout under “DTr: SLOPE, coupling”.
(13)LEVEL – Control knob.
Turning the LEVEL knob causes a different trigger point
setting (voltage). The trigger unit starts the time base when
the edge of a trigger signal (voltage) crosses the trigger point.
In mostYt modes the trigger point is displayed in the readout
by the symbol on the left vertical graticule line. If the trigger
point symbol would overwrite other readout information or
would be invisible when being set above or below the screen,
the symbol changes and an arrow indicates in which vertical
direction the trigger point has left the screen. The trigger
point symbol is automatically switched off in those modes
where there is no direct relation between the trigger signal
and the displayed signal.
The last setting in A time base mode is stored and still active
if alternate (A and B) or B time base mode are selected. This
allows for a different level setting for the B time base if the
DEL.TRIG. function is active. Under this condition the letter
“B” is added to the trigger point symbol.
(14)TR – Trigger indicator LED.
The TR LED is lit in Yt mode if the triggering conditions are
met. Whether the LED flashes or is lit constantly depends
on the frequency of the trigger signal.
(15)X-POS. – Control knob.
This control knob enables an X position shift of the signal(s)
in Yt and XY mode. In combination with X magnification x10
this function makes it possible to shift any part of the signal
on the screen.
(16)X-MAG. x10 – Pushbutton and LED.
Each time this pushbutton is pressed the x10 LED located
above is switched on or off inYt (time base) and component
test operation. If the x10 LED is lit, the signal display is
expanded 10 fold in X direction and consequently only a tenth
part of the signal curve is visible. The interesting part of the
signal can be made visible with aid of the X-POS. control.
As the X expansion results in a higher time base speed (lower
time deflection coefficient), all time and frequency relevant
information in the readout is switched over.
The X magnifier function can be operated in A and B time
base mode. In alternate time base mode (A alternate B), the
X magnification only effects the B time base and similarly
the readout.
Please note that in alternate time base mode the inten-
sified sector may become invisible due to the X position
setting.
This pushbutton is not operative in XY mode.
(17) VOLTS/DIV. – This control knob for channel I has a double
function.
The following description relates to the input attenuator
function (VAR LED dark). The vernier function is active if the
VAR LED is lit (see VAR under item (18)).
Turning the control knob clockwise increases the sensitivity
in a 1-2-5 sequence and decreases it if turned in the opposite
direction (ccw.). The available range is from 1mV/div up to
5V/div. The knob is automatically switched inactive if the
channel related to it is switched off, or if the input coupling is
set to GD (ground).
The deflection coefficients and additional information
regarding the active channels are displayed in the readout,
e.g. “Y1: deflection coefficient, input coupling”. The “:”
symbolizes calibrated measuring conditions and is replaced
by the “>” symbol in uncalibrated conditions.
(18)CH I - VAR. – Pushbutton with several functions.
CH I mode:
Briefly pressing the CHI button sets the instrument to channel
I (Mono CH I) mode. The deflection coefficient displayed in
the readout indicates the current conditions (“Y1...”).
If neither external nor line (mains) triggering was active,
the internal trigger source automatically switches over to
channel I (“TR:Y1...”).The last function setting of theVOLTS/
DIV (17) knob remains unchanged.
All channel I related controls are active if the input (28) is
not set to GD (30).
VAR.:
Pressing and holding this pushbutton selects the VOLTS/
DIV. (17) control knob function between attenuator and
vernier (variable).The current setting is displayed by the VAR-
LED located above the knob.
After switching the VAR-LED (17) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (17)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated.
The readout then displays “Y1>...” indicating the uncalibrated
condition instead of “Y1:...”. Pressing and holding the CHI
pushbutton again switches the LED off, sets the deflection
Controls and Readout
MADE IN GERMANY
A/ALT. DEL.TRIG
TRS
PUSH
BOTH
PUSH
LONG
VAR.
B
CH I CH II
DEL.POS.
ADD
CHP. VA R .VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
17 18 19 20 22 23
24 25
26 27
21
910 12
11 14 161513
8

