6
Important: Do not connect PAT leads to sensitive components
such as PCBs, control switches and the like.
If in doubt as to the most suitable connection point contact
Humboldt Service Department for assistance.
3. Procedure
3.1 10 rock specimens each weighing 40 to 60 g, giving a total
weight of 450 to 550 g, are selected to represent the sample.
3.2 The selected specimens are placed in the drum and dried to
constant weight at a temperature of 105°C (this usually
requires 3 to 12 hours in an oven).
3.3 The drum assembly for each test is placed in the related
water tank and coupled in series to the motor drive unit.
Note: the drum assemblies are marked ‘left hand’ and ‘right
hand’ and should be placed accordingly, relative to the motor
drive unit.
3.4 The tank is lled with slaking uid to within 20 mm of the
drum axle (indicated by a mark on the tank). The test is immedi-
ately set in motion for a period of 10 minutes.
3.5 At the end of the period the drive is immediately switched
off, the drum assembly removed from the tank, and the assem-
bly with the specimens dried to constant weight at 105°C.