5. Battery Backup allows for off-mains
emergency operation for up to 15 minutes
under a normal load. To ensure long battery
life, the bed should be connected to mains
power, at all possible times. Do not exceed
the duty cycle.
6. Electrical protection of wiring and cords will
differ between Class 1 (Earthed) and Class 2
(Double Insulation) Beds.
4.3 Standard features
The backrest is a large mattress panel that raises
from a supine position to an upright position,
convenient for sitting, which allows a patient
to enjoy the flexibility of changing to multiple
positions for comfort and health.
Kneebreak
The kneebreak is a double mattress panel
that splits as it electronically lifts or bends
the patient’s thigh and calf at the knee, thus
combining functionality and comfort for both
the patient and carer.
Auto-Contour
A Handset feature that uses one button to adjust
both Backrest & Kneebreak simultaneously, into
a cardiac chair position, even when the bed is in
it’s lowest height position.
Battery Backup
The Battery Backup allows for operation of the
bed, should mains power fail, or if the bed is
being operated temporarily away from mains
power. The battery can operate for a total
time of up to 15 minutes under a normal load.
Make sure not to exceed the duty cycle (2 min
operation, 18 minutes rest).
To ensure long battery life, the bed should be
connected to mains power, at all possible times.
If the backup battery becomes flat, raising
lowering functions will slow down or cease. A
replacement should be ordered and fitted.
4. Product description ENG
Each time the bed is allocated, it is
recommended that the patient is instructed in
all the functions that are important for him/her,
by the operator or user.
4.2 Structural Design
Mattress Platform: The mattress platform is
a four panel design, divided into a back rest,
a seat section and a double panel kneebreak.
The mattress base can be horizontally adjusted
in height. The bed can be adjusted to head-
low (Trendelenburg) or feet-low (Reverse
Trendelenburg ) positions.
Chassis:
The chassis is constructed of welded steel.
It features four individually locking castors
that include: 3 grey Brake Castors and 1 black
Directional Lock Castor.
Electrical Adjustment System:
The electric adjustment system comprises:
1. The Backlit Handset for patient control of
the bed’s positions. It is a ‘remote-control’
attached via a curly cord, to the bed’s control
box; consisting of a robust, easy-care, wash-
down plastic casing with a backlit membrane
keypad.
2. The Attendant Control Keypad
(only for the FloorLine-i Plus bed)
(ACK) is located at the foot end of the bed. It
is used to lock out and replace the handset,
for nurse convenience. It comprises a wash-
down membrane keypad, with indictor lights.
3. Actuators & Junction Boxes for adjusting and
controlling the backrest, kneebreak and the
bed heights.
4. The Linak Open-Bus Central (only for the
FloorLine-i Plus bed) Control Box Unit and
Battery Backup are located underneath the
Head End Cover. They contain a low-voltage,
safe 24V transformer. The electric motors/
actuators, the battery backup, the ACK and
handsets (24V) are connected to the Control
Box and Junction Boxes via dust and moisture
proof plugs and cords/cables.