15
Subject to change without notice
coefficient into calibrated condition and activates the
attenuator function. The previous vernier setting will not be
stored.
The CHI pushbutton can also be pressed simultaneously with
the DUAL(19) button.
Please note item (19)
.
(19)DUAL - XY – Pushbutton with multiple functions.
DUAL mode:
Briefly pressing this button switches over to DUAL mode.
Both deflection coefficients are then displayed.The previous
trigger setting stays as it was, but can be changed.
All controls related to both channels are active, if the inputs
(28) and (32) are not set to GD (30) (34). Whether alternated
or chopped channel switching is present depends on the
actual time base setting, and is displayed in the readout.
ALT
displayed in the readout, indicates alternate channel
switching. After each time base sweep the instrument
internally switches over from channel I to channel II and vice
versa.This channel switching mode is automatically selected
if any time coefficient from 200µs/div to 50ns/div is active.
CHP
indicates chopper mode, whereby the channel switching
occurs constantly between channel I and II during each
sweep. This channel switching mode occurs when any time
base setting between 500ms/div and 500µs/div has been
chosen.
The actual channel switching can be changed to the opposite
mode by briefly pressing both CHI (18) and DUAL (19)
simultaneously. If afterwards the time coefficient is changed,
the channel switching is automatically set to the time
coefficient related mode.
ADD mode:
Addition mode can be selected by briefly pressing the DUAL
(19) and CHII (22) buttons simultaneously. Whether the
algebraic sum (addition) or the difference (subtraction) of both
input signals is displayed, depends on the phase relationship
and the INV (30) (34) setting(s). As a result both signals are
displayed as one signal. For correct measurements the
deflection coefficients for both channels must be equal.
The readout indicates this mode by a “+” sign located
between both channel deflection coefficients. While the
trigger mode is not affected, the trigger point symbol is
switched off. The Y-position of the signal can be influenced
by both Y-POS controls (9) and (11).
XY mode:
This mode can be switched on or off by pressing and holding
the DUAL button (19).
In XY mode the deflection coefficients are displayed as “X...”
for channel I and “Y...” for channel II, followed by “XY”. Except
the cursor lines which may be active, all other readout
information including the trigger point symbol are switched
off. In addition to all trigger and time base related controls,
the Y-POS. I (9) knob and INV (30) button are deactivated.
For X position alteration, the X-POS. (15) knob can be used.
(20)TRIG. - ALT – Pushbutton with double function for trigger
source selection and associated LEDs.
The button and the LEDs are deactivated if line (mains)
triggering is selected or XY operation is chosen.
With the aid of this button, the trigger source can be chosen.
There are three trigger sources available: channel I, channel
II (both designated as internal trigger sources) and theTRIG.
EXT. (35) input for external triggering.
The availability of the internal sources depends on the actual
channel mode. The actual setting is indicated by the
associated LED(s). Briefly pressing the button switches over
in the following sequence:
I - II - EXT - I in DUAL and ADD (addition) mode,
I - EXT - I if mono channel I is present,
II - EXT - II under mono channel II conditions.
Each condition is indicated by the associated LED and
displayed by the readout (“TR:Y1...”, “TR:Y2...” and
“TR:EXT...”). The trigger point symbol is switched off in
external trigger condition.
ALT:
Pressing and holding the button selects alternate triggering
in DUAL mode. Under these conditions both I and II LEDs
are lit and the readout displays “TR:ALT...”. As alternate
triggering requires alternate channel operation, alternate
channel switching is set automatically. A change of the time
coefficient then has no affect regarding the channel switching
mode. In addition to the deflection coefficients display, “A LT ”
is displayed by the readout instead of “CHP”. In alternate
trigger mode the trigger point symbol is switched off.
Alternate triggering is not available or automatically
switched off under the following conditions:
– ADD (addition) mode,
– alternate (A & B) time base mode,
– B time base mode,
– TVL, TVF and line (mains) trigger coupling.
(21) VOLTS/DIV. – This control knob for channel II has a double
function.
The following description relates to the input attenuator
function (VAR LED dark). The vernier function is active if the
VAR LED is lit (see VAR under item (22)).
Turning the control knob clockwise increases the sensitivity
in a 1-2-5 sequence and decreases it if turned in the opposite
direction (ccw.). The available range is from 1mV/div up to
20V/div. The knob is automatically switched inactive if the
channel related to it is switched off, or if the input coupling is
set to GD (ground).
Controls and Readout
MADE IN GERMANY
A/ALT. DEL.TRIG
TRS
PUSH
BOTH
PUSH
LONG
VAR.
B
CH I CH II
DEL.POS.
ADD
CHP. VA R .VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
17 18 19 20 22 23
24 25
26 27
21
910 12
11 14 161513
8

Subject to change without notice
16
The deflection coefficients and additional information
regarding the active channels are displayed in the readout,
e.g. “Y2: deflection coefficient, input coupling”. The “:”
symbolizes calibrated measuring conditions and is replaced
by the “>” symbol in uncalibrated conditions.
(22)CH II - VAR. – Pushbutton with several functions.
CH II mode:
Briefly pressing the button sets the instrument to channel II
(Mono CH II) mode. The deflection coefficient displayed in
the readout indicates the current conditions (“Y2...). If neither
external nor line (mains) triggering was active, the internal
trigger source automatically switches over to channel II
(“TR:Y2...). The last function setting of the VOLTS/DIV (21)
knob remains unchanged.
All channel related controls are active if the input (32) is not
set to GD (34).
VAR.:
Pressing and holding this pushbutton selects the VOLTS/
DIV. (21) control knob function between attenuator and
vernier (variable).The current setting is displayed by the VAR-
LED located above the knob.
After switching the VAR-LED (21) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (21)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated.
The readout then displays “Y2>...” indicating the uncalibrated
condition instead of “Y2:...”. Pressing and holding the CHII
pushbutton again switches the LED off, sets the deflection
coefficient into calibrated condition and activates the
attenuator function. The previous vernier setting will not be
stored.
The CHII pushbutton can also be pressed simultaneously
with the DUAL (19) button.
Please note item (19)
.
(23)TRIG. MODE – Pushbuttons and indicator LEDs.
Pressing the upper or lower button selects the trigger
coupling. The actual setting is indicated by a LED and by the
readout (“TR: source, slope, AC”).
Each time the lowerTRIG. MODE pushbutton is pressed the
trigger coupling changes in the sequence:
AC (DC content suppressed),
DC (peak value detection inactive),
HF (high-pass filter cuts off frequencies below
approx. 50 kHz), trigger point symbol switched off
NR (high frequency noise rejected),
LF (low-pass filter cuts off frequencies above
approx. 1.5 kHz),
TVL (TV signal, line pulse triggering)
trigger point symbol switched off,
TVF (TV signal, frame pulse triggering)
trigger point symbol switched off.
~(line/mains triggering) trigger point symbol
and TRIG. LED (20) are switched off.
Please note:
In delay trigger mode (B time base) the instrument is
automatically set to normal triggering mode and DC trigger
coupling. Both settings are indicated by the NM- (12) and
the “DC” TRIG. MODE-LED(23). The previous trigger
settings regarding the A time base remain unchanged.
In some trigger modes such as alternate triggering, some trigger
coupling modes are automatically disabled and can not be
selected.
(24)DEL.POS. - HO – Control knob with a double function and
associated LED. This control knob has two different func-
tions depending on the time base mode.
A time base:
In A time base mode, the control knob applies to the hold off
time setting. If the HO-LED associated with the knob is dark,
the hold off time is set to minimum.
Turning the control knob clockwise switches the LED on and
extends the hold off time until the maximum is reached
(
please note “Hold Off time adjustment”
). The hold off
time is automatically set to minimum (LED dark), if the A
time base setting is changed. The hold off time setting is
stored and active if alternate (A and B) or B time base mode
is selected.
Alternate (A and B) and B time base:
In alternate (A and B) and B time base modes, the knob
controls the delay time setting.
Under alternate time base mode conditions, the delay time
is visible on the A trace, beginning at the trace start and
ending at the start of the intensified sector. In the free run
condition (delay trigger not active) of the time base, an
approximate delay time value is displayed in the readout
(“∆∆
∆∆
∆t:...”). This is an aid to find the position of the intensified
sector which may be very small.
If only the B time base is being operated, the delay time can
also be varied, but there is no intensified sector as the A
trace is not visible.
(25)TIME/DIV. – This control knob has a double function.
The following description applies to the time base switch
function (VAR LED dark).
Time Base Switch:
Turning the control knob clockwise reduces the deflection
coefficient in a 1-2-5 sequence and increases it if turned in
the opposite direction (ccw.). The time coefficient(s) is (are)
displayed in the readout.
In A time base mode, time deflection coefficients between
500ms/div and 20ns/div can be chosen in 1-2-5 sequence, if
the X-MAG. x10 function is not activated.
During alternate (A and B) and B time base operation, the
control knob changes the B time base setting in 1-2-5
sequence. The available deflection coefficient range is from
Controls and Readout
MADE IN GERMANY
A/ALT. DEL.TRIG
TRS
PUSH
BOTH
PUSH
LONG
VAR.
B
CH I CH II
DEL.POS.
ADD
CHP. VA R .VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
17 18 19 20 22 23
24 25
26 27
21
910 12
11 14 161513
8

17
Subject to change without notice
20ms/div up to 20ns/div (without X-MAG. x10) but the
availability depends on the A time base setting. The internal
control of the oscilloscope prevents the B time deflection
coefficient from becoming higher than the A deflection
coefficient, as such an operation condition would make no
sense.
If the A time base setting is 200µs/div the B time base range
from 20ms/div up to 500µs/div is not available and the
maximum time deflection coefficient for B would be 200µs/
div. In the last named condition the change of the A time
base from 200µs/div to 100µs/div switches the B time base
also to 100µs/div. However the B time base setting remains
unchanged if the A time base is set to 500µs/div.
As already mentioned under DUAL (19) the channel switching
depends on the time deflection coefficient setting. In the
time base ranges from 500ms/div to 500µs/div chopped
(CHP) channel switching is automatically selected, through
which the switching occurs constantly during the time base
sweeps. Alternate (ALT) channel switching is automatically
used in all other time deflection coefficient settings (200µs/
div - 20ns/div). In the latter case the active channel is switched
off and the previously inactive channel is switched on after
the completion of each time base sweep. To avoid
interference in chopped mode, or to make both channels
appear simultaneously visible, the actual setting (ALT or CHP)
displayed in the readout can be overwritten and changed to
the opposite mode. This is carried out by simultaneously
pressing and holding the CHI (18) and the DUAL (19)
pushbuttons.
(26)A/ALT – B
Pushbutton for time base mode selection.
The instrument contains two time bases designated A and
B. With the aid of the B time base, signal parts displayed by
the A time base can be expanded in X-direction. The
expansion ratio depends on the time deflection coefficient
ratio of both time bases (e.g. “A: 100µs”, “B:1µs” = 100).
With higher expansion ratios the B time base trace intensity
reduces.
Each time the A/ALT pushbutton is briefly pressed, the time
base mode changes in the sequence A - alternate A and B -
A. The actual setting is displayed in the readout.
A:
In A time base mode the TIME/DIV. (25) control knob is
operative only for this time base. The readout then only
displays the A time coefficient. The time base settings for
this condition are stored if the time base mode is changed.
ALT:
If alternate (A and B = ALT) time base mode is selected, the
TIME/DIV (25) knob only controls the B time base switch or
vernier function. The alternate time base mode is a
subfunction of the B time base mode where both time base
traces are displayed. Consequently the readout displays both
time deflection coefficients (e.g. “A:100µs B:1µs”). Unlike
the former A time base mode, an intensified sector is also
visible on the A trace. This sector indicates the signal part
which is displayed by the B time base.
The intensified segment can be shifted horizontally by the
DEL. POS. (24) control knob continuously (if the B time base
is operated in free run conditions). The difference between
the start of the A time base trace and the beginning of the
intensified sector shows the delay time. This information is
also displayed in the readout as an approximative value (e.g.
“∆t:2.5ms”) related on the calibrated A time coefficient
(uncalibrated e.g. “∆t>2.5ms”). The width of the segment
decreases when the B time coefficient is set to a lower value
(higher time deflection speed). For better reading, the vertical
position of the B trace position can be shifted (
please note
“TRS” (8)
).
Alternate time base mode causes the display to alter
between A and B time base after each sweep. In alternate
DUAL mode the sequence is channel I with A time base,
channel I with B time base, channel II with A time base and
channel II with B time base.
In alternate time base mode (A alternate B), the X-MAG.
x10 (16) function only effects the B time base and similarly
the readout.
B:
Pressing and holding this button switches over from A or
alternate time base mode to B time base mode. If B time
base mode is selected, pressing and holding selects alternate
time base mode. Briefly pressing this button switches over
from B time base mode to A time base mode.
In B time base mode the display of the A trace(s), the
intensified sector(s) and the A time coefficient display in the
readout are disabled. As the trace separation (TRS) is no
longer required under this circumstances, this function is
switched off too. Consequently only the B time coefficient
is displayed by the readout.
When selecting alternate or B time base mode, the triggered
or free run state depends on the last chosen setting.
(27) DEL.TRIG. - VAR. – Pushbutton with two functions.
DEL.TRIG.:
Each time the pushbutton is briefly pressed, the instrument
switches between free run (untriggered) and triggered B time
base, if alternate (A and B) or B time base mode is present.
The actual setting is displayed in the readout. Instead of the
approximate delay time (“Dt:...”) in free run mode, the
readout displays “DTr: slope, DC (trigger coupling)” in
triggered delay mode. In this mode, the former A time base
trigger settings regarding the trigger mode (automatic or
normal, -coupling, -slope and –level) are stored but still active.
With the activated delay trigger the instrument is
automatically set to normal trigger mode and DC trigger
coupling for the B time base. As the instrument contains a
separate trigger unit for the B time base, the trigger level
Controls and Readout
MADE IN GERMANY
A/ALT. DEL.TRIG
TRS
PUSH
BOTH
PUSH
LONG
VAR.
B
CH I CH II
DEL.POS.
ADD
CHP. VA R .VAR.
DUAL
X-Y
VOLTS / DIV. VOLTS / DIV. TIME / DIV.
5V 1mV 5V 1mV 0.5s 20ns
TRIG. MODE
TRIG.
ALT
CHI AC
DC
HF
NR
LF
TVL
TVF
CHII
EXT
Y-POS. I Y-POS. II LEVEL X-POS.
NM
AT
X-
MAG.
6
3
-
K
3
0
4
-
0
0
4
0
/
0
0
5
5
BWL
BW
LIMIT
VAR
VAR
TR
NM
H
O
VAR
x
1
0
17 18 19 20 22 23
24 25
26 27
21
910 12
11 14 161513
8

Subject to change without notice
18
and slope can be set independently using the same controls
used for the A time base trigger setting. The trigger point is
indicated again but has the added letter B in the readout.
In delay trigger mode, the delay time must first elapse. Then
the next suitable signal slope (direction and height) starts
the B time base. If those basic requirements are not met,
there will be no B trace visible. Under the condition that the
signal contains several suitable slopes after the delay time,
it can be seen in alternate time base mode that the delay
time setting (DEL. POS.) now causes the intensified sector
to jump from one slope to another. If there is no slope after
the delay time, the B time base will not be triggered.
VAR.:
Pressing and holding the DEL.TRIG. - VAR. pushbutton
selects the TIME/DIV. (25) control knob function between
time base switch and vernier (variable). The current setting
is displayed by the VAR-LED located above the knob.
The variable function can be activated for the A time base or
the B time base, the settings are stored separately. As
alternate time base mode is a B time base sub mode, only
the B time base is affected in this condition.
ATime Base.
After switching the VAR-LED on, the time deflection
coefficient is still calibrated until further adjustments are
made. Turning the TIME/DIV. (25) control knob counter
clockwise increases the time deflection coefficient (reduces
the deflection speed) and the deflection coefficient becomes
uncalibrated. Instead of e.g. “A:10µs”, the readout then
displays “A>10µs” indicating the uncalibrated condition.This
setting is stored if the instrument is switched to alternate (A
and B) or B time base mode.
Pressing and holding the DEL.TRIG. -VAR. pushbutton again
in A time base mode switches the LED off, the time base
switch function on, and sets the time deflection coefficient
back into the calibrated condition.
B and Alternate Time base.
In alternate (A and B) as well as in B time base mode, pressing
and holding the DEL.TRIG. - VAR. pushbutton selects the
TIME/DIV. knob function between B time base switch and
B vernier (VAR -LED on). In the latter case the TIME/DIV.
knob can be used in the same way as described before under
A time base condition.
Underneath the front panel sector described above, the BNC
sockets and four pushbuttons are located.
(28)INPUT CH I – BNC socket.
This BNC socket is the signal input for channel I. In XY mode,
signals at this input are used for the X deflection. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug.
(29)AC / DC – Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC (~
symbol) to DC (= symbol) input coupling and vice versa.The
setting is displayed in the readout with the deflection
coefficient.
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel I displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor is
thus indicated by a probe symbol displayed by the readout in
front the channel information (e.g. “probe symbol”, Y1...). In
the case of cursor voltage measurement, the probe factor is
automatically included.
Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.
(30)GD - INV. – Pushbutton with two functions.
GD:
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is displayed
in the readout as an earth (ground) symbol instead of the
deflection coefficient and the ~ (AC) or = (DC) (input coupling)
symbol.
The GD setting disables the input signal, the AC/DC (29)
pushbutton and the VOLTS/DIV (17) knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
The ”0 Volt” position is indicated by a ”⊥” symbol in the
readout. Please note Y-POS. I (9).
INV.:
Pressing and holding this pushbutton switches the channel I
invert function on or off.The invert “on” condition is indicated
by the readout with a horizontal bar above “Y1” (Yt mode).
The invert function causes the signal display of channel I to
be inverted by 180°. In XY mode the invert function is not
available.
(31)Ground socket – 4mm banana socket galvanically
connected to safety earth.
This socket can be used as reference potential connection
for DC and low frequency signal measurement purposes and
in COMPONENTTEST mode.
(32)INPUT CH II – BNC socket.
This BNC socket is the signal input for channel II. In XY mode,
signals at this input are used for the Y deflection. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug.
(33)AC / DC – Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC (~
symbol) to DC (= symbol) input coupling and vice versa.The
setting is displayed in the readout with the deflection
coefficient.
Controls and Readout
MADE IN GERMANY
INV.
INPUT CH I (HOR. INP.(X))
AC/DC
max.
250 Vp
1MΩII
15pF
INPUT CH II
max.
250 Vp
1MΩII
15pF
TRIG. EXT. INP
max.
100 Vp
(Z)
x1/x10 INV.
x1/x10
CAT
I
!!!
GD AC/DC GD
28 30 31 32 34
29 33 35

19
Subject to change without notice
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor is
thus indicated by a probe symbol displayed by the readout in
front the channel information (e.g. “probe symbol”, Y2...). In
the case of cursor voltage measurement, the probe factor is
automatically included.
Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.
(34)GD - INV. – Pushbutton with two functions.
GD:
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is displayed
in the readout as an earth (ground) symbol instead of the
deflection coefficient and the ~ (AC) or = (DC) (input coupling)
symbol.
The GD setting disables the input signal, the AC/DC (33)
pushbutton and the VOLTS/DIV (21) knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
The ”0 Volt” position is indicated by a ”⊥” symbol in the
readout. Please note Y-POS. II (11).
INV.:
Pressing and holding this pushbutton switches the channel
II invert function on or off.The invert on condition is indicated
by the readout with a horizontal bar above “Y2” (Yt mode) or
“Y” (XY mode). The invert function causes the signal display
of channel II to be inverted by 180°.
(35)TRIG. EXT. / INP. (Z) – BNC socket with two functions.
The impedance of this input is 1MΩII 20pF.The outer (ground)
connection is galvanically connected to the instrument
ground and consequently to the safety earth contact of the
line/mains plug.
TRIG. EXT.:
Signals applied at this input can be used for external
triggering. Briefly pressing the TRIG. (20) pushbutton, until
the information “TR:EXT, slope, coupling” is visible in the
readout and the TRIG. “EXT” -LED is lit, switches the input
active as an external trigger source.
Z Input:
The input can be used for Z (trace intensity) modulation, if
software setting for “INPUT Z” is ON and neither external
triggering nor component tester operation is selected.
“INPUT Z” is part of the MISCELLANEOUS menu which is a
submenu of the SETUP menu.
The trace becomes blanked by a TTL high signal (positive
logic). Voltages higher then +5V are not permissible.
Below the CRT there are the controls for the readout, the
component tester and the square wave calibrator with their
outputs.
(36)MENU – Pushbutton.
Pressing and holding this pushbutton activates the display
of the MAIN MENU. It contains the submenus TEST &
CALIBRATE and SETUP.The intensity of the display depends
on the RO-INTENS (4) setting.
Once a menu is displayed, the following pushbuttons are
of importance:
1. SAVE and RECALL (7) pushbutton.
Briefly pressing selects the submenu or an item within
the submenu.
2. SAVE –SET (7) pushbutton.
Pressing and holding the SAVE (7) pushbutton calls (sets)
the menu or the previously selected item. In those cases
where the item is marked with ON / OFF the setting
changes from ON to OFF or vice versa.
In some cases the called function is not performed at once
and a warning is displayed to protect from calling the
function inadvertently. Then the function can be called by
pressing and holding the SAVE pushbutton (SET function)
again, otherwise if the function was called inadvertently
the proceeding can be cancelled by pressing the
AUTOSET (3) pushbutton.
3. AUTOSET (3) pushbutton.
Each time the AUTOSET pushbutton is pressed the menu
is switched back one step until MAIN MENU is displayed.
Then pressing the AUTOSET pushbutton again switches
the menu operation off and the AUTOSET pushbutton is
automatically set to the normal function.
(37)ON/OFF - CHI/II - 1/∆∆
∆∆
∆t– This pushbutton has several
functions.
The following description of the cursor related controls
assumes that the readout is visible.
ON/OFF:
Pressing and holding the pushbutton switches both cursor
lines on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on. If the cursor
lines are switched off and MEAN VALUE ON is activated
(Menu: SETUP > MISCELLANEOUS) the instrument is
prepared for DC Mean Value display. Please note section
“Mean Value Display”.
CHI/II:
This function is required and available only in DUAL and
XY mode, if cursor supported voltage measurement (∆∆
∆∆
∆V (39))
is chosen, as the deflection coefficients may be different.
Briefly pressing the pushbutton switches between the
deflection coefficients of channel I and channel II and vice
versa. The measured result is displayed by the readout with
“∆∆
∆∆
∆V1:...” or “∆∆
∆∆
∆V2:...” providing the deflection coefficient is
Controls and Readout
MADE IN GERMANY
INV.
INPUT CH I (HOR. INP.(X))
AC/DC
max.
250 Vp
1MΩII
15pF
INPUT CH II
max.
250 Vp
1MΩII
15pF
TRIG. EXT. INP
max.
100 Vp
(Z)
x1/x10 INV.
x1/x10
CAT
I
!!!
GD AC/DC GD
28 30 31 32 34
29 33 35
CT
ON
OFF
∆V
∆t1
∆t
1kHz
1MHz
CAL.
0.2 Vpp
CURSOR
CH I/II
I/II
MENU
TRK
37 41
42 40 39 38 37 36

Subject to change without notice
20
calibrated.The settings of the cursors must relate to the signal
of the selected channel.
In XY-mode the instrument is automatically set to ∆∆
∆∆
∆V
measurement. The deflection coefficient selected for each
channel may be different, thus as in DUAL mode the ∆∆
∆∆
∆V
cursor measurement requires a channel selection. Under
channel I (X signal) measuring condition the cursor lines are
displayed as vertical lines and the readout displays “∆∆
∆∆
∆VX...”.
Briefly pressing the pushbutton changes to channel II (Y
signal) voltage measurement. Then the cursor lines are
displayed as horizontal lines and the readout indicates “ ∆∆
∆∆
∆VY...”.
In CH I or CH II single mode, only one deflection coefficient
is present and there is no requirement to select between
different deflection coefficients. Consequently the pushbutton
is inactivated in combination with ∆∆
∆∆
∆Vmeasurements.
ADD (addition) mode requires an equal deflection coefficient
setting for both channels. Cursor voltage measurement now
incorporates both channels, therefore no switching is possible.
1/∆∆
∆∆
∆t:
Briefly pressing the pushbutton selects between time and
frequency measurement (only when “∆∆
∆∆
∆t”or “f”is displayed
by the readout). The vertical cursor lines and the
measurement result apply to the active time base (A or B,
resp. B in alternate time base mode).
In calibrated time base condition the readout displays “∆∆
∆∆
∆t:...”
if time measurement is chosen. After switching over to 1/∆∆
∆∆
∆t
(frequency) “f:...” is displayed. If the time base is uncalibrated
the readout displays “ ∆∆
∆∆
∆t>...” or “f<...”.
(38)TRK –CURSOR-Track function.
The following description of the cursor related controls
assumes that the readout is visible.
Briefly pressing simultaneously both buttons ON/OFF - CHI/
II - 1/∆∆
∆∆
∆t (37) and I/II - ∆∆
∆∆
∆V/∆∆
∆∆
∆t (39) switches over from single
cursor line operation to track mode and vice versa.
In TRK (track) mode both cursor lines are indicated as active
(both dotted lines uninterrupted).
(39) I/II - ∆∆
∆∆
∆V/∆∆
∆∆
∆t – Pushbutton with two functions.
The following description of the cursor related controls
assumes that the readout is.
I/II:
Briefly pressing this button changes the active (controllable)
cursor in the sequence I - II - I, if TRK (track) mode is not
active.The active cursor is indicated by a continuously dotted
line. An interrupted dotted line indicates the inactive cursor.
∆∆
∆∆
∆V/∆∆
∆∆
∆t:
Pressing and holding this pushbutton changes from voltage
to time (or frequency) measurement and vice versa. In XY
mode the instrument is automatically set to ∆∆
∆∆
∆V.
∆∆
∆∆
∆V:
In all ∆V (voltage) measurement conditions, the division ratio
of the probe(s) must be taken into account. In case of x10
(10:1) probes, the probe factor can be automatically included
(see item (29) and (33)).
If a x100 (100:1) probe is used, switch to the probe symbol
(10:1) off if displayed. Then multiply the voltage value
displayed in the readout by 100.
1st: Time base modes
(CH I or CH II single channel mode, DUAL and ADD).
In the voltage measurement condition, the cursors are
displayed as horizontal lines and the result is displayed
in the readout.
Single channel mode (channel I or channel II)
The ∆∆
∆∆
∆Vmeasuring result is automatically related to the
deflection coefficient of the active channel. The readout
displays “ ∆∆
∆∆
∆V1...” or “∆∆
∆∆
∆V2...”.
Dual mode
The CURSOR lines must be set on the channel I or
channel II signal. As the deflection coefficients may be
different, it will be required to select between the
deflection coefficients of channel I and II.
Please note
item CHI/II (37)
.
Addition mode (ADD)
In ADD (addition) mode normally two input signals are
displayed as one signal (sum or difference). As the result
can only be determined if both (calibrated) deflection
coefficients are equal, the CHI/II (37) selection function
is deactivated. In that case the readout indicates “ ∆∆
∆∆
∆V...”
without any additional channel information. Different
deflection coefficient settings or uncalibrated deflection
coefficient(s) are indicated in the readout as “Y1<>Y2”.
2nd: XY mode
In XY mode the instrument is automatically set to ∆∆
∆∆
∆V
measurement. The deflection coefficient selected for
each channel may be different, thus as in DUAL mode
the ∆∆
∆∆
∆Vcursor measurement requires a channel selection.
Under channel I (X signal) measuring condition the cursor
lines are displayed as vertical lines and the readout
displays “∆∆
∆∆
∆VX...”. The cursor lines are displayed as
horizontal lines and the readout indicates “∆∆
∆∆
∆VY...” if the
channel II deflection coefficient is chosen.
∆∆
∆∆
∆t:
In time or frequency measurement condition two vertical
cursor lines are displayed.
The measurement result is indicated as “∆∆
∆∆
∆t...” (time
measurement) or “f...” (frequency measurement).
Please
note item 1/
∆∆
∆∆
∆
t (37)
.
For frequency measurement, the distance between
the cursors must equal exactly one signal period.
In XY mode the time base is switched off. Conse-
quently time or frequency measurement is disabled.
(40)CURSOR – Center biased lever.
The active cursor line (in track mode: both lines) can be shifted
in the required direction, until the graticule limits are reached.
The directions are marked on the front panel and depend on
the selected measurement (∆U, ∆t or 1/∆t). Which of two
shift speeds is used, depends on how far the CURSOR knob
is pressed. With slight knob pressure the cursor line(s) moves
Controls and Readout
CT
ON
OFF
∆V
∆t1
∆t
1kHz
1MHz
CAL.
0.2 Vpp
CURSOR
CH I/II
I/II
MENU
TRK
37 41
42 40 39 38 37 36
Table of contents
Other Hameg Test Equipment manuals

Hameg
Hameg HM 203-7 User manual

Hameg
Hameg HM303-6 User manual

Hameg
Hameg HM 204 User manual

Hameg
Hameg HM6050-2 User manual

Hameg
Hameg HM1507-3.02 User manual

Hameg
Hameg HM 407A User manual

Hameg
Hameg HM 203-6 User manual

Hameg
Hameg HM 203 User manual

Hameg
Hameg HM 1004-3.01 User manual

Hameg
Hameg HM404-2.02 User manual

Hameg
Hameg HM 605 User manual

Hameg
Hameg HMO Series User manual

Hameg
Hameg HM 203-6 User manual

Hameg
Hameg HM1000 User manual

Hameg
Hameg CombiScope HM2008 User manual

Hameg
Hameg HM1004 User manual

Hameg
Hameg HM304 User manual

Hameg
Hameg HM 1007 User manual

Hameg
Hameg HMO3000 Series Instruction manual

Hameg
Hameg HM 604 User manual